SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)
Applicant
Edan Instruments, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jul 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SE-1202 Series Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Device Story
Electrocardiograph (ECG) device; acquires ECG signals from adult/pediatric patients via body surface electrodes; simultaneous 9 or 12-lead acquisition. Features LCD touch screen, operation panel, and thermal printer. Operates on battery or AC power. Uses internal algorithms (SEMIP or Glasgow) to generate measurements, data/graphical presentations, and interpretive statements. Used in hospitals/healthcare facilities by physicians and trained healthcare professionals. Records saved to flash memory or transmitted to PC. Output provided to clinicians on an advisory basis to support heart disease diagnosis. Optional 2.4/5 GHz WiFi and 4G connectivity.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and software verification/validation. Bench testing confirmed compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-25 standards, demonstrating accuracy and performance specifications equivalent to the predicate.
Technological Characteristics
Class II electrocardiograph; 9 or 12-lead simultaneous acquisition. Frequency response 0.01-500Hz; DC offset ±960 mV. Thermal dot-matrix recorder. Connectivity: WiFi (2.4/5 GHz), 4G (optional). Power: Battery/AC. Safety: Type CF defibrillation-proof, Class I. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25.
Indications for Use
Indicated for adult and pediatric patients for the acquisition of ECG signals via body surface electrodes to assist clinicians in the analysis and diagnosis of heart disease.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 2, 2026
Edan Instruments, Inc.
Tracy Yue
Regulatory Engineer
15 Jinhui Rd., Jinsha Community, Kengzi Sub-District,
Pingshan District,
Shenzhen, Guangdong 518122
China
Re: K252463
Trade/Device Name: SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)
Regulation Number: 21 CFR 870.2340
Regulation Name: Electrocardiograph
Regulatory Class: Class II
Product Code: DPS
Dated: June 5, 2026
Received: June 5, 2026
Dear Tracy Yue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252463 - Tracy Yue
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252463 - Tracy Yue
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# KIMBERLY N.
CROWLEY -S
For: Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252463 | ? |
| Please provide the device trade name(s). | | ? |
| SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E) | | |
| Please provide your Indications for Use below. | | ? |
| The SE-1202 Series Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Electrocardiograph
Page 8 of 50
{4}
Traditional 510(k) of Electrocardiograph
# 510(K) Summary
Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
# 1. Submitter:
Applicant:
Edan Instruments, Inc.
#15 Jinhui Road, Jinsha Community,
Kengzi Sub-District, Pingshan District,
Shenzhen, 518122 P.R.China.
Tel: +86(0755) 26858736
Fax: +86(0755) 26882223
Contact person:
Tracy Yue
Preparing date:
August 6th, 2025
# 2. Device name and classification:
Trade name:
SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)
Common/Usual Name:
Electrocardiograph
Classification Name:
21 CFR 870.2340 Electrocardiograph
Regulatory class:
Class II
Primary product code:
DPS - Electrocardiograph
# 3. Predicate Device(s):
Primary Predicate Device: Edan Instruments, Inc., Electrocardiograph, Model: SE-1202 (K210140)
# 4. Device Description:
The SE-1202 Series Electrocardiograph feature an LCD touch screen, an operation panel, user-programmable reports, and the ability to operate on either battery or AC power. It is capable of simultaneous acquisition, display, and print of electrocardiogram. It uses algorithm to generate measurements, data presentations, graphical presentations and interpretative statements. The record can be saved in flash memory or send to PC.
# 5. Indication for Use
The SE-1202 Series Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
1 / 5
{5}
Traditional 510(k) of Electrocardiograph
## 6. Predicate Device Comparison
The table below compares the indication for use and key technological feature of the subject devices to the predicate device.
