Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable Implantable Port; PowerPort™ duo M.R.I. ™ Implantable Port
K252478 · Bard Access Systems, Inc. · LJT · Sep 5, 2025 · General Hospital
Device Facts
Record ID
K252478
Device Name
Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable Implantable Port; PowerPort™ duo M.R.I. ™ Implantable Port
Applicant
Bard Access Systems, Inc.
Product Code
LJT · General Hospital
Decision Date
Sep 5, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bard Power-Injectable Implantable Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy), I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with the PowerLoc™ Safety Infusion Set, the Bard Power-Injectable Implantable Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.
Device Story
Implantable vascular access device; consists of injection port with self-sealing silicone septum and radiopaque catheter. Used for long-term repeated vascular access; infusion of medications, fluids, blood products; blood sampling. Power injection of contrast media enabled via PowerLoc™ Safety Infusion Set. Operated by clinicians in clinical settings; percutaneous needle insertion into port septum. Output is direct vascular access for therapy delivery or sampling. Benefits include reduced need for repeated venipuncture; reliable long-term access for oncology and chronic therapy patients.
Clinical Evidence
Bench testing only. Verification included port subassembly air burst and tensile strength testing. Functional testing with chemical conditioning (simulating oncology drugs, TPN, and locking solutions) confirmed device performance. No clinical data presented.
Technological Characteristics
Materials: Titanium, Delrin plastic, silicone septum, ChronoFlex™ polyurethane catheter. Principle: Percutaneous access via non-coring needle to self-sealing septum. Energy: None (mechanical). Form factor: Triangular port body, single or dual lumen. Connectivity: None. Sterilization: Ethylene Oxide. Shelf life: 2 years.
Indications for Use
Indicated for patients requiring repeated vascular access for infusion of medications (including chemotherapy), IV fluids, parenteral nutrition, blood products, and blood withdrawal. When used with PowerLoc™ Safety Infusion Set, indicated for power injection of contrast media at max 5 ml/s.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PowerPort™ Slim Implantable Port; PowerPort™ M.R.I.™ isp Implantable Port (K251253)
PowerPort™ Duo M.R.I.™ Implanted Port with 9.5 Fr. Dual Lumen ChronoFlex™ Polyurethane Catheter (K090512)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 5, 2025
Bard Access Systems, Inc.
Kaitlyn Nielsen
Regulatory Affairs Specialist
605 North 5600 West
Salt Lake City, Utah 84116
Re: K252478
Trade/Device Name: Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable Implantable Port; PowerPort™ duo M.R.I.™ Implantable Port
Regulation Number: 21 CFR 880.5965
Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port And Catheter
Regulatory Class: Class II
Product Code: LJT
Dated: August 6, 2025
Received: August 7, 2025
Dear Kaitlyn Nielsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252478 - Kaitlyn Nielsen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252478 - Kaitlyn Nielsen
Page 3
Sincerely,
David Wolloscheck -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252478 | |
| Device Name Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable Implantable Port; PowerPort™ duo M.R.I.™ Implantable Port | |
| Indications for Use (Describe) The Bard Power-Injectable Implantable Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy), I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with the PowerLoc™ Safety Infusion Set, the Bard Power-Injectable Implantable Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 1 of 15
K252478 - 510(k) SUMMARY
## Submitter Information
Applicant: Bard Access Systems, Inc
605 North 5600 West
Salt Lake City, UT 84116
Phone: 602-830-5681
Contact: Kaitlyn Nielsen, Regulatory Affairs Specialist
Date: August 25, 2025
## Subject Device Name
Device Trade Name: Vaccess™ CT Low-Profile Power-Injectable Implantable Port
Vaccess™ CT Power-Injectable Implantable Port
PowerPort™ duo M.R.I.™ Implantable Port
Common Name: Subcutaneous, implanted, intravascular Port and Catheter
Classification: Class II
Regulation: 21 CFR 880.5965, Subcutaneous, implanted, intravascular infusion port and catheter
Review Panel: General Hospital
Product Code: LJT
## Predicate Devices
PowerPort™ Slim Implantable Port; PowerPort™ M.R.I.™ isp Implantable Port (K251253, cleared 06/18/2025)
PowerPort™ Duo M.R.I.™ Implanted Port with 9.5 Fr. Dual Lumen ChronoFlex™ Polyurethane Catheter (K090512, cleared 03/27/2009)
## Reference Device
PowerPort™ ClearVUE™ isp Implantable Ports (K242328, cleared 10/31/2024)
{5}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 2 of 15
