K252482 · Medcognetics, Inc. · QFM · Dec 11, 2025 · Radiology
Device Facts
Record ID
K252482
Device Name
CogNet AI-MT+
Applicant
Medcognetics, Inc.
Product Code
QFM · Radiology
Decision Date
Dec 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252482 · Dec 11, 2025
CogNet AI-MT+
Medcognetics, Inc.
Retrospective clinical mammography datasets; Biopsy-proven clinical outcomes; Longitudinal medical record follow-up (2-year negative diagnosis)
Retrospective clinical data was used to train the AI algorithm and to conduct a standalone retrospective performance validation study comparing the device's triage performance against biopsy-proven ground truth and standard-of-care benchmarks.
Standalone retrospective study of device performance; Retrospective performance validation; Follow-up/Duration: 2-year follow-up for benign cases
806 women of screening age; Sample Size: 806
Not applicable for this study
AUROC, Sensitivity, Specificity
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Breast cancer triage
Deep Learning
AUROC >= 0.95
AUROC = 0.9548
Development Source 1: 1292 patients; Development Source 2: 498 patients
—
Retrospective study: 806 women of screening age (403 benign, 403 malignant)
—
Indications for Use
The MedCognetics CogNet AI-MT+ software is a passive notification for prioritization-only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize viewing patients with suspicious findings in the medical care environment. CogNet AI-MT+ utilizes an artificial intelligence algorithm to analyze DBT screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. CogNet AI-MT+ produces an exam level output to a PACS/Workstation for flagging the suspicious study and allows for worklist prioritization. MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The CogNet AI-MT+ device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis. The CogNet AI-MT+ device is intended for use with DBT mammography exams acquired using validated DBT equipment only.
Device Story
CogNet AI-MT+ is a cloud-based SaMD that analyzes digital DBT screening mammograms to triage studies for radiologists. Input consists of DICOM-formatted DBT studies from Hologic Selenia Dimensions systems. The device uses a deep learning algorithm to extract features and identify suspicious lesions. Output is a DICOM secondary capture image and textual flags (Suspicious/Processed) sent to a PACS/workstation. The software operates in parallel to standard clinical workflows; it does not remove studies from worklists or provide diagnostic confirmation. MQSA-qualified physicians use the AI-generated flags to prioritize their reading list. The device aims to improve clinical efficiency by highlighting potentially malignant cases. Data is anonymized and encrypted in transit and at rest. The system is licensed and cloud-hosted, requiring no local installation.
Clinical Evidence
Retrospective standalone study of 806 women (403 malignant, 403 benign). Primary endpoint: AUROC ≥ 0.95. Results: AUROC 0.9548 (95% CI: 0.9364-0.9699), Sensitivity 0.8809 (95% CI: 0.8511-0.9032), Specificity 0.9156 (95% CI: 0.8933-0.9380). Performance compared favorably to predicate device and BCSC standard of care benchmarks.
Technological Characteristics
Cloud-based SaMD; utilizes deep learning algorithm for image analysis. Inputs: DICOM DBT mammography images. Outputs: DICOM secondary capture and metadata tags. Connectivity: Networked via DICOM transfer. Standards: IEC 62304 (Software Lifecycle), DICOM PS3.1. Validated for use with Hologic Selenia Dimensions equipment.
Indications for Use
Indicated for female patients aged 22 and older undergoing DBT screening mammography. Excludes patients with breast implants and post-surgical studies. Used by MQSA qualified interpreting physicians for triage and worklist prioritization of studies suggestive of suspicious findings.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 11, 2025
Medcognetics, Inc.
John Jenkins
Chief Quality Officer
17217 Waterview Parkway
Suite 1.202E
Dallas, Texas 75252 USA
Re: K252482
Trade/Device Name: CogNet AI-MT+
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QFM
Dated: November 10, 2025
Received: November 10, 2025
Dear John Jenkins:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252482 - John Jenkins
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252482 - John Jenkins
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**YANNA S. KANG -S**
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
1/1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252482 | ? |
| Please provide the device trade name(s). | | ? |
| CogNet AI-MT+ | | |
| Please provide your Indications for Use below. | | ? |
| The MedCognetics CogNet AI-MT+ software is a passive notification for prioritization-only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize viewing patients with suspicious findings in the medical care environment. CogNet AI-MT+ utilizes an artificial intelligence algorithm to analyze DBT screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level.