| Item | <Subject Device> Model: SE-1202, SE-1202E | <Predicate Device> Model: SE-1202 | Comparison Result |
| --- | --- | --- | --- |
| K# | Current Submission | K210140 | —— |
| Regulation/ Product code | 21 CFR 870.2340 Electrocardiograph DPS - Electrocardiograph | 21 CFR 870.2340 Electrocardiograph DPS - Electrocardiograph | Same |
| **Intended Use/Indications for Use** | | | |
| Intended Use/Indications for Use | The SE-1202 Series Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. | The SE-1202 12-lead electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. | Same |
| Intended Patient Population | Adult and pediatric patients | Adult and pediatric patients | Same |
| Algorithm | SEMIP or Glasgow | SEMIP or Glasgow | Same |
| **Physical Specifications** | | | |
| Dimension | SE-1202: 400*360*90mm, ±2mm | SE-1202: 400*360*90mm, ±2mm | Similar |
| | SE-1202E: 400*360*183mm, ±5mm | / | |
| Weight | SE-1202: ≤6.5kg, ±0.3 kg | SE-1202: ≤6.5kg, ±0.3 kg | |
2 / 5
{6}
Traditional 510(k) of Electrocardiograph
| Item | <Subject Device> Model: SE-1202, SE-1202E | <Predicate Device> Model: SE-1202 | Comparison Result |
| --- | --- | --- | --- |
| | SE-1202E: ≤5.6kg, ±0.3 kg | / | |
| **Display& Touchscreen** | SE-1202: 10.1", 1280×800 multicolor LCD Screen (flippable screen with 0° or 25° forward tilt) | SE-1202: 10.1", 1280×800 multicolor LCD Screen (flippable screen with ≤25°forward tilt) | |
| | SE-1202E: 13.3", 1920×1080 multicolor LCD Screen (fixed 25° screen) | / | |
| **Safety Specifications** | | | |
| **Safety Standards** | IEC 60601-1:2015+A1:2012+A2:2020 IEC 60601-1-2:2014+A1:2020 IEC 60601-2-25:2011 | IEC 60601-1:2005/A1:2012 IEC 60601-1-2:2014 IEC 60601-2-25:2011 | Similar |
| **Anti-electric-shock type** | Class I with internal power supply | Class I with internal power supply | Same |
| **Anti-electric-shock degree** | Type CF with defibrillation-proof | Type CF with defibrillation-proof | Same |
| **Working mode** | Continuous operation | Continuous operation | Same |
| **Performance Specifications** | | | |
| **Recorder** | Thermal dot-matrix recorder | Thermal dot-matrix recorder | Same |
| **HR Range** | 30 bpm ~300 bpm, ±1 bpm | 30 bpm ~300 bpm, ±1 bpm | Same |
| **Leads** | 9 or 12 standard leads | 9 or 12 standard leads | Same |
| **Acquisition Mode** | 9 or 12 leads acquisition simultaneously | 9 or 12 leads acquisition simultaneously | Same |
| **Frequency Response** | 0.01~500Hz | 0.01~350Hz | Similar |
| **DC Offset Voltage** | ±960 mV, ±5% | ±900 mV, ±5% | Similar |
| **Wireless Function** | 2.4 GHz & 5 GHz WiFi (Optional) 4G (Optional) | 2.4 GHz & 5 GHz WiFi (Optional) | Similar |
| **Environmental Specifications** | | | |
3 / 5
{7}
Traditional 510(k) of Electrocardiograph
| Item | <Subject Device> Model: SE-1202, SE-1202E | <Predicate Device> Model: SE-1202 | Comparison Result |
| --- | --- | --- | --- |
| Working Temperature | +5°C (+41°F) ~ +40°C (+104°F) | +5°C (+41°F) ~ +40°C (+104°F) | Same |
| Transport /Storage Temperature | -20°C (-4°F) ~ +55°C (+131°F) | -20°C (-4°F) ~ +55°C (+131°F) | Same |
| Working Relative Humidity | 15% RH~95% RH Non-Condensing | 15% RH~95% RH Non-Condensing | Same |
| Transport /Storage Relative Humidity | 15% RH~95% RH Non-Condensing | 15% RH~95% RH Non-Condensing | Same |
| Working Atmospheric Pressure | 70kPa ~106kPa | 70kPa ~106kPa | Same |
| Transport /Storage Atmospheric Pressure | 70kPa ~106kPa | 70kPa ~106kPa | Same |
As seen in the comparison tables, the subject and predicate device have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
### 7. Performance Data:
#### Non-clinical data:
Electrical safety and electromagnetic compatibility (EMC)
SE-1202 Series Electrocardiograph were assessed for conformity with the relevant requirements of the following standards and found to comply:
- IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Performance testing-Bench
4 / 5
{8}
Traditional 510(k) of Electrocardiograph
Edan has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 60601-2-25:2011 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions".
Clinical data: Not applicable.
# Summary
The non-clinical performance testing showed that the subject devices are as safe and effective as the predicate device.
# 8. Conclusion
The bench testing data and software verification and validation demonstrate that the SE-1202 Series Electrocardiograph are substantially equivalent to the predicate device.
5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.