Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria and characteristics of the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
Device Story
Sonio Detect is a SaaS-based concurrent reading aid for fetal ultrasound exams; used by sonographers, OB/GYN MFMs, and fetal surgeons. Input: real-time fetal ultrasound images/clips from GE, Samsung, Canon, or Philips machines. Operation: Edge software on local server receives DICOM instances; uploads to cloud for AI analysis. AI automatically detects views, anatomical structures, and quality criteria; outputs bounding boxes for views/structures. Highlights unverified items in yellow; allows user to override/edit assignments. Benefit: assists HCPs in ensuring exam completeness and protocol adherence; improves workflow efficiency during routine fetal ultrasound.
Clinical Evidence
Bench testing only. No clinical study performed. Performance validated on an independent dataset of 22,496 fetal ultrasound images. Results show high sensitivity (0.87-0.94) and specificity (0.80-0.99) for view/structure detection across trimesters. Localization performance measured by mIoU (0.679-0.777). Subgroup analysis performed for ultrasound manufacturer, BMI, maternal age, image quality, and gestational age.
Technological Characteristics
SaaS solution; cloud-based and edge software. Compatible with GE, Samsung, Canon, and Philips ultrasound systems. Uses AI/ML for image analysis. Outputs include view detection, anatomical structure detection, quality criteria verification, and bounding box localization. No specific hardware materials or sterilization required as it is a software-only accessory.
Indications for Use
Indicated for use by healthcare professionals (sonographers, OB/GYN MFMs, fetal surgeons) to assist in the acquisition and interpretation of fetal ultrasound images during routine examinations for trimesters 1, 2, and 3 (gestational age 11 to 41 weeks).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 16, 2026
Sonio
Florian Akpakpa
Director of Regulatory Affairs and Compliance Officer
147 rue d'Aboukir
Paris, 750002
France
Re: K252433
Trade/Device Name: Sonio Detect (v3)
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, QIH
Dated: February 14, 2026
Received: February 17, 2026
Dear Florian Akpakpa:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252433 - Florian Akpakpa
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252433 - Florian Akpakpa
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252433 | ? |
| Please provide the device trade name(s). | | ? |
| Sonio Detect (v3) | | |
| Please provide your Indications for Use below. | | ? |
| Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria and characteristics of the views.
The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Sonio
510(k) Premarket Notification Submission
S
sonio
K252433
# 510(k) Summary
In accordance with 21 CFR 807.92 the 510(k) summary for Sonio Detect v3 is provided below.
# I. Submitter
| Applicant: | Sonio
147 rue d’Aboukir,
75002, Paris France |
| --- | --- |
| Primary Contact Person: | Florian Akpakpa
Director Regulatory Affairs and Compliance Officer
Sonio
Phone: +33 6 66 66 19 89
Email: florian.akpakpa@sonio.ai |
| Date Prepared: | February 13, 2026 |
# II. Device
| Device Trade Name: | Sonio Detect |
| --- | --- |
| Classification Name: | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System
21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System
21 CFR 892.2050 - Medical Image Management and Processing System |
| Regulatory Class: | Class II |
| Product Code: | IYN (primary)
IYO, QIH (Secondary) |
# III. Predicate Device
Sonio Detect v2 cleared in K240406.
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
{5}
Sonio
510(k) Premarket Notification Submission
sonio
# IV. Device Description
Sonio Detect is a Software as a Service SaaS solution that aims at helping sonographers, OB/GYN MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP in the following) to perform their routine fetal ultrasound examinations in real-time. Sonio Detect can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 1, Trimester 2 and Trimester 3 of the fetus (GA: from 11 weeks to 41 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol.
Sonio Detect requires the following:
- Edge Software (described below) to install on a server on the same network as the Ultrasound Machine;
- SaaS accessibility from any internet browser (recommended browser: Google Chrome).
Sonio’s Edge Software is a light-weight application that runs on a server (computer) connected to the same network as the Ultrasound Machine. Sonio Edge Software is installed on the HCP server (computer) and network and the main purpose is to receive DICOM instances from the Ultrasound Machine and upload them to Sonio’s Cloud to be used by Sonio Detect.