# Device Description
The Bard Power-Injectable Implantable Port is an implantable access device designed to provide repeated access to the vascular system. Port access is performed by percutaneous needle insertion using a non-coring needle. Power injection is performed using a PowerLoc™ Safety Infusion Set only. The Bard Power-Injectable Implantable Port consists of two primary components: an injection port with a self-sealing silicone septum and a radiopaque catheter. Single lumen Bard Power-Injectable Implantable Ports can be identified subcutaneously by feeling the top of the septum, which may include three palpation bumps arranged in a triangle, and by palpating the sides of the port, which is also triangular. Dual lumen Bard Power-Injectable Implantable Ports can be identified subcutaneously by feeling the top of each septum; each septum may feature three palpation bumps arranged in a triangle.
# Indications for Use of Device
The Bard Power-Injectable Implantable Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy), I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the Bard Power-Injectable Implantable Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.
# Comparison to Predicate Device and Reference Device
The technological characteristics of the subject Bard Power-Injectable Implantable Port devices share many technological characteristics with the predicate devices. Among these shared technological characteristics are:
- Same intended use
- Same indications for use
- Same patient population
- Same principle of operation
- Same insertion site
- Same catheter tip termination location
- Same duration of use
- Same sterility assurance level and method of sterilization
{6}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 3 of 15
- Same shelf life
- Same catheter locking solution options (Vaccess™ CT Implantable Ports)
Compared to the reference device, the subject devices have the following similarities:
- Device materials
- Additional catheter locking solution options (PowerPort™ duo Implantable Port)
Compared to the predicate devices, the subject devices have the following differences:
- Device dimensions
{7}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Table 1. Subject and Predicate Device Comparison – Vaccess™ CT Low-Profile Power-Injectable Implantable Port and Vaccess™ CT Power-Injectable Implantable Port
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ Slim Implantable Port (K251253) | Vaccess™ CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ M.R.I.™ isp Implantable Port (K251253) | Vaccess™ CT Power-Injectable Implantable Port | |
| Note | 1. Bold Font indicates a difference between the subject device and predicate devices
2. Plain Type indicates that the attribute of the subject device is the same as that of the predicate device(s). | | |
| Device Identification | | | |
| Manufacturer | Bard Access Systems, Inc. | Bard Access Systems, Inc. | Same as Predicate |
| Device Classification Name | Port & Catheter, Implanted, Subcutaneous, Intravascular | Port & Catheter, Implanted, Subcutaneous, Intravascular | Same as Predicate |
| Regulation Number | 21 CFR §880.5965 | 21 CFR §880.5965 | Same as Predicate |
| FDA Product Code | LJT | LJT | Same as Predicate |
| Device Use | | | |
| Intended Use | The PowerPort™ Implanted Port is a totally implantable vascular access device designed to provide long-term, repeated access to the vascular system. | The Vaccess™ CT Implanted Port is a totally implantable vascular access device designed to provide long-term, repeated access to the vascular system. | Same as Predicate |
| Indications for Use | The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy), I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. | The Vaccess™ CT Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy), I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the Vaccess™ CT device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. | Same as Predicate |
| Patient Population | Patients requiring repeated access to the vascular system | Patients requiring repeated access to the vascular system | Same as Predicate |
{8}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 5 of 15
| Attribute | Predicate Device | | Subject Device | | Discussion |
| --- | --- | --- | --- | --- | --- |
| | PowerPort™ | Slim Implantable Port (K251253) | Vaccess™ | CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ | M.R.I.™ isp Implantable Port (K251253) | Vaccess™ | CT Power-Injectable Implantable Port | |
| Duration of Use | Long term (>30 days) | | Long term (>30 days) | | Same as Predicate |
| Insertion Site | Most commonly on upper chest | | Most commonly on upper chest | | Same as Predicate |
| Visualization Techniques | Fluoroscopy | | Fluoroscopy | | Same as Predicate |
| Principle of Operation | The device’s primary components consist of a triangular injection port with self-sealing silicone septum and a radiopaque catheter. A simple sliding lock collar secures the catheter to the port’s stem.
