Electronic Medical Records (EMR); Right Heart Catheterization (RHC) reports; Echocardiogram reports; 12-lead ECG digital waveforms; ICD-9/ICD-10 diagnostic codes
Retrospective clinical study used to establish diagnostic performance (sensitivity, specificity, PPV, NPV) of the ECG-AI PH 12-Lead algorithm in a diverse population of adults presenting with dyspnea.
Retrospective study; Diagnostic performance; Pulmonary hypertension; Real-world clinical data
Adults (≥ 18 years) presenting with dyspnea (ICD codes) who had an RHC or echocardiogram and a digital 12-lead ECG.; Sample Size: 21,066 patient records; Number of Sites: 5 health systems
Not applicable for this study
Sensitivity and specificity for detection of pulmonary hypertension (mPAP >20 mmHg or TRV >3.4 m/s)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Pulmonary Hypertension
—
PPV and NPV greater than predetermined acceptance criteria
Diagnostic dataset: 259,227 patients (39,823 PH patients, 219,404 control patients). Sub-divided into train (n=18,987 PH; n=105,260 controls), tune (n=4,661 PH; n=26,146 controls), and test (n=16,175 PH; n=87,998 controls).
—
Retrospective study of 21,066 patient records across 5 health systems in the United States.
—
Indications for Use
The ECG-AI PH 12-Lead algorithm is software intended to aid in earlier detection of elevated mean pulmonary arterial pressure (mPAP), an indicator of pulmonary hypertension, in adults presenting with dyspnea. ECG-AI PH 12-Lead algorithm is not intended to be a stand-alone diagnostic device for pulmonary hypertension or replace current clinical practice guidelines. A positive result may suggest the need for further clinical evaluation. Additionally, if the patient is at high risk for pulmonary hypertension, a negative result should not rule out additional clinical evaluation. ECG-AI PH 12-Lead algorithm should be applied jointly with clinician judgment. ECG-AI PH 12-Lead algorithm should not be used on ECGs from adults with an implanted pacemaker or on ECGs with a paced rhythm.
Device Story
Software-based AI tool; analyzes 10+ seconds of 12-lead ECG voltage data; detects likelihood of pulmonary hypertension (PH). Integrated into clinical workflows via EMR or ECG Management Systems; accessed by clinicians via PC or mobile devices. Inputs: digital 12-lead ECG waveform. Outputs: binary result (Detected/Not Detected/Error). Used in clinical settings to support physician decision-making regarding further workup or specialist referral. Benefits: provides non-invasive, accessible PH screening; facilitates earlier detection of underdiagnosed conditions. Risks: false negatives (delayed diagnosis) or false positives (unnecessary testing); mitigated by clinical judgment requirement and automated signal quality checks.
Clinical Evidence
Retrospective study of 21,066 patient records across 5 US health systems. Primary endpoints: sensitivity and specificity for PH detection. Results: 73.0% sensitivity (95% CI: 72.1-73.9%) and 74.4% specificity (95% CI: 73.5-75.2%). PPV and NPV adjusted for 4.36% prevalence were 11.5% and 98.4%, respectively. Validation included diverse demographic subgroups (age, sex, race, BMI).
Technological Characteristics
Software module in Docker container; no GUI; integrates with EMR/EMS via API. Analyzes resting 12-lead ECGs. Includes automated signal quality checks for noise/artifacts. Compliant with IEC 62304 (software lifecycle) and FDA cybersecurity guidance.
Indications for Use
Indicated for adults presenting with dyspnea to aid in detection of elevated mean pulmonary arterial pressure (mPAP) as an indicator of pulmonary hypertension. Contraindicated for patients with implanted pacemakers or paced rhythms.
Regulatory Classification
Identification
Viz HCM is a cardiovascular machine learning-based notification software intended to be used in parallel to the standard of care to analyze 12-lead ECG recordings from patients 18 years of age or older. It detects signs associated with hypertrophic cardiomyopathy (HCM) and allows the user to view the ECG and analysis results. It is not intended for use on patients with implanted pacemakers, does not replace standard diagnostic methods, and is not intended to rule out HCM or be used in lieu of a full patient evaluation.
