The surgical microscope is intended for use by qualified dental professionals in hospital operating rooms and hospital clinical settings to provide illumination and magnification of the surgical area and for the support of visualization in surgical procedures or examinations (except ophthalmic surgeries). It optionally enables handling of light curing dental composite resins and aids in reducing specular light reflections and enhances contrast between red tissue and blood. The fluorescence mode is intended for use by qualified dental professionals in hospital clinical settings as an adjunctive aid to assist in the detection of dental caries.
Device Story
Surgical microscope (OMS2360, OMS2380) provides illumination and magnification for dental procedures; features fluorescence mode using 405nm blue LED light to excite bacteria for caries detection. Operated by dental professionals in hospital settings; provides visual output via microscope optics. Fluorescence mode acts as adjunctive diagnostic aid; enhances contrast between red tissue and blood; reduces specular reflections. Device includes super balancing arm and microscope body; installed via floor, ceiling, wall, or fixed ground mounts. Benefits include improved visualization and uninterrupted workflow during dental examinations and surgeries.
Clinical Evidence
Bench testing only. Includes photobiological safety (IEC 62471), electrical safety/EMC (IEC 60601-1/2), and performance testing (ISO 10936-1). Concurrence evaluation performed on clinical images from phantoms/patients demonstrated equivalent diagnostic capability to predicate.
Technological Characteristics
Fluorescence-based dental microscope; 405nm LED light source. Materials/specs comply with ISO 10936-1, IEC 60601-1, IEC 62471, IEC 62366-1, IEC 80601-2-60. Working distance 200-450mm. Power 40VA. Mechanical control for light/filters; no wireless connectivity.
Indications for Use
Indicated for qualified dental professionals in hospital operating rooms and clinical settings for surgical visualization, illumination, and magnification (excluding ophthalmic surgery); optional light curing of dental resins; and as an adjunctive aid for dental caries detection.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 28, 2026
Zumax Medical CO., LTD
% Mike Gu
RA Director
Suzhou Device Innovation Medical Consulting Co., Ltd
Room 1001, Building 19, No. 3188 Renmin Road
Suzhou, Jiangsu 215000
CHINA
Re: K252464
Trade/Device Name: Surgical Microscope (OMS2360, OMS2380)
Regulation Number: 21 CFR 872.1745
Regulation Name: Laser fluorescence caries detection device
Regulatory Class: Class II
Product Code: NBL, EPT
Dated: June 25, 2025
Received: August 6, 2025
Dear Mike Gu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252464 - Mike Gu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K252464 - Mike Gu
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252464 | |
| Device Name Surgical Microscope | |
| Indications for Use (Describe) The surgical microscope is intended for use by qualified dental professionals in hospital operating rooms and hospital clinical settings to provide illumination and magnification of the surgical area and for the support of visualization in surgical procedures or examinations (except ophthalmic surgeries). It optionally enables handling of light curing dental composite resins and aids in reducing specular light reflections and enhances contrast between red tissue and blood. The fluorescence mode is intended for use by qualified dental professionals in hospital clinical settings as an adjunctive aid to assist in the detection of dental caries. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
zumax
K252464
Section 038: 510(k) Summary
Surgical Microscope
1 of 6
{5}
zumax
# 510(k) Summary
K number: K252464
## I, Submitter Information
Name: Zumax Medical CO., LTD.
Address: 5 Zhiying Street, Suzhou New District, Suzhou, Jiangsu, China 215129
Telephone: +86-512-66650512-504
Fax: +86-512-66909655
Contact person: Alfred Hee
Email: alfred.hee@zumaxmedical.com
## II, Date Prepared
June 30, 2025
## III, Device Information
Trade name: Fluorescence Mode
Common name: Dental Fluorescence Examination Device
Regulation class: Class II
Regulation number: 21 CFR 872.1745
Review Panel: Dental
Product code: NBL
## IV, Predicative Device
510(k): K171007
Device name: Fluorescence Mode
Manufacturer: Carl Zeiss Suzhou Co., Ltd
## V, Device Description
Surgical microscopes OMS2360, OMS2380 can be installed in floor stand ceiling mounted, wall mounted and fixed ground. The installation methods are electrically identical, and the components include microscope main body, super balancing arm, second arm, and first arm, and the difference is that the installed structural parts are different.
The Fluorescence Mode is an accessory to the OMS2360, OMS2380 surgical microscope to utilize a kind of LED light that illuminates the tooth surfaces in the visible domain in the blue light region with wavelength 405nm.
2 of 6
{6}
zumax
## VI, Indication for Use
The surgical microscope is intended for use by qualified dental professionals in hospital operating rooms and hospital clinical settings to provide illumination and magnification of the surgical area and for the support of visualization in surgical procedures or examinations (except ophthalmic surgeries). It optionally enables handling of light curing dental composite resins and aids in reducing specular light reflections and enhances contrast between red tissue and blood.
The fluorescence mode is intended for use by qualified dental professionals in hospital clinical settings as an adjunctive aid to assist in the detection of dental caries.