CogNet AI-MT+ produces an exam level output to a PACS/Workstation for flagging the suspicious study and allows for worklist prioritization. MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The CogNet AI-MT+ device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis.
The CogNet AI-MT+ device is intended for use with DBT mammography exams acquired using validated DBT equipment only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(K) Summary: K252482
# I. SUBMITTER
Medcognetics, Inc
17217 Waterview Parkway
Suite 1.202 E
Dallas, Texas 75252 USA
Phone: (214) 264-5612
Contact Person: John Jenkins
Date Prepared: June 10, 2025
# II. DEVICE
Name of Device: CogNet AI-MT+
Common or Usual Name: CogNet AI-MT+
Classification Name: 21 CFR 892.2080 - Radiological computer aided triage and notification software
Regulatory Class: II
Product Code: QFM
# III. PREDICATE DEVICE
CogNet QmTRIAGE (Renamed to CogNet AI-MT), K220080
This predicate has not been subject to a design-related recall.
# IV. DEVICE DESCRIPTION
The MedCognetics CogNet AI-MT+ is a non-invasive computer-assisted triage and notification software as a medical device (SaMD) that analyzes DBT screening mammograms using a machine learning algorithm and notifies a PACS/workstation of the presence of findings suspicious of cancer in a study. The passive-notification enables interpreting physicians to prioritize their worklist and assists them in viewing prioritized studies using the standard PACS or workstation viewing software. The device aim is to aid in the prioritization and triage of radiological medical images only. It is a software tool for MQSA interpreting physicians reading mammograms and does not replace complete evaluation according to the standard of care.
The software modules that compose the CogNet AI-MT+ Deep Learning software are:
The Qualification Module - The requirement for acceptance into the CogNet AI-MT+ analysis is a completed Mammogram DICOM image. In the Qualification Module, the image arrives from the Mammogram modality and is "read" to
Page 1 of 9
{5}
510(K) Summary: K252482
determine if this qualification applies.
The Mammogram Pre-Processing module – The DBT pixel brightness, image size, and shape is adjusted for consistency in this module. After the DICOM image has been qualified, the Pre-Processing module assures that the images are from a mammogram device and then validates that the DICOM is properly formed and consists of "For Presentation" image pixel data.
Mammogram Learning Module – This module accepts the normalized image data from the pre-processing module and uses Deep Learning techniques to extract features to determine if any lesions suspicious for cancer exist in the mammogram study
Failures in any of the above modules will generate error messages that are recorded in an accessible log file and, if user specific issues are encountered, sent to the user in a secondary capture report.
CogNet AI-MT+ has no viewing capability, but the results data are sent via a secure network function to the PACS/workstation, and the PACS/workstation "reads" the necessary DICOM tags and matches it with the original mammogram study images as a normal function of a PACS or Workstation.
When the study data is fed into the configured reading worklist, the results are merged as part of the mammogram study. This process allows an AI Result to be ready for prioritization of the study prior to the interpreting physician's review.
A reading worklist is a listing of available studies for reading and diagnosis. The worklist is populated by the parsing of a DICOM file of a completed mammogram study, using the demographic and study fields to fill in the designated columns of the worklist. The columns are sortable by study, based on the column headings. CogNet AI-MT+ provides an API for adding an AI Results column with 0 to 1 response per study. If an analysis was not performed on that study, the AI Results indication is 0. If an analysis was performed on that study, then the AI Results column indicates either Suspicious (red diamond icon) or Processed (blue circle icon). The AI Results column may be sorted by the interpreting physicians by clicking an up or down arrow next to the column heading. This sort would allow the studies that contain suspicious findings to be brought to the top of the viewing list.
Device Inputs
Page 2 of 9
{6}
510(K) Summary: K252482
Inputs to the CogNet AI-MT+ analysis software are digital copies of completed mammogram studies in DICOM format. Current or pre-existing DBT studies are uploaded as DICOM data into CogNet AI-MT+ via a DICOM network transfer from the facility's mammography imaging system, the facilities PACS, or DICOM router, depending on network configuration.
