← Product Code [LZA](/productcode/LZA) · K252549

# Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Green), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid; Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid (K252549)

_Kossan International Sdn Bhd · LZA · Jan 13, 2026 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252549

## Device Facts

- **Applicant:** Kossan International Sdn Bhd
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jan 13, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid as per ASTM D6978-05 (Reapproved 2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

## Device Story

Disposable non-sterile powder-free nitrile examination gloves; dual-color (White-Green or White-Blue). Worn by healthcare examiners to provide barrier protection against contaminants. Tested for chemical permeation resistance per ASTM D6978-05 (2023) against various chemotherapy agents, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid. Provides barrier protection; does not include electronic components or software. Used in clinical settings; protects examiner from patient-borne pathogens and chemical exposure. Warning provided for specific drugs (Carmustine, Thiotepa) with insufficient breakthrough times.

## Clinical Evidence

Bench testing only. No clinical data provided. Permeation testing conducted per ASTM D6978-05 (Reapproved 2023) to determine breakthrough detection times for chemotherapy drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid.

## Technological Characteristics

Material: Nitrile. Form factor: Non-sterile, powder-free, dual-color examination glove. Testing standard: ASTM D6978-05 (Reapproved 2023) for chemical permeation. No electronic, software, or energy-based components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 13, 2026

Kossan International sdn Bhd
Cho Sow Fong
Assistant General Manager, Regulatory Affairs
KOSSAN Tower No. 6D, Persiaran Setia Dagang, Setia Alam
Seksyen U13
Shah Alam, Selangor 40170
Malaysia

Re: K252549

Trade/Device Name: Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Green), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid; Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid

Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZA, LZC, OPJ, QDO
Dated: January 5, 2026
Received: January 5, 2026

Dear Cho Sow Fong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252549 - Cho Sow Fong
Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K252549 - Cho Sow Fong
Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, Ph.D., M.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252549

Device Name
Non-sterile Powder Free Nitrile Patient Examination Glove, Dual Color (White-Green), tested for use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid.

Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid as per ASTM D6978-05 (Reapproved 2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

|  Chemotherapy Drugs Permeation Test | Minimum Breakthrough Detection Time in Minutes  |
| --- | --- |
|  Chemotherapy Drugs (Concentration) | White-Green  |
|  Carmustine (BCNU) (3.3 mg/ml) | 30.0 (40.0, 30.0, 30.0)  |
|  Cisplatin (1.0 mg/ml) | >240  |
|  Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240  |
|  Cytarabine HCl (100 mg/ml) | >240  |
|  Dacarbazine (DTIC) (10.0 mg/ml) | >240  |
|  Doxorubicin Hydrochloride (2.0 mg/ml) | >240  |
|  Etoposide (20.0 mg/ml) | >240  |
|  Fluorouracil (50.0 mg/ml) | >240  |
|  Ifosfamide (50.0 mg/ml) | >240  |
|  Methotrexate (25.0 mg/ml) | >240  |
|  Mitomycin C (0.5 mg/ml) | >240  |
|  Mitoxantrone (2.0 mg/ml) | >240  |
|  Paclitaxel (Taxol) (6.0 mg/ml) | >240  |
|  Thiotepa (10.0 mg/ml) | 60.0 (100.0, 80.0, 60.0)  |
|  Vincristine Sulfate (1.0 mg/ml) | >240  |

Please note that Carmustine (BCNU) and Thiotepa has low permeation times of 30.0 and 60.0 minutes respectively.
Warning: Do not use with Carmustine and Thiotepa

|  Opioid Drugs and Others Permeation Test | White-Green  |
| --- | --- |
|  Fentanyl Citrate Injection (100mcg/2mL) | >240  |
|  Simulated Gastric Acid | >240  |
|  Xylazine HCl (100 mg/ml) | >240  |

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740

{4}

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 2 of 2

{5}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252549

Device Name
Non-sterile Powder Free Nitrile Patient Examination Glove, Dual Color (White-Blue), tested for use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid.

Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid as per ASTM D6978-05 (Reapproved 2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

|  Chemotherapy Drugs Permeation Test | Minimum Breakthrough Detection Time in Minutes  |
| --- | --- |
|  Chemotherapy Drugs (Concentration) | White-Blue  |
|  Carmustine (BCNU) (3.3 mg/ml) | 30.0 (40.0, 30.0, 40.0)  |
|  Cisplatin (1.0 mg/ml) | >240  |
|  Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240  |
|  Cytarabine HCl (100 mg/ml) | >240  |
|  Dacarbazine (DTIC) (10.0 mg/ml) | >240  |
|  Doxorubicin Hydrochloride (2.0 mg/ml) | >240  |
|  Etoposide (20.0 mg/ml) | >240  |
|  Fluorouracil (50.0 mg/ml) | >240  |
|  Ifosfamide (50.0 mg/ml) | >240  |
|  Methotrexate (25.0 mg/ml) | >240  |
|  Mitomycin C (0.5 mg/ml) | >240  |
|  Mitoxantrone (2.0 mg/ml) | >240  |
|  Paclitaxel (Taxol) (6.0 mg/ml) | >240  |
|  Thiotepa (10.0 mg/ml) | 80.0 (100.0, 80.0, 80.0)  |
|  Vincristine Sulfate (1.0 mg/ml) | >240  |

Please note that Carmustine (BCNU) and Thiotepa has low permeation times of 30.0 and 80.0 minutes respectively. Warning: Do not use with Carmustine and Thiotepa

|  Opioid Drugs and Others Permeation Test | White-Blue  |
| --- | --- |
|  Fentanyl Citrate Injection (100mcg/2mL) | >240  |
|  Simulated Gastric Acid | >240  |
|  Xylazine HCl (100 mg/ml) | >240  |

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740

{6}

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 2 of 2

---

**Source:** [https://fda-staging.innolitics.com/device/K252549](https://fda-staging.innolitics.com/device/K252549)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).