K252347 · Dongguan Smarfody Mobility Technology Co., Ltd. · INI · Jan 29, 2026 · Physical Medicine
Device Facts
Record ID
K252347
Device Name
Electric Scooter (Mojo-T580)
Applicant
Dongguan Smarfody Mobility Technology Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Jan 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
Electric scooter (Mojo-T580) provides mobility for users restricted to sitting position; features metal frame, four wheels, seat, LCD display, electric motor, electromagnetic brake, and rechargeable lithium battery. Rider operates via control lever and speed knob; electromagnetic brake engages automatically when not in use. Device is foldable manually and intended for use on flat surfaces. Provides mobility assistance to patients; improves independence. Operated by patient.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance demonstrated through compliance with ISO 7176 series (stability, braking, distance, speed, dimensions, fatigue, climatic, obstacle-climbing), ISO 10993-1 (biocompatibility), ISO 7176-21/IEC 60601-1-2 (EMC), and IEC 62133-2 (battery safety).
Technological Characteristics
Motorized four-wheeled vehicle; metal frame; electric motor; electromagnetic brake; rechargeable lithium battery; LCD display. Complies with ISO 7176 series (1-11, 13-16, 21), ISO 16840-10 (ignition resistance), and IEC 62133-2 (battery). Manual folding mechanism. Standalone operation.
Indications for Use
Indicated for adults restricted to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 29, 2026
Dongguan Smarfody Mobility Technology Co., Ltd.
% Eva Li
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave.
Shanghai, 200122
China
Re: K252347
Trade/Device Name: Electric Scooter (Mojo-T580)
Regulation Number: 21 CFR 890.3800
Regulation Name: Motorized Three-Wheeled Vehicle
Regulatory Class: Class II
Product Code: INI
Dated: July 29, 2025
Received: July 29, 2025
Dear Eva Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252347 - Eva Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252347 - Eva Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252347 | |
| Device Name Electric Scooter (Mojo-T580) | |
| Indications for Use (Describe) The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252347
# 510(K) Summary
## Application Information:
Dongguan Smarfody Mobility Technology Co., Ltd.
Address: Room 202, Building2, No.9 Junpu Industry 1st Road, Houjie Town, Dongguan City, Guangdong Province, P.R. China
Telephone Number: 0086 13798786794
Contact Person: Alex Wu
## Device Name: Electric Scooter
Model: Mojo-T580
Classification: II
Product Code: INI
Regulation: 890.3800
Panel: Motorized three-wheeled vehicle
Date Prepared: Jan 27,2026
## Predicate Device
510K number: K232692
Device Name: Powered Mobility Scooter
Model: PMS101, GUT112, PMS103, GUT140
Manufacturer: Intradin (Shanghai) Machinery Co., Ltd
Indication for use: The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Patient Population: Adult only
## 1. Device Description
The Electric Scooter, has a base with metal frame, two front wheels, two rear wheels, a seat, a LCD Display, electric motor, electromagnetic brake, 1 rechargeable Lithium Battery with an off-board charger. The movement of the scooter is controlled by the rider who operates the control lever, speed control turn "speed" Knob. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter only can be operated on the flat surface. The device can be folded from up to down by manual.
{5}
# 2. Comparison with Predicate Device
Table 1 General Comparison
| Elements of Comparison | Subject Device | Predicate Device (K232692) | Remark |
| --- | --- | --- | --- |
| Manufacturer | Dongguan Smarfody Mobility Technology Co., Ltd. | Intradin (Shanghai) Machinery Co., Ltd. | NA |
| Device name | Electric Scooter | Powered Mobility Scooter | NA |
| Model(s) | Mojo-T580 | PMS101 | NA |
| Product Code | INI | INI | Same |
| Device Classification Regulation | 890.3800 | 890.3800 | Same |
| Indications for use | The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. | The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. | Same |
| Number of Wheels | 4 | 3 | Similar |
| Overall dimension (Length x Width x Height) | 1030mm x 530mm x 960mm | 1025mm x 595mm x 940mm | Similar |
| Folded Dimensions (Length x Width x Height) | 1000mm x 440mm x 385mm | 640mm x 595mm x 405mm | Similar |
{6}
Table 2 safety comparison
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| Biocompatibility | All user directly contacting materials are compliance with ISO 10993-1 | All user directly contacting materials are compliance with ISO 10993-5, ISO 10993-10, and ISO 10993-23 requirements | S.E. |
| EMC | ISO7176-21& IEC 60601-1-2:2014+A1:2020 | ISO7176-21& IEC 60601-1-2 | S.E. |
| Biotransf. (SPE) | SPE | SPE | S.E. |
| Biotransf. (SPE) | SPE | SPE | S.E. |
| Biotransf. (SPE) | SPE | SPE | S.E. |
| Biotransf. (SPE) | SPE | SPE | S.E. |
| Biotransf. (SPE) | SPE | SPE | S.E. |
{7}
| Performance | ISO7176 series | ISO7176 series | S.E. |
| --- | --- | --- | --- |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. |
4 / 7
{8}
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| ISO7176-1 | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | S.E. |
| ISO7176-2 | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | S.E. |
| ISO7176-3 | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | S.E. |
| ISO7176-4 | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | S.E. |
| ISO7176-5 | The dimensions, mass has been determined after the testing according to the ISO 7176-5, | The dimensions, mass has been determined after the testing according to the ISO 7176-5, | S.E. |
| ISO7176-6 | The maximum speed has been determined after the testing according to the ISO 7176-6, | The maximum speed has been determined after the testing according to the ISO 7176-6, | S.E. |
| ISO7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7, | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7, | S.E. |
| ISO7176-8 | All test results meet the requirements in Clause 4 of ISO 7176-8 | All test results meet the requirements in Clause 4 of ISO 7176-8 | S.E. |
| ISO7176-9 | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9 | The test results shown that the device under tests could continue to function according to manufacturer’s specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9 | S.E. |
| ISO7176-10 | The obstacle-climbing ability of device has been determined after the | The obstacle-climbing ability of device has been determined after | S.E. |
{9}
6 / 7
| | testing according to the ISO 7176-10, | the testing according to the ISO 7176-10, | |
| --- | --- | --- | --- |
| ISO7176-11 | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11 | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11 | S.E. |
| ISO7176-13 | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved | S.E. |
| ISO7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14 | S.E. |
| ISO7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15 | S.E. |
| 16840-10 | The performance of resistance to ignition meet the requirements of 16840-10 | The performance of resistance to ignition meet the requirements of 16840-10 | S.E. |
| ISO 7176-21 | The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020 | The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020 | S.E. |
| | The performance of batteries and charger of device meet the Requirements of ISO 7176-31 and IEC 62133-2 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25 | S.E. |
## Substantial Equivalence Discussion
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 7176-16, ISO 7176-21, ISO 16840-10, and IEC 62133-2 and FDA guidance for Scooter.
The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the
{10}
test results show that the subject product is substantially equivalent to the predicate device in performance.
The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (Scooter tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.
The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
## 3. Substantially Equivalency Conclusion
Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate devices, Powered Mobility Scooter from Intradin (Shanghai) Machinery Co., Ltd under K232692.
7 / 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.