K252579 · Ziehm-Orthoscan, Inc. · OXO · Jan 14, 2026 · Radiology
Device Facts
Record ID
K252579
Device Name
Orthoscan TAU MVP Mini C-Arm System
Applicant
Ziehm-Orthoscan, Inc.
Product Code
OXO · Radiology
Decision Date
Jan 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Orthoscan TAU MVP system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments. The scope used with Orthoscan TAU MVP system is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.
Device Story
Mobile mini C-arm system; integrates fluoroscopic X-ray imaging with arthroscopic video imaging. Inputs: X-ray signals from C-arm; video signals from sterile, single-use arthroscopic scopes (60mm, 90mm, 120mm lengths) with LED illumination. Transformation: Video processing unit and imaging control board process inputs for concurrent display on workstation monitor. Used in hospitals, surgery centers, and clinics by physicians. Output: Real-time fluoroscopic and arthroscopic images; stored video/images. Clinical utility: Enables simultaneous visualization of internal joint structures and fluoroscopic guidance during orthopedic/arthroscopic procedures; facilitates intra-articular assessment and surgical navigation. Benefits: Standardized platform combining two imaging modalities; improved procedural workflow; real-time visual feedback for minimally invasive interventions.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical safety (IEC 60601-1, 60601-2-54, 60601-2-28), EMC, and software validation. Image quality analysis (sharpness, color, SNR) demonstrated performance equivalent or superior to the MIDASVu predicate. Human factors/usability testing (IEC 62366-1) confirmed safe operation of dual-mode workflows. Latency testing verified end-to-end display latency remained within acceptable limits for fluoroscopy, arthroscopy, and dual-mode operation.
Technological Characteristics
Mobile mini C-arm with integrated video processing unit. X-ray imaging via C-arm; arthroscopic imaging via single-use scopes with LED light source. Connectivity: DICOM wired/wireless network interfaces. User interface: Touch screen, wired/wireless footswitch. Compliance: IEC 60601-1, 60601-2-54, 60601-2-28, 21 CFR Subchapter J. Software: Integrated control board for concurrent fluoroscopic/arthroscopic display.
Indications for Use
Indicated for patients of all ages, including pediatric, requiring fluoroscopic visualization of limbs, extremities, and shoulders during diagnostic, surgical, and emergency procedures. Also indicated for diagnostic and operative arthroscopic and endoscopic visualization of interior cavity joints and body cavities.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 14, 2026
Ziehm-Orthoscan, Inc.
% Kevin Bridgman
VP of Regulatory Affairs and Quality Assurance
14555 N. 82nd St.
SCOTTSDALE, AZ 85260
Re: K252579
Trade/Device Name: Orthoscan TAU MVP Mini C-Arm System
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-Intensified Fluoroscopic X-Ray System
Regulatory Class: Class II
Product Code: OXO, JAA, MQB, HRX, GCJ
Dated: December 15, 2025
Received: December 15, 2025
Dear Kevin Bridgman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252579 - Kevin Bridgman
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252579 - Kevin Bridgman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252579 | |
| Device Name Orthoscan TAU MVP Mini C-Arm System | |
| Indications for Use (Describe) The Orthoscan TAU MVP system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments. The scope used with Orthoscan TAU MVP system is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Traditional 510(k) Summary
Traditional 510 (k) Premarket Notification Submission- Ziehm-Orthoscan, Inc.
Orthoscan TAU MVP Mini C-Arm System (K252579)
Submitter: Ziehm-Orthoscan, Inc.