Sonio Detect receives fetal ultrasound images and clips from the ultrasound machine, that are submitted through the edge software by the performing healthcare professional, in real-time and performs the following:
- Automatically detect views;
- Automatically detect anatomical structures within the supported views;
- Automatically verify quality criteria and characteristics of the supported views by checking whether they conform to standardized quality criteria
- Automatically output bounding box for views and structures
Quality criteria are related to:
- The presence of an anatomical structure;
- The absence of an anatomical structure;
Characteristics are related to other items than quality criteria:
- Location of the placenta
- Fetus sex
Sonio Detect then automatically associates the image to its detected view. It also highlights in yellow the view and/or the corresponding quality criteria or characteristics if there are unverified items: quality criteria or characteristics not verified or view not detected.
The end user can interact with the software to override the Sonio Detect’s outputs (reassign the image to another view or unassign it or assign it if it was not assigned, changes the status
Page 29-3
{6}
510(k) Premarket Notification Submission
sonio
of a quality criteria from verified to unverified or from unverified to verified) and manually set the characteristics of the views. The user has the ability to review and edit/override the matching at any time during or at the end of the exam.
The list of views, anatomical structures, quality criteria and characteristics that can be automatically detected and verified by the software are detailed in tables 1, 2, 3 and 4 below.
Table 1: List of views per trimester that can be automatically detected by Sonio Detect
| Trimester | View |
| --- | --- |
| First Trimester | 1. Choroid Plexus
2. Profile/Nuchal translucency
3. Abdominal circumference
4. Hand
5. Foot
6. Crown Rump Length
7. Doppler 3VT
8. Lower Arm
9. Upper Leg
10. Lower Leg
11. Doppler 4CH
12. Adnexa
13. Ovaries
14. Umbilical cord insertion |
| Second and Third trimester | 1. Transthalamic or Cavum septum pellucidum or Midline falx/Transventricular or Choroid Plexus
2. Transcerebellar view
3. Profile
4. Lips and Nose
5. Orbits
6. 4 Chambers
7. LVOT
8. Sagittal Spine
9. Abdominal circumference
10. Axial Bladder
11. Axial Kidneys
12. Cervix/Placenta
13. Hand
14. Foot
15. External genitalia (female or male)
16. 3 Vessels/3 Vessels and Trachea (3VV)
17. 3 Vessels and Trachea (3VT)
18. Doppler 3VT |
Page 29-4
{7}
sonio
# 510(k) Premarket Notification Submission
| Trimester | View |
| --- | --- |
| | 19. Humerus
20. Femur
21. Radius/Cubitus
22. Tibia/Fibula
23. Upper arm
24. Lower arm
25. Upper leg
26. Adnexa
27. Ovaries
28. Placenta Cord Insertion
29. Coronal Face
30. Doppler 4CH
31. Umbilical Cord Insertion
32. Coronal Kidneys
33. Diaphragm
34. Maxilla
35. Mandible
36. Aortich Arch
37. Ductal Arch
38. Bicaval View
39. Palate
40. Corpus Callosum |
Page 29-5
{8}
Sonio
510(k) Premarket Notification Submission
sonio
Table 2: List of anatomical structures that can be automatically detected by Sonio Detect
| View name | Structures to be detected | First Trimester T1 | Second/Third Trimesters T2/T3 |
| --- | --- | --- | --- |
| Brain views & structures | | | |
| Transthalamic
Transventricular
Transcerebellar
Choroid plexus | Thalami on the transthalamic view | X | X |
| | Cavum septum pellucidum | - | X |
| | Sylvian fissure | - | X |
| | Choroid Plexus | - | X |
| | Cisterna Magna | - | X |
| | Cerebellum | - | X |
| | Vermis | - | X |
| Thorax and Heart views & structures | | | |
| 4 chambers
3VV
3VT
RVOT
LVOT
Abdominal circumference
Axial view of the kidneys
Diaphragm
Crown Rump length
Abdominal cord insertion | Adrenal gland | - | - |
| | Apex of the heart | - | X |
| | Descending aorta | - | X |
| | Interatrial septum | - | X |
| | Interventricular septum | X | X |
| | Kidneys | - | - |
| | Left atrium | X | X |
