K252361 · Accurkardia, Inc. · DPS · Dec 22, 2025 · Cardiovascular
Device Facts
Record ID
K252361
Device Name
AccurECG Analysis System (v2.0)
Applicant
Accurkardia, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The AccurECG® Analysis System v2.0 is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing. The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG® Analysis System v2.0 is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads). The AccurECG® Analysis System v2.0 is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors. The AccurECG® Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion. The AccurECG® Analysis System v2.0 does not interface with data management systems or hardware and is device independent.
Device Story
AccurECG Analysis System v2.0 is a server-side software tool for analyzing ambulatory ECG recordings. Input consists of single-lead ECG files (.DAT, .HEA, .ECG, .EDF) uploaded via a web interface. The system uses proprietary algorithms to perform signal delineation (QRS/T-wave detection) and automated arrhythmia classification. Output is a standard report providing heart rate, rhythm analysis, and ectopic beat detection for physician review. Used in clinical/office settings by healthcare professionals (technicians, MDs). The system is device-independent and does not interface with hardware or telemetry. Physicians use the report as an adjunct to clinical history and ECG patterns to support diagnostic conclusions. Benefits include automated, standardized analysis of ambulatory ECG data to facilitate efficient clinical review.
Clinical Evidence
Bench testing only. Performance evaluated per AAMI/ANSI EC 57:2012 and IEC 60601-2-47:2012. Testing included unit, module, system, and cybersecurity verification. Comparative performance testing against the predicate device demonstrated statistical equivalence, with 95% confidence intervals for all measured parameters falling within specified equivalence margins.
Technological Characteristics
Cloud-based software system; server-side proprietary interpretation algorithm; client-side web interface. Compatible with single-lead ECG (200-1000 Hz, 11/16-bit). Connectivity via web API. Software environment: Ubuntu Linux virtual machines; Python-based. Complies with AAMI/ANSI EC 57:2012 and IEC 60601-2-47:2012.
Indications for Use
Indicated for adults aged 22 and older for assessment of recorded ambulatory ECG data; intended for use by qualified healthcare professionals as an adjunct for clinical decision-making; not for use in life-supporting systems, ECG monitors, alarms, or cardiac telemetry.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
AccurKardia Inc.
% Kwame Ulmer
Regulatory Consultant, Official Correspondent
MedTech Impact Partners
16844 Margate Street
Encino, California 91436
Re: K252361
Trade/Device Name: AccurECG Analysis System (v2.0)
Regulation Number: 21 CFR 870.2340
Regulation Name: Electrocardiograph
Regulatory Class: Class II
Product Code: DPS, DQK
Dated: November 26, 2025
Received: November 26, 2025
Dear Kwame Ulmer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252361 - Kwame Ulmer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252361 - Kwame Ulmer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
KIMBERLY N. CROWLEY-S
Digitally signed by KIMBERLY N. CROWLEY-S
For: Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K252361
Device Name
AccurECG Analysis System (v2.0)
Indications for Use (Describe)
The AccurECG® Analysis System v2.0 is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.
The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG® Analysis System v2.0 is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads).
The AccurECG® Analysis System v2.0 is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.
The AccurECG® Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion.
The AccurECG® Analysis System v2.0 does not interface with data management systems or hardware and is device independent.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
# 510(k) Summary
## 1. General Information
### 1.1. Applicant/Submitter
AccurKardia, Inc.
