The Fusion Bioline Vascular Grafts are designed to repair or replace peripheral arteries.
Device Story
Fusion Bioline Vascular Graft is a single-use, permanent implantable synthetic graft used to restore blood flow in peripheral arteries. The device consists of an inner layer of extruded, expanded polytetrafluoroethylene (ePTFE) and an outer layer of knit polyester (PET) textile, fused together with a proprietary polycarbonate-urethane adhesive. The luminal surface features a heparin and recombinant human albumin coating. Some models include a removable, continuous spiral support coil and a Guideline stripe for alignment. The device is implanted by a surgeon in a clinical setting. It functions as a mechanical conduit to bypass or replace diseased arterial segments, thereby improving perfusion and potentially preventing complications associated with peripheral arterial disease.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by non-clinical performance testing (bond strength, tensile strength, burst pressure, suture retention, etc.), biocompatibility testing (cytotoxicity, sensitization, irritation, hemocompatibility, etc.), and design validation (user evaluation of handling and suturability).
Technological Characteristics
Two-layer synthetic vascular graft: inner ePTFE, outer knit PET. Fused with polycarbonate-urethane adhesive. Luminal surface coated with porcine heparin and recombinant human albumin. Includes optional removable spiral support coil. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring repair or replacement of peripheral arteries.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 27, 2026
Maquet Cardiovascular, LLC
Suwah Amara
Sr. Regulatory Affairs Specialist
45 Barbour Pond Drive
Wayne, New Jersey 07470
Re: K252445
Trade/Device Name: Fusion Bioline Vascular Graft (Various model numbers)
Regulation Number: 21 CFR 870.3450
Regulation Name: Vascular Graft Prosthesis
Regulatory Class: Class II
Product Code: DSY, DYF
Dated: March 20, 2026
Received: March 20, 2026
Dear Suwah Amara:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252445 - Suwah Amara
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252445 - Suwah Amara
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rohini Retarekar -S
for Carmen Gacchina Johnson, Ph.D.
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252445 | |
| Device Name Fusion Bioline Vascular Graft | |
| Indications for Use (Describe) The Fusion Bioline Vascular Grafts are designed to repair or replace peripheral arteries. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
GETINGE
# 510(k) Summary
(in accordance with 21 CFR part 807.92)
510(K) Number: K252445
Submitter Name and Address: Maquet Cardiovascular, LLC
45 Barbour Pond Rd
Wayne NJ 07470
United States
Contact Person: Suwah Amara
Sr. Regulatory Affairs Specialist
Tel.: (973)-692-5496
Email: suwah.amara@getinge.com
Date Prepared: April 24, 2026
Device Information
Device Trade Name: Fusion Bioline Vascular Graft
Common/Generic Name (s): Vascular Graft
Classification Name: Prosthesis, Vascular Graft, of 6mm and Greater Diameter
Prosthesis, Vascular Graft, of Less Than 6mm Diameter
Regulation Number: 870.3450
Product Code: DSY, DYF
Device Class: Class II according to 21 CFR 870.3450
| Legally Marketed Predicate Device(s) | 510(k) Number | Product Code |
| --- | --- | --- |
| Fusion Bioline Vascular Graft | K131778 | DSY |
Fusion Bioline Perseus Project Traditional 510(k) K252445
{5}
GETINGE
# Device Description:
The Fusion Bioline Vascular Grafts are single-use, permanent implantable devices that restore blood flow by repairing or replacing peripheral arteries. They are synthetic vascular grafts constructed of two layers. The inner layer is comprised of extruded, expanded polytetrafluoroethylene (ePTFE). The outer layer is comprised of knit polyester (PET or Polyethylene Terephthalate) textile. These two layers are fused together with a proprietary polycarbonate – urethane adhesive. Externally Supported Fusion Bioline Vascular Grafts have a removable, continuous spiral support coil. The graft features a Guideline stripe to facilitate proper graft alignment. The Fusion Bioline Vascular Grafts include medicinal and recombinant products.
The Fusion Bioline Vascular Graft features a heparin and albumin coating on the graft's luminal surface. The heparin is of porcine origin and the albumin is animal-free recombinant human albumin.
# Indications for Use:
The Fusion Bioline Vascular Grafts are designed to repair or replace peripheral arteries.
# Indications for Use Comparison:
The subject Fusion Bioline Vascular Grafts share the same Indications For Use as the predicate device of the same name.
# Comparison of Technological Characteristics with the Predicate Device:
The subject device, Fusion Bioline Vascular Grafts, and the predicate device have the following similarities:
- The same Indications for Use
- The same operating principles
- The same design (two layer graft, with the layers fused together by an adhesive., and Bioline Coating on the surface of the graft inner layer)
- The same Bioline Coating applied on the surface of the inner layer
- The same sterilization method
- The same sterile barrier packaging material and process
- The same shelf-life
- The same patient population
- The same user population
- The same anatomical site
- The same use environment
The subject device incorporates the following modifications:
- Comparable outer layer
- Comparable adhesive that fuse the outer and inner layers
Design verification and validation tests results demonstrated that the subject Fusion Bioline Vascular Graft do not raise new or different concerns of safety or effectiveness. The tests conducted to support this conclusion are included in the submission and are summarized below.
# Non-Clinical and/or Clinical Test Summary:
Non-clinical performance test, design validation (user evaluation), and biocompatibility tests were conducted on the subject device to support a determination of substantial equivalence. Non-clinical performance tests were also performed to establish shelf-life; these tests are marked with an asterisk “*”. There were no animal or clinical studies of the modified device.
Fusion Bioline Perseus Project Traditional 510(k) K252445
{6}
GETINGE
## Performance Tests
- Bond Adhesive Strength*
- Longitudinal Tensile Strength (LTS)*
- Water Entry Pressure (WEP)*
- Wall Thickness*
- Suture Retention Strength – Oblique*
- Suture Retention Strength – Longitudinal*
- Kink Diameter (unsupported graft only)*
- Relaxed Internal Diameter – RID*
- Guideline*
- Radial Burst Strength (Burst Pressure)*
- Graft Profile – OD*
- Longitudinal Axial Stretch*
- Graft Appearance*
- Bead Peel Strength (Support Graft only)*
## Biocompatibility Tests
- Physical and chemical information
- Cytotoxicity
- Sensitization
- Irritation
- Material Mediated Pyrogenicity
- Acute Systemic Toxicity
- Subacute, Subchronic, Chronic toxicity
- Implantation
- Hemocompatibility
- Genotoxicity / Carcinogenicity
## User Evaluation (Design Validation)
- Cut to length
- Suturability
- Graft appearance and handling
- Use of commercially available sheath tunneler
## Conclusion:
The subject Fusion Bioline Vascular Graft and its predicate device have the same indication for use and device design. There are no performance or functional differences between the predicate and subject device. The difference between the subject device and the predicate device do not raise new or different questions of safety or effectiveness. The subject device test results demonstrate substantial equivalence to the predicate device.
Fusion Bioline Perseus Project Traditional 510(k) K252445
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.