14555 N 82nd St.
Scottsdale, AZ 85260
Phone: (480) 503-8010
Fax: (480) 503-8011
Primary Contact: Kevin Bridgman; Kevin.bridgman@ziehm-orthoscan.com
Secondary Contact: TuAnh Ngo; tuanh.ngo@ziehm-orthoscan.com
Date: January 14, 2026
In accordance with the requirements of 21 CFR 807.92 the following 510(k) summary of information is provided:
Device Trade Name: Orthoscan TAU MVP Mini C-Arm System
510(k) Number: K252579
Common/Usual Names: Fluoroscopic X-Ray System, Mobile Mini Mobile C-arm, Mini C-arm, Arthroscope, Endoscope and Accessories
Regulation: 21 CFR 892.1650
Device Class: Class II
Regulation Names: Image-intensified fluoroscopic x-ray system
Device: Image-Intensified Fluoroscopic X-ray System, Mobile
Product Codes: OXO, JAA, MQB, HRX, GCJ
Primary Predicate: Orthoscan TAU Mini C-Arm
510(k) Number: K250587
Regulation: 21 CFR 892.1650
Device Class: Class II
Regulation Names: Image-intensified fluoroscopic x-ray system
Product Code: OXO, JAA, MQB
Secondary Predicate: MIDASVu
510(k) Number: K243020
## General Description:
The proposed Orthoscan TAU MVP (Multi Viewing Platform) Mini C-Arm system models 1000-0022, 1000-0023 retain identical function as the predicate Orthoscan TAU Mini C-Arm (250587) as a mobile fluoroscopic mini C-arm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment and surgical procedures involving anatomical regions such as but not limited to that of extremities, limbs, shoulders and knees and hips. The system consists of C-Arm support attached to the image workstation.
The proposed Orthoscan TAU MVP Mini C-Arm system models 1000-0022, 1000-0023 represent a modification of our presently legally marketed devices Orthoscan TAU Mini C-Arm (K250587). The proposed Orthoscan TAU MVP Mini C-Arm System will consist of the Orthoscan TAU Mini C-Arm X-ray System with additional arthroscopic imaging capability to further standardize the product platform by incorporating components currently available in predicate devices into the Orthoscan TAU MVP system.
{5}
The proposed new Orthoscan TAU MVP system is a collaborative effort to display the fluoroscopic and arthroscopic image concurrently.
The proposed modifications to the predicate encompass the incorporation of hardware and software from MIDASVu arthroscopic imaging system reusable tablet (K243020) utilizing a modified mechanical design to incorporate a video processing unit imaging control board, umbilical cord and arthroscopic imaging system comprised of sterile, single-use scopes (K243020). The scopes include camera and image capture features with LED light source. The distal tip of the scopes contains the camera, illumination, and imaging optics. The scopes are available in three lengths: 60mm, 90mm, and 120mm. The scopes and the TAU MVP Mini C-Arm system work in concert as a system to acquire, display and record an intra-articular image as well as store images and video taken during the procedure.
For both the predicate TAU (K250587), MIDASVu (K243020) and proposed device Orthoscan TAU MVP, the following are unchanged: identical C-arm support and mechanical connections, balancing, locking, rotations, work-station platform, monitor display and main user interface controls, touch screen interface, selectable imaging, X-ray technique control, entry of patient information, wired or wireless footswitch operation, interface connection panel and DICOM fixed wire and wireless network interfaces, single use disposable scope, optics and components, mechanical dimensions and materials, imaging optics and connection into system. The proposed Orthoscan TAU MVP device uses a similar and substantially equivalent arthroscopic image acquisition and display type.
## Indications for Use:
The Orthoscan TAU MVP system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments. The scope used with Orthoscan TAU MVP system is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.
## Indications for Use Comparison:
The proposed Orthoscan TAU MVP Mini C-Arm System has the same indications for use as the predicate devices Orthoscan TAU (K250587) and MIDASVu (K243020).
## Technology:
The changes to the proposed device Orthoscan TAU MVP Mini C-Arm System represent a modification of our presently legally marketed device Orthoscan TAU Mini C-Arm (K250587). With the introduction of the MIDASVu arthroscopic imaging system comprised of sterile, single-use scopes (K243020), arthroscopic imaging capability is established to further standardize the product platform by incorporating features currently available in predicate devices into the proposed Orthoscan TAU MVP.
## Summary of Technological Characteristics:
The comparisons of the proposed Orthoscan TAU MVP Mini C-Arm demonstrate that the scientific and technology characteristics indicate substantial equivalence to the predicate device Orthoscan TAU Mini C-Arm (K250587) and MIDASVu (K243020).