101 Avenue of the Americas
Suite 931
New York, NY 10013
Phone Number: +1-718-708-3522
Website: www.accurkardia.com
Contact Person: Juan Jiménez
Phone: +1 (718) 708-3522
Email: jcj@accurkardia.com
### 1.2. Correspondent Information
Official Correspondent: Kwame Ulmer
Correspondent Phone: +1 (202) 596-2492
Correspondent Email: kwame.ulmer@medtechimpactpartners.com
### 1.3. Date of Preparation
November 26, 2025
## 2. Device Information
### 2.1. Subject Device
| Device Name | AccurECG® Analysis System v2.0 |
| --- | --- |
| Model Number | v2.0 |
| Regulation Name | Electrocardiograph |
| Regulation | 21 CFR 870.2340 |
| Product Code | DPS, DQK |
| Regulatory Panel | Cardiovascular |
| Submission Number | K252361 |
510(k) Summary
{5}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
## 2.2. Predicate Device
| Device Name | AccurECG® Analysis System |
| --- | --- |
| Model Number | v1.3.0 |
| Regulation Name | Electrocardiograph |
| Regulation | 21 CFR 870.2340 |
| Product Code | DPS, DQK |
| Regulatory Panel | Cardiovascular |
| Submission Number | K223013 |
## 3. Device Description
The AccurKardia ECG Analysis System consists of: (1) A client-side web interface which provides tools to upload and review ECG recordings via a web application programming interface (API), and (2) A server-side automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide physicians supportive information for ECG analysis.
AccurECG v2.0 is intended to analyze recordings performed on adults aged 22 and older and works in the following sequence:
i. The web interface allows the user to select files and upload to the AccurECG secure database.
ii. The proprietary AccurECG algorithm analyzes and labels the ECG:
- Delineation – detection of QRS complexes and T waves on the ECG signal through advanced signal processing and image-based techniques.
- Abnormality labels – automated arrhythmia interpretation and statistical classification.
iii. AccurECG® Analysis System v2.0 displays the resulting analysis and original ECG signal with reviewing tools in the web interface.
iv. AccurECG® Analysis System v2.0 allows for preliminary comments summarizing the review before generating a report.
## 4. Intended Use / Indications for Use
The AccurECG® Analysis System v2.0 is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.
The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG® Analysis System v2.0 is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads).
K252361
2 of 10
510(k) Summary
{6}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
The AccurECG® Analysis System v2.0 is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.
The AccurECG® Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion.
The AccurECG® Analysis System v2.0 does not interface with data management systems or hardware and is device independent.
# 5. Comparison of Technological Characteristics with Predicate
The software features have been modified in comparison to the predicate device to support enhanced device functionality. The modifications include changes to the operating system and programming language; however, the intended use, indications for use, technological characteristics, usability, and principles of operation of the subject device remain identical to those of the predicate device cleared under Premarket Notification K223013. The following table (Table 1) demonstrates the substantial equivalence comparison between the subject and predicate device.
There is no impact on the substantial equivalence of the AccurECG® Analysis System v2.0. Based on these considerations, the subject device is considered substantially equivalent to the predicate device in accordance with FDA Guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" (Jul. 28, 2014).
Table 1 – Substantial Equivalence Table
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| Manufacturer | AccurKardia Inc. | AccurKardia Inc | Same |
| 510(k) | K223013 | K252361 | Same |
| Classification | 21 CFR 870.2340 | 21 CFR 870.2340 | Same |
| Regulation | 21 CFR 870.1425 | 21 CFR 870.1425 | |
| Product Codes | DPS, DQK | DPS, DQK | Same |
| Regulatory Class | II | II | Same |
| Classification Name | Electrocardiograph | Electrocardiograph | Same |
| Indications for Use | The AccurECG® Analysis System is intended for use by qualified | The AccurECG® Analysis System v2.0 is intended for use by | Same |
K252361
510(k) Summary
{7}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE
AccurECG® Analysis System
(K223013) | SUBJECT DEVICE
AccurECG® Analysis System
v2.0 | COMPARISON |
| --- | --- | --- | --- |
| | healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.
The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG® Analysis System is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads).
The AccurECG® Analysis System is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.
The AccurECG® Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to | qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.
The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG® Analysis System v2.0 is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads).
The AccurECG® Analysis System v2.0 is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.
The AccurECG® Analysis System v2.0's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other | |
4 of 10
K252361
510(k) Summary
{8}
ACCURKARDIA
101 Avenue of Americas,
Suite 931,
New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| | arrive at a diagnostic conclusion.