Special Controls
In combination with the general controls of the FD&C Act, cardiovascular machine learningbased notification software is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 28, 2026
Anumana, Inc.
Taylor Gold West
Director, Regulatory Affairs
One Main Street, Suite 400
East Arcade, 4th Floor
Cambridge, Massachusetts 02142
Re: K252360
Trade/Device Name: ECG-AI Pulmonary Hypertension 12-Lead algorithm
Regulation Number: 21 CFR 870.2380
Regulation Name: Cardiovascular machine learning-based notification software
Regulatory Class: Class II
Product Code: SAT
Dated: July 29, 2025
Received: July 29, 2025
Dear Taylor Gold West:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252360 - Taylor Gold West
Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K252360 - Taylor Gold West
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN C. BROWNING -S**
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252360 | |
| Device Name ECG-AI Pulmonary Hypertension 12-Lead algorithm | |
| Indications for Use (Describe) The ECG-AI PH 12-Lead algorithm is software intended to aid in earlier detection of elevated mean pulmonary arterial pressure (mPAP), an indicator of pulmonary hypertension, in adults presenting with dyspnea. ECG-AI PH 12-Lead algorithm is not intended to be a stand-alone diagnostic device for pulmonary hypertension or replace current clinical practice guidelines. A positive result may suggest the need for further clinical evaluation. Additionally, if the patient is at high risk for pulmonary hypertension, a negative result should not rule out additional clinical evaluation. ECG-AI PH 12-Lead algorithm should be applied jointly with clinician judgment. ECG-AI PH 12-Lead algorithm should not be used on ECGs from adults with an implanted pacemaker or on ECGs with a paced rhythm. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
K252360
510(k) Summary
ECG-AI Pulmonary Hypertension 12-Lead Algorithm
Applicant Name:
Anumana, Inc.
One Main Street, Suite 400
East Arcade 4th Floor
Cambridge, MA 02142
Primary Contact:
Taylor Gold West, MBA, RAC
Director, Regulatory Affairs
Anumana, Inc.
tgoldwest@anumana.ai
Alternate Contact:
Tyler Wagner, PhD
Vice President, Biomedical Research
Anumana, Inc.
Tyler@anumana.ai
Date Summary Prepared: March 27, 2026
Proprietary Name of Device: ECG-AI Pulmonary Hypertension 12-Lead algorithm
Common or Usual Name: ECG-AI analysis tool
Classification Panel: Cardiology
Regulation Number: 21 CFR § 870.2380
Regulation Name: Cardiovascular machine-learning notification software
Regulation Class: Class II
Product Code: SAT
Predicate Device(s):
Device Name: ECG-AI Low Ejection Fraction 12-Lead algorithm
Manufacturer: Anumana, Inc.
Application Number: K250652
Device Description
ECG-AI PH 12-Lead algorithm interprets 12-lead ECG voltage time series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds or longer duration of a single 12-lead ECG acquisition, and within seconds detects the presence of pulmonary hypertension in the intended patient population.
ECG-AI PH 12-Lead algorithm is integrated into existing clinical workflows to support overall assessment and to enable physicians to determine whether further workup or referral to a PH specialist is clinically indicated. ECG-AI PH 12-Lead algorithm is intended to be used by clinicians across clinical care settings.
The software module can be integrated into a client application to be accessed by clinicians and results viewed through an Electronic Medical Record (EMR) system or an ECG Management System (EMS)
Page 1 of 11
{5}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
accessed via a PC, mobile device, or another medical device. In each case, the physician imports 12-lead ECG data in a digital format. The tool analyzes the 10 seconds or longer duration of voltage data collected during a standard 12-lead ECG and outputs a binary result of the likelihood of pulmonary hypertension as an API result.
