It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. It is intended for use by patients 18 years or older.
Device Story
Rhythm Master ECG Patch (HM-15BB-AX, HM-15BW-AX) is a lightweight (6.8g) wearable cardiac monitor for continuous ECG recording (up to 3 or 7 days). Device captures ECG signals via single-use electrodes; processes signals through an AC ECG channel to filter baseline drift. HM-15BB-AX transmits data via Bluetooth Low Energy to Android app; HM-15BW-AX transmits via USB-C to PC. Device computes heart rate (HR) and heart rate variability (HRV) from QRS detection; includes accelerometer for motion reference. Healthcare professionals initialize the device; patients wear it in home settings. Output (ECG waveforms, HR, HRV, motion data) is viewed by clinicians on paired software for diagnostic interpretation. Device does not provide automated ECG analysis or real-time monitoring; intended for retrospective diagnostic review to aid clinical decision-making regarding cardiac symptoms.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1, 60601-2-47), wireless coexistence (ANSI C63.27), and software verification/validation (IEC 62304).
Technological Characteristics
Type CF cardiac floating device; 3.6V Li-ion battery; 0.5-40Hz frequency response; 15-bit resolution; >10GOhms input impedance. Dimensions: 37.0x22.5x7.7mm. Connectivity: BLE (HM-15BB-AX) or USB 2.0 (HM-15BW-AX). Standards: IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-47, ISO 10993-1/5/10/23, ANSI/AAMI EC12. Software: Basic documentation level, non-automated signal processing.
Indications for Use
Indicated for patients 18+ years old, asymptomatic or experiencing transient symptoms (palpitations, dyspnea, dizziness, lightheadedness, pre-syncope, syncope, fatigue, chest pain, anxiety).
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 23, 2026
Smwmed Inc.
% Youshan Gong
RA Specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K252389
Trade/Device Name: Rhythm Master ECG Patch (HM-15BB-AX, HM-15BW-AX, HM-15BW-DX)
Regulation Number: 21 CFR 870.2800
Regulation Name: Medical Magnetic Tape Recorder
Regulatory Class: Class II
Product Code: DSH
Dated: July 29, 2025
Received: July 31, 2025
Dear Youshan Gong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252389 - Youshan Gong
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252389 - Youshan Gong
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252389 | |
| Device Name Rhythm Master ECG Patch (HM-15BB-AX, HM-15BW-AX) | |
| Indications for Use (Describe) It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. It is intended for use by patients 18 years or older. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Smwmed Inc.
K252389 -510(k) Summary
# K252389 510(k) Summary
## I. Submitter
Company name: Smwmed Inc.
Address: 9920 Pacific Heights Blvd Ste 410, San Diego CA 92121, United States of America
Contact person: Bo Pi
Title: CEO
Tel: +1 760 803 9092
Email: bosmwmed@smwmed.com
Date: April 22, 2026
## II. Subject Device
Name of Device: Rhythm Master ECG Patch
Model(s): HM-15BB-AX, HM-15BW-AX
Regulatory Class: II
Common Name: recorder, magnetic tape, medical
Regulation Name: Medical Magnetic Tape Recorder
Regulation Number: 870.2800
Product Code: DSH
Review Panel: Cardiovascular
## III. Predicate Devices
| No. | Manufacturer | Device name | Product code | 510(k) Number | Cleared Date |
| --- | --- | --- | --- | --- | --- |
| 1. | Wellysis Corp. | S-Patch Ex Wearable ECG Patch | DSH | K231289 | 08/30/2023 |
| 2. | Braemar Manufacturing, LLC | ePatch® | DSH | K171410 | 01/04/2018 |
## IV. Device Description
The Rhythm Master ECG Patch is a lightweight, compact cardiac monitoring device designed for continuous ECG recording for up to 3 days for the HM-15BB-AX model while the HM-15BW-AX can record for 7 days. This Type CF cardiac floating device features high-precision ECG monitoring with a dynamic range of $\pm 10\mathrm{mV}$ and excellent signal quality with a frequency response of $0.5 - 40\mathrm{Hz}$ for HM-15BB-AX, HM-15BW-AX.
The device weighs only 6.8 grams and measures $37.0 \times 22.5 \times 7.7 \mathrm{~mm}$, making it comfortable for extended wear. The device is powered by a rechargeable $3.6 \mathrm{~V}$ lithium-ion battery and HM-15BB-AX utilizes Bluetooth Low Energy technology for seamless data transmission to smartphones, HM-15BW-AX utilizes Type-C cable for seamless data transmission to computers. The device does not provide and is not intended for automated ECG analysis and/or use with third party analysis software. The device is not intended for real time monitoring.
The Rhythm Master ECG Patch is intended to be only applied and initialized by healthcare professionals prior to the patient wearing period. The patients can then wear the patch in home
{5}
Smwmed Inc.
K252389-510(k) Summary
environment. Before patients use the device in home setting, they shall receive training from healthcare professionals regarding the device intended use, contraindications, warnings and cautions with the device use.
