K252396 · Avanutri Equipamentos DE Saúde · IRP · May 29, 2026 · Physical Medicine
Device Facts
Record ID
K252396
Device Name
Cryo Sport 2.0 (AVA 501)
Applicant
Avanutri Equipamentos DE Saúde
Product Code
IRP · Physical Medicine
Decision Date
May 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cryo Sport 2.0 combines cold and compression therapies. It is intended to treat post-surgical and acute injuries to reduce edema, swelling and pain for which cold and compression is indicated. It is intended to treat post-traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy are indicated. It is intended to be used by, or on the order of, licensed healthcare professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
Device Story
Cryo Sport 2.0 is a portable, AC/DC-powered physiotherapy device for musculoskeletal injury recovery. It circulates ice-chilled water through a flexible wrap applied to the patient's body while simultaneously delivering pneumatic compression. The system comprises a Water Circulation Unit (WCU) with a reservoir, pump, air compressor, and lithium-ion battery; a connection hose; and a wrap. Operated by healthcare professionals in clinics, the device uses a 2.8-inch LCD panel to control treatment time, compression pressure (0–75 mmHg), and water flow. By extracting heat from tissue via cold water and applying compression, the device aims to reduce edema, swelling, and pain. The portable design allows for flexible clinical use, while the battery enables cordless operation. Healthcare providers monitor the patient's response during therapy sessions, which are adjustable up to 120 minutes.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software V&V, biocompatibility (ISO 10993-1), and environmental/transport testing. Performance metrics verified: compression accuracy (±25%), water flow rate (±25%), thermometer accuracy (±2°C), and battery runtime (~4 hours). All tests passed.
Technological Characteristics
Class II, Type B applied part. Materials: biocompatible wrap/hose (ISO 10993). Cooling: ice-water reservoir. Compression: pneumatic pump (0–75 mmHg). Power: 110–220 VAC or 11.1V Li-ion battery. Connectivity: standalone. Software: firmware-controlled (Moderate Level of Concern). Sterilization: non-sterile, applied over intact skin/dressings.
Indications for Use
Indicated for patients with post-surgical, post-traumatic, or acute musculoskeletal injuries requiring cold and compression therapy to reduce edema, swelling, and pain. For use by licensed healthcare professionals in clinical settings. Prescription use only.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
Avanutri Equipamentos DE Saúde
% Bruno Milhoci
Consultant
Bhp Consulting, LLC
Rua Santo Antônio, Nº 2262
Tres Rios, RJ 25820-185
Brazil
Re: K252396
Trade/Device Name: Cryo Sport 2.0 (AVA 501)
Regulation Number: 21 CFR 890.5650
Regulation Name: Powered Inflatable Tube Massager
Regulatory Class: Class II
Product Code: IRP
Dated: May 5, 2026
Received: May 5, 2026
Dear Bruno Milhoci:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252396 - Bruno Milhoci
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252396 - Bruno Milhoci
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
| Submission Number (if known) |
| --- |
| K252396 |
| Device Name |
| Cryo Sport 2.0 (AVA 501) |
| Indications for Use (Describe) |
| The Cryo Sport 2.0 combines cold and compression therapies. It is intended to treat post-surgical and acute injuries to reduce edema, swelling and pain for which cold and compression is indicated. It is intended to treat post-traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy are indicated. It is intended to be used by, or on the order of, licensed healthcare professionals in rehabilitation facilities, outpatient clinics, and athletic training settings. |
| Type of Use (Select one or both, as applicable) |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) Summary
510(k) Number: K252396
Prepared on: May 20, 2026
# Contact Details 21 CFR 807.92(a)(1)
Applicant