Port access is performed by percutaneous needle insertion using a non-coring needle. | | The device’s primary components consist of a triangular injection port with self-sealing silicone septum and a radiopaque catheter. A simple sliding lock collar secures the catheter to the port’s stem.
Port access is performed by percutaneous needle insertion using a non-coring needle. | | Same as Predicate |
| Catheter Tip Termination Location | Central venous system - lower 1/3 of superior vena cava preferred | | Central venous system - lower 1/3 of superior vena cava preferred | | Same as Predicate |
| Design Characteristics | | | | | |
| Device Measurements | PowerPort™ Slim Implantable Port:
Port Body Dimensions:
Height: 9.8 mm
Base Width: 21.2 mm x 25.5 mm
Reservoir Volume: 0.5 mL
PowerPort™ M.R.I.™ isp Implantable Port:
Port Body:
Height: 11.7 mm
Width: 23.7 mm x 26.6 mm | | Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
Port Body Dimensions:
Height: 9.8 mm
Width: 21.29 x 24.64
Reservoir Volume: 0.5 mL
Vaccess™ CT Power-Injectable Implantable Port:
Port Body:
Height: 11.7mm
Width: 23.6 mm x 26.7 mm | | Different than Predicate
Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
The subject devices use the same material (titanium) as the predicate device, so all material evaluations for the predicate device are applicable to the subject device.
Vaccess™ CT Power-Injectable Implantable Port:
The subject devices use the same material (plastic) as the predicate device, so all material evaluations for the |
{9}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 6 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ Slim Implantable Port (K251253) | Vaccess™ CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ M.R.I.™ isp Implantable Port (K251253) | Vaccess™ CT Power-Injectable Implantable Port | |
| | Reservoir Volume: 0.6 mL | Reservoir Volume: 0.6 mL | predicate device are applicable to the subject device.
Sterilization and microbiology studies for both subject devices indicated that no new risks were introduced. The new dimensional specifications for the subject devices were qualified and fulfilled performance criteria. No changes were made to the intended use, indications, contraindications, warnings, precautions, or packaging of the final device. |
| | PowerPort™ Slim Implantable Port:
Catheter Dimensions:
6 Fr x 60 cm ChronoFlex, 1.3 mm ID
8 Fr x 45 cm ChronoFlex, 1.6 mm ID | Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
Catheter Dimensions:
6 Fr x 60 cm ChronoFlex, 1.3 mm ID
8 Fr x 45 cm ChronoFlex, 1.6 mm ID | Same as Predicate |
| | PowerPort™ M.R.I. ™ isp Implantable Port:
Catheter Dimensions:
6 Fr x 45 cm ChronoFlex, 1.3 mm ID
8 Fr x 45 cm ChronoFlex, 1.6 mm ID
9.6 Fr x 45cm Silicone, 1.6 mm ID | Vaccess™ CT Power-Injectable Implantable Port:
Catheter Dimensions:
6 Fr x 45 cm ChronoFlex, 1.3 mm ID
8 Fr x 45 cm ChronoFlex, 1.6 mm ID
9.6 Fr x 45 cm Silicone, 1.6 mm ID | |
{10}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 7 of 15
| Attribute | Predicate Device | | Subject Device | | Discussion |
| --- | --- | --- | --- | --- | --- |
| | PowerPort™ | Slim Implantable Port (K251253) | Vaccess™ | CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ | M.R.I.™ isp Implantable Port (K251253) | Vaccess™ | CT Power-Injectable Implantable Port | |
| Device Description and Materials | PowerPort™ Slim Implantable Port:
Port Body:
- Port Base: Titanium with engraved CT identification
- Port Top: Titanium
- One piece silicone septum with 3 raised palpation bumps
- 3 suture holes with or without 3 suture plugs | Vaccess™ ST Low-Profile Power-Injectable Implantable Port:
Port Body:
- Port Base: Titanium with engraved CT identification
- Port Top: Titanium
- One piece silicone septum with **no** raised palpation bumps
- 3 suture holes with or without 3 suture plugs | Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
Port Body:
- Triangular Delrin plastic port (**with no colorant**)
- radiopaque identifying feature displaying power injection symbol and the letters "C T"
- One piece silicone septum with **no** raised palpation bumps | | Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
Same as Reference
The reference device does not have raised palpation bumps. The removal of palpation bumps does not introduce new hazards or failure modes. Palpation bumps are one of several identification methods for a power injectable port.