The device offers signal processing mode: AC ECG channel. The AC ECG channel focuses on processing the dynamic components of ECG signals, enhancing the dynamic parts of the signal through a series of signal conditioning steps while filtering out unwanted baseline drift, making it suitable for routine ECG monitoring scenarios.
## Specific device features:
Heart Rate (HR): The device provides a calculated heart rate value as a derived parameter obtained from the detection of QRS complexes in the acquired ECG signal. It is intended solely for general reference and long-term trend observation, offering a summary indication of average heart rate over the recording period.
Heart Rate Variability (HRV): HRV is a secondary, computed parameter derived from the time intervals between successive QRS complexes (RR intervals). It provides a quantitative indication of variations in heart rate over time and is displayed for general reference and trend observation only. The HRV value is not intended for diagnostic or interpretive purposes.
Accelerometer (ACC): The ACC waveform provides motion reference data to allows users and physicians to understand whether the patient is in motion or at rest for reference to signal quality. It is for reference only and has no diagnostic function.
Viewing of ECG: The device allows for the viewing of recorded ECG waveforms on a paired mobile application or PC software. This is the primary functionality that enables a qualified healthcare professional to perform diagnostic interpretation.
The above mentioned specific device features are standard features in modern ambulatory ECG monitoring devices and are verified to comply with IEC 60601-2-47 requirements as evidence for substantial equivalence.
## V. Comparison of Intended use/ Indications for Use With the Predicate Device
It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. It is intended for use by patients 18 years or older. This intended use is similar to the predicate devices. Therefore, the Rhythm Master ECG Patch may be found substantially equivalent to its predicate devices.
## VI. Comparison of Technological Characteristics With the Predicate Device
The Rhythm Master ECG Patch has the same intended use as the predicate devices. The technological characteristics, features, specifications, materials are similar to the predicate devices. Any minor differences between the subject device and the listed predicate devices do not raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use.
Therefore, the Rhythm Master ECG Patch may be found substantially equivalent to its predicate devices.
{6}
Smwmed Inc.
K252389-510(k) Summary
| | Subject device | Primary predicate device | Predicate device 1 | Remark |
| --- | --- | --- | --- | --- |
| K number | K252389 | K231289 | K171410 | / |
| Trade name | Rhythm Master ECG Patch | S-Patch Ex Wearable ECG Patch | ePatch® | / |
| Manufacturer | Smwmed Inc. | Wellysis Corp. | Braemar Manufacturing, LLC | / |
| Model | HM-15BB-AX
HM-15BW-AX | S-Patch Ex | ePatch® | / |
| Device Classification Name | Medical Magnetic Tape Recorder | Medical Magnetic Tape Recorder | Medical Magnetic Tape Recorder | Same |
| Review panel | Cardiovascular | Cardiovascular | Cardiovascular | Same |
| Regulation number | 21 CFR 870.2800 | 21 CFR 870.2800 | 21 CFR 870.2800 | Same |
| Device class | II | II | II | Same |
| Product code | DSH | DSH | DSH | Same |
| Indication for use/Intended use | It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. Rhythm Master ECG Patch is intended for use by patients 18 years or older. | S-Patch Ex wearable ECG patch is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. S-Patch Ex wearable ECG patch is intended for use by patients 18 years or older. | The ePatch® is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. The ePatch® is intended for use by adolescents 18-21 and adults. | Same |
| Location for use | Prescription Use | Prescription Use | Prescription Use | Same |
| Target population | patients 18 years or older. | S-Patch Ex wearable ECG patch is intended for use by patients 18 years or older. | The ePatch® is intended for use by adolescents 18-21 and adults. | Same |
| Single Use | Reusable / Rechargeable Monitor; Single Use Electrodes | S-Patch Ex is multi-patient, multi-use; compatible 3rd-party ECG electrode is single use | Reusable / Rechargeable Monitor, Single Use Electrodes | Same |
| Intended Use | Home & Healthcare settings | Home & Healthcare settings | Home & Healthcare settings | Same |
3 / 8
{7}
Smwmed Inc.
K252389-510(k) Summary
| | Subject device | Primary predicate device | Predicate device 1 | Remark |
| --- | --- | --- | --- | --- |
| K number | K252389 | K231289 | K171410 | / |
| Environment | | | | |
| Measured Parameters | Yes | Yes | Yes | Same |
| ECG Dynamic Range | -10mV to + 10mV | -10mV to + 10mV | -10mV to + 10mV | Same |
| Applied Part Category | Type CF (cardiac floating) | Type CF (cardiac floating) | Type BF (body floating) | Same |
| Battery | LIR1654 3.6V 110mAh
Rechargeable lithium-ion button battery | DC 3V, Coin Battery (CR2032) - 100 hours | DC3.7V, rechargeable lithium-ion polymer battery – 48 hours
Nominal voltage: 3.7V
Charging voltage: 4.2V
Capacity: Typical 350 mAh | Different |
| Data Stored and Transfer | Yes | Yes | Yes | Same |
| Communication Protocol | HM-15BB-AX: Bluetooth Low Energy (2402MHz – 2480 MHz)
HM-15BW-AX: USB2.0 | Bluetooth Low Energy (2402 – 2480 MHz) | USB 2.0 | Same |
| Viewing Software Platform | HM-15BB-AX: Android APP
HM-15BW-AX: Windows APP | A compatible 3rd-party ECG viewing software | Compatible 3rd party ECG viewing software including Cardiologs | Different |
| Data Encryption | Huffman Encoding encryption | Advanced Encryption Standard-CCM mode | Not publicly available | Different |
| Type of ECG Recorder | Patch | Not publicly available | Patch | Same |
| Number of ECG Channels | 1 | Not publicly available | 1, 2 or 3 | Same |
| Wear Time | Up to 7 days
The two models use the same medical electrode patch which can be used up to 7 days, since the patch is for single use, the actual wear time for each specific model shall correspondent to its recording | Not publicly available | Up to 5 days | Similar |
4 / 8
{8}
Smwmed Inc.