| Applicant Name | Avanutri Equipamentos de Saude |
| --- | --- |
| Applicant Address | Rua Santo Antonio, no 2262, Triangulo, CEP 25820-185, Tres Rios, RJ, Brazil |
| Applicant Telephone | +55 24 22522468 |
| Applicant Contact | Mrs. Daiana Vieira |
| Applicant Contact Email | regulatorio@avanutri.com.br |
Correspondent
| Correspondent Name | BHP Consulting LLC |
| --- | --- |
| Correspondent Address | Antonio Machado Santtanna, Ribeirao Preto, SP 14022310, Brazil |
| Correspondent Telephone | +55 14 981229294 |
| Correspondent Contact | Mr. Bruno Milhoci |
| Correspondent Email | bruno@bhpconsulting.net |
Device Name 21 CFR 807.92(a)(2)
| Device Trade Name | Cryo Sport 2.0 (AVA501) |
| --- | --- |
| Common Name | Powered inflatable tube massager |
| Classification Name | Massager, Powered Inflatable Tube |
| Regulation Number | 21 CFR 890.5650 |
| Product Code(s) | IRP, ILO |
| Regulatory Class | Class II |
| Classification Panel | Physical Medicine (89) |
{5}
Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
| 510(k) # | Trade Name (Primary listed first) | Owner / Sponsor | Product Code |
| --- | --- | --- | --- |
| K171685 | Med4 Elite™ (Primary Predicate) | CoolSystems, Inc. (dba Game Ready) | IRP, ILO |
None of the predicate devices listed above have been subject to design recalls.
# Device Description Summary 21 CFR 807.92(a)(4)
The Cryo Sport 2.0 (Model AVA501) is a portable, AC/DC-powered physiotherapy device that combines cryotherapy (cold water therapy) with adjustable pneumatic compression for the treatment of musculoskeletal injuries, post-surgical recovery, and acute soft tissue conditions. The device circulates cooled water through a flexible wrap applied to the affected anatomical area, simultaneously delivering controlled compression via an integrated air compressor.
The system consists of three primary components: (1) a Water Circulation Unit (WCU) housing the reservoir, water pump, air compressor, control electronics, and internal lithium-ion battery; (2) a Connection Hose linking the WCU to the wrap; and (3) a flexible Wrap (heat exchanger sleeve) applied directly to the patient's body part. The WCU incorporates a 2.8-inch color LCD control panel with push-button controls for treatment time, compression pressure, water flow rate, and session start/pause.
Cooling is achieved by filling the reservoir with ice and distilled or filtered water. The circulating cold water extracts heat from the treated tissue via the wrap. Temperature is modulated by adjusting the ice-to-water ratio. Compression is delivered via an integrated pneumatic pump that inflates the wrap bladder to the selected pressure level, in continuous or intermittent mode.
# Key Technical Specifications
- Supply voltage: 110–220 VAC, 50/60 Hz (external 12 VDC / 4.0 A adapter)
- Internal battery: 11.1 V, 4400 mAh lithium-ion (approx. 4 hours autonomy)
- Treatment time: 0–120 minutes (5-minute increments); Default: 30 minutes
- Compression range: 0–75 mmHg (continuous and intermittent modes)
- Water flow settings: Low / Medium / High (53–64 L/h)
- Dimensions (WCU): 449 mm x 102 mm x 215 mm; Weight: 5.0 kg (system, without water/ice)
- Electrical classification: Class II, Type B applied part (IEC 60601-1)
- Operating conditions: 0°C to 40°C, 30–90% RH non-condensing
The device is intended for use in professional healthcare environments under the supervision of a licensed healthcare professional. It is not intended for home use and is classified as Prescription Use Only.
Intended Use / Indications for Use 21 CFR 807.92(a)(5)
{6}
# Indications for Use Statement
The Cryo Sport 2.0 combines cold and compression therapies. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post-traumatic and post-surgical medical and/or surgical conditions for which localized cold thermal therapy is indicated. It is intended to be used by, or on the order of, licensed healthcare professionals in rehabilitation facilities, outpatient clinics, and athletic training settings.
Prescription Use Only (21 CFR 801 Subpart D).