ECG-AI PH 12-Lead algorithm is not intended to diagnose PH, or to replace current clinical practice guidelines.
ECG-AI Pulmonary Hypertension 12-Lead algorithm is not intended for assessing disease progression or monitoring response to therapy. The ECG-AI Pulmonary Hypertension 12-Lead algorithm is not intended for repeated testing in a short time interval, serial use during the same hospitalization or for serial monitoring.
## Indications for Use
The ECG-AI PH 12-Lead algorithm is software intended to aid in earlier detection of elevated mean pulmonary arterial pressure (mPAP), an indicator of pulmonary hypertension, in adults presenting with dyspnea. ECG-AI PH 12-Lead algorithm is not intended to be a stand-alone diagnostic device for pulmonary hypertension or replace current clinical practice guidelines.
A positive result may suggest the need for further clinical evaluation. Additionally, if the patient is at high risk for pulmonary hypertension, a negative result should not rule out additional clinical evaluation.
ECG-AI PH 12-Lead algorithm should be applied jointly with clinician judgment.
ECG-AI PH 12-Lead algorithm should not be used on ECGs from adults with an implanted pacemaker or on ECGs with a paced rhythm.
## Technological Characteristics
ECG-AI PH 12-Lead algorithm is provided as a software module packaged in a Docker container to facilitate installation. Technical installation details including access to docker hub, docker hub path, software upgrades, and associated access rights can be found in the Installation Manual ECG-AI Pulmonary Hypertension 12-Lead algorithm. ECG-AI PH 12-Lead algorithm does not provide a graphical user interface (GUI) of its own. It is integrated with other medical systems such as Electronic Medical Record (EMR) systems or ECG Management Systems (EMS). The third-party integrating software will furnish a 12-lead ECG digital waveform as input to ECG-AI PH 12-Lead algorithm and record the output from the device for display or for printing in an offline report.
ECG-AI Pulmonary Hypertension 12-Lead algorithm is solely intended to analyze resting 12-lead ECGs acquired by compatible ECG device models. Automated quality checks are included to ensure that the signal is acquired with appropriate characteristics (e.g., filter, resolution, duration, etc.) from compatible devices, and do not contain excessive noise/artifacts.
ECG-AI Pulmonary Hypertension 12-Lead algorithm is not intended to process other ECGs, such as 6L or 2L ECGs, synthesized 12-lead ECG, data from continuous ECG monitoring, Holter monitoring, telemetry data, data from an ambulatory patient, or ECGs captured when patient is not in resting position.
Page 2 of 11
{6}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
# Predetermined Change Control Plan
This device has been cleared by the U.S. Food and Drug Administration (FDA) with a Predetermined Change Control Plan (PCCP). As part of this authorization, the device software may be updated periodically to enhance performance, including for higher sensitivity and/or specificity.
This modification, related to device performance improvement, will be validated through a multi-center retrospective clinical study that uses a combination of new and existing data. This approach involves supplementing the validation dataset with at least 10% new data for each iteration to ensure it remains current and representative. To be implemented, a modified version must demonstrate improved performance by meeting pre-specified acceptance criteria.
Users will be informed of each applicable update through revised labeling, release notes, or other appropriate communication channels. It is important to review updated instructions and performance summaries accompanying each software version to ensure continued proper use of the device.
# Summary of Non-Clinical Performance Data
The performance characteristics for the ECG-AI PH 12-Lead Algorithm have been evaluated with the following non-clinical testing: software verification and validation (per IEC 62304), cybersecurity, labeling verification and validation, and human factors.
# Summary of Clinical Performance Data
The performance characteristics for the ECG-AI PH 12-Lead Algorithm have been clinically validated for detection of pulmonary hypertension in adult patients presenting with dyspnea, with PPV and NPV greater than the study's predetermined acceptance criteria, and clinically useful sensitivity and specificity.
