K252337 · Edgecare, Inc. · MUI · Apr 24, 2026 · Radiology
Device Facts
Record ID
K252337
Device Name
EdgeFlow Gel Pad
Applicant
Edgecare, Inc.
Product Code
MUI · Radiology
Decision Date
Apr 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The EdgeFlow Gel Pad is a non-sterile, single-use accessory intended for use exclusively with the EdgeFlow UW20 wearable bladder scanner. The device provides acoustic coupling between the ultrasound transducer and the patient's intact abdominal skin during bladder ultrasound procedures, enabling transmission of ultrasound signals required for bladder imaging and bladder volume measurement. It is intended for use in professional healthcare facilities, such as hospitals and clinics, by qualified and trained healthcare professionals.
Device Story
EdgeFlow Gel Pad is a single-use, solid silicone elastomer pad; functions as an acoustic interface between the EdgeFlow UW20 wearable bladder scanner transducer and patient's intact abdominal skin. Device facilitates ultrasound signal transmission for bladder imaging and volume measurement. Used in professional healthcare facilities (hospitals/clinics) by trained professionals. Pad is applied to the transducer; provides stable coupling during 24-hour wearable monitoring. Benefits include consistent acoustic contact, enabling accurate bladder volume assessment without liquid gel mess. Output is raw ultrasound signal transmission to the scanner; clinical decision-making relies on the resulting bladder volume measurements.
Clinical Evidence
Prospective, single-center in-vivo study (n=10 adults) at Severance Hospital evaluated 24-hour acoustic coupling stability. 1,454 images analyzed; 1,414 maintained valid coupling (97.24% retention ratio), exceeding 80% acceptance criterion. Minimal skin discomfort/slippage reported; no adverse events. Bench testing confirmed axial/lateral distance accuracy (≤1 mm) and bladder volume accuracy (≤7.5%) using phantoms.
Technological Characteristics
Solid silicone elastomer gel pad; non-sterile; single-use. Dimensions/form factor: patch. Acoustic properties: 970 m/s sound velocity, 0.97 MRayl impedance, 0.080 dB/mm/MHz attenuation. Hardness: 55.4 Shore A. Biocompatibility: ISO 10993-5, -10. Packaging: sealed pouch, ASTM F1929 (dye penetration) and ASTM F88/F88M (seal strength) compliant. Shelf life: 17 months.
Indications for Use
Indicated for use as an acoustic coupling medium for patients undergoing bladder ultrasound procedures using the EdgeFlow UW20 wearable bladder scanner; intended for application on intact abdominal skin.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
AQUASONIC 100 Ultrasound Gel (K802146), PARKER LABORATORIES
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Edgecare Inc.
Beom Ki Cha
RA Senior Manager
12F, 8 Yangpyeong-ro 25-gil
Yeongdeungpo-gu, Seoul 07207
SOUTH KOREA
Re: K252337
Trade/Device Name: EdgeFlow Gel Pad
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: MUI
Dated: March 25, 2026
Received: March 25, 2026
Dear Beom Ki Cha:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
April 24, 2026
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252337 - Beom Ki Cha
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252337 - Beom Ki Cha
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.04.24 10:21:54 -04'00' For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (303) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252337 | |
| Device Name EdgeFlow Gel Pad | |
| Indications for Use (Describe) The EdgeFlow Gel Pad is a non-sterile, single-use accessory intended for use exclusively with the EdgeFlow UW20 wearable bladder scanner. The device provides acoustic coupling between the ultrasound transducer and the patient's intact abdominal skin during bladder ultrasound procedures, enabling transmission of ultrasound signals required for bladder imaging and bladder volume measurement. It is intended for use in professional healthcare facilities, such as hospitals and clinics, by qualified and trained healthcare professionals. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
(K252337)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: April 22, 2026
# 1. Information
## 1.1 Submitter Information
Submitter Name: Edgecare Inc.
Address: 12F, 8 Yangpyeong-ro 25-gil, Yeongdeungpo-gu, Seoul, 07207, Korea, South
Telephone: +82-10-6229-0668
Website: www.edgecare.co.kr
## 1.2 Official Correspondent
Name: Beom Ki Cha
Address: 12F, 8 Yangpyeong-ro 25-gil, Yeongdeungpo-gu, Seoul, 07207, Korea, South
E-mail: beomki.cha@edgecare.kr
U.S. Agent: KMC USA Inc., R5450 Astor Lane #203, Rolling Mdws, IL 60008, U.S.A.