The AccurECG® Analysis System does not interface with data management systems or hardware and is device independent. | diagnostic information to arrive at a diagnostic conclusion.
The AccurECG® Analysis System v2.0 does not interface with data management systems or hardware and is device independent. | |
| Device Description | The AccurKardia ECG Analysis System consists of: (1) A client-side web interface which provides tools to upload and review ECG recordings via a web application programming interface (API), and (2) A server-side automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide physicians supportive information for ECG analysis.
AccurECG is intended to analyze recordings performed on adults over the age of 22 and works in the following sequence:
i. The web interface allows the user to select files and upload to the AccurECG secure database.
ii. The proprietary AccurECG algorithm analyzes and labels the ECG:
○ Delineation – detection of QRS complexes and T | The AccurKardia ECG Analysis System v2.0 consists of: (1) A client-side web interface which provides tools to upload and review ECG recordings via a web application programming interface (API), and (2) A server-side automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide physicians supportive information for ECG analysis.
AccurECG v2.0 is intended to analyze recordings performed on adults aged 22 and older and works in the following sequence:
i. The web interface allows the user to select files and upload to the AccurECG secure database.
ii. The proprietary AccurECG algorithm analyzes and labels the ECG:
○ Delineation – detection of QRS complexes and T | Same |
K252361
5 of 10
510(k) Summary
{9}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| | waves on the ECG signal through advanced signal processing and image-based techniques.
○ Abnormality labels – automated arrhythmia interpretation and statistical classification. | waves on the ECG signal through advanced signal processing and image-based techniques.
○ Abnormality labels – automated arrhythmia interpretation and statistical classification. | |
| iii. The AccurECG® Analysis System displays the resulting analysis and original ECG signal with reviewing tools in the web interface. | iii. The AccurECG® Analysis System v2.0 displays the resulting analysis and original ECG signal with reviewing tools in the web interface. | |
| iv. The AccurECG® Analysis System allows for preliminary comments summarizing the review before generating a report. | iv. The AccurECG® Analysis System v2.0 allows for preliminary comments summarizing the review before generating a report. | |
| Patient Population | Adults aged 22 and older
Prescribed for out-patient cardiac monitoring for up to ten days | Adults aged 22 and older
Prescribed for out-patient cardiac monitoring for up to ten days | Same |
| Use Environment | Office or clinical setting at a workstation equipped with a PC and high-speed internet access | Office or clinical setting at a workstation equipped with a PC and high-speed internet access | Same |
| Intended User | Qualified healthcare professionals (including ECG technicians, medical supervisors (MDs), and healthcare providers) | Qualified healthcare professionals (including ECG technicians, medical supervisors (MDs), and healthcare providers) | Same |
| Electrocardiograph Signal Requirements |
| ECG File Format | .DAT, .HEA, .ECG (ISHINE-formatted Holter ECG file) or .EDF format from single lead ambulatory ECG device | .DAT, .HEA, .ECG (ISHINE-formatted Holter ECG file) or .EDF format from single lead ambulatory ECG device | Same |
6 of 10
K252361
510(k) Summary
{10}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| Duration | 10 - 32.9 seconds | 10 - 32.9 seconds | Same |
| Sampling frequency | 200 Hz – 1000 Hz | 200 Hz – 1000 Hz | Same |
| Resolution | 11Bit and 16 Bit | 11Bit and 16 Bit | Same |
| Lead positioning | Lead II | Lead II | Same |
| Supported electrode types | Wet Silver (Ag) / Silver Chloride (AgCl) ECG electrodes with conductive gel | Wet Silver (Ag) / Silver Chloride (AgCl) ECG electrodes with conductive gel | Same |
| AccurECG System Requirements | | | |
| Device Execution Environment (Cloud-Based) | Windows Virtual Machines | Ubuntu Linux Virtual Machines | Substantially Equivalent: ISO 14971 risk analysis and verification / validation testing have confirmed differences do not introduce new questions of safety and effectiveness. Device inputs/outputs and user interface remain the same between both the subject and the predicate device. |
| Programming Language | LabVIEW (G Programming Language) | Python | Substantially Equivalent: The different programming language does not impact the core functionality, principles of operation, or performance of the device. ISO 14971 risk analysis and verification / validation testing have confirmed differences do not introduce new |
7 of 10
K252361
510(k) Summary
{11}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| | | | questions of safety and effectiveness.
