The GEN 4 JOY Wheelchair is intended to provide mobility to persons restricted to a seated position.
Device Story
Manual, four-wheeled mechanical wheelchair; provides mobility for persons restricted to seated position. Supports self-propulsion and attendant-assisted propulsion; suitable for indoor, outdoor, and rural environments. Features foldable frame for transport/storage; adjustable center of gravity via rear wheel repositioning; quick-release wheels; padded seat/backrest; wheel locks; armrests; adjustable leg rests. Operated by user or attendant. Output is physical mobility support. Benefits include improved patient mobility and maneuverability.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per ISO 7176 series (static stability, brakes, dimensions, static/impact/fatigue strength, labeling) and ISO 16840-10 (ignition resistance). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-23 (irritation).
Technological Characteristics
Manual mechanical wheelchair. Materials: fire-retardant upholstery, metal frame. Features: foldable frame, adjustable center of gravity, quick-release wheels. Standards: ISO 7176 (1, 3, 5, 7, 8, 15), ISO 16840-10, ISO 10993 (5, 10, 23). No energy source or software.
Indications for Use
Indicated for persons restricted to a seated position requiring mobility assistance. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 24, 2026
Free Wheelchair Mission
Bonnie Gonzalez
Mechanical Engineer
15279 Alton Pkwy. Suite 300, Irvine
CA, 92618, USA
Re: K252370
Trade/Device Name: GEN 4 JOY Wheelchair
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: June 17, 2026
Received: June 18, 2026
Dear Bonnie Gonzalez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252370 - Bonnie Gonzalez
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252370 - Bonnie Gonzalez
Page 3
Sincerely,

Digitally signed by
ZACHARY MCKINNEY -S
Date: 2026.07.24
08:35:59 -04'00'
for Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K252370
Device Name
GEN 4 JOY Wheelchair
Indications for Use (Describe)
The GEN 4 JOY Wheelchair is intended to provide mobility to persons restricted to a seated position.
Type of Use (Select one or both, as applicable)
☐
Prescription Use (Part 21 CFR 801 Subpart D)
☑
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) #: K252370
# 510(k) Summary
Prepared on: 2026-06-17
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Free Wheelchair Mission |
| --- | --- |
| Applicant Address | 15279 Alton Parkway Suite 300 Irvine CA 92618 United States |
| Applicant Contact Telephone | 949-273-8470 |
| Applicant Contact | Mrs. Nuka Hart |
| Applicant Contact Email | nhart@freewheelchairmission.org |
| Correspondent Name | Free Wheelchair Mission |
| Correspondent Address | 15279 Alton Parkway Suite 300 Irvine CA 92618 United States |
| Correspondent Contact Telephone | 949-273-8470 |
| Correspondent Contact | Ms. Bonnie Gonzalez |
| Correspondent Contact Email | bgonzalez@freewheelchairmission.org |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | GEN 4 JOY Wheelchair |
| --- | --- |
| Common Name | Mechanical wheelchair |
| Classification Name | Wheelchair, Mechanical |
| Regulation Number | 890.3850 |
| Product Code(s) | IOR |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K141691 | GEN_3 Mechanical Wheelchair | IOR |
| | | |
Device Description Summary
21 CFR 807.92(a)(4)
The GEN 4 JOY Wheelchair (Project name: "Wheely") by Free Wheelchair Mission is a manual, four-wheeled wheelchair intended to provide mobility to persons restricted to a seated position. It supports both self-propulsion and attendant-assisted propulsion and is suitable for indoor and outdoor use, including rural environments. The wheelchair features an adjustable center of gravity, achieved by repositioning the rear wheels, to improve maneuverability. It includes a foldable frame for ease of transport and storage. Key components include two large rear wheels, two front casters, a padded seat and backrest, wheel locks, armrests, and adjustable leg rests.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
{5}
The GEN 4 JOY Wheelchair is intended to provide mobility to persons restricted to a seated position.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The Indications for Use is identical to that of the primary predicate.
## Technological Comparison
21 CFR 807.92(a)(6)
The GEN 4 JOY Wheelchair is substantially equivalent in design, materials, and intended use to GEN_3 Mechanical Wheelchair. The indications for use for the GEN 4 JOY and the GEN_3 mechanical wheelchairs are identical. The size ranges for the GEN 4 JOY in terms of frame width, seat depth, back depths, chair weight and weight limit overlap those of the GEN_3. The GEN 4 JOY is provided in 6 sizes compared to 4 sizes for the GEN_3. All other features are identical with the exception that the wheels for the GEN 4 JOY are quick-release. The materials used in the seat, backrest, fire retardant upholstery, and wheel locks are also identical between the GEN 4 JOY and the GEN_3. The fundamental technology used in the GEN 4 JOY chairs is identical to that of the GEN_3. The information provided supports a substantial equivalence decision based on the repeat non-clinical testing of the static, impact and fatigue strengths. Any differences in features to the predicate do not affect the performance of the device and do not raise new types of safety or effectiveness questions.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
ISO Standard 7176 for Wheelchairs was used as the basis for performance testing of the GEN 4 JOY Mechanical Wheelchair. FDA recognizes the following applicable parts of ISO 7176 as consensus standards:
1) ISO 7176-1:1999 Determination of static stability
2) ISO 7176-3:2006 Determination of effectiveness of brakes
3) ISO 7176-5:2008 Determination of dimensions, mass and maneuvering space
4) ISO 7176-7:1998 Measurement of seating and wheel dimensions
5) ISO 7176-8:1998 Requirements and test methods for static, impact and fatigue
6) ISO 7176-15:1996 Requirements for information disclosure, documentation and labeling
7) ISO 16840-10:2021 Resistance to ignition of upholstered parts- Requirements and test methods
The ISO tests were conducted with methods that met the following standards:
1) ISO 7176-11:1992 Test dummies
2) ISO 7176-13: 1989 Determination of coefficient of friction of test surfaces
### Biocompatibility
The biocompatibility of the subject device is based on the use of low-biocompatibility risk materials and supporting information in accordance with Attachment G of FDA's 2023 Biocompatibility Guidance, and cytotoxicity, sensitization, and irritation testing per ISO 10993-5:2009, ISO 10993-10:2021, and 10993-23:2021, respectively.
Clinical tests are Not Applicable.
The non-clinical test results demonstrate the GEN 4 JOY Wheelchair does not raise any different questions of safety and effectiveness. The GEN 4 JOY, like the predicate, meets the ISO Standard 7176 for wheelchairs. The testing supports a determination of substantial equivalence between the GEN 4 JOY Wheelchair and predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.