The criteria for the CogNet AI-MT+ software excludes patients with breast implants and post-surgical studies. Only tomographic studies are included in this submission as CogNet AI-MT+ is designed for analyzing DBT mammogram studies.
The imaging equipment manufacturer's model that has been validated with CogNet AI-MT+ is Hologic's Selenia Dimensions. The Hologic equipment was validated in tests executed in the CogNet AI-MT+ Design Validation (VAL-COG-060) and results discussed in the Validation Report.
## Device Outputs
The outputs of CogNet AI-MT+ software is a DICOM secondary capture image that is not intended for diagnostic image review, and textual data of Suspicious or Processed results, as well as company information in DICOM private tags. These are the data that notifies the reading interpreting physicians of suspicious results.
## Software Installation and Data Workflow
The CogNet AI-MT+ software is cloud based so installation is not required. The use of the product is by license only so the login for the user and User Instruction manual will be sent directly to the licensed individuals that will be accessing or responsible for the CogNet AI-MT+ device. MedCognetics quality and technical staff will coordinate and facilitate the initial usage.
## Cybersecurity
MedCognetics is attentive to cybersecurity issues in medical devices. CogNet AI-MT+ is HIPAA compliant and assures that Personal Health Information is protected by promoting anonymization of data prior to analysis. This is accomplished by requiring de-identification as part of the data transfer to the CogNet AI-MT+ algorithm. DICOM data retains the necessary DICOM tags using these to merge the secondary capture image containing the CogNet analysis results into the original mammogram study for final viewing by a MQSA interpreting physician. DICOM data in network and in transfer to the CogNet AI-MT+ algorithm is encrypted in transit and at rest. User access is strictly password protected. The transferred data is subject to existing firewall solutions, auditing, and all interactions are logged to facilitate review of potential issues.
## User Instruction
{7}
510(K) Summary: K252482
A User instruction manual is provided for the software. The proposed labeling for the proposed device is included in the submission.
## Standards Applied
The standards applied for the development of the system are listed below in Table 1.
Table 1: Standards Applied
| Standard | Title of Standard |
| --- | --- |
| IEC 62304 | Software Development Life Cycle |
| 21 CFR Part 820 | FDA Regulatory Requirements and Design Controls |
| DICOM PS3.1 | Digital Image and Communications in Medicine |
## V. INDICATIONS FOR USE
The MedCognetics CogNet AI-MT+ software is a passive notification for prioritization- only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize patients with suspicious findings in a medical care environment. CogNet AI- MT+ utilizes an artificial intelligence algorithm to analyze DBT screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. CogNet AI-MT+ produces an exam level output to a PACS/Workstation for flagging the suspicious study and allows for worklist prioritization.
MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The CogNet AI-MT+ device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis.
The CogNet AI-MT+ device is intended for use with DBT mammography exams acquired using validated equipment systems, only.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The predicate device is CogNet AI-MT (formerly QmTRIAGE) for analysis of 2D FFDM mammography studies. CogNet AI-MT+, the proposed device, is designed to process only Digital Breast Tomosynthesis (DBT) Mammography images. These studies contain multiple "slices" of images so they must be evaluated individually, taking more time but offering a potentially better view of any lesions present in the breast. The difference between these two image modalities is in the separation and handling of the multi-frame images, but beyond that, the AI processing is virtually the same. On the next two pages is a list of technological characteristics and comparison between the predicate and the proposed device:
{8}
| | New Device | Predicate Device (1) | Status |
| --- | --- | --- | --- |
| Trade Name | CogNet AI-MT+ | CogNet QmTRIAGE | |
| 510(k) Submitter [Number] | MedCognetics [K252482] | MedCognetics [K220080] | |