# 2. Device Information
2.1 Device Name: EdgeFlow Gel Pad
2.2 Model Name: 00274
2.3 Common Name: Ultrasound Coupling Media
2.4 Classification Name: Media, Coupling, Ultrasound
2.5 Product Code: MUI
2.6 Classification Regulation: 21 CFR 892.1570
2.7 Device Class: Class II
2.8 Classification Panel: Radiology
{5}
2
# 3. Predicate Device
- Primary Predicate Device: SOLID GEL PAD (K131905), BLUEMTECH
- Reference Device: AQUASONIC 100 Ultrasound Gel (K802146), PARKER LABORATORIES
# 4. Subject Device Description
The EdgeFlow Gel Pad is a single-use, non-sterile gel pad composed of solid silicone elastomer. The device is intended to provide acoustic coupling between the ultrasound transducer of the EdgeFlow UW20 wearable bladder scanner and the patient's intact skin. The device is supplied as a silicone gel pad protected by two removable release films (blue and transparent) and packaged in a sealed pouch. The release films are provided only to protect the gel surface during storage and handling and are removed prior to clinical use.
# 5. Indications for Use
The EdgeFlow Gel Pad is a non-sterile, single-use accessory intended for use exclusively with the EdgeFlow UW20 wearable bladder scanner. The device provides acoustic coupling between the ultrasound transducer and the patient's intact abdominal skin during bladder ultrasound procedures, enabling transmission of ultrasound signals required for bladder imaging and bladder volume measurement. It is intended for use in professional healthcare facilities, such as hospitals and clinics, by qualified and trained healthcare professionals.
# 6. Substantial Equivalence
Comparison of the technical characteristics of the subject device and predicate devices is shown in the table below.
| | Subject Device | Primary Predicate | Reference Predicate | SE Decision |
| --- | --- | --- | --- | --- |
| 510(K) Number | - | K131905 | K802146 | - |
| Manufacturer | Edgecare Inc. | BLUEMTECH | PARKER LABORATORIES, INC. | - |
| Trade Name | EdgeFlow Gel Pad | SOLID GEL PAD | AQUASONIC 100 ULTRASOUND TRANS. GEL | - |
| Product code | MUI | MUI | IYO | - |
| Classification | Class II | Class II | - | - |
{6}
| Intended Use | The EdgeFlow Gel Pad is a non-sterile, single-use accessory intended for use exclusively with the EdgeFlow UW20 wearable bladder scanner. The device provides acoustic coupling between the ultrasound transducer and the patient's intact abdominal skin during bladder ultrasound procedures, enabling transmission of ultrasound signals required for bladder imaging and bladder volume measurement. It is intended for use in professional healthcare facilities, such as hospitals and clinics, by qualified and trained healthcare professionals. | Non-sterile ultrasound couplant for use with medical diagnostic ultrasound. It is intended to be used during non-invasive medical diagnostic ultrasound procedures to couple sound waves between a patient and the medical imaging electronics. The gel is intended for use in all diagnostic ultrasound procedures which require ultrasound coupling gel or liquid or fluid. | Parker Laboratories Ultrasound Gel can be used on intact, unbroken skin and on all patients in facilities where cross contamination is of minimal concern. Discard empty dispenser after use. Refill containers and refillable dispensers (product numbers 01-50, 03-50 and 03-54) are for use on low-risk patients and procedures in non-acute care facilities. Follow the enclosed instructions for use or your facility's protocol, if it is different. | - |
| --- | --- | --- | --- | --- |
| Device Form | Solid Silicone Gel Pad | Solid Gel Pad | Liquid Type | Similar to primary predicate |
| Single Use | Yes | Yes | Yes | Same |
| Sterilization | Non-sterile | Non-sterile | Non-sterile | Same |
| Biocompatibility | ISO 10993-5, -10 tested | ISO 10993-5:2009 ISO 10993-10:2010 | ISO 10993-5:2009 ISO 10993-10:2010 | Same |
| Reusability | No (single use only) | No (single use only) | No (single use only) | Same |
| Packaging | Individually sealed pouches to maintain cleanliness | Individually sealed pouches to maintain cleanliness | Tube type | Same to primary predicate |
{7}
| Shelf Life | Supported by accelerated aging test (17 months) | Manufacturer-claimed shelf life | Manufacturer-claimed shelf life | - |
| --- | --- | --- | --- | --- |
| Application Method | Applied as a patch between probe and skin | Applied as a patch between probe and skin | Liquid gel applied directly to skin | Same to primary predicate |
# Technological Differences:
The primary difference is the base material composition (silicone vs. aqueous gel). However, this difference does not raise new questions of safety or effectiveness because both materials provide equivalent acoustic coupling, meet the same biocompatibility standards, and demonstrate similar imaging performance.
