AIO Breathe

K252525 · Aiomega, LLC · LRK · Nov 6, 2025 · Dental

Device Facts

Record IDK252525
Device NameAIO Breathe
ApplicantAiomega, LLC
Product CodeLRK · Dental
Decision DateNov 6, 2025
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The AIO Breathe is intended to reduce snoring and to treat mild and moderate obstructive sleep apnea in children 6-17 years of age who are diagnosed with snoring and/or mild or moderate obstructive sleep apnea.

Device Story

AIO Breathe is a custom-fabricated, traction-based mandibular repositioning device consisting of two intraoral trays (upper and lower) worn during sleep. The device is manufactured via stereolithography (SLA) 3D printing using biocompatible resins based on dentist-provided intraoral scans. It features buccal flanges and occlusal plateaus that guide the mandible into a prescribed anterior advancement while allowing for vertical and lateral jaw movement. This mechanical protrusion increases pharyngeal space and encourages tongue migration, maintaining an open airway to reduce snoring and treat obstructive sleep apnea. The device is prescribed by a dentist and used by the patient at home. Because the device is non-adjustable, dentists must periodically assess fit; if growth or dental changes occur, a new device must be ordered. The device does not fix the jaw in a single position, facilitating air exchange throughout mouth opening and closing.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance testing (biocompatibility per ISO 7405/ISO 10993, mechanical integrity, and risk analysis) and the technological identity to the predicate device.

Technological Characteristics

Custom-fit intraoral trays manufactured via SLA 3D printing using Formlabs Dental LT Clear V2 resin. Features right/left buccal protrusive flanges and occlusal mandibular plateaus. Non-sterile, prescription-only, patient-worn device. Operates via mechanical mandibular advancement. Biocompatibility testing performed per ISO 7405 and ISO 10993.