K252389-510(k) Summary
| | Subject device | Primary predicate device | Predicate device 1 | Remark |
| --- | --- | --- | --- | --- |
| K number | K252389 | K231289 | K171410 | / |
| | period. | | | |
| Recording Period | HM-15BB-AX: Up to 3 days
HM-15BW-AX: Up to 7 days | No related information | Up to 9 days | Similar |
| Recording Format | Continuous | Not publicly available | Continuous | Same |
| Sensor Dimensions | 37.0x22.5x7.7 (H) mm | Not publicly available | 10 x 40 x 49mm | Different |
| Sensor Weight | 6.8 grams | Not publicly available | 16 grams | Different |
| Frequency Response | 0.5-40Hz | Not publicly available | 0.05-215 Hz | Similar |
| Input Impedance | >10GOhms | Not publicly available | >10MOhms | Different |
| Resolution | 15bit | Not publicly available | 12 bit or 16 bit, depending on customer preference | Similar |
| Recording Standard | Holter | Not publicly available | Holter | Same |
| Patch Placement | Left chest, a finger's distance below the clavicle, tilted toward the left nipple at a 30-45° degree | Not publicly available | Sternum or appropriate position on torso | Same |
| Heart rate measurement function | Yes for model HM-15BB-AX.
Measurement range: 30-200bpm
Accuracy of heart rate measurement: ±2 | Not publicly available | Not publicly available | Different |
| Heart Rate Variability (HRV) function | Yes | Not publicly available | Not publicly available | Different |
| Accelerometer (ACC) function | Yes | Not publicly available | Not publicly available | Different |
| Viewing of ECG waveform | Yes | Not publicly available | Yes | Same |
5 / 8
{9}
Smwmed Inc.
K252389-510(k) Summary
| | Subject device | Primary predicate device | Predicate device 1 | Remark |
| --- | --- | --- | --- | --- |
| K number | K252389 | K231289 | K171410 | / |
| Signal processing mode | AC ECG channel | Not publicly available | Not publicly available | Different |
| Defibrillation-proof | No | Not publicly available | No | Same |
| Performance Standard | IEC 60601-1
IEC 60601-1-2, IEC 60601-4-2
IEC 60601-1-11
IEC 60601-2-47
ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-23
Cybersecurity assessment and testing | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
Cybersecurity assessment and testing | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-47
ISO 10993-1 | Same |
{10}
Smwmed Inc.
K252389-510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the Rhythm Master ECG Patch was conducted in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process'”, as recognized by FDA. The following testing was performed to, and passed, including:
- ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23, Biological evaluation of medical devices - Part 23: Tests for irritation
### 2) Electrical Safety
- IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11, Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-47, Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
- IEC 62133-2, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
### 3) Software Verification and Validation
Software documentation consistent with Basic documentation level was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement
{11}
Smwmed Inc.
K252389-510(k) Summary
specifications are met and all software hazards have been mitigated to acceptable risk levels. The following testing was performed to, and passed:
- IEC 62304, Medical device software - Software life cycle processes
- FDA’s Guidance for Industry and FDA Staff, Content of Premarket Submissions for Device Software Functions
4) Cybersecurity
Cybersecurity assessment and testing were conducted, according to FDA’s Guidance for Industry and FDA Staff, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
5) ECG Electrodes performance test
ECG Electrode was performed according to FDA Standard
- ANSI/AAMI EC12:2000 (R2020) Disposable ECG Electrodes
6) Energy Reduction
Energy reduction test was performed according to the test method of IEC 60601-1 Clause 8.5.5.2. The device was tested and demonstrated to be safe in case of defibrillation. However, the device is not defibrillation-proof.
7) Wireless coexistence test
Wireless coexistence test was performed according to ANSI/USEMCSC C63.27:2021 and ANSI TIR 69:2017/(R) 2020.
8) Human Factors and Usability
Usability test were performed according to FDA guidance. - Applying Human Factors and Usability Engineering to Medical Device
9) Patch attachment position
The Patch attachment position test was performed according to the internal criteria.
VIII. Conclusions
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is substantially equivalent to the selected predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.