Device inputs/outputs and user interface remain the same between both the subject and the predicate device. |
| **End-User System Requirements for Viewing** | | | |
| End-user Compatible Workstation PC | Microsoft® Windows® 10 version 21H2 or later | Microsoft® Windows® 11 Version 24H2 for x64-based Systems (KB5055523) as the minimum desktop OS | Same - Both configurations specify currently supported Microsoft Windows operating systems consistent with vendor-supported release versions |
| Display Resolution | 1920 X 1080 (1080p Monitor recommended) | 1920 X 1080 (1080p Monitor recommended) | Same |
| Web Browser | Google Chrome Web Browser (Version 81.0 or later) | Google Chrome Web Browser (Version 134.0 or later) | Same |
| Pdf Report Viewer | Adobe Acrobat Reader DC (2020.012 or later) with Adobe Acrobat Extension for Chrome (Version 17 or later), or
Adobe Acrobat Pro DC (2020.012 or later) with Adobe Acrobat Extension for Chrome (Version 15 or later) | Adobe Acrobat Reader DC (25.001.20458 or later) with Adobe Acrobat Extension for Chrome (Version 17 or later), or
Adobe Acrobat Pro DC (25.001.20458 or later) with Adobe Acrobat Extension for Chrome (Version 15 or later) | Same |
| **Device Outputs** | | | |
| Heart rate determination (non-paced adult) | Yes | Yes | Same |
| QRS Detection | Yes | Yes | Same |
K252361
8 of 10
510(k) Summary
{12}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
| Technological Characteristics | PREDICATE DEVICE AccurECG® Analysis System (K223013) | SUBJECT DEVICE AccurECG® Analysis System v2.0 | COMPARISON |
| --- | --- | --- | --- |
| Non-paced arrhythmia interpretation | Yes | Yes | Same |
| Non-paced ventricular arrhythmia calls | Yes | Yes | Same |
| Intervals measurements | Yes | Yes | Same |
| Ventricular ectopic beat detection | Yes | Yes | Same |
## 6. Summary of Performance Data
A summary of the performance data, which was based upon well-established test methods and FDA-recognized consensus standards, demonstrate conformity to the intended use. Performance testing applicable to the subject device was completed to ensure it performs as intended per the product specifications and requirements. The following testing has been completed in support of the AccurECG® Analysis System v2.0, and all acceptance criteria were met:
- Unit testing
- Code review
- Software module verification testing
- Field level verification testing
- System verification testing
- Performance evaluation
- Cybersecurity testing
Performance tests have been conducted in conformance with the following FDA-recognized consensus standards:
- AAMI/ANSI EC 57:2012 Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms
- IEC 60601-2-47:2012 Medical Electrical Equipment, Part 2-47: Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems (Section 201.12.1.101 – Algorithm Testing)
Comparative performance testing demonstrated statistical equivalence between the subject device and the predicate device. For every measured parameter, the 95% confidence interval lay within the
9 of 10
K252361
510(k) Summary
{13}
ACCURKARDIA
101 Avenue of Americas, Suite 931, New York, NY 10013
specified equivalence margin. These results support that no new questions of safety or effectiveness are raised, and the subject device is considered substantially equivalent to the predicate device based on the performance data collected.
## 7. Conclusion
Based upon the Intended Use, Indications for Use, product technical information, performance testing and evaluation, and standards compliance provided in this premarket notification, the AccurECG® Analysis System v2.0 has been shown to be substantially equivalent to the cited predicate.
K252361
10 of 10
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.