K252570 · Qpix Solutions, Inc. · MUH · Nov 10, 2025 · Dental
Device Facts
Record ID
K252570
Device Name
EzSensor HD, EzSensor UHD
Applicant
Qpix Solutions, Inc.
Product Code
MUH · Dental
Decision Date
Nov 10, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
EzSensor HD / UHD is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.
Device Story
Digital intraoral sensor system; captures dental x-ray photons using Cesium Iodide (CsI) scintillator and silicon sensor; converts photons to electronic impulses; transmits data via integrated USB 2.0 module to computer. Used in dental clinics by dentists; provides near-instantaneous image acquisition. Output viewed via EzDent-i software; enables diagnostic visualization of bony and soft tissue. Ergonomic design incorporates stainless-steel internal frame and soft silicon exterior for patient comfort and impact protection. Device facilitates dental diagnosis through high-resolution imaging; replaces traditional film-based radiography.
Clinical Evidence
Bench testing only. Comparative evaluation of clinical images between subject and predicate devices performed. Metrics included DQE, MTF, and NPS. Results demonstrated equivalent image quality, contrast, and resolution. Electrical, mechanical, and environmental safety verified per IEC 60601-1 and IEC 60601-1-2 standards.
Indicated for use by dentists to collect, store, view, and manipulate dental x-ray images for diagnostic purposes.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 10, 2025
Qpix Solutions Inc
% Dave Kim
President
Mtech Group LLC
7505 Fannin St. Suite 610
HOUSTON, TX 77054
Re: K252570
Trade/Device Name: EzSensor HD, EzSensor UHD
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral Source X-Ray System
Regulatory Class: Class II
Product Code: MUH
Dated: August 11, 2025
Received: October 22, 2025
Dear Dave Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252570 - Dave Kim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K252570 - Dave Kim
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252570 | |
| Device Name EZ Sensor HD, EZ Sensor UHD | |
| Indications for Use (Describe) | |
| EzSensor HD / UHD is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Submission-EzSensor HD, EzSesnsor UHD
# 510(k) Summary K252570
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510K summary prepared: October 11, 2025
## 1. Submitter's Name, address, telephone number, a contact person:
Submitter's Name : Qpix Solutions Inc
Submitter's Address: 1001 Aviation Parkway, Ste 200, Morrisville, North Carolina, 27560 USA
Submitter's Telephone: +1-919-908-6917
Contact person: Mr. Seungman Yun / CEO / +1-919-908-6917
Official Correspondent: Mtech Group LLC
Mr. Dave Kim, MBA
Address: 7505 Fannin St. Ste 610, Houston, TX 77054
Telephone: +713-467-2607
Email: davekim@mtechgroupllc.com
## 2. Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
Trade/proprietary name: EzSensor HD, EzSensor UHD
Model Name: IOS-U20IF, IOS-U15IF, IOS-U10IF
Common Name: Digital Dental Intra Oral Sensor
Regulatoin number: 21 CFR 872.1800
Classification Name: Extraoral source X-ray system
Regulatory Class: Class 2
Product Code: MUH
{5}
510(k) Submission-EzSensor HD, EzSesnsor UHD
## 3. Predicate device
Trade Name: X Sensor
510(k) Number: K242778 (Decision Date –10/11/2024)
The Regulation Number: 21 CFR 872.1800
Classification Name: Extraoral Source X-ray System
Regulatory Class: Class 2
Primary Product Code: MUH
## 4. Device Description
EzSensor HD / UHD is a digital intraoral sensor which acquires intra oral images with a sensor that is connected to a computer to produce an image almost instantaneously following exposure. The primary advantage of Indirect sensor systems is the speed with which images are acquired. The ergonomic design based on human intraoral anatomy improves patient comfort. EzSensor HD / UHD includes the software (firmware) with basic level of concern.
## 5. Indication for use
EzSensor HD / UHD is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.
## 6. Summary of Design Control Risk management
The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.
## 7. Summary of the technological characteristics of the device compared to the predicate device
Digital intraoral sensor described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device.
These differences do not raise the questions of safety or effectiveness. Based on the laboratory testing results submitted in this 510k, we conclude that the subject device is substantially equivalent to the predicate device.
Qpix Solutions Inc
{6}
510(k) Submission-EzSensor HD, EzSesnsor UHD
The potential risks for a new sensor size were analyzed by conducting complete verification for IEC 60601-1 and drop & vibration test which included electronic shock, leakage current, etc.
