ChestView US is a radiological Computer-Assisted Detection (CADe) software device that analyzes frontal and lateral chest radiographs of patients presenting with symptoms (e.g. dyspnea, cough, pain) or suspected for findings related to regions of interest (ROIs) in the lungs, airways, mediastinum/hila and pleural space. The device uses machine learning techniques to identify and produces boxes around the ROIs. The boxes are labeled with one of the following radiographic findings: Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, and Consolidation. ChestView US is intended for use as a concurrent reading aid for radiologists and emergency medicine physicians. It does not replace the role of radiologists and emergency medicine physicians or of other diagnostic testing in the standard of care. ChestView US is for prescription use only and is indicated for adults only.
Device Story
ChestView US is a CADe software device analyzing frontal/lateral chest radiographs for nodules, consolidation, pleural space abnormalities, and mediastinum/hila abnormalities. It functions as a concurrent reading aid for radiologists and emergency medicine physicians. The device receives DICOM images from PACS or X-ray systems via intermediate nodes (gateways/APIs). An AI algorithm processes images to identify ROIs, generating DICOM overlays with bounding boxes (solid/dotted based on confidence) and text labels. If no ROI is detected, it outputs 'NO CHESTVIEW ROI'; if processing is impossible, it outputs 'OUT OF SCOPE'. The device does not alter original images. It is deployed via cloud-based processing. By highlighting suspicious regions, it assists clinicians in diagnosis, potentially improving detection accuracy and efficiency. It does not replace standard diagnostic procedures.
Clinical Evidence
Bench testing only. Standalone clinical performance study on 3,837 chest radiographs. Primary endpoint: AUC of ROC curve. Results showed non-inferiority to predicate (K241620) for AUC, sensitivity, and specificity across all ROI types (Nodule, Consolidation, Pleural space, Mediastinum/Hila) at exam/patient and ROI levels. AUC values for version 1.1 ranged from 0.899 to 0.966.
Technological Characteristics
Supervised deep learning CADe software. Cloud-based deployment. Inputs: DICOM frontal/lateral chest radiographs. Outputs: DICOM secondary capture with bounding box overlays. Architecture: Multi-stage pipeline (object detection + two false-positive classifier stages).
Indications for Use
Indicated for adults presenting with symptoms (e.g., dyspnea, cough, pain) or suspected findings related to lungs, airways, mediastinum/hila, and pleural space on frontal and lateral chest radiographs.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 9, 2026
Gleamer Sas
% Antoine Tournier
Chief Compliance Officer
47bis Rue Des Vinaigriers
PARIS, 75010
FRANCE
Re: K252486
Trade/Device Name: ChestView US (1.1)
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN
Dated: August 10, 2026
Received: August 10, 2026
Dear Antoine Tournier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252486 - Antoine Tournier
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K252486 - Antoine Tournier
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252486 | ? |
| Please provide the device trade name(s). | | ? |
| ChestView US (1.1) | | |
| Please provide your Indications for Use below. | | ? |
| ChestView US is a radiological Computer-Assisted Detection (CADe) software device that analyzes frontal and lateral chest radiographs of patients presenting with symptoms (e.g. dyspnea, cough, pain) or suspected for findings related to regions of interest (ROIs) in the lungs, airways, mediastinum/hila and pleural space. The device uses machine learning techniques to identify and produces boxes around the ROIs. The boxes are labeled with one of the following radiographic findings: Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, and Consolidation. | | |
| ChestView US is intended for use as a concurrent reading aid for radiologists and emergency medicine physicians. It does not replace the role of radiologists and emergency medicine physicians or of other diagnostic testing in the standard of care. ChestView US is for prescription use only and is indicated for adults only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
GLEAMER – ChestView US
Gleamer
Date prepared: July 26th, 2026
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for ChestView US 1.1 is provided below.
