EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-Q190); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-H190 ); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-1TH190)
K252341 · Olympus Medical Systems Corp. · EOQ · Apr 23, 2026 · Ear, Nose, Throat
Device Facts
Record ID
K252341
Device Name
EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-Q190); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-H190 ); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-1TH190)
Applicant
Olympus Medical Systems Corp.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Apr 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Device Story
Flexible bronchovideoscope (models BF-Q190, BF-H190, BF-1TH190) used for visualization, diagnosis, and treatment within airways and tracheobronchial tree. Device consists of control section (angulation knobs), insertion section (image/light guides, instrument channel), and connector section (to video system center). Physician operates device in clinical settings; uses manual angulation to navigate distal tip; utilizes suction and air/water valves for lens cleaning and fluid removal. Therapeutic instruments (e.g., biopsy forceps) pass through instrument channel. Dual focus mechanism (near/normal) uses Voice Coil Motor (VCM) actuator. Output is real-time endoscopic video displayed on monitor for clinical decision-making. This submission updates the operation manual regarding safe use with laser, high-frequency, and Argon Plasma Coagulation (APC) systems.
Clinical Evidence
No clinical or animal testing performed. Evidence consists of bench testing, specifically ignition factor safety studies for laser, high-frequency, and APC systems, and human factors validation testing.
Technological Characteristics
Flexible bronchovideoscope; CCD-based imaging; Voice Coil Motor (VCM) for dual-focus mechanism; NBI (Narrow Band Imaging) capability. Sterilization via Ethylene Oxide or STERRAD NX. Connectivity via proprietary one-touch connector to Olympus video system center. No changes to materials, dimensions, or software from predicate.
Indications for Use
Indicated for use within the airways and tracheobronchial tree for endoscopy and endoscopic surgery in patients requiring diagnostic or therapeutic bronchoscopy.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
EVIS EXERA III BRONCHOVIDEOSCOPE BF-Q190, BF-H190, BF-1TH190 (K121959)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 23, 2026
Olympus Medical Systems Corp.
Teffany Hutto
Project Manager, Regulatory Affairs
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581
Re: K252341
Trade/Device Name: Evis Exera III Bronchovideoscope (Olympus BF-Q190); Evis Exera III Bronchovideoscope (Olympus BF-H190); Evis Exera III Bronchovideoscope (Olympus BF-1TH190)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: March 24, 2026
Received: March 24, 2026
Dear Teffany Hutto:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K252341 - Teffany Hutto
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K252341 - Teffany Hutto
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOYCE C. LIN -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252341 | |
| Device Name EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-Q190); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-H190); EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-1TH190) | |
| Indications for Use (Describe) This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
Date Prepared: April 23, 2026
# 510(k) Summary
# K252341
## 1. GENERAL INFORMATION
■ 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan
■ Contact Person: Teffany Hutto
Olympus Corporation of the Americas
3500 Corporate Parkway PO Box 610
Center Valley, PA 18034-0610, USA
Phone: 512-508-6550
Fax: 484-896-7128
Email: teffany.hutto@olympus.com
■ Manufacturing site: Aizu Olympus Co., Ltd.,
3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan
## 2. DEVICE IDENTIFICATION
| Trade/Device Name | EVIS EXERA III Bronchovideoscope BF-Q190 |
| --- | --- |
| Model Name | BF-Q190 |
| Common Name | Bronchovideoscope |
| Regulation Number | 874.4680 |
| Regulation Name | Bronchoscope (flexible or rigid) and accessories |
| Regulatory Class | II |
| Product Code | EOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum) |
| Classification Panel | Ear, Nose and Throat |
510(k) Summary
Page 1 of 12
{5}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
| Trade/Device Name | EVIS EXERA III Bronchovideoscope BF-H190 |
| --- | --- |
| Model Name | BF-H190 |
| Common Name | Bronchovideoscope |
| Regulation Number | 874.4680 |
| Regulation Name | Bronchoscope (flexible or rigid) and accessories |
| Regulatory Class | II |
| Product Code | EOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum) |
| Classification Panel | Ear, Nose and Throat |
| Trade/Device Name | EVIS EXERA III Bronchovideoscope BF-1TH190 |
| --- | --- |
| Model Name | BF-1TH190 |
| Common Name | Bronchovideoscope |
| Regulation Number | 874.4680 |
| Regulation Name | Bronchoscope (flexible or rigid) and accessories |
| Regulatory Class | II |
| Product Code | EOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum) |
| Classification Panel | Ear, Nose and Throat |
# 3. PREDICATE DEVICES
| Device name | 510(k) Submitter | 510(k) No. |
| --- | --- | --- |
| EVIS EXERA III BRONCHOVIDEOSCOPE
BF-Q190, BF-H190, BF-1TH190 | OLYMPUS MEDICAL SYSTEMS CORP. | K121959 |
# 4. DEVICE DESCRIPTION
The EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-Q190, OLYMPUS BF-H190, and BF-1TH190) are used for endoscopic diagnosis and treatment within the respiratory organs. These endoscopes consist of three parts: the control section, the insertion section, and the connector section.