Table 1: Performance Characteristics
| Performance Characteristic | Value |
| --- | --- |
| Sensitivity | 73.0% |
| Specificity | 74.4% |
| Positive Predictive Value (Validation Study) | 28.7% |
| Negative Predictive Value (Validation Study) | 95.1% |
| Positive Predictive Value (Estimated Prevalence of US adults with dyspnea: 4.36%) | 11.5% |
| Negative Predictive Value (Estimated Prevalence of US adults with dyspnea: 4.36%) | 98.4% |
{7}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
# Algorithm Training
For the development and testing of the algorithm, data were obtained retrospectively from adult patients. Eligible patients were adults who had undergone RHC or had an echocardiogram with results available in digitized form and voltage–time data from at least one standard 10-s, 12-lead ECG. Patients with PH were defined as those with mean pulmonary arterial pressure (mPAP) >20 mmHg at rest if the patient had a right heart catheterization (RHC) performed, or if a patient could not be classified as PH positive or PH negative by RHC, a high probability of PH by echocardiogram, defined as tricuspid regurgitation velocity (TRV) >3.4 m/s at baseline. Control patients were defined as those with mPAP ≤20 mmHg on all available RHCs, or if no RHC was performed, TRV ≤2.8 m/s on all available echocardiograms.
A total of 282,927 adult patients (PH patients n=46,157; control patients n=236,770) with paired ECGs were retrospectively identified from Mayo Clinic. Among these patients, 259,227 had an ECG within 30 days before or after the date of PH diagnosis, or an ECG on or before last screening (for control patients), and were included in the diagnostic dataset. The diagnostic dataset included 39,823 PH patients (17,688 patients with dyspnea code) and 219,404 control patients (40,251 with dyspnea code). This dataset was sub-divided into the train (n=18,987 PH patients; n=105,260 controls), tune (n=4,661 PH; n=26,146 controls) and test (n=16,175 PH; n=87,998 controls) sets. The ECG-AI PH 12-Lead algorithm was trained on ECGs from the Mayo Clinic database taken +/-1 month of PH diagnosis (diagnostic dataset). Within the diagnostic test set, a total of 23,310 patients (n=7,155 PH patients, n=16,155 control patients) had a dyspnea ICD code recorded on or prior to an ECG and the diagnostic RHC or echocardiogram, and were used to test performance within the intended use population. Performance analyses were also conducted in subgroups of age, sex, and race.
{8}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
# Summary of Clinical Validation
## Study Design
The performance profile of the Anumana ECG-AI PH 12-Lead algorithm was validated in a retrospective study of 21,066 patient records across 5 health systems across the United States. The objective of the study was to establish the diagnostic performance of the device for the purpose of detecting the presence of a pulmonary hypertension in a clinically and demographically diverse population. The inclusion criteria for a subject was: age ≥ 18 with at least one dyspnea ICD code, at least one of either a right heart catheterization (RHC) with mean pulmonary arterial pressure (mPAP) measured or an echocardiogram (echo) with tricuspid regurgitation velocity (TRV) measured any time on or after a dyspnea ICD code; and the availability of at least one digital 12-lead ECG taken on or after a dyspnea ICD code and prior to or up to 30 days following an RHC or echo. The primary analysis was conducted using RHC and echo criteria to define cases and controls.
## Summary of Study Results
The 5 sites contributed to the final pool of 21,066 patient records. The study sample was representative of the US population and was 70.4% White, 16.6% Black/African American, 0.4% American Indian or Alaska Native, 2.1% Asian, 0.3% Native Hawaiian or Pacific Islander, 0.1% Multiracial, and 10.0% Other or Unknown, with 5.6% Hispanic or Latino. The sample consisted of 50.1% male and 49.9% female participants. 19.6% of participants were under age 50, 17.4% were 50-59, 26.8% were 60-69, and 36.2% were over 70.
Within this enriched study sample, a total of 10,036 PH Positive cases were identified from 21,066 records. The Anumana ECG-AI PH 12-Lead device achieved a sensitivity of 73.0% (95% CI of 72.1% to 73.9%), and a specificity of 74.4% (95% CI of 73.5% to 75.2%).