# Indications for Use Comparison with Predicate
The Cryo Sport 2.0 and the predicate device Med4 Elite™ (K171685) share the same fundamental intended use: both devices are indicated for the treatment of post-surgical and acute musculoskeletal injuries through the delivery of cold therapy and pneumatic compression to reduce edema, swelling, and pain. Both devices are intended for use exclusively by licensed healthcare professionals in professional rehabilitation and clinical settings, and both are Prescription Use Only.
The primary difference in indications for use is that the Med4 Elite™ additionally offers heat therapy and contrast (alternating hot/cold) therapy, while the Cryo Sport 2.0 is limited to cold and compression therapy. This represents a narrower scope of indications — the Cryo Sport 2.0 indications are a proper subset of the predicate's cleared indications. The absence of heat and contrast therapy modalities does not broaden the indicated population, change the indicated anatomical areas, or introduce any new patient safety considerations.
Based on the comparison above, the indications for use of the Cryo Sport 2.0 are substantially equivalent to those of the predicate device.
# Technological Comparison 21 CFR 807.92(a)(6)
The Cryo Sport 2.0 and the Med4 Elite™ (K171685) are both Class II physiotherapy devices indicated for cold and compression therapy. Both devices deliver cold water through a flexible wrap applied to the patient's body via a circulating pump system, and both incorporate an air compressor to deliver adjustable pneumatic compression. The core therapeutic mechanism is identical between subject and predicate.
The table below provides a complete comparison of all key technological characteristics, identifying differences and providing justification that no new safety or effectiveness concerns are raised:
| Characteristic | Cryo Sport 2.0 (AVA501) [Subject Device] | Med4 Elite™ (K171685) [Predicate Device] | Comparison and Justification |
| --- | --- | --- | --- |
| 1. Indications for Use (IFU) | | | |
| Intended Use | "The CRYO SPORT 2.0 system combines cold and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling and pain for which cold and compression are indicated. It is | Combines cold, heat, contrast, and compression therapies. Intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. Intended to treat post- | The Cryo Sport 2.0 shares the same core intended use as the predicate: cold and compression therapy for post-surgical and acute injuries in a professional healthcare setting. The predicate additionally provides heat and contrast therapy; the Cryo Sport 2.0 |
{7}
| | intended to treat post-traumatic and post-surgical conditions for which localized cold therapy is indicated." | traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot, cold, or contrast) are indicated. Intended for use by licensed healthcare professionals in rehabilitation facilities, outpatient clinics, and athletic training settings. Prescription use only. | does not include these modalities, making it a subset of the predicate's indications. The absence of heat/contrast therapy does not raise new safety concerns since the device is limited in scope. Both devices are prescription-only and intended for the same user population and use environment. The IFU is substantially equivalent. |
| --- | --- | --- | --- |
| Intended Users | Licensed healthcare professionals only (prescription use only). | Licensed healthcare professionals only (prescription use only). | Identical — substantially equivalent. |
| Use Environment | Indoor professional healthcare facilities: rehabilitation clinics, outpatient clinics, athletic training settings. Not suitable for home use. | Indoor professional healthcare facilities: rehabilitation clinics, outpatient clinics, athletic training settings. | Identical — substantially equivalent. |
| 2. Cooling System | | | |