## VII, Predicate Comparison
Table 1: Substantial Equivalence Table
| Attribute | Subject device | Predicate device | Comments |
| --- | --- | --- | --- |
| 510(k) number | Not applicable | K171007 | / |
| Intended use | The surgical microscope is intended for use by qualified dental professionals in hospital operating rooms and hospital clinical settings to provide illumination and magnification of the surgical area and for the support of visualization in surgical procedures or examinations (except ophthalmic surgeries). It optionally enables handling of light curing dental composite resins and aids in reducing specular light reflections and enhances contrast between red tissue and blood. The fluorescence mode is intended for use by qualified dental professionals in hospital clinical settings as an adjunctive aid to assist in the detection of dental caries. | The Fluorescence Mode is intended to be used by dentist as an aid in the detection of dental caries. | Same |
| Indication for use | The surgical microscope is intended for use by qualified dental professionals in hospital operating rooms and hospital clinical settings to provide illumination and magnification of the surgical area and for the support of visualization in surgical procedures or examinations (except ophthalmic surgeries). It optionally enables handling of light curing dental composite resins and aids in reducing specular light | The Fluorescence Mode is intended to be used by dentist as an aid in the detection of dental caries. | Same |
3 of 6
{7}
zumax
| | reflections and enhances contrast between red tissue and blood. The fluorescence mode is intended for use by qualified dental professionals in hospital clinical settings as an adjunctive aid to assist in the detection of dental caries. | | |
| --- | --- | --- | --- |
| Device Classification Name | Laser Fluorescence Caries Detection | Laser Fluorescence Caries Detection | Same |
| Generic Common Name | Dental Fluorescence Examination Device | Dental Fluorescence Examination Device | Same |
| Classification Product Code | NBL | NBL | Same |
| Class | II | II | Same |
| Intended user | No special requirements | No special requirements | Same |
| Technology | Fluorescence technology to aid in the detection of carious lesions | Fluorescence technology to aid in the detection of carious lesions | Same |
| Mode of Operation | Excites bacteria to fluoresce | Excites bacteria to fluoresce | Same |
| Detection Wavelength | 405 nm | 405 nm | Same |
| Device Operating Feature | Microscope with built-in feature provides visualization and an uninterrupted workflow | Dental microscope with built-in feature provides visualization and an uninterrupted workflow | Same |
| Light intensity | ≥90,000 lx | 97540 lx | Similar, analysis |
4 of 6
{8}
zumax
| Attribute | Subject device | | Predicate device | | Comments |
| --- | --- | --- | --- | --- | --- |
| | (at the working distance of 200mm) | | (at the working distance of 200mm) | | 1 |
| theWorking distance | 200-450 mm | | 200-430 mm | | Similar, analysis 2 |
| Power source | 40VA | | 120VA | | Similar, analysis 3 |
| Environmental conditions | Ambient requirements for operation | | Ambient requirements for operation | | Similar, analysis 4 |
| | Designation | Value | Designation | Value | |
| | Temperature | + 10°C ... +40°C | Temperature | + 10°C ... +40°C | |
| | Rel. humidity | 30% ... 75% | Rel. humidity | 30% ... 75% | |
| | Air pressure | 700 hPa ... 1060 hPa | Air pressure | 700 hPa ... 1060 hPa | |
| | Ambient requirements for transport and storage | | Ambient requirements for transport and storage | | |
| | Designation | Value | Designation | Value | |
| | Temperature | - 40°C ... + 55°C | Temperature | - 40°C ... + 70°C | |
| | Rel. humidity | 10% ... 80% | Rel. humidity | 10% ... 90% | |
| | Air pressure | 500 hPa ... 1060 hPa | Air pressure | 500 hPa ... 1060 hPa | |
| Standards | ISO 10936-1:2000
IEC 60601-2:2014
IEC 60601-1:2005
IEC 62471:2006
IEC 62366-1:2015
IEC 60601-1-6:2020
IEC80601-2-60:2019 | | ISO 10936-1:2000
IEC 60601-2:2014
IEC 60601-1:2005
IEC 62471:2006
IEC 62366-1:2015 | | Same |
| Software | Software of Surgical Microscope:
• Video Control (Only applicable to the video accessory which is optional before leaving the factory) | | Software of Surgical Microscope:
• Light Source Control
• Filter Control
• Video Control
• Wi-Fi Module: Firmware processes network communication | | Similar, analysis 5 |
Analysis 1: Test shows that the light intensity of the subject product is slightly higher than predicate one. The light source of the subject product and predicate product both meet the requirements of IEC 62471:2006/EN 62471: 2008 Photobiological safety of lamps and lamp systems. Also, the performance of the two products meet the requirements of ISO 10936- 1:2000 Optics and photonics – Operation microscopes – Part 1: Requirements and test methods.
The light source is intended for illumination of the surgical area, the difference in
5 of 6
{9}
zumax
light intensity does not affect the performance and clinical use.
Analysis 2: The working distance of the subject product contains and is better than predicate product. This difference does not affect the performance of the product.
Analysis 3: Compared with subject device and predicate device, there is a difference in the power of the devices, but both can drive the devices to achieve their respective functions. The difference does not affect the safety and effectiveness of the device.
Analysis 4: Compared with subject device and predicate device, there is a difference in the temperature for transport and storage. The difference does not affect the safety and effectiveness of the device.
Analysis 5: Software of the subject product is used to control the video, the light source and filter are controlled are mechanically controlled, and there is no wireless module in the subject device.
## VIII, Non-clinical Performance Testing
- Light safety testing in accordance with IEC 62471:2006.
- Electrical safety Testing and EMC in accordance with IEC60601-1 and IEC60601-1-2.
- Performance test for the fluorescence mode base on surgical microscope in accordance with ISO 10936-1:2000.
Additionally Conducting a concurrence evaluation on a certain number of clinical images from the same phantom or patients and reporting the results. The results show the ability of the device to provide images with equivalent diagnostic capability to those of the cleared predicate devices.
The test results of non-clinical tests performed on the subject device supported that it is substantially equivalent to the predicate devices despite the differences.
## IX, Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.