Indications for Use

Indicated for children 6-17 years of age diagnosed with snoring and/or mild or moderate obstructive sleep apnea.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION November 6, 2025 Aiomega, LLC Raghavendra Ghuge President and CEO 3187 Paluxy Drive Tyler, Texas 75701 Re: K252525 Trade/Device Name: AIO Breathe Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ Dated: August 6, 2025 Received: November 6, 2025 Dear Raghavendra Ghuge: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252525 - Raghavendra Ghuge Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252525 - Raghavendra Ghuge Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252525 | ? | | Please provide the device trade name(s). | | ? | | AIO Breathe | | | | Please provide your Indications for Use below. | | ? | | The AIO Breathe is intended to reduce snoring and to treat mild and moderate obstructive sleep apnea in children 6-17 years of age who are diagnosed with snoring and/or mild or moderate obstructive sleep apnea. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} 510(k): K252525 # 510(k) Summary Sponsor Information: AIOMEGA, LLC 3187 Paluxy Drive Tyler, TX 75701-8303 Contact Person Dr. Raghavendra Ghuge Founder, CEO raghu.ghuge@aiomd.com (903) 316-1269 Date of Summary November 5, 2025 ## Device Name and Classification Proprietary Name: AIO Breathe Common or Usual Name: Device, Jaw Repositioning; Device, Anti-Snoring Classification Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea. (21 CFR § 872.5570) Classification Product Code: LQZ, LRK Predicate Device: AIO BREATHE (K233754) ## Device Description AIO Breathe consists of two separate intraoral trays (upper, lower) that are customized to fit over all the teeth. The device is manufactured at AIOMEGA facilities using additive manufacturing with Stereolithography (SLA) 3D printing technology that builds the device from biocompatible resins. The customized trays are fabricated based on intraoral scans provided by the dentist and the dentist's prescription. AIO Breathe features right and left protrusive flanges on the buccal sides of the upper tray. These flanges engage with corresponding right and left vertical flanges featured on the buccal sides of the lower tray. This engagement repositions the jaw to reflect the dentist's prescribed anterior mandibular advancement. Additionally, mandibular plateaus, as prescribed by the dentist, featured on the right, and left occlusal cranial surfaces of lower tray, guide the mandible downward, thus opening the anterior airway. The plateaus and flanges allow vertical opening of the jaw (jaw is not fixed in a single position) and work together to maintain advancement in open and closed mouth positions. This design feature allows more room and creates traction for the tongue to migrate forward. The resulting mechanical protrusion increases the patient's pharyngeal space, improving their ability to exchange air, thereby reducing the tendency to snore and alleviating signs of obstructive sleep apnea. ## Operating Principle AIO Breathe is a traction-based mandibular repositioning device that features two separate trays worn on 1 of 3 {5} 510(k): K252525 the maxilla and mandible. The device is custom fabricated to the patient's unique dentition and is intended to maintain an open airway during sleep by advancing the mandible and tongue. Because patients in the intended age range (6-17 years) may experience changes in dentition, jaw alignment, and overall craniofacial development, it is the dentist's responsibility to periodically assess fit and function. If growth or dental changes alter the fit or effectiveness of the device, the AIO Breathe shall not be adjusted or modified. Instead, the dentist is required to obtain new dental impressions or digital scans and order a replacement device to maintain proper fit, function, and therapeutic advancement. The device permits anterior, vertical, and lateral jaw movements, and is designed to maintain an open airway intended to increase air exchange: - by anterior advancement of the mandible and tongue. - throughout opening and closing movements of the mouth. - throughout lateral movements of the mandible. # Indications for Use The AIO Breathe is intended to reduce snoring and to treat mild and moderate obstructive sleep apnea in children 6-17 years of age who are diagnosed with snoring and/or mild or moderate obstructive sleep apnea. # Biocompatibility Biocompatibility testing was completed on the materials to be used in accordance with ISO 7405 and ISO 10993. The manufacturing process and materials of the subject device are identical to the predicate device AIO Breathe in K233754. Predicate and Reference Comparison Summary | Feature | Subject Device AIO Breathe | Predicate (K233754) AIO Breathe | Reference (K234089) DNA Appliance | Comparison Summary | | --- | --- | --- | --- | --- | | Regulation | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | Identical | | Product Code | LRK, LQZ | LRK, LQZ | LRK, LQZ | Identical | | Intended Use | Treatment of snoring and mild to moderate OSA in children ages 6–17 | Treatment of snoring and mild to moderate OSA in adults | Treatment of snoring and OSA in children ages 6–17 | Intended use aligns with predicate and reference; only population differs | | Prescription Use | Yes | Yes | Yes | Identical | | Design | Two custom-fit trays with mandibular advancement | Two custom-fit trays with mandibular advancement | Similar (with mechanical adjustment) | Technologically equivalent | | Materials | Formlabs Dental LT Clear V2 | Formlabs Dental LT Clear V2 | Orthodontic acrylics | Biocompatible and functionally equivalent materials | | Operation Principle | Mandibular advancement to maintain open airway | Mandibular advancement to maintain open airway | Mandibular advancement to maintain open airway | Identical | | Manufacturing | CAD/CAM 3D printed | CAD/CAM 3D printed | Cast | Equivalent performance | | Sterility | Non-sterile | Non-sterile | Non-sterile | Identical | | Use Environment | Dentist-fitted, home use | Dentist-fitted, home use | Dentist-fitted, home use | Identical | {6} 510(k): K252525 ## Substantial Equivalence Discussion The AIO Breathe Pediatric device is substantially equivalent to the predicate AIO Breathe (K233754) with the DNA Appliance (K234089) serving as a reference device for the pediatric indication. ## Intended Use: The only difference from the predicate device is the target population (children ages 6–17). The fundamental intended use—to reduce or alleviate snoring and treat mild to moderate obstructive sleep apnea by advancing the mandible—is unchanged. ## Technological Characteristics: The subject device shares identical materials, design features, manufacturing process, and operating principles with the predicate device. These similarities ensure equivalent safety and performance. ## Reference Device Support: The DNA Appliance (K234089) provides FDA precedent for the safe and effective use of mandibular advancement devices in children ages 6–17, supporting that the change in patient population does not raise new questions of safety or effectiveness. ## Risk Mitigation and Labeling: Pediatric-specific risks (tooth movement, occlusal changes, or growth-related fit issues) are addressed through: - Custom fabrication and professional oversight - Periodic dental follow-up - Non-adjustable design and labeling to replace if fit changes ## Clinical Data: Clinical performance data are not required, as the subject device is technologically identical to the predicate and supported by a cleared reference device for the same pediatric indication. Non-clinical testing (biocompatibility, mechanical integrity, and risk analysis) adequately demonstrates safety and performance. ## Non-Clinical Performance Testing: Non-clinical performance testing in accordance with recommendations in Special Controls Guidance Document: Intraoral Devices for Snoring and /or Obstructive Sleep Apnea – Class II Special Controls Guidance Document for Industry and FDA (November 2002). ## Conclusion The AIO Breathe for pediatric is substantially equivalent to the predicate AIO Breathe (K233754). The difference in patient population does not introduce new safety or effectiveness concerns and is supported by the reference device DNA Appliance (K234089), which demonstrates established FDA acceptance of similar technology in pediatric patients. 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...