| | Left ventricle | X | X |
| | Right atrium | - | X |
| | Right ventricle | - | X |
Page 29-6
{9}
sonio
# 510(k) Premarket Notification Submission
| View name | Structures to be detected | First Trimester T1 | Second/Third Trimesters T2/T3 |
| --- | --- | --- | --- |
| | Spine | - | - |
| | Stomach (axial / sagittal) | X | X |
| | Superior vena cava | - | X |
| | Tricuspid valve | - | X |
| | Umbilical vein | - | X |
| | Aorta on LVOT View | - | X |
| | Aorta on RVOT or 3 vessels view | - | X |
| | Pulmonary artery trunk on 3 vessels View | - | X |
| | Pulmonary artery with visible bifurcation | - | X |
| | Portal sinus | - | X |
| | Bowels | X | X |
| | Spleen | - | X |
| | Gallbladder | - | X |
| | Bladder | X | X |
| | Trachea | - | - |
| | Liver | - | X |
| | Lungs | - | X |
| | Ribs | X | X |
| | Sagittal heart | X | - |
| | CRL/NT/Profile views & structures | | |
| CRL Profile/NT Corpus Callosum | Nasal bone | - | - |
| | Diencephalon | X | - |
Page 29-7
{10}
sonio
510(k) Premarket Notification Submission
| View name | Structures to be detected | First Trimester T1 | Second/Third Trimesters T2/T3 |
| --- | --- | --- | --- |
| | Fourth ventricle on NT view | X | - |
| | Nuchal translucency | - | - |
| | Palate | X | X |
| | Corpus Callosum | - | - |
| | Liquid space under the chin | X | - |
| | Midbrain tectum | - | - |
| | Choroid plexus on sagittal plane | - | - |
| | Cisterna magna on NT view | X | - |
| | Brainstem on NT view | X | - |
| | Tip of the mandible | - | - |
| Placenta views | | | |
| Placenta / Cervix | Cervix | - | X |
| | Maternal bladder | - | X |
| | Placenta | - | X |
| Spine views | | | |
| Sagittal spine | Cervical Spine | - | X |
| | Lumbar Spine | - | X |
| | Sacral Spine | - | X |
| Facial views | | | |
| Lips and nose
Orbits
Coronal face | Lenses | - | - |
| | Orbits | - | X |
| | Nose tip | - | X |
Page 29-8
{11}
sonio
# 510(k) Premarket Notification Submission
| View name | Structures to be detected | First Trimester T1 | Second/Third Trimesters T2/T3 |
| --- | --- | --- | --- |
| | Nare | - | X |
| | Upper lip | - | X |
| | Lower lip | - | X |
# List of localization that can be automatically identified (bounding box)
Bounding boxes can be automatically detected and displayed for the views and anatomical structures.
| View name | Structure name |
| --- | --- |
| Brain views & structures | |
| Choroid plexus | Thalami/cerebral peduncles on axial view |
| Thorax and Heart views & structures | |
| - | Left Ventricle |
| | Left Atrium |
| | Ribs |
{12}
Sonio
510(k) Premarket Notification Submission
sonio
Quality criteria are related to:
- The presence of an anatomical structure;
- The absence of an anatomical structure;
For instance the transthalamic view, at the second and third trimester, Sonio Detect can evaluate the following criteria.
Table 3: List of quality criteria that can be automatically verified by Sonio Detect
| Quality criteria | 1st trimester | 2nd/3rd Trimester |
| --- | --- | --- |
| Presence of the cavum septum pellucidum or of the pillars of the fornix | - | X |
| Presence of the cavum septum pellucidum | - | X |
| Presence of the Sylvian fissure | - | X |
| Absence of the cerebellum | - | X |
| Presence of the thalami | X | X |
Table 4 : List of characteristics that can be automatically verified by Sonio Detect
| View name | Characteristics | First trimester T1 | Second/Third Trimester T2/T3 |
| --- | --- | --- | --- |
| Genitalia view | Male | - | X |
| | Female | - | X |
| Placenta location | Anterior | - | X |
| | Posterior | - | X |
Page 29-10
{13}
Sonio
510(k) Premarket Notification Submission
sonio
# V. Indications for Use
Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria and characteristics of the views.
The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
Sonio Detect v3 and the predicate have similar intended use. Both devices are used as a concurrent aid to automatically detect fetal ultrasound views, automatically detect fetal anatomical structures within the views.