Vaccess™ CT Power-Injectable Implantable Port:
Same as Reference
The Triangular Delrin body (with no colorant) has been previously used for the reference device. Since there was no substantial change in the material and design, pre-existing biocompatibility and sterilization evaluations were applicable to this change. There is no change in sterilization or biocompatibility. The reference device does not have raised palpation bumps. The removal of |
| | PowerPort™ M.R.I.™ isp Implantable Port:
Port Body:
- Purple Triangular Delrin plastic port
- radiopaque identifying feature displaying power injection symbol and the letters "C T"
- One piece silicone septum with 3 raised palpation bumps | | | | |
{11}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 8 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ Slim Implantable Port (K251253) | Vaccess™ CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ M.R.I.™ isp Implantable Port (K251253) | Vaccess™ CT Power-Injectable Implantable Port | |
| | | | palpation bumps does not introduce new hazards or failure modes. Palpation bumps are one of several identification methods for a power injectable port. |
| | PowerPort™ Slim Implantable Port:
Catheter:
- White ChronoFlex polyurethane
- open-ended
- with depth markings
- single lumen
- attachable | Vaccess™ CT Low-Profile Power-Injectable Implantable Port:
Catheter:
- White ChronoFlex polyurethane
- open-ended
- with depth markings
- single lumen
- attachable | Same as Predicate |
| | PowerPort™ M.R.I. ™ isp Implantable Port:
Catheter:
- White ChronoFlex polyurethane
- open-ended
- with depth markings
- single lumen
- attachable | Vaccess™ CT Power-Injectable Implantable Port:
Catheter:
- White ChronoFlex polyurethane
- open-ended
- with depth markings
- single lumen
- attachable | |
| | Silicone Catheter
- open-ended
- with depth markings
- single lumen
- attachable | Silicone Catheter
- open-ended
- with depth markings
- single lumen
- attachable | |
| Locking Solution | Cathlock:
Polycarbonate | Cathlock:
Polycarbonate | Same as Predicate |
| | Open-ended catheters: | Open-ended catheters: | Same as Predicate |
{12}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 9 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ Slim Implantable Port (K251253) | Vaccess™ CT Low-Profile Power-Injectable Implantable Port | |
| | PowerPort™ M.R.I.™ isp Implantable Port (K251253) | Vaccess™ CT Power-Injectable Implantable Port | |
| | To help prevent clot formation and catheter blockage, each lumen of the implanted ports with open-ended catheters should be flushed with sterile normal saline then filled with sterile heparinized saline, sterile normal saline, or other approved locking solution per institutional protocol after each use. If the port remains unused for long periods of time, the lock should be changed at least once every 28 days. | To help prevent clot formation and catheter blockage, each lumen of the implanted ports with open-ended catheters should be flushed with sterile normal saline then filled with sterile heparinized saline, sterile normal saline, or other approved locking solution per institutional protocol after each use. If the port remains unused for long periods of time, the lock should be changed at least once every 28 days. | |
| | Warning: Alcohol should not be used to soak or declot polyurethane catheters because alcohol is known to degrade polyurethane catheters over time with repeated or prolonged exposure. | Warning: Alcohol should not be used to soak or declot polyurethane catheters because alcohol is known to degrade polyurethane catheters over time with repeated or prolonged exposure. | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same as Predicate |
