The device is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals.
Device Story
BruxZir® NOW is a pre-shaded, fully sintered yttria-stabilized zirconia (YSZ) block used for fabricating single or multiple-unit dental restorations. The device is processed by dental laboratories or professionals using CAD/CAM or manual milling machines to create anatomical restorations based on patient tooth rendering. The device is provided in various VITA Classical shades. The output is a finished dental restoration; the clinician uses this to restore patient dentition. The device is a passive material; it does not perform active sensing or computation. The addition of a contraindication for bridge block design (max 4 units, 2 pontics) serves as a practitioner reminder regarding block size limitations.
Clinical Evidence
No clinical data included. Substantial equivalence is based on non-clinical bench testing previously submitted under K220816, including flexural strength, solubility, shade evaluation, biocompatibility (ISO 10993-1), radioactivity, and packaging validation (ASTM D4169-16).
Technological Characteristics
Material: Yttria-stabilized zirconia (YSZ) with pigments. Form: Fully sintered block. Standards: ISO 6872:2024 (Type II, Class 4/5), ISO 10993-1:2018. Mechanical: Flexural strength >800 MPa, solubility <100 μg/cm². Radioactivity: U-238 ≤ 1.0 Bq/g. Connectivity: None (passive material).
Indications for Use
Indicated for dental restorations using CAD/CAM or manual milling machines. Processed by dental laboratories or professionals. Contraindication: Bridge block design limited to max 4 units with max 2 pontics between pillars.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 3, 2025
Prismatik Dentalcraft, Inc.
So Hyun Park
Sr. Regulatory Affairs Manager, MS
2144 Michelson Drive
Irvine, California 92612
Re: K252446
Trade/Device Name: BruxZir® NOW
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain Powder For Clinical Use
Regulatory Class: Class II
Product Code: EIH
Dated: August 1, 2025
Received: August 4, 2025
Dear So Hyun Park:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252446 - So Hyun Park
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252446 - So Hyun Park
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252446 | |
| Device Name BruxZir® NOW | |
| Indications for Use (Describe) The device is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals. Contraindication: BruxZir® NOW Bridge Block design is limited to a maximum of four units with a maximum of two pontics between pillars. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
PRISMATIK DENTALCRAFT, INC.
K252446
# 510(k) Summary
Date Prepared: August 1, 2025
## CONTACT DETAILS (21 CFR 807.92(a)(1))
Applicant Name: Prismatik Dentalcraft, Inc.
Applicant Address: 2144 Michelson Drive, Irvine, CA 92612, USA
Applicant Contact Telephone: 949-863-5479
Applicant Contact: So Hyun Park, Sr. Regulatory Affairs Manager
Applicant Contact Email: so.park@glidewelldental.com
Secondary Contact: Shelly Gallup, Sr.VP – RA/QA
Secondary Contact Email: shelly.gallup@glidewelldental.com
Secondary Contact Phone: 949-222-3590
## DEVICE NAME (21 CFR 807.92(a)(2))
Device Trade Name: BruxZir® NOW
Common Name: Zirconia Milling Block or Dental CAD/CAM Block
Classification Name: Porcelain powder for clinical use
Regulation Number: 872.6660
Product Code: EIH
## LEGALLY MARKETED PREDICATE DEVICE (21 CFR 807.92(a)(3))
Predicate 510(k) #: K220816
Predicate Trade Name: BruxZir® NOW
Product Code: EIH
## DEVICE DESCRIPTION SUMMARY (21 CFR 807.92(a)(4))
BruxZir® NOW is a pre-shaded, fully sintered zirconia CAD/CAM block used for fabricating single-unit or multiple-unit dental restorations. BruxZir® NOW is used for fabricating single-unit dental restorations, while BruxZir® NOW Bridge Block is used for fabricating multiple-unit dental restorations. BruxZir® NOW is available in the following shades: Bleach White, Bleach 1, Bleach 3, and 16 VITA Classical shades (A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4). BruxZir® NOW Bridge Block is available in the following 6 VITA Classical shades: A1, A2, A3, B1, C2, D2. The method of fabricating the restoration is with CAD/CAM milling systems.
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® NOW
August 2025
{5}
PRISMATIK DENTALCRAFT, INC.
# INTENDED USE/INDICATIONS FOR USE (21 CFR 807.92(a)(5))
The device is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals.
Contraindication: BruxZir® NOW Bridge Block design is limited to a maximum of four units with a maximum of two pontics between pillars.
# INDICATIONS FOR USE COMPARISON (21 CFR 807.92(a)(5))
The subject device, BruxZir® NOW, has the same indications for use as the predicate device, BruxZir® NOW (K220816). Both devices are indicated for use for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals. The subject device, BruxZir® NOW, also has the addition of a contraindication for use. BruxZir® NOW Bridge Block design is limited to a maximum of four units with a maximum of two pontics between pillars.
# TECHNOLOGICAL COMPARISON (21 CFR 807.92(a)(6))
The subject device, BruxZir® NOW, is substantially equivalent in technical characteristics to the predicate device, BruxZir® NOW (K220816), in terms of overall design principles and performance.
The subject device, BruxZir® NOW, is unchanged in terms of overall design to the predicate device, BruxZir® NOW (K220816). Both devices are offered in a fully sintered block form in two sizes for single and multiple-unit restorations based on the anatomical rendering of the patient's teeth using CAD/CAM equipment.