The PPV and NPV were analyzed using an adjusted prevalence of 4.36% based on a data set comprised of 183,348 unique adult patients who presented with an ICD code for dyspnea (ICD9: 786.0, 786.05; ICD10: R06.00, R06.01, R06.02, R06.09) and had received a standard 12-lead ECG. This dataset is from the Mayo Clinic, which includes tertiary specialty centers as well as community care facilities and is intended to represent an estimate of prevalence of PH in a generalized dyspnea adult population in a broad real-world setting. This resulted in a PPV of 11.5% (95% CI of 11.2% to 11.8%) and NPV of 98.4% (95% CI of 98.3% to 98.4%). Performance is shown in Table 1.
Subgroup analyses results provided in Table 2: Subgroup Analyses.
## Subgroup Analyses
Subgroup assessments of diagnostic performance were conducted to evaluate the device across varying clinical and demographic populations. Table 2 presents the sensitivity and specificity of the ECG-AI PH 12-Lead algorithm across key patient subgroups. Note: Additional performance evaluations across specific clinical sites, ECG manufacturers and device models, patient comorbidities, and underlying ECG rhythms were conducted and maintained comparable safety profiles. Full details are available in the Clinical Study Report.
Page 5 of 11
{9}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
Table 2: Subgroup Analyses
| Subgroup | Sensitivity (95% CI) (n/N) | Specificity (95% CI) (n/N) |
| --- | --- | --- |
| Overall Performance | 73.0% (72.1%, 73.9%) (7,328/10,036) | 74.4% (73.5%, 75.2%) (8,202/11,030) |
| PH Screen Type | | |
| Echo | 80.8% (79.0%, 82.5%) (1,629/2,017) | 75.8% (74.9%, 76.7%) (6,922/9,128) |
| Right Heart Catheterization (RHC) | 71.1% (70.1%, 72.1%) (5,699/8,019) | 67.3% (65.1%, 69.4%) (1,280/1,902) |
| Biological Sex | | |
| Female | 68.9% (67.5%, 70.2%) (3,226/4,685) | 78.2% (77.1%, 79.3%) (4,550/5,818) |
| Male | 76.7% (75.5%, 77.8%) (4,102/5,351) | 70.1% (68.8%, 71.3%) (3,652/5,212) |
| Age at Encounter | | |
| < 50 Years | 77.5% (75.5%, 79.4%) (1,409/1,819) | 79.4% (77.7%, 81.0%) (1,835/2,312) |
| 50 – 59 Years | 72.7% (70.6%, 74.8%) (1,300/1,787) | 76.4% (74.4%, 78.3%) (1,437/1,882) |
| 60 – 69 Years | 69.9% (68.1%, 71.5%) (1,958/2,803) | 74.6% (73.0%, 76.2%) (2,117/2,838) |
| 70+ Years | 73.4% (71.9%, 74.8%) (2,661/3,627) | 70.4% (68.9%, 71.8%) (2,813/3,998) |
| Race | | |
| American Indian/Alaska Native | 80.9% (66.7%, 90.9%) (38/47) | 77.1% (59.9%, 89.6%) (27/35) |
| Asian | 76.6% (69.5%, 82.8%) (128/167) | 75.9% (70.4%, 80.9%) (208/274) |
| Black | 80.4% (78.6%, 82.1%) (1,585/1,971) | 71.6% (69.2%, 73.8%) (1,098/1,534) |
| Multiracial | 54.5% (23.4%, 83.3%) (6/11) | 78.9% (54.4%, 93.9%) (15/19) |
| Native Hawaiian or Other Pacific Islander | 80.6% (64.0%, 91.8%) (29/36) | 66.7% (43.0%, 85.4%) (14/21) |
| White | 69.8% (68.7%, 70.9%) (4,725/6,769) | 74.7% (73.7%, 75.6%) (6,027/8,068) |
| Other or Unknown | 78.9% (76.3%, 81.4%) (817/1,035) | 75.3% (72.7%, 77.9%) (813/1,079) |
| Ethnicity | | |
| Hispanic or Latino | 76.9% (73.1%, 80.5%) (407/529) | 76.6% (73.2%, 79.8%) (498/650) |
| Not Hispanic or Latino | 72.8% (71.9%, 73.7%) (6,844/9,400) | 74.1% (73.2%, 74.9%) (7,501/10,124) |
| Unknown | 72.0% (62.5%, 80.2%) (77/107) | 79.3% (73.8%, 84.1%) (203/256) |