| Indication for Use | The MedCognetics CogNet AI-MT+ software is a passive notification for prioritization-only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize patients with suspicious findings in a medical care environment. CogNet AI-MT+ utilizes an artificial intelligence algorithm to analyze DBT screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. CogNet AI-MT+ produces an exam level output to a PACS/Workstation for flagging the suspicious study and allows for worklist prioritization. MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The CogNet AI-MT+ device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis. The CogNet AI-MT+ device is intended for use with DBT mammography exams acquired using validated equipment systems, only. | The MedCognetics (CogNet) QmTRIAGE software is a passive notification for prioritization-only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize patients with suspicious findings in the medical care environment. QmTRIAGE utilizes an artificial intelligence algorithm to analyze 2D FFDM screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. QmTRIAGE produces an exam level output to a PACS/Workstation for flagging suspicious study and allows for worklist prioritization. MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The QmTRIAGE device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis. The QmTRIAGE device is intended for use with 2D FFDM mammography exams acquired using validated FFDM systems, only. | Similar |
| Product Code(s) | QFM | QFM | Same |
| Regulation(s) | 892.2080 | 892.2080 | Same |
| Notification Only | Yes | Yes | Same |
| Parallel Workflow | Yes | Yes | Same |
| User | MQSA Interpreting physician | MQSA Interpreting physician | Same |
| Alert to finding | Yes. Passive notification flagged for review | Yes. Passive notification flagged for review | Same |
| Independent of SoC workflow | Yes. No cases are removed from | Yes. No cases are removed from worklist | Same |
{9}
510(K) Summary: K252482
| | New Device | Predicate Device (1) | Status |
| --- | --- | --- | --- |
| Trade Name | CogNet AI-MT+ | CogNet QmTRIAGE | |
| 510(k) Submitter [Number] | MedCognetics [K252482] | MedCognetics [K220080] | |
| | worklist | | |
| Modality | DBT screening mammograms | FFDM screening mammograms | Different |
| Equipment Manufacturer | Hologic | Hologic | Same |
| Body Part | Breast | Breast | Same |
| AI algorithm | Yes | Yes | Same |
| Limited to analysis of imaging data | Yes | Yes | Same |
| Inclusion Criteria | • Standard DBT screening mammograms
• Biopsy proven cancer studies (soft tissues and microcalcifications)
• BIRADS 1 and 2 normal/benign cases with 2-year follow-up of a negative diagnosis
• Female patients 22 and older
• Bilateral Studies with 4 standard views (LCC, LMLO, RCC, RMLO) | • Standard 2D FFDM screening mammograms
• Biopsy proven cancer studies (soft tissues and microcalcifications)
• BIRADS 1 and 2 normal/benign cases with 2-year follow-up of a negative diagnosis
• Female patients 22 and older
• Bilateral Studies with 4 standard views (LCC, LMLO, RCC, RMLO) | Different |
| Aids in prompt identification of cases with indicated findings | Yes | Yes | Same |
| Results Preview | Secondary Capture stored with original DICOM study and may be viewed with the study. The device operates in parallel with the standard of care, which remains the default option. | Encapsulated PDF stored with original DICOM study and may be downloaded and viewed as a PDF. The device operates in parallel with the standard of care, which remains the default option. | Different |
| Deployment | Cloud based | Cloud based | Same |
| Where results are received | PACS / Workstation | PACS / Workstation | Same |
| Patient Contact | No direct or indirect contact | No direct or indirect contact | Same |
{10}
510(K) Summary: K252482
| | New Device | Predicate Device (1) | Status |
| --- | --- | --- | --- |
| Trade Name | CogNet AI-MT+ | CogNet QmTRIAGE | |
| 510(k) Submitter [Number] | MedCognetics [K252482] | MedCognetics [K220080] | |
| Cleaning | Not applicable | Not applicable | Same |
| Sterilization | Not applicable | Not applicable | Same |
# VII. PERFORMANCE DATA
The algorithm developed by MedCognetics is trained with samples both suspicious of cancer and not suspicious of cancer to build a model for predicting mammograms as "Suspicious" or Processed. During training, a batch of training examples are passed through the MLN which produces a batch of predictions (forward pass). The difference between the predictions and the ground truth labels is then measured using a loss function. Reverse-mode automatic differentiation is applied to determine how each model parameter should be updated to reduce the loss (backward pass). The parameter gradients produced during this process are then used by the optimizer to update the model parameters. This process is repeated for a fixed number of iterations, over which time the model's AUROC on both the training and development datasets are monitored to ensure that the model is not overfitting. Mammographic modalities utilized during training include FFDM, SFM, synthetic view, and DBT. Although the trained model will only process the DBT modality, we find that a diverse training set of modalities is beneficial to improving generalization.