The subject device, BruxZir® NOW, is unchanged in terms of material composition to the predicate device, BruxZir® NOW (K220816). The main ceramic component is composed of yttria-stabilized zirconia with varying trace amounts of colorants. There is no difference in overall chemical composition and therefore, does not affect the biocompatibility and performance characteristics.
The performance specifications, fundamental scientific technology, and intended use of the subject device, BruxZir® NOW, are unchanged from the predicate device, BruxZir® NOW (K220816).
The technological comparison table below outlines and provides the similarities between the subject device, BruxZir® NOW, and the predicate device, BruxZir® NOW (K220816). Both the subject device and the predicate device have the same physical/mechanical and biocompatibility properties that met the requirements of ISO 6872:2024 (Type II, Class 4) and ISO 10993. Any differences between the subject device and the predicate device do not raise any new concerns of safety and effectiveness.
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® NOW
August 2025
Page 2 of 5
{6}
PRISMATIK DENTALCRAFT, INC.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS (21 CFR 807.92(a)(6))
| Technological Characteristics | | Subject Device (TBD) | Predicate Device (K220816) | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | | BruxZir® NOW | BruxZir® NOW | N/A |
| Manufacturer | | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | Same |
| Product Code | | EIH | EIH | Same |
| Prescription Device | | Yes | Yes | Same |
| Indications for Use | | BruxZir® NOW is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals. | BruxZir® NOW is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals. | Same |
| Contraindication | | BruxZir® NOW Bridge Block design is limited to a maximum of four units with a maximum of two pontics between pillars. | None. | Different; the subject device includes the addition of a contraindication for use of the bridge block design. |
| Design Characteristics | Chemical Composition | The device is composed of Yttria-stabilized zirconia (YSZ) and pigments to achieve the desired shades. | The device is composed of Yttria-stabilized zirconia (YSZ) and pigments to achieve the desired shades. | Same |
| | Design | Fully sintered zirconia block available in block form in 2 sizes for single and multiple-unit restorations | Fully sintered zirconia block available in block form in 2 sizes for single and multiple-unit restorations | Same |
| | Shades | BruxZir® NOW is available in the following shades: Bleach White, Bleach 1, Bleach 3, and 16 VITA Classical shades (A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4). BruxZir® NOW Bridge Block is available in the following 6 VITA Classical shades: A1, A2, A3, B1, C2, D2. | BruxZir® NOW is available in the following shades: Bleach White, Bleach 1, Bleach 3, and 16 VITA Classical shades (A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4). BruxZir® NOW Bridge Block is available in the following 6 VITA Classical shades: A1, A2, A3, B1, C2, D2. | Same |
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® NOW
August 2025
{7}
PRISMATIK DENTALCRAFT, INC.
| Technological Characteristics | | Subject Device (TBD) | Predicate Device (K220816) | Comparison |
| --- | --- | --- | --- | --- |
| | Flexural Strength | >800 MPa
Type II Class 5 per ISO 6872:2024 | >800 MPa
Type II Class 5 per ISO 6872:2024 | Same |
| | Biocompatibility | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 | Same |
| | Solubility | <100 μg/cm² | <100 μg/cm² | Same |
| | Radioactivity | The activity concentration of Uranium-238 is no more than 1.0 Bq/g. | The activity concentration of Uranium-238 is no more than 1.0 Bq/g. | Same |
# NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS 21 CFR 807.92(b))
No changes were made to any physical component of BruxZir® NOW. The performance specifications, fundamental scientific technology, and intended use of the subject device, BruxZir® NOW, are unchanged from the predicate device, BruxZir® NOW (K220816).
The proposed device is BruxZir® NOW with a labeling modification. Specifically, the modification involves the addition of a contraindication pertaining to the use of the device in the BruxZir® NOW Bridge Block design, which led to a Change Being Effected (CBE) 510(k). A risk-based assessment was conducted and identified a specific use of BruxZir® NOW Bridge Blocks to a maximum of four units with a maximum of two pontics between pillars. The labeling change was added to help remind the practitioner of the limitation in bridge design imposed by the size of the block, rather than to address any substantial safety risks of the device that restrict its use. It is not associated with any new safety concerns or risk mitigation strategies, nor does it impact the previously cleared indications for use under the existing 510(k) for BruxZir® NOW (K220816).
Following the FDA Guidance, Deciding When to Submit a 510(k) for a Change to an Existing Device, a Change Being Effected (CBE) 510(k) is appropriate when the change is only the addition of a contraindication. Per the FDA Guidance, the Special 510(k) Program, a Special 510(k) is appropriate when it is a design or labeling change to the manufacturer's own device and performance data are not needed to evaluate the change.
Design verification for BruxZir® NOW (K220816) was previously conducted by Prismatik Dentalcraft, Inc. and is not required for this submission. The following non-clinical data has already been submitted under BruxZir® NOW (K220816) to demonstrate substantial equivalence:
- Flexural Strength according to ISO 6872:2024
- Solubility according to ISO 6872:2024
- Shade evaluation
- Biocompatibility according to ISO 10993-1:2018
- Radioactivity according to ISO 6872:2024
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® NOW
August 2025
{8}
PRISMATIK DENTALCRAFT, INC.
- Packaging Validation according to ASTM D4169-16
No additional verification or validation is necessary to evaluate this change. No clinical data is included in this submission.
The subject device, BruxZir® NOW, is substantially equivalent to the predicate device, BruxZir® NOW (K220816), in regard to intended use, indications for use, fundamental technology including design, materials, manufacturing methods and operational principles. Based on this information, the modification does not raise any new issues regarding the safety and effectiveness of the device when compared to its predicate.
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® NOW
August 2025
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.