| Cooling System Type | Ice-based system: requires ice and water placed in an onboard reservoir. Water is chilled by the ice and circulated through the wrap. Temperature is controlled by the user-adjusted water-to-ice ratio. Thermometer reading range: 1°C to 40°C. | Vapor compression (refrigeration) system. Two separate onboard reservoirs (cold and hot), each 1 gallon (3.8 L). Cold reservoir range: 34–50°F (1.1–10°C). Device maintains temperature automatically without ice. | The Cryo Sport 2.0 uses ice to chill water, while the predicate uses vapor compression refrigeration. Both systems generate cold water that is circulated through a wrap applied to the patient. The functional outcome is equivalent — cold water therapy applied via a wrap. The ice-based approach is well-established in clinical cold therapy (e.g., Game Ready Classic GR2, K072620, also uses ice). The difference in cooling mechanism does not raise new safety or effectiveness concerns. |
| Coolant / Fluid | Filtered or distilled water with ice. | Distilled water (cold reservoir); distilled water (hot reservoir). | Both devices use water as the circulating medium. No safety impact from this difference. |
| Temperature Range (Cold) | Temperature depends on ice/water ratio. Thermometer reads 1°C–40°C. Clinical literature guidance: greatest benefit in 4.5°C–15.5°C range. | Cold reservoir set-point range: 34–50°F (1.1–10°C). | Both devices deliver cold water in a clinically relevant range. The Cryo Sport 2.0 temperature is user-controlled (ice ratio), consistent with other ice-based predicates (e.g., K072620). No new safety concerns. |
{8}
3. Chilling Mechanism
| Chilling Mechanism | Ice. The reservoir is filled with water and ice by the operator. The melting ice chills the water, which is then pumped through the wrap to deliver cold therapy. | Vapor compression (mechanical refrigeration unit). An onboard compressor automatically maintains the set cold temperature without ice. | The technological approach differs (ice vs. vapor compression), but both produce chilled water for circulation through the wrap. Ice-based cold therapy is well-established and safe. The primary predicate device (Game Ready Classic, K072620) also uses ice as its chilling mechanism, confirming substantial equivalence. No new issues of safety or effectiveness are raised. |
| --- | --- | --- | --- |
| Heating Mechanism | Not applicable — no heat therapy function. | Resistance heaters (hot water reservoir, set-point range: 95–113°F / 35–45°C). | The Cryo Sport 2.0 does not offer heat therapy and therefore does not require a heating mechanism. Deliberate design choice consistent with the device's cold-and-compression-only indications for use. No safety concern. |
4. Power Source
| Mains Power Supply | External AC/DC adapter: input 110–220 VAC, 50/60 Hz; output 12 VDC, 4.0 A (42 VA max). Power supply provided by manufacturer. | Direct AC power: 100–240 VAC, 50/60 Hz. Large internal electronics including compressor, resistance heaters, pumps, and touch-screen computer. | Both devices accept universal input voltage and operate at 50/60 Hz. The Cryo Sport uses a low-voltage external adapter (12 VDC), which reduces electrical risk at the device level. The lower power design (42 VA maximum) is an improvement in electrical simplicity, not a safety concern. |
| --- | --- | --- | --- |
| Battery (Internal) | Yes — Lithium-ion battery (11.1 V nominal, 4400 mAh, ~48.84 Wh). Provides approximately 4 hours of operation. Charging time: ~3 hours. Enables cordless / portable use. | No internal battery. AC power required at all times. | The Cryo Sport 2.0 includes an internal rechargeable battery to enable portable/cordless operation. The predicate does not have this feature. The addition of a battery does not raise safety concerns — the battery (LYS 18650 lithium cell) is a standard component, and the device complies with IEC 60601-1 for battery-powered operation. This feature adds user convenience without affecting the therapeutic function or safety profile. |
| Electrical Safety Classification | Class II equipment, Type B applied part (IEC 60601-1). IPX0. | Class II equipment, Type B applied part (ANSI/AAMI ES60601-1:2005/(R)2012; CAN/CSA C22.2 No. 60601-1:2014). | Both devices are Class II, Type B — substantially equivalent. Both comply with IEC 60601-1 (3rd edition) and its national deviations. |
{9}
| 5. Technical Standards Compliance | | | |