The indications for use of Sonio Detect v3 and the predicate differ in the clinical outcomes. Sonio Detect v3 outputs localization for views and anatomical structures whereas the predicate Sonio Detect v2 does not output localization.
However, these differences do not raise new questions regarding safety and effectiveness of the device when used as labeled.
# VI. Comparison of Technological Characteristics with the Predicate Device
Table 5 provides a comparison of the Technological Characteristics of Sonio Detect v3 to the predicate Sonio Detect v2 cleared in K240406.
Table 5: Comparison of technological characteristics
| Items | Predicate device: Sonio Detect v2 | Proposed device : Sonio Detect v3 |
| --- | --- | --- |
| Manufacturer name | Sonio | Sonio |
| Device name | Sonio Detect | Sonio Detect |
| Regulation Number | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System
21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System
21 CFR 892.2050 - Medical Image Management and Processing System | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System
21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System
21 CFR 892.2050 - Medical Image Management and Processing System |
| Product code | IYN (primary)
IYO, QIH (Secondary) | IYN (primary)
IYO, QIH (Secondary) |
Page 29-11
{14}
sonio
# 510(k) Premarket Notification Submission
| Items | Predicate device: Sonio Detect v2 | Proposed device : Sonio Detect v3 |
| --- | --- | --- |
| Features | - Sonio Detect automatically detects views
- Sonio Detect automatically detects anatomical structures within the supported views
- Sonio Detect automatically verifies the quality criteria and characteristics of the supported views. | - Sonio Detect automatically detect views
- Sonio Detect automatically detect anatomical structures within the supported views
- Sonio Detect automatically verifies the quality criteria and characteristics of the supported views.
- Sonio Detect automatically localizes views and anatomical structures |
| Algorithm Methodology | Artificial Intelligence
Lecture of biometrics
Colorimetry for 3D and Doppler | Artificial Intelligence |
| Platform | Secure cloud-based and stand-alone software compatible with ultrasound system from GE Medical, Samsung, Canon and Philips | Secure cloud-based and stand-alone software compatible with ultrasound system from GE Medical, Samsung, Canon and Philips |
Sonio Detect v3 and its predicate differs in the following:
- Features : Sonio Detect v3 automatically localizes views and anatomical structures.
- Algorithm Methodology : Sonio Detect v3 uses Artificial Intelligence and does not include lectures of biometrics and colorimetry for 3D and Doppler.
However, these differences do not raise new questions regarding safety and effectiveness of the device when used as labeled.
# VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
# Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions."
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Design Reviews
- Software Development Lifecycle
- Algorithm Verification (Algorithm internal validation)
- Software units verification
- Software verification
- Simulated use testing (Validation)
- Performance testing
- Cybersecurity testing
Page 29-12
{15}
Sonio
510(k) Premarket Notification Submission
^{}[]
sonio
## Bench Testing
Sonic conducted a standalone performance testing on a dataset of 22496 fetal ultrasound images. This global validation dataset was independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points.