{10}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
| Body Mass Index (BMI) | | |
| --- | --- | --- |
| < 30 kg/m² | 76.7% (75.5%, 77.9%)
(3,624/4,725) | 74.5% (73.3%, 75.6%)
(4,135/5,552) |
| 30 – 34 kg/m² | 70.3% (68.2%, 72.3%)
(1,372/1,953) | 76.5% (74.6%, 78.4%)
(1,459/1,906) |
| 35+ kg/m² | 64.2% (62.3%, 66.2%)
(1,526/2,376) | 76.6% (74.5%, 78.6%)
(1,298/1,695) |
| Unknown | 82.1% (79.5%, 84.4%) (806/982) | 69.8% (67.7%, 71.9%)
(1,310/1,877) |
Page 7 of 11
{11}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
# Substantial Equivalence
The subject device, Anumana ECG-AI PH 12-Lead algorithm, is substantially equivalent to the predicate, ECG-AI LEF 12-Lead algorithm (K250652). The devices have equivalent intended uses, principles of operation, and technical characteristics. Where differences occur between the subject device and the predicates, results of clinical performance and results of non-clinical verification and validation demonstrate that the subject device does not raise new questions of safety and effectiveness.
Table 3: Substantial Equivalence Comparison of Subject Device to the Predicate
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana Inc. | Anumana Inc. | |
| Product Name | ECG-AI PH 12-Lead algorithm (v3.1.0) | ECG-AI LEF 12-Lead algorithm (v2.5.0) | |
| Application No. | K252360 | K250652 | |
| Product Code | SAT | QYE | Equivalent. The SAT product code is for Pulmonary Hypertension, specifically. QYE is for low ejection fraction. Both Product Codes are assigned to the Regulation number 21 CFR 870.2380. |
| Regulation No. | 21 CFR 870.2380 | 21 CFR 870.2380 | Identical. |
| Regulation Name | Cardiovascular machine learning-based notification software | Cardiovascular machine learning-based notification software | Identical. |
| Intended Use/Indications for Use | The ECG-AI PH 12-Lead algorithm is software intended to aid in earlier detection of elevated mean pulmonary arterial pressure (mPAP), an indicator of pulmonary hypertension, in adults presenting with dyspnea. ECG-AI PH 12-Lead algorithm is not intended to be a stand-alone | The ECG-AI LEF 12-Lead algorithm is software intended to aid in earlier detection of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults (aged 18 and above) at risk for heart failure. This population includes, but is not limited to: • patients with cardiomyopathies • patients who are post-myocardial infarction • patients with aortic stenosis | Equivalent. The subject device intended use is equivalent to the predicate in that both are intended to aid in earlier detection of a cardiovascular disease or condition. The subject device and predicate have equivalent language to address the need for further clinical evaluation to establish diagnosis since neither device is intended for standalone use. |
{12}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana Inc. | Anumana Inc. | |
| Product Name | ECG-AI PH 12-Lead algorithm (v3.1.0) | ECG-AI LEF 12-Lead algorithm (v2.5.0) | |
| | diagnostic device for pulmonary hypertension or replace current clinical practice guidelines.
A positive result may suggest the need for further clinical evaluation. Additionally, if the patient is at high risk for pulmonary hypertension, a negative result should not rule out additional clinical evaluation.