| --- | --- | --- | --- |
| **Electrical Safety** | IEC 60601-1: Medical electrical equipment — General requirements for basic safety and essential performance (3rd edition and applicable national deviations). | ANSI/AAMI ES60601-1:2005/(R)2012; CAN/CSA C22.2 No. 60601-1:2014 (U.S. and Canadian national deviations of IEC 60601-1 3rd edition). | Both devices comply with IEC 60601-1 (3rd edition) and applicable national variants. Substantially equivalent. |
| **Electromagnetic Compatibility** | IEC 60601-1-2: Electromagnetic disturbances — Requirements and tests. | IEC 60601-1-2 (collateral standard for EMC, as required by IEC 60601-1 3rd edition). | Both comply with IEC 60601-1-2 for EMC. Substantially equivalent. |
| **Biocompatibility** | ISO 10993-1: Biological evaluation of medical devices. Patient-contacting components evaluated per applicable ISO 10993 parts (-5, -10, -12). | ISO 10993-1, -5 (cytotoxicity), -10 (sensitization), -12. Wraps and connector hose exteriors tested and found biocompatible. | Both devices evaluate biocompatibility per ISO 10993-1 framework. The Cryo Sport wraps are surface devices contacting intact skin; the same biocompatibility approach applies as for the predicate. Substantially equivalent. |
| **Quality Management System** | ISO 13485: Medical devices — Quality management systems. 21 CFR Part 820 (FDA Quality System Regulation). | 21 CFR Part 820 (FDA Quality System Regulation) / ISO 13485. | Both devices operate under a compliant quality management system. Substantially equivalent. |
| **Software** | Firmware Version AVA501 V.1.0.0. Controls timing, pressure, and flow. Moderate level of concern (per FDA software guidance). | Software-controlled multimodality device with touch-screen interface. Moderate level of concern per FDA guidance. Full software V&V submitted. | Both devices contain embedded software controlling therapy delivery. Both classified as Moderate Level of Concern. No new safety concerns from software differences. |
| 6. Physical Characteristics | | | |
| **Dimensions** | Water Circulation Unit: 449 mm (L) x 102 mm (W) x 215 mm (H). Compact portable form factor. Supplied with carrying bag. | 83 cm (L) x 63 cm (W) x 109 cm (H). Large floor-standing unit with caster wheels. | The Cryo Sport 2.0 is significantly more compact than the predicate. The size difference is attributable to the predicate’s dual-patient capability, onboard refrigeration/heating system, and larger water reservoirs. The Cryo Sport is designed for single-patient use with ice-based cooling, requiring no large mechanical systems. The difference in size does not affect safety or clinical performance. |
| **Weight** | 5.0 kg (water circulation unit + wrap + hose + power supply + cable, without water/ice). Portable. | 78 kg (172 lbs). Non-portable; unit has caster wheels for repositioning. | The Cryo Sport 2.0 is substantially lighter, reflecting its portable, single-patient, ice-based design. The weight difference does not affect the therapeutic function (cold/compression |
{10}
| | | | delivery via wrap). Portability is a design feature, not a safety concern. The lightweight design may reduce ergonomic risk for operators. |
| --- | --- | --- | --- |
| Mobility | Stationary during use (stable surface required), but portable — can be carried to different locations. Supplied with carrying bag. | Clinic-stationary unit on caster wheels. Not designed for portability outside the clinical setting. | Both devices are intended for use in a fixed position during operation. The Cryo Sport's portability allows transport between settings. No safety impact. |
| Number of Patients | Single-patient use. | One or two patients simultaneously, or two anatomical locations on one patient. | The Cryo Sport 2.0 treats one patient at a time, consistent with its design as a single-channel device. The predicate's dual-patient capability is an additional feature not claimed by the Cryo Sport. This difference does not raise safety concerns — it is a scope limitation, not a technological deficiency. |