The results of the standalone performance testing demonstrated that Sonio Detect v3 performs the following, as summarized in table below:
Page 29-13
{16}
Sonio
510(k) Premarket Notification Submission
sonio
Table 6: results of the standalone performance testing
| Items (fetal ultrasound views, anatomical structures and characteristics automatically detected) | Sensitivity PE (95%CI) | Specificity PE (95%CI) |
| --- | --- | --- |
| Automatic detection of 14 T1 fetal ultrasound images | 0.886 (0.876-0.898) | 0.981 (0.979-0.982) |
| Automatic detection of 40 T2/T3 fetal ultrasound images | 0.901 (0.896-0.905) | 0.993 (0.993-0.994) |
| Automatic localization of 1 view (T1) | mIoU: 0.777 (0.743-0.811) | |
| Automatic detection of 1 fetal brain anatomical structure on the view “Transthalamic” at T1 | 0.815 (0.751-0.871) | 0.938 (0.901-0.973) |
| Automatic detection of 7 fetal brain anatomical structures on the views “Transthalamic”, “Transventricular”, “Transcerebellar” at T2/T3 | 0.941 (0.934-0.947) | 0.951 (0.943-0.958) |
| Automatic detection of 8 fetal thorax and heart anatomical structures on the views “4 chambers”, “LVOT”, “RVOT”, “Three vessels”, “Three vessels and trachea”, “Abdominal Circumference”, “Axial view of the kidneys”, “Diaphragm”, “Abdominal cord insertion” at T1 | 0.872 (0.848-0.892) | 0.921 (0.913-0.928) |
| Automatic detection of 24 fetal thorax and heart anatomical structures on the views “Four chambers”, “LVOT”, “RVOT”, “Three vessels”, “Three vessels and trachea”, “Abdominal Circumference”, “Axial view of the kidneys”, “Diaphragm”, “Abdominal cord insertion” at T2/T3 | 0.914 (0.906-0.920) | 0.963 (0.961-0.965) |
| Automatic detection of 3 fetal placenta anatomical structures on the view “Placenta / Cervix” at T2/T3 | 0.887 (0.874-0.899) | 0.962 (0.950-0.972) |
Page 29-14
{17}
Page 29-15
# 510(k) Premarket Notification Submission
| Items (fetal ultrasound views, anatomical structures and characteristics automatically detected) | Sensitivity PE (95%CI) | Specificity PE (95%CI) |
| --- | --- | --- |
| Automatic detection of 6 fetal Sagittal Fetus anatomical structures on the views “Crown Rump Length”, “Profile” at T1 | 0.869 (0.849-0.896) | 0.848 (0.822-0.875) |
| Automatic detection of 1 fetal Sagittal Fetus anatomical structures on the views “Profile” at T2/T3 | 0.883 (0.852-0.913) | 0.800 (0.754-0.842) |
| Automatic detection of 5 fetal Coronal Face anatomical structures on the views “Lips and nose”, “Orbits”, “Coronal face” at T2/T3 | 0.922 (0.897-0.947) | 0.901 (0.883-0.923) |
| Automatic detection of 3 fetal Spine anatomical structures on the view “Sagittal Spine” at T2/T3 | 0.839 (0.818-0.862) | 0.852 (0.829-0.873) |
| Automatic localization of 1 brain anatomical structures (on T1) | mIoU: 0.683 (0.632-0.734) | |
| Automatic localization of 3 thorax and heart anatomical structures (on T1) | mIoU: 0.679 (0.653-0.705) | |
| Automatic detection of the Anterior placenta location for the view “Placenta / Cervix” at T2/T3 | 0.925 (0.894-0.949) | 0.918 (0.885-0.948) |
| Automatic detection of the Posterior placenta location for the views “Placenta / Cervix”at T2/T3 | 0.918 (0.885-0.948) | 0.925 (0.894-0.949) |
| Automatic detection of the “Female sex” for fetal sex for the view “External Genitalia” at T2/T3 | 1.000 (1.000-1.000) | 0.985 (0.969-1.000) |
| Automatic detection of the “Male sex” for fetal sex for the view “External Genitalia” at T2/T3 | 0.985 (0.969-1.000) | 1.000 (1.000-1.000) |
{18}
Sonio
510(k) Premarket Notification Submission
sonio
Additionally, the performance for the detection of views and structures was also validated for subgroups including: Ultrasound machine manufacturer, BMI, maternal age, confounding cases, Image quality, geography, gestational age and race/ethnicity when appropriate.
Sonic Detect was validated only with GE, Canon, Philips and Samsung Ultrasound devices and is intended only to be used with these Ultrasound vendors.
The results of verification and performance testing demonstrate the safe and effective use of Sonio Detect.
## Clinical Study
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
Page 29-16
{19}
Sonio
510(k) Premarket Notification Submission
sonio
# VIII. Conclusions
Sonio Detect intended users, intended use and clinical applications are similar to those of the predicate device Sonio Detect v2, the cleared version in K240406.
The clinical outcome differences identified and discussed in Section II do not raise different questions of safety and effectiveness of the device.
The features and algorithm methodology differences identified and discussed in Section III do not raise different questions of safety and effectiveness of the device.
Furthermore, results of successful verification and validation activities and additional bench performance testing do not raise any new issue regarding the safety and effectiveness of the device.
Thus, Sonio Detect is substantially equivalent to its predicate Sonio Detect v2 (K240406).
Page 29-17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.