## Adverse Effects on Health:
The proposed Orthoscan TAU MVP's potential radiation, mechanical, and electrical hazards are identified and analyzed as part of risk management and controlled by meeting the applicable CDRH 21 CFR subchapter J performance requirements, Recognized Consensus Standards, designing and manufacturing under Ziehm-Orthoscan, Inc. Quality System, and system verification and validation testing ensure the device performs to the product specifications and its intended use. The adherence to these applicable regulations and certification to Recognized Consensus Standards that apply to this product provides the assurance of device safety and effectiveness.
{6}
Non-Clinical Testing Summary:
Verification and Validation including hazard mitigations executed resulted in demonstrated system that met Design Input and user needs. The device was tested by certified test laboratory resulting in device being certified compliant with IEC 60601-1 ED 3.2 series, including IEC 60601-2-54 and IEC 60601-2-28. Further, the device met all applicable sections of 21 CFR Subchapter J performance standards. The proposed Orthoscan TAU MVP Mini C-Arm development occurred under our design control processes, software development processes, and overall quality management system. They included but are not limited to:
- Risk Analysis
- Required reviews
- Design reviews
- Component testing
- Integration testing
- Performance testing
- Safety testing
- Product use testing
The Orthoscan TAU MVP system arthroscopic imaging underwent non-clinical testing to confirm safety, effectiveness, and performance. Electrical safety and EMC testing met IEC 60601 standards, while software validation ensured intended functionality. Software and cybersecurity measures were validated. Results demonstrate the device is as safe and effective as the predicate for its intended use.
Ziehm-Orthoscan, Inc. conducted testing using a leading industry software utility to provide objective image quality statistical analysis of camera systems and compare the displayed arthroscopic image presented by the Orthoscan TAU MVP system to the predicate IntraVu TabletVu display system. Sharpness, color analysis and signal to noise were evaluated and showed that the Orthoscan TAU MVP system performs with equivalent or better arthroscopic image quality when compared to than the predicate IntraVu TabletVu.
Additionally, human factors and usability validation testing was conducted in accordance with IEC 62366-1 and FDA human factors guidance to confirm that intended users can safely and effectively operate the device. The usability validation evaluated representative clinical workflows, including standard fluoroscopic operation, arthroscopic imaging, and dual-mode operation combining fluoroscopy and arthroscopy, to ensure that mode transitions, user interface controls, and system feedback are clear and intuitive. Testing demonstrated that critical tasks, including those performed during dual-mode operation, can be completed without use-related errors that could result in unacceptable risk, and that identified hazards are adequately mitigated through the user interface design, labeling, and training.
Image latency testing was also performed to evaluate absolute end-to-end display latency in fluoroscopy-only, arthroscopy-only, and dual-mode operation, ensuring latency remained below industry-standard criteria established in the scientific literature. Comparative latency was also assessed between fluoroscopy-only mode and the fluoroscopy function in dual-mode operation, and between arthroscopy-only mode and the arthroscopy function in dual-mode operation, to verify that dual-mode operation did not introduce additional risk. Testing was conducted under representative and worst-case operating conditions. Results demonstrated that display latency remained within established acceptance criteria and dual mode functionality did not introduce unacceptable risk related to visualization delay, hand-eye coordination, or procedural workflow.
Therefore, Ziehm-Orthoscan, Inc. believes the proposed Orthoscan TAU MVP Mini C-arm safety and effectiveness to be substantially equivalent to that of the predicate device Orthoscan TAU (K250587) and MidasVu (K243020).
Substantial Equivalence Conclusion:
{7}
The changes and differences of the proposed Orthoscan TAU MVP Mini C-arm system do not change the control mechanism, operating principle, energy type, or intended use found on the predicate devices Orthoscan TAU Mini C-Arm (K250587) and MIDASVu (K243020).
Ziehm-Orthoscan, Inc. considers the proposed Orthoscan TAU MVP Mini C-Arm System to be as safe as effective and performs substantially equivalent to the predicate devices Orthoscan TAU Mini C-Arm (K250587) and MIDASVu (K243020) in accordance with its labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.