| Chilling Reservoir Capacity | Single reservoir for ice and water (user-filled). Filled to labeled fill line. Consumables: ice and filtered/distilled water. | Cold reservoir: 1 gallon (3.8 L). Hot reservoir: 1 gallon (3.8 L). Iceless — filled with distilled water. | Both devices hold sufficient fluid for a standard therapy session. The Cryo Sport replenishes cold by adding ice; the predicate maintains temperature via mechanical refrigeration. No safety difference. |
| 7. Compression Levels and Settings | | | |
| Compression Available | Yes — intermittent and continuous pneumatic compression via wrap. | Yes — intermittent pneumatic compression via wrap. | Both devices deliver compression therapy through wraps applied to the body. Substantially equivalent. |
| Compression Pressure Levels | No pressure — 0 mmHg Low (continuous) — 25 mmHg Average (continuous) — 50 mmHg High (continuous) — 75 mmHg Low (intermittent) — 25/12.5 mmHg (30s on/off) Average (intermittent) — 50/25 mmHg (30s on/off) High (intermittent) — 75/37.5 mmHg (30s on/off) Range: 0–75 mmHg. Resolution: 25 mmHg. Accuracy: ±25%. | Low: 5–15 mmHg Medium-Low: 5–30 mmHg Medium: 5–50 mmHg High: 5–75 mmHg (No-compression option also available) | Both devices offer compression pressure levels up to a maximum of 75 mmHg. The Cryo Sport 2.0 provides both continuous and intermittent compression modes with discrete preset levels (0, 25, 50, 75 mmHg), while the predicate offers a variable range (5–75 mmHg) across four levels. Both devices offer equivalent maximum pressure (75 mmHg) and include a no-compression option. The discrete preset structure of the Cryo Sport simplifies clinical use and does not raise new safety concerns. Maximum compression of 75 mmHg is consistent |
{11}
| | | | with both the primary and secondary predicates. |
| --- | --- | --- | --- |
| Compression Mode | Both continuous and intermittent modes available. Intermittent: 30 seconds on at set pressure, 30 seconds at 50% of set pressure. | Intermittent pneumatic compression (IPC). Compression cycle not specified in 510(k) summary. | The Cryo Sport 2.0 explicitly supports both continuous and intermittent compression, providing greater clinical flexibility. This does not raise new safety concerns — similar modes are present in referenced devices (e.g., VascuTherm, K061866). No safety impact. |
## 8. Treatment Time
| Cold Therapy Treatment Time | Adjustable: 0 to 120 minutes. Resolution: 5-minute increments. Default: 30 minutes. | 5 to 60 minutes. Default: 15 minutes. | The Cryo Sport 2.0 allows up to 120 minutes of cold therapy, which exceeds the predicate's 60-minute maximum. The extended time range is consistent with clinical practice guidelines for cold therapy (literature recommends up to 30 minutes per session, multiple times daily). The device is used only by licensed healthcare professionals who monitor patient response, and the user manual includes explicit warnings about monitoring skin during cold therapy. A 120-minute maximum does not introduce new safety concerns. |
| --- | --- | --- | --- |
| Compression-Only Treatment Time | Adjustable: 0 to 120 minutes. Default: 30 minutes. | 5 to 60 minutes. Default: 15 minutes. | The longer maximum treatment time for compression is consistent with established clinical practice for post-surgical compression. The device is intended for use under healthcare professional supervision. No new safety concerns. |
| Heat Therapy Treatment Time | Not applicable — no heat therapy function. | 5 to 30 minutes. Default: 15 minutes. | Not applicable to Cryo Sport 2.0, which does not include heat therapy. |
| Contrast Therapy Treatment Time | Not applicable — no contrast therapy function. | Total 15–90 minutes (default 30 min); Heat cycle: 1–10 min; Cold cycle: 1–10 min. | Not applicable to Cryo Sport 2.0, which does not include contrast therapy. |
| Timer Resolution / Default | 5-minute resolution. Default: 30 minutes. | Varies by modality (5-minute increments). Default: 15 minutes. | Both devices use incrementally adjustable timers. The Cryo Sport's default of 30 minutes aligns with clinical recommendations for cold therapy. No safety concern. |
{12}
| 9. Operating and Storage Conditions | | | |
| --- | --- | --- | --- |