## 1. Submitter
| Submitter | GLEAMER SAS 47bis, rue des Vinaigriers 75010, Paris 10, FRANCE |
| --- | --- |
| Primary Contact Person | Antoine Tournier Chief Compliance Officer Tel: 0033 6 15 81 23 45 Email: antoine.tournier@gleamer.ai Alternate email: qara@gleamer.ai |
## 2. Device
| Trade Name | ChestView US (1.1) |
| --- | --- |
| 510(k) Number | K252486 |
| Common Name | Medical Image Analyzer |
| Classification Name | Medical Image Analyzer |
| Regulation | 21 CFR 892.2070 |
| Product Code | MYN |
| Classification | Class II |
## 3. Predicate Device
| Predicate Device | ChestView US |
| --- | --- |
| 510(k) Number | K241620 |
## 4. Device Description
ChestView US is a radiological Computer-Assisted Detection (CADe) software device intended to analyze frontal and lateral chest radiographs for suspicious regions of interest (ROIs): Nodule, Consolidation, Pleural Space Abnormality and Mediastinum/Hila Abnormality.
The nodule ROI category was developed from images with focal nonlinear opacity with a generally spherical shape situated in the pulmonary interstitium.
The consolidation ROI category was developed from images with area of increased attenuation of lung parenchyma due to the replacement of air in the alveoli.
The pleural space abnormality ROI category was developed from images with:
- Pleural Effusion that is an abnormal presence of fluid in the pleural space
{5}
510(k) Summary
GLEAMER – ChestView US
Gleamer
- Pneumothorax that is an abnormal presence of air or gas in the pleural space that separates the parietal and the visceral pleura
The mediastinum/hila abnormality ROI category was developed from images with enlargement of the mediastinum or the hilar region with a deformation of its contours.
ChestView US can be deployed on the cloud and be connected to several computing platforms and X-ray imaging platforms such as radiographic systems, or PACS. More precisely, ChestView US can be deployed in the cloud connected to a DICOM Source/Destination with a DICOM Viewer, i.e. a PACS.
After the acquisition of the radiographs of the patient and their storage in the DICOM Source, the radiographs are automatically received by ChestView US from the user's DICOM Source through intermediate DICOM node(s) (for example, a specific Gateway, or a dedicated API). The DICOM Source can be the user's image storage system (for example, the Picture Archiving and Communication System, or PACS), or other radiological equipment (for example X-ray systems).
Once received by ChestView US, the radiographs are automatically processed by the AI algorithm to identify regions of interest. Based on the processing result, ChestView US generates result files in DICOM format. These result files consist of annotated images with boxes drawn around the regions of interest on a copy of all images (as an overlay). ChestView US does not alter the original images, nor does it change the order of original images or delete any image from the DICOM Source.
Once available, the result files are sent by ChestView US to the DICOM Destination through the same intermediate DICOM node(s). Similar to the DICOM Source, the DICOM Destination can be the user's image storage system (for example, the Picture Archiving and Communication System, or PACS), or other radiological equipment (for example X-ray systems). The DICOM Source and the DICOM Destination are not necessarily identical.
The DICOM Destination can be used to visualize the result files provided by ChestView US or to transfer the results to another DICOM host for visualization. The users are then able to use them as a concurrent reading aid to provide their diagnosis.
For each exam analyzed by ChestView US, a DICOM Secondary Capture is generated.
If any ROI is detected by ChestView US, the output DICOM image includes a copy of the original images of the study and the following information:
- Above the images, a header with the text "CHESTVIEW ROI(s)" and the list of the findings detected in the image.
- Around the ROI(s), a bounding box with a solid or dotted line depending on the confidence of the algorithm and the type of ROI written above the box:
- Dotted-line Bounding Box: Identified region of interest when the confidence degree of the AI algorithm associated with the possible finding is above "high-sensitivity operating point" and below "high specificity operating point" displayed as a dotted bounding box around the area of interest.
{6}
510(k) Summary
GLEAMER – ChestView US
Gleamer
- ○ **Solid-line Bounding Box:** Identified region of interest when the confidence degree of the AI algorithm associated with the finding is above “high-specificity operating point” displayed as a solid bounding box around the area of interest.
- ● Below the images, a footer with:
- ○ The scope of ChestView US to allow the user to always have available the list of ROI type that are in the indications for use of the device and avoid any risk of confusion or misinterpretation of the types of ROI detected by ChestView US.
- ○ The total number of regions of interest identified by ChestView US on the exam (sum of solid-line and dotted-line bounding boxes)
If no ROI is detected by ChestView US, the output DICOM image includes a copy of the original images of the study and the text “NO CHESTVIEW ROI” with the scope of ChestView US to allow the user to always have available the list of ROI type that are in the indications for use of the device and avoid any risk of confusion or misinterpretation of the types of ROI detected by ChestView US. Finally, if the processing of the exam by ChestView US is not possible because it is outside the indications for use of the device or some information is missing to allow the processing, the output DICOM image includes a copy of the original images of the study and, in a header, the text “OUT OF SCOPE” and a caution message explaining the reason why no result was provided by the device.