# 7. Non-clinical and Clinical Performance Data
# Bench Performance Testing
Bench testing was conducted to evaluate the performance of the EdgeFlow Gel Pad in comparison with predicate ultrasound coupling media. Testing included axial and lateral distance measurement accuracy using a multi-purpose ultrasound phantom and bladder volume measurement accuracy using a bladder phantom. The predefined acceptance criteria were axial distance accuracy of $\leq 1$ mm, lateral distance accuracy of $\leq 1$ mm, and bladder volume measurement accuracy of $\leq 7.5\%$ . The test results demonstrated that the EdgeFlow Gel Pad provided comparable acoustic coupling and imaging performance relative to the predicate devices.
# Acoustic and Physical Characterization
Acoustic and physical characterization was performed as part of the shelf-life verification program. The evaluated parameters included sound velocity, acoustic impedance, and attenuation coefficient. These characteristics were assessed to confirm that the gel pad maintains appropriate ultrasound transmission properties throughout its labeled shelf life.
{8}
5
# Acoustic and Physical Properties
Acoustic characterization was performed to verify ultrasound transmission properties and stability over the labeled shelf life.
| Test Item | Results |
| --- | --- |
| Density | 995.60 kg/m³ |
| Sound Velocity | 970.00 m/s |
| Acoustic Impedance | 0.97 MRayl |
| Attenuation Coefficient | 0.080 dB/mm/MHz |
| Hardness (Shore A) | 55.4 (average) |
| Adhesive Strength | 4.40 N (average) |
| Release Film Peeling Force | 0.027 N/24mm (average) |
# Biocompatibility
Biocompatibility testing was conducted in accordance with ISO 10993-5 and ISO 10993-10 for a skin-contacting device intended for contact of up to 24 hours. The device was evaluated for cytotoxicity, sensitization, and irritation. The results support that the EdgeFlow Gel Pad is biocompatible for its intended use.
# Packaging and Microbiological Testing
Packaging and microbiological evaluations were performed as part of the real-time shelf-life verification program. Testing included bioburden evaluation, dye penetration testing in accordance with ASTM F1929, and seal strength testing in accordance with ASTM F88/F88M. These evaluations were conducted to confirm maintenance of packaging integrity and freedom from microbial contamination over the labeled shelf life. Dye penetration testing showed no evidence of package seal leakage, seal strength testing met the predefined acceptance criteria, and bioburden testing supported maintenance of acceptable bioburden levels.
{9}
6
# Shelf Life
Real-time stability testing was conducted to evaluate mechanical, acoustic, imaging, microbiological, and packaging integrity characteristics over time. These data support a shelf life of 17 months for the EdgeFlow Gel Pad.
# In-Vivo Wearable Performance Testing
A prospective, single-center in-vivo study was conducted at Severance Hospital to evaluate acoustic coupling stability during 24-hour wearable use of the EdgeFlow UW20 system with the designated EdgeFlow Gel Pad. Ten adult subjects participated in the study. A total of 1,454 ultrasound images were evaluated, and 1,414 images maintained valid acoustic coupling, resulting in an overall acoustic coupling retention ratio of 97.24%, which exceeded the predefined acceptance criterion of 80%. Self-reported scores for skin discomfort, sweat, and gel pad slippage remained minimal on a 5-point scale, and no adverse events or early device removals were reported. These results support that the designated gel pad can maintain stable acoustic coupling during the intended 24-hour wear period when used with the EdgeFlow UW20 wearable bladder scanner under representative clinical use conditions.
# 8. Conclusion
The EdgeFlow Gel Pad and the predicate device have similar intended use and technological characteristics. Non-clinical testing demonstrates that the EdgeFlow Gel Pad is biocompatible, maintains functional performance over its intended shelf life, and is safe and effective for use as intended. These results support a finding of substantial equivalence. Therefore, Edgecare Inc. considers the subject device to be as safe and effective as, and substantially equivalent to, the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.