| Shelf Life | 2 years | 2 years | Same as Predicate |
| Packaging Configuration | Triple tray packaging; two sterile barriers in the form of nested, sealed trays | Triple tray packaging; two sterile barriers in the form of nested, sealed trays | Same as Predicate |
{13}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Table 2. Subject and Predicate Device Comparison – PowerPort™ duo M.R.I.™ Implantable Port
| Attribute | Predicate Device
PowerPort™ duo M.R.I.™ Implanted Port
with 9.5 Fr. Dual Lumen ChronoFlex™
Polyurethane Catheter (K090512) | Subject Device
PowerPort™ duo M.R.I.™ Implantable Port
with Attachable 9.5F ChronoFlex™ Open-Ended Dual-Lumen Venous Catheter | Discussion |
| --- | --- | --- | --- |
| Note | 1. Bold Font indicates a difference between the subject device and predicate devices
2. Plain Type indicates that the attribute of the subject device is the same as that of the predicate device(s). | | |
| Device Identification | | | |
| Manufacturer | Bard Access Systems, Inc. | Bard Access Systems, Inc. | Same as Predicate |
| Device Classification Name | Port & Catheter, Implanted, Subcutaneous, Intravascular | Port & Catheter, Implanted, Subcutaneous, Intravascular | Same as Predicate |
| Regulation Number | 21 CFR §880.5965 | 21 CFR §880.5965 | Same as Predicate |
| FDA Product Code | LJT | LJT | Same as Predicate |
| Device Use | | | |
| Intended Use | The PowerPort™ Implanted Port is a totally implantable vascular access device designed to provide long-term, repeated access to the vascular system. | The PowerPort™ Implanted Port is a totally implantable vascular access device designed to provide long-term, repeated access to the vascular system. | Same as Predicate |
| Indications for Use | The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. | The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications including anti-cancer medicines (chemotherapy)¹, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. | Different than Predicate*
*The italicized clarifying statement was cleared for all Bard Power-Injectable Implantable Ports in K181446, clearance date 07/08/2019. |
{14}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 11 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ duo M.R.I. ™ Implanted Port with 9.5 Fr. Dual Lumen ChronoFlex™ Polyurethane Catheter (K090512) | PowerPort™ duo M.R.I. ™ Implantable Port with Attachable 9.5F ChronoFlex™ Open-Ended Dual-Lumen Venous Catheter | |
| | | ¹The italicized clarifying statement was cleared for all Bard Power-Injectable Implantable Ports in K181446, clearance date 07/08/2019. | |
| Patient Population | Patients requiring repeated access to the vascular system | Patients requiring repeated access to the vascular system | Same as Predicate |
| Duration of Use | Long term (>30 days) | Long term (>30 days) | Same as Predicate |
| Insertion Site | Most commonly on upper chest | Most commonly on upper chest | Same as Predicate |
| Visualization Techniques | Fluoroscopy | Fluoroscopy | Same as Predicate |
| Principle of Operation | The device’s primary components consist of a triangular injection port with self-sealing silicone septum and a radiopaque catheter. A simple sliding lock collar secures the catheter to the port’s stem.
Port access is performed by percutaneous needle insertion using a non-coring needle. | The device’s primary components consist of a triangular injection port with self-sealing silicone septum and a radiopaque catheter. A simple sliding lock collar secures the catheter to the port’s stem.