## 5. Intended use/ Indications for Use
ChestView US is a radiological Computer-Assisted Detection (CADe) software device that analyzes frontal and lateral chest radiographs of patients presenting with symptoms (e.g. dyspnea, cough, pain) or suspected for findings related to regions of interest (ROIs) in the lungs, airways, mediastinum/hila and pleural space. The device uses machine learning techniques to identify and produces boxes around the ROIs. The boxes are labeled with one of the following radiographic findings: Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, and Consolidation.
ChestView US is intended for use as a concurrent reading aid for radiologists and emergency medicine physicians. It does not replace the role of radiologists and emergency medicine physicians or of other diagnostic testing in the standard of care. ChestView US is for prescription use only and is indicated for adults only.
## 6. Substantial equivalence
The predicate device for ChestView US in its version 1.1 is ChestView US in its version 1.0 (K241620). ChestView US has the following FDA-cleared Indications for Use:
*ChestView US is a radiological Computer-Assisted Detection (CADe) software device that analyzes frontal and lateral chest radiographs of patients presenting with symptoms (e.g. dyspnea, cough, pain) or suspected for findings related to regions of interest (ROIs) in the lungs, airways, mediastinum/hila and pleural space. The device uses machine learning techniques to identify and produces boxes around the ROIs. The boxes are*
{7}
510(k) Summary
GLEAMER – ChestView US
Gleamer
labeled with one of the following radiographic findings: Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, and Consolidation.
ChestView US is intended for use as a concurrent reading aid for radiologists and emergency medicine physicians. It does not replace the role of radiologists and emergency medicine physicians or of other diagnostic testing in the standard of care. ChestView US is for prescription use only and is indicated for adults only.
Table 1 provides a comparison of the Indications for Use and Technological Characteristics of ChestView US in its version 1.1 to the predicate ChestView US in its version 1.0.
{8}
510(k) Summary
GLEAMER – ChestView US
[LOGO]
Gleamer
| Features and Characteristics | Subject Device Gleamer ChestView US – version 1.1 | Predicate Device Gleamer ChestView US – version 1.0 |
| --- | --- | --- |
| **Regulation Information** | | |
| Classification regulation | 21 CFR 892.2070 – Medical image analyzer | Same |
| Product Code | MYN | Same |
| Regulation Description | Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. | Same |
| **Indications for Use** | | |
| Image Modality | X-ray | Same |
| Study type | Chest | Same |
| Clinical Output | When regions of interest (ROIs) are detected: Identify and mark ROIs on chest radiographs and label the box around the ROI as one of the following: Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, and Consolidation. When no ROI is detected: Mark the chest radiographs as “NO CHESTVIEW ROI” and provide the scope of the indications for use of ChestView US. | Same |
| Intended Users | Radiologists/emergency medicine physicians | Same |
{9}
510(k) Summary
GLEAMER – ChestView US
[Non-Text]
Gleamer
| Intended User Workflow | Device intended for use as a concurrent reading aid for users interpreting chest radiographs | Same |
| --- | --- | --- |
| Patient population | Adults with Chest Radiographs | Same |
| Technological Information | | |
| Machine Learning Methodology | Supervised Deep learning | Same |
| Deployment Platform | Secure cloud-based processing and delivery of chest radiographs | Same |
| Image Source | Digital X-ray | Same |
| Image Viewing | Image displayed on PACS system | Same |
Table 1: Indications for Use and Technological Comparison between ChestView US 1.1 and ChestView US 1.0
As outlined in the comparison table above, ChestView US version 1.1 shares identical technological features, indications for use, intended users, and workflow with its predicate device, ChestView US version 1.0 (K241620). Both versions analyze chest radiographs, identify and categorize regions of interest (ROIs) as Nodule, Pleural space abnormality, Mediastinum/Hila abnormality, or Consolidation, and are intended for use as concurrent reading aids to assist radiologists and emergency medicine physicians interpreting chest radiographs.