| Region | Total Patients | Positives | Negatives |
| --- | --- | --- | --- |
| Europe | 2,372 | 40 | 2,332 |
| South Asia | 126 | 125 | 1 |
| South America | 4,066 | 577 | 3,489 |
| South Asia | 1,024 | 374 | 650 |
| Europe | 23,036 | 8,651 | 14,385 |
| Africa | 102 | 6 | 96 |
| United States | 1,566 | 723 | 843 |
| Totals | 32,292(≈129,168 images) | 10,496(≈41,984 images) | 21,796(≈87,184 images) |
Training Data Sources
{11}
510(K) Summary: K252482
A standalone retrospective study of device performance was conducted for the proposed device. The primary objective of the study was to assess the performance of the MedCognetics CogNet AI-MT+ algorithm for triage of DBT digital mammograms and validate the Area Under Receiver Operating Characteristics (AUROC) as $\geq 0.95$ with a confidence interval of $\pm 0.02$ at a confidence level of $95\%$ . The secondary objective is to assess the sensitivity and specificity performance of the algorithm that will be comparable to the standard of care as reported in the Breast Cancer Surveillance Consortium (BCSC) study, which is 0.869 sensitivity and 0.889 specificity. The same parameters were used in the previous study for the predicate device.
| Source | Data Pool (Patients) |
| --- | --- |
| Development Source 1 | 1292 |
| Development Source 2 | 498 |
To ensure generalizability for all data sources, mammography images used for testing were obtained from a site or facility that was not used to source the training or development data. The training, development, and test set data sources were all separate and independent of each other. The study consisted of 806 women of screening age, grouped according to BI-RADS distribution. The study consisted of 403 cases with an assigned label of "benign" based on a negative diagnosis (BI-RADS 1 or 2 assessment) throughout 2-years of follow-up and 403 studies with a label of "malignant" based on a positive biopsy result. The predicate device's study had similar inclusions and exclusions as the proposed device.
| Biopsy Outcome | Biopsy Proven Benign | Malignant | Screening Benign |
| --- | --- | --- | --- |
| Samples | 21 | 403 | 382 |
| Positive | 0 | 403 | 0 |
| Negative | 21 | 0 | 382 |
The performance of the CogNet AI-MT+ algorithm was analyzed across various subgroups of the test dataset. All stratified analysis was conducted using only Hologic Selenia Dimensions machines. The data was sorted into subgroups, including cohorts of Age; BIRADS; Breast Density; Lesion type; Pathology, Benign biopsy outcome, and Lesion size. The performance by subgroup was good but typically, dense breasts and small lesions are the most difficult to analyze and that is reflected in AUC being slightly less than $95\%$ . Other than that, the subgroups all fell in the target range AUC of $95\%$ . Similar results in the same subgroup breakdown from the predicate device.
Overall, CogNet AI-MT+ achieved an AUROC = 0.9548 (95% CI: 0.9364 - 0.9699)
Sensitivity $= 0.8809$ (95% CI: 0.8511 - 0.9032) and Specificity $= 0.9156$
(95% CI: 0.8933 - 0.9380) was achieved, meeting the stated criteria. This compares
{12}
510(K) Summary: K252482
favorably with the predicate device performance of (AUROC) of 0.9569 with 95% CI: 0.9364-0.9738.
The proposed device and the predicate device's performance met or exceeded the primary objective of ≥95 AUC and the secondary objective of Sensitivity ≥ 0.89 and Specificity ≥ .889.
## VIII. CONCLUSION
The regulatory, usage, and process similarities are extensive between the proposed device and the predicate device, and the technical characteristics of the two devices are similar, as discussed in this submission. Any differences have been addressed and demonstrated to have no impact on the equivalence, safety, or effectiveness. In conclusion, as demonstrated through the supporting evidence contained within this submission, CogNet AI-MT+ is substantially equivalent to the identified predicate device and does not raise new or different questions of safety or effectiveness.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.