| Operating Temperature | 0°C to 40°C (32°F to 104°F). | 50°F to 90°F (10°C to 32°C). | The Cryo Sport 2.0 operates across a wider range (0–40°C) than the predicate (10–32°C). Both are adequate for indoor professional clinical use. No safety concern. |
| Storage Temperature | 1°C to 50°C. | 33°F to 122°F (1°C to 50°C). | Identical — substantially equivalent. |
| Relative Humidity | Operating: 30–90% non-condensing. Storage: 30–90% non-condensing. | Operating: 30–90% non-condensing. Storage: 10–95% non-condensing. | Operating humidity ranges are identical. Storage range of Cryo Sport is slightly narrower than predicate; both are adequate for standard healthcare storage environments. No safety concern. |
| Atmospheric Pressure | 80 kPa to 102 kPa (transport/storage). | 0–9,842 ft (0–3,000 m) altitude equivalent. | Equivalent operating ranges suitable for standard healthcare facility environments. Substantially equivalent. |
| 10. Accessories / Patient-Contacting Components | | | |
| Wraps / Garments | One wrap included; various wrap types available for different anatomical areas. Applied to intact skin or over dressing. | Various anatomical wraps: knee (straight/articulated), elbow (standard/flexed), ankle, shoulder, back, hip-groin, hand-wrist, half-leg boot. Applied to intact skin or over sterile dressings. | Both devices use flexible garments/wraps to deliver cold and compression therapy. Patient-contacting materials must demonstrate biocompatibility per ISO 10993-1. Both are provided non-sterile and applied over intact skin or dressings. Substantially equivalent approach. |
| Sterility | Non-sterile. Not intended to be user-sterilized. Applied over intact skin or dressings. | Non-sterile. Not intended to be user-sterilized. Applied over intact skin or sterile dressings. | Identical approach — substantially equivalent. |
| Biocompatibility of Patient-Contacting Parts | Evaluated per ISO 10993-1. Surface contact with intact skin (limited/prolonged duration). Cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and other applicable tests performed. | ISO 10993-1, -5, -10, -12. Wraps and connector hose exteriors tested — found biocompatible for skin contact. | Both devices follow the same biocompatibility evaluation framework for patient-contacting components. Substantially equivalent. |
None of the performance or technological differences identified between the Cryo Sport 2.0 and the predicate Med4 Elite™ raise any new questions of safety or effectiveness. Each difference has been reviewed in the context of the device's intended use, design, applicable standards, and existing predicate data.
{13}
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The Cryo Sport 2.0 (AVA501) has been subjected to a comprehensive program of design verification and validation testing covering electrical safety, electromagnetic compatibility (EMC), software validation, system/bench performance testing, biocompatibility, and environmental/transport testing. The results confirm that the device is safe and effective for its intended use and is substantially equivalent to the predicate device.
## A. Electrical Safety Testing
Tested for compliance with IEC 60601-1 (3rd edition) and applicable national deviations. Testing included:
- Dielectric strength (hipot) testing of mains insulation
- Leakage current measurements (earth leakage, enclosure leakage, patient auxiliary current)
- Protective earth resistance verification
- Temperature rise testing under normal and single-fault conditions
- Battery safety testing per IEC 60601-1 Clause 14
Result: All electrical safety tests passed. The device meets IEC 60601-1 requirements for Class II, Type B applied part equipment.
## B. Electromagnetic Compatibility (EMC)
EMC testing conducted per IEC 60601-1-2 (4th edition). Tests included radiated/conducted emissions (CISPR 11), ESD immunity (IEC 61000-4-2), radiated RF immunity (IEC 61000-4-3), EFT/burst (IEC 61000-4-4), surge (IEC 61000-4-5), conducted RF immunity (IEC 61000-4-6), magnetic field immunity (IEC 61000-4-8), and voltage dips/interruptions (IEC 61000-4-11).
Result: The device met all applicable emissions limits and demonstrated acceptable performance under all immunity tests.