The sole difference between ChestView US version 1.0 and version 1.1 is the retraining of the underlying machine learning algorithm, together with an update of the model architecture and its initialization: the object-detection (OD) backbone was changed to a vision foundation model, and the number of False-Positive Classifiers (FPC) was reduced. This modification is a backbone initialization followed by task-specific fine-tuning performed according to standard, previously established practices. It does not alter the device's intended use, indications, clinical outputs, or technological characteristics, and does not raise new questions regarding safety and effectiveness.
Therefore, ChestView US version 1.1 is considered substantially equivalent to its predicate device, ChestView US version 1.0, maintaining the same level of safety and effectiveness.
{10}
510(k) Summary
GLEAMER – ChestView US
Gleamer
## 7. Performance data
### 7.1. Software Verification and Validation Testing
Product verification and validation testing were conducted and documented as per the requirements of the FDA guidance “Content of Premarket Submissions for Device Software Functions” for a Basic Documentation Level. Nonclinical tests include unit, integration and system levels testing for the final software version. ChestView US performed as intended and all results observed were as expected. All software requirements and risk analysis have been successfully verified and traced.
### 7.2. Bench testing
Gleamer conducted a standalone clinical performance study on 3,837 chest radiograph cases representative of the intended use population. The results of the standalone testing demonstrated that ChestView US detects ROIs with high Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve, high sensitivity and high specificity:
Table 2: AUC of the ROC Curve for ChestView US Model Predictions by ROI – Comparison of performances
| ROI Type | AUC at the exam/patient level - 95% Bootstrap CI | | |
| --- | --- | --- | --- |
| | Previous | New | Difference |
| PLEURAL SPACE ABNORMALITY | 0.959 [0.954, 0.964] | 0.966 [0.962, 0.970] | 0.007 [-0.002, 0.016] |
| CONSOLIDATION | 0.943 [0.938, 0.948] | 0.958 [0.953, 0.964] | 0.015 [0.005, 0.026] |
| NODULE | 0.906 [0.898, 0.914] | 0.921 [0.913, 0.929] | 0.015 [-0.001, 0.030] |
| MEDIASTINUM/HILA ABNORMALITY | 0.881 [0.871, 0.890] | 0.899 [0.889, 0.908] | 0.018 [-0.001, 0.037] |
Table 3: Specificity (with 95% Bootstrap CI, at the high specificity operating point) of ChestView US at the exam/patient level on the standalone dataset – Comparison of performances
| ROI Type | Specificity at the exam/patient level - 95% Bootstrap CI | | |
| --- | --- | --- | --- |
| | Previous | New | Difference |
| PLEURAL SPACE ABNORMALITY | 0.959 [0.950, 0.965] | 0.967 [0.960, 0.972] | 0.008 [-0.005, 0.022] |
| CONSOLIDATION | 0.983 [0.980, 0.987] | 0.981 [0.976, 0.986] | -0.003 [-0.011, 0.006] |
| NODULE | 0.994 [0.990, 0.997] | 0.984 [0.979, 0.988] | -0.009 [-0.018, -0.002] |
| MEDIASTINUM/HILA ABNORMALITY | 0.993 [0.991, 0.996] | 0.994 [0.991, 0.996] | 0.001 [-0.005, 0.005] |
Table 4: Sensitivity (with 95% Bootstrap CI, at the high sensitivity operating point) of ChestView US at the exam/patient level on the standalone dataset – Comparison of performances
| ROI Type | Sensitivity at the exam/patient level - 95% Bootstrap CI | | |
| --- | --- | --- | --- |
| | Previous | New | Difference |
| PLEURAL SPACE ABNORMALITY | 0.875 [0.850, 0.894] | 0.910 [0.891, 0.923] | 0.035 [-0.003, 0.073] |
| CONSOLIDATION | 0.831 [0.809, 0.864] | 0.853 [0.829, 0.874] | 0.023 [-0.035, 0.065] |
| NODULE | 0.750 [0.711, 0.772] | 0.804 [0.768, 0.824] | 0.054 [-0.004, 0.113] |
| MEDIASTINUM/HILA ABNORMALITY | 0.620 [0.588, 0.674] | 0.693 [0.645, 0.732] | 0.073 [-0.029, 0.143] |
These standalone performances of the updated version of ChestView US 1.1 were compared to the performances of the predicate device, ChestView US 1.0. Based on pre-specified acceptance criteria, the
{11}
510(k) Summary
GLEAMER – ChestView US
Gleamer
comparative standalone performance evaluation demonstrated that ChestView US 1.1 is non-inferior to ChestView US 1.0 for the primary endpoint (AUC) and the secondary endpoints (Sensitivity and Specificity at both operating points), at both the Exam/Patient and ROI levels, using pre-specified non-inferiority margins. Side-by-side performance comparisons at the Exam/Patient level are presented in Table 2, 3, and 4 for the primary endpoint and for the secondary endpoints at their respective operating points; the complete comparative results are reported in the Standalone Performance Study. Descriptive comparison of point estimates indicates that point estimates of version 1.1 are generally favorable to those of version 1.0 across the majority of evaluated regions of interest, with the limited exceptions visible in the tables above; this descriptive observation is not interpreted as a superiority claim, as superiority was not formally tested as a hypothesis.