510(k) Summary
Page 2 of 12
{6}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
# 5. PRINCIPLE OF OPERATION
The endoscope consists of three parts: the control section, the insertion section, and the connector section. The basic principle including user interface and operation for the procedure of the endoscope is identical to that of the predicate device.
## 1) Control section
The UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob on the control section is connected to the tip of the bending section by a series of wires. By operating the UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob, the bending section at the distal end bends vertically or parallel to guide the distal end for insertion and observation.
The observation mode can be selected by focus switching function, “near focus mode” featuring ground-breaking resolving power for close observation or “normal focus mode” for normal observation. To realize the dual focus mechanism, Voice Coil Motor (VCM) is incorporated as an actuator.
The endoscope contains a cylinder to attach a suction valve for suction and air/water valve. Depressing the suction valve will allow the physician to use the endoscope to suction any fluids which are obscuring a good view of the tissue. Therapeutic instruments can be passed through the instrument channel for performing endoscopic biopsy and other therapies. Depressing the air/water valve will allow the doctor to feed water through the endoscope for lens washing. It also can be operated to feed air for removing any fluids or debris adhering to the objective lens.
## 2) Insertion section
The insertion section has main parts including the image guide, light guides that bring light from the video system center through the endoscope, and instrument channel where therapeutic tools can be pushed in and out (also the suction channel).
## 3) Connector section
The connector section connects the endoscope with the video system center (CV-1500) through the universal cord.
510(k) Summary
Page 3 of 12
{7}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
## 6. INDICATIONS FOR USE
EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-Q190, OLYMPUS BF-H190, and OLYMPUS BF-1TH190) is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
This instrument is indicated for use within the airways and tracheobronchial tree.
## INDICATIONS FOR USE COMPARISON
The Indications for Use for the Subject Device is exactly the same as the Indications for Use for the Predicate Device.
## 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject devices have the same technological characteristics and design as the applicable predicate devices. There have been no modifications from the prior clearances of these devices including:
- Design
- Materials
- Sterilization
- Shelf Life
- Reprocessing
- Packaging
- Software
This 510(k) is a notification to communicate an update to the Operation Manual to provide new instructions on the proper use of the subject devices with laser, high frequency, and Argon Plasma Coagulation (APC) systems.
A side-by-side comparison of the subject device and the predicate device is provided below.
510(k) Summary
Page 4 of 12
{8}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
BF-Q190
| Specifications | Subject device BF-Q190 | Predicate device BF-Q190 |
| --- | --- | --- |
| Indications for Use | This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
This instrument is indicated for use within the airways and tracheobronchial tree. | |
| Common name | BRONCHOVIDEOSCOPE | BRONCHOVIDEOSCOPE |
| Regulation name | Bronchoscope (flexible or rigid) and accessories | Bronchoscope (flexible or rigid) and accessories |
| Regulation number | 874.4680 | 874.4680 |
| Product code | EOQ, EOB, NWB | EOQ, EOB, NWB |
| Classification panel | Ear Nose & Throat | Ear Nose & Throat |
| Sterilization | Ethylene oxide sterilization
STERRAD NX | Ethylene oxide sterilization
STERRAD NX |
Specifications