## C. Software Verification and Validation (V&V)
Firmware AVA501 V.1.0.0 classified as Moderate Level of Concern per FDA guidance. Documentation included software requirements specification (SRS), software design specification (SDS), unit/integration/system testing, hazard analysis (ISO 14971), boundary and stress testing of all therapy control algorithms, and alarm/fault condition verification.
Result: All software V&V activities passed.
## D. System and Bench Performance Testing
- Compression accuracy: Pressure delivery verified at all preset levels (0, 25, 50, 75 mmHg continuous; intermittent modes). Measured values within ±25% of set value as specified.
- Water flow rate: Flow measured at Low, Medium, and High settings. Values confirmed within ±25% of nominal (53, 56, 64 L/h respectively).
- Temperature thermometer accuracy: Confirmed at ±2°C across the 1°C–40°C operating range.
- Battery runtime: Approximately 4 hours of continuous operation on full charge confirmed.
- Timer accuracy: Confirmed ±30 seconds accuracy across the 5–120 minute operating range.
- Reservoir integrity: No leakage observed under fill-line conditions during extended operation.
Result: All system performance parameters met specification.
{14}
E. Biocompatibility
Patient-contacting components (wrap/heat exchanger sleeve) evaluated per ISO 10993-1 for surface contact with intact skin (limited/prolonged duration). Testing performed:
- Cytotoxicity (ISO 10993-5): Pass
- Sensitization (ISO 10993-10, Guinea Pig Maximization Test): Pass
- Irritation / skin irritation (ISO 10993-10): Pass
- Test sample preparation per ISO 10993-12
Result: All patient-contacting components demonstrated biocompatibility for the intended contact type and duration.
F. Cleaning and Disinfection Validation
Cleaning and disinfection procedures validated for compatibility with specified cleaning agents over the expected device service life (5 years). Wraps confirmed to tolerate repeated cleaning with mild detergent/antibacterial soap and cold water without degradation of materials or performance.
Result: Cleaning and disinfection procedures are adequate and validated.
G. Environmental and Transport Testing
Device tested under environmental stress conditions consistent with specified storage and operating conditions per IEC 60068 (temperature cycling, humidity exposure, transport/vibration simulation). The device performed within specification following all environmental tests.
H. Risk Management
A comprehensive risk management process was conducted per ISO 14971. All identified hazards were analyzed, and residual risks were determined to be acceptable in the context of the intended clinical benefit. No residual risks requiring special labeling beyond those already included in the Instructions for Use were identified.
Conclusion of Substantial Equivalence 21 CFR 807.92
Based on the information provided in this 510(k) Summary, Avanutri Equipamentos de Saude concludes that the Cryo Sport 2.0 (AVA501) is substantially equivalent to the predicate device Med4 Elite™ (K171685, CoolSystems, Inc. dba Game Ready) with respect to both indications for use and technological characteristics, in accordance with Section 513(i) of the Federal Food, Drug, and Cosmetic Act and 21 CFR 807.92.
- The Cryo Sport 2.0 has the same intended use as the predicate: treatment of post-surgical and acute injuries to reduce edema, swelling, and pain through cold and compression therapy, for use by licensed healthcare professionals in professional clinical settings.
- The technological differences identified (ice-based cooling, portable/battery-powered design, compact form factor, single-patient use, cold-and-compression-only modalities) do not raise new questions of safety or effectiveness.
- The device has been subjected to comprehensive non-clinical testing (electrical safety, EMC, software V&V, bench performance, biocompatibility, environmental) with all tests passing applicable requirements.
- All patient-contacting components have been demonstrated to be biocompatible for the intended contact type and duration.
- The device complies with IEC 60601-1 (Class II, Type B), IEC 60601-1-2, ISO 10993-1, ISO 13485, and ISO 14971.
{15}
The Cryo Sport 2.0 (AVA501) is safe and effective for its intended use and is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.