Port access is performed by percutaneous needle insertion using a non-coring needle. | Same as Predicate |
| Catheter Tip Termination Location | Central venous system - lower 1/3 of superior vena cava preferred | Central venous system - lower 1/3 of superior vena cava preferred | Same as Predicate |
| Design Characteristics | | | |
| Device Measurements | Port Body:
Length: 38.5 mm
Width: 28.5 mm
Height: 12.6 mm
Reservoir Volume: 0.6 mL per reservoir | Port Body:
Length: Max 39.1 mm
Width: Max 29.2 mm
Height: 12.5 mm
Reservoir Volume: 0.6 mL per reservoir | Different than Predicate
The change in product dimensions did not change product performance and met the same product performance specifications. Labeling was updated for clarity only to ensure safer or more effective use. There were no changes to the indications for use, warnings, precautions or contraindications for the PowerPort™ duo M.R.I.™ |
{15}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 12 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ duo M.R.I. ™ Implanted Port with 9.5 Fr. Dual Lumen ChronoFlex™ Polyurethane Catheter (K090512) | PowerPort™ duo M.R.I. ™ Implantable Port with Attachable 9.5F ChronoFlex™ Open-Ended Dual-Lumen Venous Catheter | Implantable Port. There was no change in material composition, sterilization, or biocompatibility, so there was no change in safety or efficacy of the device. |
| | Catheter: Outside Diameter: 9.5 Fr
Inner Diameter: 1.5 mm
Catheter Length: 45 cm | Catheter: Outer Diameter: 9.5 Fr
Inner Diameter: 1.5 mm
Catheter Length: 45 cm | Same as Predicate |
| Device Description and Materials | Port Body: - Purple Triangular Delrin plastic port
- radiopaque identifying feature displaying power injection symbol and the letters "C T"
- Two silicone septa
- septum with 3 raised palpation bumps | Port Body: - Purple Triangular Delrin plastic port
- radiopaque identifying feature displaying power injection symbol and the letters "C T"
- Two silicone septa
- septum with 3 raised palpation bumps | Same as Predicate |
| | Catheter: - ChronoFlex™ polyurethane (with no colorant)
- Open-ended
- Dual lumen
- Attachable | Catheter: - ChronoFlex™ polyurethane (with no colorant)
- Open-ended
- Dual lumen
- Attachable | Same as Predicate |
| | Cathlock: Polycarbonate | Cathlock: Polycarbonate | Same as Predicate |
| Locking Solution | To help prevent clot formation and catheter blockage, each lumen of the implanted ports with open-ended catheters should be filled | To help prevent clot formation and catheter blockage, each lumen of the implanted ports with open-ended catheters should be | Same as Reference
Additional locking solutions to align with Infusion Therapy |
{16}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 13 of 15
| Attribute | Predicate Device | Subject Device | Discussion |
| --- | --- | --- | --- |
| | PowerPort™ duo M.R.I. ™ Implanted Port with 9.5 Fr. Dual Lumen ChronoFlex™ Polyurethane Catheter (K090512) | PowerPort™ duo M.R.I. ™ Implantable Port with Attachable 9.5F ChronoFlex™ Open-Ended Dual-Lumen Venous Catheter | |
| | with sterile heparinized saline after each use. If the port remains unused for long periods of time, the heparin lock should be changed at least once every four weeks. | flushed with sterile normal saline then filled with sterile heparinized saline, sterile normal saline, or other approved locking solution per institutional protocol after each use. If the port remains unused for long periods of time, the lock should be changed at least once every 28 days.
Warning: Alcohol should not be used to soak or declot polyurethane catheters because alcohol is known to degrade polyurethane catheters over time with repeated or prolonged exposure. | Standards of Practice, 9^{th} Edition (2024). The device has the same intended use and this change does not raise questions of safety and effectiveness, so it is deemed that there is no impact to substantial equivalence due to this change. |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same as Predicate |
| Shelf Life | 3 years and 4 months shelf life | 3 years and 4 months shelf life | Same as Predicate |
| Packaging Configuration | Triple tray packaging; two sterile barriers in the form of nested, sealed trays | Triple tray packaging; two sterile barriers in the form of nested, sealed trays | Same as Predicate |
{17}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 14 of 15
# Discussion of Differences
## Alternative locking solutions
Vascular access devices (VAD) including implantable ports are flushed (usually with saline) prior to each use to assess catheter function and to help prevent complications. After use is concluded and the final flush, each VAD catheter lumen is locked with a locking solution to decrease the risk of lumen occlusion. The VAD lumen is filled with enough locking solution to fill the entire port system including the full length of the catheter and deaccessed with positive pressure to reduce the potential for blood backflow into the catheter.