| Image capture system | 1 Field of View | 120° | 120° |
| --- | --- | --- | --- |
| | 2 Depth of Field | 3-100mm | 3-100mm |
| | 3 Standard Focal Length | 0.558mm | 0.558mm |
| | 4 Optimum Working Distance | 5.7mm | 5.7mm |
| | 5 Direction of View | 0°(Forward viewing) | 0°(Forward viewing) |
| | 6 Highest Resolution | 15.85 line air/mm(Typical) | 15.85 line pair/mm(Typical) |
| | 7 Distortion | | |
| | 8 Magnification of Objective Lens (at Optimum) | 0.093 | 0.093 |
| | 9 F# of Objective Lens (at Optimum) | 5.066 | 5.066 |
| | 10 Total Number of Pixels | 168,096 | 168,096 |
510(k) Summary
Page 5 of 12
{9}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
| Specifications | | Subject device BF-Q190 | Predicate device BF-Q190 |
| --- | --- | --- | --- |
| | 11 Pixels per square mm | 226,757 | 226,757 |
| | 12 Size of Pixel | 2.1μm(H) x 2.1μm(V) | 2.1μm(H) x 2.1μm(V) |
| | 13 Active Area of CCD Chip | 0.8652mm(H) x 0.8568mm(V) | 0.8652mm(H) x 0.8568mm(V) |
| | 14 Type of CCD Chip | Color | Color |
| | Outer Diameter of Distal End | φ4.8mm | φ4.8mm |
| | Outer Diameter of Insertion Tube | φ6.0mm | φ6.0mm |
| | Angulation UP/DOWN | UP:210°, DOWN:130° | UP:210°, DOWN:130° |
| | Working Length | 600mm | 600mm |
| | Inner Diameter of Instrument Channel | φ2.0mm | φ2.0mm |
| | Number of Light guide fibers | 820 | 820 |
| | NBI observation | Available | Available |
| | Insertion tube rotation | 240 deg. | 240 deg. |
| Scope Connector | | Water Resistant One Touch Connector | Water Resistant One Touch Connector |
510(k) Summary
Page 6 of 12
{10}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
BF-H190
| Specifications | | Subject device BF-H190 | Predicate device BF-H190 |
| --- | --- | --- | --- |
| Indications for Use | | This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
This instrument is indicated for use within the airways and tracheobronchial tree. | |
| Common name | | BRONCHOVIDEOSCOPE | BRONCHOVIDEOSCOPE |
| Regulation name | | Bronchoscope (flexible or rigid) and accessories | Bronchoscope (flexible or rigid) and accessories |
| Regulation number | | 874.4680 | 874.4680 |
| Product code | | EOQ, EOB, NWB | EOQ, EOB, NWB |
| Classification panel | | Ear Nose & Throat | Ear Nose & Throat |
| Sterilization | | Ethylene oxide sterilization
STERRAD NX | Ethylene oxide sterilization
STERRAD NX |
| Specifications | | | |
| Image capture system | 1 Field of View | 120° | 120° |
| | 2 Depth of Field | 3-100mm | 3-100mm |
| | 3 Standard Focal Length | 0.946mm | 0.946mm |
| | 4 Optimum Working Distance | 6.5mm | 6.5mm |
| | 5 Direction of View | 0°(Forward viewing) | 0°(Forward viewing) |
| | 6 Highest Resolution | 17.78 line pair/mm(Typical) | 17.78 line pair/mm(Typical) |
| | 7 Distortion | | |
| | 8 Magnification of Objective Lens (at Optimum) | 0.138 | 0.138 |
| | 9 F# of Objective Lens (at Optimum) | 7.88 | 7.88 |
| | 10 Total Number of Pixels | 492,980 | 492,980 |
510(k) Summary
Page 7 of 12
{11}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
| Specifications | | Subject device BF-H190 | Predicate device BF-H190 |
| --- | --- | --- | --- |
| | 11 Pixels per square mm | 226,757 | 226,757 |
| | 12 Size of Pixel | 2.1μm(H) x 2.1μm(V) | 2.1μm(H) x 2.1μm(V) |
| | 13 Active Area of CCD Chip | 1.674mm(H) x 1.336mm(V) | 1.674mm(H) x 1.336mm(V) |
| | 14 Type of CCD Chip | Color | Color |
| | Outer Diameter of Distal End | φ5.5mm | φ5.5mm |
| | Outer Diameter of Insertion Tube | φ6.6mm | φ6.6mm |
| | Angulation UP/DOWN | UP:210°, DOWN:130° | UP:210°, DOWN:130° |
| | Working Length | 600mm | 600mm |
| | Inner Diameter of Instrument Channel | φ2.0mm | φ2.0mm |
| | Number of Light guide fibers | 1200 | 1200 |
| | NBI observation | Available | Available |
| | Insertion tube rotation | 240 deg. | 240 deg. |
| Scope Connector | | Water Resistant One Touch Connector | Water Resistant One Touch Connector |
510(k) Summary
Page 8 of 12
{12}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
BF-1TH190
