K252367 · Inbella Medical, Inc. · GEI · Oct 15, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252367
Device Name
InbellaMulti System
Applicant
Inbella Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InbellaMulti System with the Non-invasive Applicators employs RF energy for various applications: • i-BellaForma, BellaForma, BellaPlus and BellaPlus90 for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation. • BellaWMface is intended for use in dermatologic procedures for non-invasive treatment of mild to moderate facial wrinkles and rhytids. • BellaBodyFX /BellaMiniFX for Relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement in local blood circulation and temporary reduction in the appearance of cellulite. The InbellaMulti System with the Fractional Applicators employs RF energy for various applications: • BellaFractora Applicator with 60 pins tip is designed for use in dermatological procedures requiring ablation and resurfacing of the skin. • BellaFractora Applicator with 24 pins tip is intended for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62mJ/pin, use of the applicator is limited to skin types I-IV • BellaM8 for dermatologic procedures where coagulation/contraction of soft tissue or hemostasis is needed. At higher energy levels greater than 62mJ/pin, use of the applicator is limited to skin types I-IV
Device Story
InbellaMulti System is a radiofrequency (RF) electrosurgical platform for dermatological and surgical applications. System comprises a console with RF generator, measuring circuit, controller, touchscreen interface, and footswitch. Connects to various non-invasive and fractional RF applicators (e.g., i-BellaForma, BellaFractora, BellaM8). Physician-operated in clinical settings. Device delivers RF energy to tissue via electrodes/pins to induce heating, coagulation, contraction, or ablation. Output effects include muscle pain relief, improved circulation, cellulite reduction, wrinkle treatment, and hemostasis. Clinical decision-making relies on physician selection of applicator and energy parameters based on procedure type and patient skin type. Benefits include non-invasive or minimally invasive tissue treatment and surgical coagulation.
Clinical Evidence
No clinical data provided; substantial equivalence based on identical technological characteristics and indications for use to cleared predicate devices.
Technological Characteristics
RF-based electrosurgical platform. Components: console, AC/DC power supply, RF generator, RF measuring circuit, controller, touchscreen, footswitch. Applicators: non-invasive and fractional (pin-based). Standards: ANSI AAMI ES 60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC 60601-2-2. Energy source: RF. Connectivity: wired applicator connection.
Indications for Use
Indicated for relief of minor muscle aches, pain, and muscle spasms; temporary improvement of local blood circulation; temporary reduction in cellulite appearance; non-invasive treatment of mild to moderate facial wrinkles/rhytids; dermatological ablation and resurfacing; and electrocoagulation/hemostasis in dermatological and general surgical procedures. Contraindicated for skin types V-VI at energy levels >62mJ/pin for specific applicators.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
InMode System with Morpheus8 applicators (K231790)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 15, 2025
Inbella Medical Inc.
% Janice Hogan
Partner
Hogan Lovells US LLP
1735 Market Street, Floor 23
Philadelphia, Pennsylvania 19103
Re: K252367
Trade/Device Name: InbellaMulti System
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: July 29, 2025
Received: July 30, 2025
Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252367 - Janice Hogan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252367 - Janice Hogan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.10.15
13:36:54 -04'00'
Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K252367
Page 1 of 1
| Indications for Use | |
| --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252367 |
| Please provide the device trade name(s). | |
| InbellaMulti System | |
| Please provide your Indications for Use below. | |
| The InbellaMulti System with the Non-invasive Applicators employs RF energy for various applications:
• i-BellaForma, BellaForma, BellaPlus and BellaPlus90
for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local
blood circulation.
• BellaWMface is intended for use in dermatologic procedures for non-invasive treatment of mild to
moderate facial wrinkles and rhytids.
• BellaBodyFX /BellaMiniFX for Relief of minor muscle aches and pain, relief of muscle spasm, temporary
improvement of local blood circulation and temporary reduction in the appearance of cellulite.
The InbellaMulti System with the Fractional Applicators employs RF energy for various applications:
• BellaFractora Applicator with 60 pins tip is designed for use in dermatological procedures requiring
ablation and resurfacing of the skin.
• BellaFractora Applicator with 24 pins tip is intended for use in dermatological and general surgical
procedures for
electrocoagulation and hemostasis. At higher energy levels greater than 62mJ/pin, use of the applicator is
limited to skin types I-IV
• BellaM8 for dermatologic procedures where coagulation/contraction of soft tissue or hemostasis is
needed. At higher energy levels greater than 62mJ/pin, use of the applicator is limited to skin types I-IV | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
InbellaMulti System
{4}
K252367
Page 1 of 3
510(K) Summary
INBELLA MEDICAL INC.'S INBELLAMULTI SYSTEM
Applicant Name:
Company Name: Inbella Medical Inc.