ECG-AI PH 12-Lead algorithm should be applied jointly with clinician judgment.
The ECG-AI PH 12-Lead algorithm should not be used on ECGs from adults with an implanted pacemaker or on ECGs with a paced rhythm. | • patients with chronic atrial fibrillation
• patients receiving pharmaceutical therapies that are cardiotoxic, and
• postpartum women.
The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.
A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.
The ECG-AI LEF 12-Lead algorithm should be applied jointly with clinical judgement. | |
| Diagnostic Application | Pulmonary Hypertension | Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% | Equivalent. The predicate device is designed for a different disease population but the subject device similarly helps to detect Pulmonary Hypertension. Additional clinical validation was conducted to support equivalence. |
Page 9 of 11
{13}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana Inc. | Anumana Inc. | |
| Product Name | ECG-AI PH 12-Lead algorithm (v3.1.0) | ECG-AI LEF 12-Lead algorithm (v2.5.0) | |
| Target Population | Adults presenting with dyspnea | Adults at risk for heart failure | Equivalent. Both devices are intended to target adults with cardiovascular symptoms and/or risk factors. |
| Intended User | Clinicians | Clinicians | Identical. |
| Principal of Operation | Machine learning-based algorithm | Machine learning-based algorithm | Identical. |
| Data Acquisition | Acquires data from 12-L ECGs | Acquires data from 12-L ECGs | Identical. Subject device and predicate both use 12-L ECGs for data acquisition. |
| Device Output Format | Algorithm output is provided to third party software that displays a binary result to clinicians. Output provided for each ECG is “Detected,” “Not Detected” or “Error”. | Algorithm output is provided to third party software that displays a binary result to clinicians. Output provided for each ECG is “Detected,” “Not Detected” or “Error”. | Identical. Subject and predicate devices provide binary output based on calculations from data for a given disease state to indicate likelihood of disease. |
| Cybersecurity | Compliant with FDA Guidance on Cybersecurity | Compliant with FDA Guidance on Cybersecurity | Identical. Subject and predicate device algorithms are compliant with FDA Guidance and Standards for Cybersecurity. |
Page 10 of 11
{14}
eSTAR Traditional 510(k) Submission – ECG-AI PH 12-Lead Algorithm
K252360 510(k) Summary
# Benefit-Risk Determination
The probable benefits of the ECG-AI PH outweigh the probable risks when used as an aid in the early detection of elevated mean pulmonary arterial pressure (mPAP), an indicator of pulmonary hypertension (PH), in adult patients with dyspnea.
**Probable Benefits**: The primary benefit of the device is the ability to provide an immediate, non-invasive, and accessible result for PH detection using a standard 12-lead ECG, as PH is frequently underdiagnosed or diagnosed late in the disease course. In the clinical validation study, the device demonstrated Sensitivity (73.0%) and Specificity (74.4%), as well as a Positive (11.5%) and Negative Predictive Value (98.4%) at the projected intended use prevalence (4.36%).
**Probable Risks and Mitigation**: The primary risks associated with the device are False Negative and False Positive results.
- **False Negatives**: A false negative may result in a delayed diagnosis of PH. The device is indicated as an adjunct to clinical judgment; a negative result does not preclude further workup if clinical suspicion remains high.
- **False Positives**: A false positive result may lead to unnecessary follow-up testing or referral to a specialist. The device is intended to be used as an adjunct and should be used in context of the clinician's clinical suspicion for PH.
- **Signal Quality**: The device incorporates a robust quality assessment module that prevents analysis of poor-quality signals, outputting an "ERROR" rather than an unreliable prediction, further mitigating the risk of incorrect outputs due to noise or artifacts.
**Overall Conclusion**: The clinical and non-clinical data demonstrate that the ECG-AI PH is substantially equivalent to the predicate device for its intended use. The performance profile supports its use as a detection aid to prioritize patients for downstream diagnostic evaluation.
Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.