| Specifications | | Subject device BF-1TH190 | Predicate device BF-1TH190 |
| --- | --- | --- | --- |
| Indications for Use | | This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
This instrument is indicated for use within the airways and tracheobronchial tree. | |
| Common name | | BRONCHOVIDEOSCOPE | BRONCHOVIDEOSCOPE |
| Regulation name | | Bronchoscope (flexible or rigid) and accessories | Bronchoscope (flexible or rigid) and accessories |
| Regulation number | | 874.4680 | 874.4680 |
| Product code | | EOQ, EOB, NWB | EOQ, EOB, NWB |
| Classification panel | | Ear Nose & Throat | Ear Nose & Throat |
| Sterilization | | Ethylene oxide sterilization STERRAD NX | Ethylene oxide sterilization STERRAD NX |
| Specifications | | | |
| Image capture system | 1 Field of View | 120° | 120° |
| | 2 Depth of Field | 3-100mm | 3-100mm |
| | 3 Standard Focal Length | 0.946mm | 0.946mm |
| | 4 Optimum Working Distance | 6.5mm | 6.5mm |
| | 5 Direction of View | 0°(Forward viewing) | 0°(Forward viewing) |
| | 6 Highest Resolution | 17.78 line pair/mm(Typical) | 17.78 line pair/mm(Typical) |
| | 7 Distortion | | |
| | 8 Magnification of Objective Lens (at Optimum) | 0.138 | 0.138 |
| | 9 F# of Objective Lens (at Optimum) | 7.88 | 7.88 |
| | 10 Total Number of Pixels | 492,980 | 492,980 |
510(k) Summary
Page 9 of 12
{13}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
| Specifications | | Subject device BF-1TH190 | Predicate device BF-1TH190 |
| --- | --- | --- | --- |
| | 11 Pixels per square mm | 226,757 | 226,757 |
| | 12 Size of Pixel | 2.1μm(H) x 2.1μm(V) | 2.1μm(H) x 2.1μm(V) |
| | 13 Active Area of CCD Chip | 1.674mm(H) x 1.336mm(V) | 1.674mm(H) x 1.336mm(V) |
| | 14 Type of CCD Chip | Color | Color |
| | Outer Diameter of Distal End | φ6.2mm | φ6.2mm |
| | Outer Diameter of Insertion Tube | φ7.2mm | φ7.2mm |
| | Angulation UP/DOWN | UP:180°, DOWN:130° | UP:180°, DOWN:130° |
| | Working Length | 600mm | 600mm |
| | Inner Diameter of Instrument Channel | φ2.8mm | φ2.8mm |
| | Number of Light guide fibers | 1200 | 1200 |
| | NBI observation | Available | Available |
| | Insertion tube rotation | 240 deg. | 240 deg. |
| Scope Connector | | Water Resistant One Touch Connector | Water Resistant One Touch Connector |
510(k) Summary
Page 10 of 12
{14}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
# 8. PERFORMANCE DATA
The following performance data were provided in support of the labeling change to address the proper use of the subject devices with laser, high frequency, and APC systems.
## 1) Performance testing - Bench
Bench testing as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.
- Ignition Factors Safety Study (Laser)
- Ignition Factors Safety Study (High-Frequency)
- Ignition Factors Safety Study (APC)
- Ignition Factors Safety Study (Scope Comparison)
- Human Factors Validation Testing
## 2) Performance testing – Animal/Clinical
Animal and clinical testing were not performed.
## 3) Risk management
Risk management was performed in accordance with ISO 14971:2019. The design verification tests and their acceptance criteria were performed and identified as a result of this risk management.
## 4) Testing not completed
Since there were no design, material, sterilization, reprocessing, packaging, shelf life, or software changes, the following testing was not performed:
- Biocompatibility Testing
- Sterilization / Reprocessing
- Shelf Life
- Stability Testing
- Package Integrity Test
- Software Testing and Cybersecurity
- Electrical Safety and EMC
- Residual Toxicity of Reprocessing Chemicals
- Clinical
510(k) Summary
Page 11 of 12
{15}
OLYMPUS
EVIS EXERA III BRONCHOSCOPE
OLYMPUS BF-Q190, BF-H190 and BF-1TH190
# 9. CONCLUSIONS
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-Q190, OLYMPUS BF-H190, and BF-1TH190) raise no new questions of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, effectiveness, and performance.
510(k) Summary
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.