Address: 100 Leek Crescent, Unit 13
Richmond Hill, Ontario, Canada
Contact Person: Janice M. Hogan
Partner
janice.hogan@hoganlovells.com
T +1 267 675 4611
F +1 267 675 4601
Date Prepared: July 29, 2025
Name of Device: InBellaMulti System
Common or Usual Name: Radiofrequency-based electrosurgical coagulation device
Classification Name: 21 CFR 878.4400
Classification: Class II Medical Device
Product Code: GEI
Predicate Device:
The InBellaMulti System manufactured by Inbella Medical, Inc. (platform and all applicators other than BellaM8) is substantially equivalent to the InMode RF Pro System cleared in K210492 by InMode Ltd. For the BellaM8 applicator, the subject device is substantially equivalent to the InMode System with Morpheus8 applicators cleared in K231790 by InMode Ltd.
Purpose of the 510(k) Notice:
The InBellaMulti System manufactured by Inbella Medical, Inc. is identical in all aspects to the listed predicate devices. The only difference are the manufacturer and 510(k) holder, as well as system and related handpiece names.
Intended Use:
The InbellaMulti System with the Non-invasive Applicators employs RF energy for various applications:
- i-BellaForma, BellaForma, BellaPlus and BellaPlus90 for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
{5}
K252367
Page 2 of 3
- BellaWMface is intended for use in dermatologic procedures for non-invasive treatment of mild to moderate facial wrinkles and rhytids.
- BellaBodyFX /BellaMiniFX for relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and temporary reduction in the appearance of cellulite.
The InbellaMulti System with the Fractional Applicators employs RF energy for various applications:
- BellaFractora Applicator with 60 pins tip is designed for use in dermatological procedures requiring ablation and resurfacing of the skin.
- BellaFractora Applicator with 24 pins tip is intended for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62mJ/pin, use of the applicator is limited to skin types I-IV
- BellaM8 for dermatologic procedures where coagulation/contraction of soft tissue or hemostasis is needed. At higher energy levels greater than 62mJ/pin, use of the applicator is limited to skin types I-IV
Except for the difference in device name and applicator names, the indications for use of the subject device are identical to those of the predicate applicators in K210492 for all applicators except for the BellaM8 applicators. The indications for the BellaM8 applicators are identical to those cleared more recently in K231790 for the Morpheus8 applicators.
## Device Description:
The InbellaMulti System supports multiple radiofrequency (RF) applications and accessories. The InbellaMulti System is compatible with the Fractional RF Applicators and the Non-Invasive RF Applicators, and employs RF energy for various applications. The InbellaMulti System consists of a platform console with an AC/DC power supply unit, two applicator connectors, RF generator, RF measuring circuit, controller, footswitch and user interface including a touch screen. The RF Applicator is connected to the console via a cable and a footswitch activates the energy delivery to the applicator. The applicators are comprised of a handle and electrodes, and some of them are used with a single-use tip. The below list comprises the set of applicators to be registered under the subject device:
- i-BellaForma
- BellaForma
- BellaPlus
- BellaPlus90
- BellaWMFace
- BellaBodyFX
- BellaMiniFX
- BellaFractora with 60 pins tip
- BellaFractora with 24 pins tip
- BellaM8 with the following tip heads: Resurfacing (T), Prime (12-pin), 24-pin, 40-pin (Body)
The technological characteristics of the subject device are identical to those of the predicate applicators in K210492 for all applicators except for the BellaM8 applicators. The technological
{6}
K252367
Page 3 of 3
characteristics for the BellaM8 applicators are identical to the Morpheus8 applicators cleared more recently in K231790.
## Performance Standards:
The InBellaMulti System has been tested and complies with the following voluntary recognized standards:
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
## Non-Clinical (Bench) Performance Data:
Not Applicable
## Animal Performance Data / Histology Data:
Not Applicable
## Clinical Performance Data:
Not Applicable
## Substantial Equivalence:
The indications for use and technological characteristics of the InBellaMulti System manufactured by Inbella Medical, Inc. are the same as those of the InMode System manufactured by InMode, Ltd. The only difference are the manufacturer and 510(k) holder, as well as system and related handpiece names.
Thus, the InbellaMulti System is substantially equivalent to the cleared predicate and may, therefore, be legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.