### 7.3. Clinical Testing
No clinical studies were conducted in support of the 510(k) submission of ChestView US 1.1.
### 8. Predetermined Change Control Plan
This submission is including a Predetermined Change Control Plan (PCCP). Modifications to the algorithm of ChestView US will be made in accordance with the guidelines outlined in the “Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions”. The PCCP describes the modifications that can occur in a controlled manner ensuring the continued safety, effectiveness and performance of the device.
There are two modifications outlined in the PCCP, and accordingly, any change will only occur after training, testing and validating the algorithm after which the change will be locked. When the modification is locked this will be communicated to all users through updated labelling via the Instructions for use. Verification and Validation activities will be conducted in accordance with the PCCP to compare the performance of the modified ChestView US algorithms to both the original cleared model and the most recently deployed model to ensure the safety and effectiveness.
The allowed modifications are detailed in Table 5 below.
Table 5: Details of modifications of ChestView US allowed in the PCCP
| Modification #1: Model Re-training with New or Expanded Data | |
| --- | --- |
| Characteristics | Retrain or fine-tune the ChestView US AI model using an expanded training and/or tuning dataset of chest X-ray images limited to the data categories enumerated in the PCCP, without any change to the model architecture or pre-/post-processing pipeline. The algorithmic pipeline and output remain the same: the software processes frontal/lateral chest X-rays and outputs the likelihood and location of existing ROI categories only. No new imaging modality or hardware is introduced, and input image requirements (DICOM chest radiographs in specified orientation) remain unchanged. |
| Rationale | This modification allows Gleamer to leverage new data and clinical experience to improve the algorithm's accuracy and robustness over time. By training on larger |
{12}
510(k) Summary
GLEAMER – ChestView US
Gleamer
| | and more representative datasets (for instance, including images from additional hospitals or increasing dataset sizes), the AI can better recognize subtle findings and reduce false positives/negatives. The rationale is to continuously enhance clinical performance within the cleared indication, thereby improving radiologists' detection support without changing the device's fundamental function. New data becomes available through ongoing use and research, and retraining ensures the algorithm keeps pace with real-world variability (e.g. imaging technique differences or emerging patterns) while staying within the original intended use. |
| --- | --- |
| Performance Specifications | The modified model is deployed only if it meets all comparison non-inferiority criteria against both comparators and all primary-endpoint standalone criteria; failure of any of these is recorded as an unresolvable failure and the model is not deployed. If a standalone secondary-endpoint minimum or a subgroup criterion is not met, deployment is governed by pre-specified risk-benefit assessment: the model is not deployed if the deviation may compromise safety or effectiveness. |
| Implementation | Manual update: model retraining is performed by the manufacturer's development team; no autonomous/self-learning occurs on deployed units) Homogeneous deployment: the updated model version is rolled out universally to all users/platforms of ChestView US, so that every user operates the same algorithm version. |
| Impact Assessment | Benefits: Improved detection performance for existing ROIs; increased robustness through exposure to a broader dataset; insurance that the algorithm stays current with any shifts in imaging practices or patient populations over time Risks: Potential unintended performance degradation, overfitting, or uneven performance across subgroups Risk mitigation: Use of representative datasets, comparative performance testing against both the original cleared version and the most recently deployed version, predefined acceptance criteria, subgroup analyses, locked models post-validation |
| Modification #2: Algorithm Architecture or Parameter Updates | |
| Characteristics | Modify the AI algorithm's model architecture or tuning parameters in order to maintain or improve performance on the existing detection tasks. These changes are limited to the specific modifications enumerated in the PCCP, each mapped to a component of the current cleared multi-stage detection pipeline, which consists of an object detection (OD) stage and two False Positive Classifier (FPC) stages, whose ensemble yields the final detection confidence. An architecture modification might involve replacing/improving one of these components (such as using a more powerful object detection network, using an additional FPC with a different design, replacing the backbone of one component) to boost accuracy. Importantly, these changes will not alter the device's inputs or outputs: ChestView US will continue to take the same chest X-ray images and output the same four types of ROI findings, with the only difference being how the software's internal model processes the images to identify abnormalities. Any such architectural |