While the predicate devices are limited to being locked with heparinized saline, the subject devices with open-ended catheters allow the option to use sterile normal saline or other approved locking solution per institutional protocol. Infusion Therapy Standards of Practice, 9th Edition (2024), which provides standards of practice for infusion nurses, states that "In adults, randomized controlled trials (RCTs) and systematic reviews have shown equivalent outcomes with heparin and sodium chloride lock solutions for multilumen, nontunneled CVADs, peripherally inserted central catheters (PICCs), and implanted vascular access ports while accessed and when the access needle is removed." Also described in the Infusion Therapy Standards of Practice, 9th Edition (2024) is the use of other locking solutions: "Change to an alternative locking solution when the heparin lock solution is thought to be the cause of adverse drug reactions from heparin; when heparin-induced thrombocytopenia and thrombosis (HITT) develops; and when there are spurious laboratory studies drawn from the CVAD that has been locked with heparin." Examples of alternative locking solutions listed in this article are antimicrobial and antiseptic locking solutions (sodium bicarbonate, taurolidine, citrate, concentrated sodium chloride, and ethylenediaminetetraacetic acid). It is ultimately the physicians' judgement on what locking solution is best for the patient. It is important to note that although alcohol has been known to be used as an antiseptic locking solution, some of our catheters are incompatible with this solution and a clear warning to avoid using alcohol to soak or declot polyurethane catheters has been added to the lock procedure in the IFU.
The intended users of Bard Power-Injectable Implantable Ports rely on these standards of practice for their institutional protocols. The goal is to ensure that Bard Power-Injectable Implantable Ports can continue to be used by users with the most up to date locking standard of practices.
This exact locking solution procedure update was cleared for the reference device PowerPort™ ClearVUE™ Implantable Ports (K242328, cleared 10/31/2024) and the predicate device
{18}
Bundled Special 510(k): Device Modification
Bard Power-Injectable Implantable Port
Page 15 of 15
PowerPort™ Slim Implantable Port and PowerPort™ M.R.I.™ isp Implantable Port (K251253, cleared 18 June 2025).
## Dimensional Changes
The subject devices use the same materials as the predicate devices, so all material evaluations for the predicate devices are applicable to the subject devices. Sterilization and microbiology studies indicated that no new risks were introduced. The new dimensional specifications for the product were qualified and fulfilled performance criteria. No changes were made to the intended use, indications, contraindications, warnings, precautions, or packaging of the final device.
## Design Verification
Historical testing has been conducted and consisted of simulants used to account for the FDA approved oncology drugs, total parenteral nutrition (TPN) and vesicants used in intravascular access devices approved up to 2019. Chemical conditioning was performed prior to testing to simulate the worst-case solutions used with ports and their components. The testing evaluated the effects of the chemical simulants on silicone catheters as well as polyurethane catheters. It captures both medicines administrated, flushing solutions, and locking solutions. The functional testing with chemical conditioning demonstrates acceptable function of port devices. Studies evaluating these proposed solutions as well as guidelines of the use of these solutions in the field by health care providers are captured and summarized in the most recent Infusion Nurse Society guidelines, "Infusion Therapy Standards of Practice, revision 9".
Table 3. List of Verification/Validation Methods and Standards
| Verification/Validation Method(s) | Standard/Guidance |
| --- | --- |
| Port Subassembly Air Burst | FDA Port Guidance |
| Port Subassembly Tensile Strength | BAS Internal |
## Conclusion
Bard Access Systems, Inc. concludes that the subject devices, Vaccess™ CT Low-Profile Power-Injectable Implantable Port, Vaccess™ CT Power-Injectable Implantable Port, and PowerPort™ duo M.R.I.™ Implantable Port are substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.