While applying the stainless-steel material to the inside of the frame, soft silicon material surrounds the exterior of the USB connector to protect the sensors from a potential external impact. Additional risk analysis was conducted to mitigate the potential risks that may arise with respect to leakage current, sensor fracture or breakage, and cable disconnection. The risk mitigation measures were satisfactory to manage the new risks identified and the residual risks were within acceptable limits.
| Characteristic | Proposed | Predicate |
| --- | --- | --- |
| Manufacturer | Qpix solutions Inc | Qpix solutions Inc |
| Device's name | EzSensor HD, EzSensor UHD (Model: IOS-U20IF, IOS-U15IF, IOS-U10IF) | X Sensor (Model: IOS-A15IF, HDI-15DGF) |
| Feature | | |
| 510(k) number | K252570 | K242778 |
| X-ray converter | CsI | CsPbBr3 |
| Indications for use | EzSensor HD / UHD is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists. | Digital Dental Intra Oral Sensor is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. |
| Sensor Dimension(mm) (±10%) | Size 1.0: 36.8 x 25.4 Size 1.5: 39.5 x 29.2 Size 2.0: 42.9 x 31.3 | Size 1.5: 41.1 x 30.4 |
| Sensor Thickness(mm) | 4.8 | 4.8 |
| Active Area(mm) | Size 1.0: 30.01 x 20.01 Size 1.5: 33.00 x 23.98 Size 2.0: 35.99 x 25.99 | Size 1.5: 23.98 x 33.00 |
| USB Module | Integrated USB 2.0 module | Integrated USB 2.0 module |
| Max. Resolution (lp/mm) | 33.8 | 33.8 |
Qpix Solutions Inc
{7}
510(k) Submission-EzSensor HD, EzSesnsor UHD
| Pixel Pitch (μm) | Full Resolution | 14.8 | 14.8 |
| --- | --- | --- | --- |
| DQE (6 lp/mm) | Full Resolution | 0.318 | 0.278 |
| MTF (3 lp/mm) | Full Resolution | 0.668 | 0.716 |
| Viewer Software | | EzDent-i (K241114) dental imaging viewer by EWOO SOFT | EzDent-i (K241114) dental imaging viewer by EWOO SOFT |
# 8. Summary of Performance Testing
The intended use, application of EzSensor HD(EzSensor UHD) is the same as that of the predicate device, X Sensor.
EzSensor HD(EzSensor UHD) is a Indirect conversion sensor that utilizes a photoconductor (CsI) and single crystal Silicon as the sensing means whereas X Sensor, the predicate device, utilizes a photoconductor(CsPbBr3) and single crystal Silicon as the sensing means.
The performance test result indicates that the EzSensor HD(EzSensor UHD) intra oral sensor performed equally to the X Sensor, the predicated device, as both sensors have the same pixel pitch, thereby providing the same maximum line-pair resolution. No additional safety risk is identified in the bench test: Non-clinical report.
Non-clinical test was performed according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X- ray Imaging Devices".
Due to differences in the utilization a photoconductor between the EzSensor(EzSensor UHD) CsI and the predicate device X Sensor CsPbBr3, there is a slight variation in EzSensor HD MTF(0.668) for X Sensor MTF(0.716). However, this difference does not affect their practical equivalence performance.
Summary of EzSensor HD for Image Quality Test
The image quality test of EzSensor HD (Report No.: Q-20250613-01) confirmed that there were no special issues with the resolution and image quality.
DQE, MTF, and NPS test results demonstrated that EzSensor HD(EzSensor UHD) has equivalent performance outcome than X Sensor, the predicate sensor.
Qpix Solutions Inc
{8}
510(k) Submission-EzSensor HD, EzSesnsor UHD
Non-clinical test was performed according to FDA Guidance “Guidance for the Submissions of 510(k)’s for Solid State X- ray Imaging Devices”.
Electrical, mechanical, environmental safety and performance testing were performed according to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance). EMC testing was conducted in accordance with standard IEC 60601-1-2:2014+AMD1:2020.
The clinical images obtained from the EzSensor HD(EzSensor UHD) and X Sensor were reviewed and rated comparatively.
The image quality in terms of contrast and resolution are overall similar for the EzSensor HD(EzSensor UHD), the proposed new device and X Sensor, the predicate device
There are no observable radiographic findings and no quality issues with intra oral diagnostic images provided by both sensors.
The proposed device produces overall better definition and grayscale of bony and soft tissue images.
The expert evaluation of the representative clinical images demonstrated that the proposed device achieved equivalent image quality to the predicate device.
# 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Qpix Solutions Inc concludes EzSensor HD(EzSensor UHD) is safe and effective and substantially equivalent to predicate device as described herein.
Qpix Solutions Inc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.