{13}
510(k) Summary
GLEAMER – ChestView US
[LOGO]
Gleamer
| | update will be designed to fit within the existing algorithm pipeline and adhere to the same input/output format and operating principles. Architectural updates under this PCCP are limited to modifications that do not change the fundamental model paradigm, output semantics, or clinical behavior of the device. The algorithm will remain a CADe detection system and will not introduce classification-only outputs, generative inference mechanisms, self-supervised inference at deployment, or changes to confidence score interpretation. |
| --- | --- |
| Rationale | Advancements in machine learning techniques occur rapidly. This modification allows ChestView US to incorporate technological improvements (such as more effective network architectures or improved training methodologies) to enhance accuracy, robustness, and generalization over time. By integrating architectural refinements, the algorithm can better recognize subtle findings, adapt to variability in imaging techniques, and reduce false positives and false negatives. For example, introducing a new deep learning architecture may improve the recognition of complex radiographic patterns or enable more refined analysis of object detection proposals. The rationale is to continuously strengthen clinical performance within the cleared indication, ensuring the device keeps pace with real-world variability while maintaining its fundamental function and rigorously controlling for safety. |
| Performance Specifications | The modified model is deployed only if it meets all comparison non-inferiority criteria against both comparators and all primary-endpoint standalone criteria; failure of any of these is recorded as an unresolvable failure and the model is not deployed. If a standalone secondary-endpoint minimum or a subgroup criterion is not met, deployment is governed by pre-specified risk-benefit assessment: the model is not deployed if the deviation may compromise safety or effectiveness. |
| Implementation | Manual update: model retraining is performed by the manufacturer's development team; no autonomous/self-learning occurs on deployed units) Homogeneous deployment: the updated model version is rolled out universally to all users/platforms of ChestView US, so that every user operates the same algorithm version. |
| Impact Assessment | Benefits: Improved detection performance for existing ROIs; increased robustness through the use of state-of-the-art AI trainings Risks: Potential unintended performance degradation, unintended changes in algorithm behavior or impact on integrated software operation Risk mitigation: Use of representative datasets, comparative performance testing against both the original cleared version and the most recently deployed version, predefined acceptance criteria, re-execution of relevant system-level V&V testing, locked models post-validation |
{14}
510(k) Summary
GLEAMER – ChestView US
Gleamer
## 9. Conclusion
ChestView US version 1.1 is substantially equivalent to its predicate device, ChestView US version 1.0 (K241620).
A comparative standalone study demonstrated that version 1.1 performs as well as the predicate device across the evaluated regions of interest, with non-inferiority to version 1.0 demonstrated for the primary endpoint (AUC) and the secondary endpoints (Sensitivity and Specificity at both operating points) at both the Exam/Patient and ROI levels, using pre-specified non-inferiority margins.
Descriptive comparison of point estimates indicates that for the primary endpoint (AUC) at the Exam/Patient level, point estimates of version 1.1 are favorable to those of version 1.0 across the evaluated regions of interest. For the secondary endpoints (Sensitivity and Specificity) at both the Exam/Patient and ROI levels, point estimates of version 1.1 are generally favorable to those of version 1.0 across the majority of evaluated regions of interest, with the limited exceptions presented for full transparency in Table 3 and 4. This descriptive observation is presented for transparency and is not interpreted as a superiority claim, as superiority was not formally tested as a hypothesis.
The modification implemented in ChestView US 1.1 does not affect the device's intended use, clinical outputs, or technological characteristics, confirming continued compliance with FDA regulatory requirements.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.