← Product Code [OLZ](/productcode/OLZ) · K252330

# DeepRESP (K252330)

_Nox Medical Ehf · OLZ · Nov 17, 2025 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252330

## Device Facts

- **Applicant:** Nox Medical Ehf
- **Product Code:** [OLZ](/productcode/OLZ.md)
- **Decision Date:** Nov 17, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K252330 · Nov 17, 2025 | DeepRESP | Nox Medical Ehf | Retrospective sleep study recordings from U.S. sleep clinics | Retrospective clinical data were used to validate the performance of the DeepRESP software in scoring sleep recordings (Type I, II, and III) by comparing automatic scoring results against manual clinical scoring. | Retrospective study; Sleep study recordings; Clinical performance validation; Routine clinical care |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective validation of DeepRESP v2.0; Retrospective comparative study | Adults (18 years and older) suspected of suffering from sleep disorders, representative of U.S. clinical populations.; Sample Size: 4,030 PSG recordings (Type I/II); 5,771 sleep recordings (Type III); Number of Sites: Multiple sleep clinics in the United States | Manual scoring and predicate devices (DeepRESP K241960, Nox Sleep System K192469) | Agreement in scoring and clinical indices (AHI, CAHI, sleep staging, respiratory events); PPA, NPA, OPA. |

## Indications for Use

DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, a patient's home, or an ambulatory setting. It is indicated for use with adults (18 years and above) in a clinical environment by or on the order of a medical professional. DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotations of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); and infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional.

## Device Story

DeepRESP is a cloud-based SaMD for sleep study analysis. It ingests sleep study recordings (EEG, EMG, EOG, RIP, SpO2, nasal pressure, body position) and uses AI and rule-based models to perform automated sleep staging, arousal scoring, and respiratory event detection (including AHI and CAHI). Operated by medical professionals in clinical or home settings, the device provides automated outputs for clinician review and verification. By streamlining the scoring process, it aids in the diagnosis of sleep-related disorders. The system architecture includes a web API for third-party integration, protocol-based analysis services, and object storage for results. It does not include hardware.

## Clinical Evidence

Retrospective study of 4,030 PSG (Type I/II) and 5,771 Type III recordings. Validated agreement between DeepRESP automated scoring and manual scoring against predicate performance. Metrics included PPA, NPA, and OPA for AHI, CAHI, sleep stages, and respiratory events. Subgroup analysis confirmed performance consistency across age groups (including 18-21) and BMI categories. No significant impact of age on performance observed.

## Technological Characteristics

Cloud-based SaMD; serverless stack architecture. Uses AI and rule-based models for signal analysis. Inputs: EEG, EMG, EOG, RIP, SpO2, nasal pressure, body position. Standards: IEC 62304:2006/A1:2015, ISO 14971:2019, ANSI/AAMI SW96:2023. Connectivity: Web API for third-party client integration.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- DeepRESP / Nox Medical ([K241960](/device/K241960.md))
- Nox Sleep System / Nox Medical ([K192469](/device/K192469.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 17, 2025

Nox Medical ehf
% Hrishikesh Gadagkar
Senior Principal
Rqm+
2790 Mosside Blvd #800
Monroeville, Pennsylvania 15146

Re: K252330
Trade/Device Name: DeepRESP
Regulation Number: 21 CFR 882.1400
Regulation Name: Electroencephalograph
Regulatory Class: Class II
Product Code: OLZ
Dated: August 15, 2025
Received: August 19, 2025

Dear Hrishikesh Gadagkar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}

K252330 - Hrishikesh Gadagkar
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252330 - Hrishikesh Gadagkar
Page 3

Sincerely,

Patrick Antkowiak -S

for

Jay Gupta

Assistant Director

DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices

OHT5: Office of Neurological and
Physical Medicine Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252330  |   |
|  Device Name DeepRESP  |   |
|  Indications for Use (Describe) DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, a patient's home, or an ambulatory setting. It is indicated for use with adults (18 years and above) in a clinical environment by or on the order of a medical professional. DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotations of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); and infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

nox medical

510(k) Summary DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040 Revision Number: 04

# DATE PREPARED

November 13, 2025

# MANUFACTURER AND 510(k) OWNER

Nox Medical ehf
Katrinartuni 2,
IS - 105 Reykjavik,
Iceland
Telephone: +354 570 7170
Official Contact: Elisabet Finnbogadóttir, Director, Regulatory Affairs

# REPRESENTATIVE/CONSULTANT

Hrishikesh Gadagkar,
Sr. Principal, RQM+
2790 Mosside Blvd #800, Monroeville, PA
Telephone: +1 (410) 245-0501
Email: hgadagkar@rqmplus.com

# DEVICE INFORMATION

Proprietary Name/Trade Name: DeepRESP
Common Name: Electroencephalograph, Automatic Event Detection Software for Polysomnograph with Electroencephalograph
Regulation Number: 882.1400
Class: II
Product Code: OLZ
Premarket Review: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices (DHT5A), Office of Neurological and Physical Medicine Devices (OHT5)
Review Panel: Neurology

# PREDICATE DEVICE IDENTIFICATION

The DeepRESP (subject device) is substantially equivalent to the following predicate:

|  510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate | Additional Predicate  |
| --- | --- | --- | --- |
|  K241960 | DeepRESP / Nox Medical | x |   |
|  K192469 | Nox Sleep System / Nox Medical |  | x  |

# DEVICE DESCRIPTION

DeepRESP is a cloud-based software as a medical device (SaMD), designed to perform analysis of sleep study recordings, with and without EEG signals, providing data for the assessment and diagnosis of sleep-related disorders. Its algorithmic framework provides the derivation of sleep staging, including arousals, scoring of respiratory events, and key parameters such as the Apnea-Hypopnea Index (AHI) and Central Apnea-Hypopnea Index (CAHI).

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{5}

nox medical

510(k) Summary DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040 Revision Number: 04

DeepRESP (K252330) is hosted on a serverless stack. It consists of:

- A web Application Programming Interface (API) intended to interface with a third-party client application, allowing medical professionals to access DeepRESP's analytical capabilities.
- Predefined sequences called Protocols that run data analyses, including artificial intelligence and rule-based models for the scoring of sleep studies, and a parameter calculation service.
- A Result storage using an object storage service to temporarily store outputs from the DeepRESP Protocols.

![img-0.jpeg](img-0.jpeg)
Figure 1 – DeepRESP dataflow

The following constitute changes in DeepRESP (K252330) in comparison with DeepRESP (K241960) device:

a) Patient population extended to include patients 18 of age and above.
b) Automated Central Sleep Apnea (CSA) scoring.
c) Respiratory and Sleep staging model update to improve performance.

# INDICATIONS FOR USE

DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, a patient's home, or an ambulatory setting. It is indicated for use with adults (18 years and above) in a clinical environment by or on the order of a medical professional.

DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotations of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); and infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional.

# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Nox Medical believes that the DeepRESP (K252330) is substantially equivalent to the predicate devices listed below, based on the information summarized here:

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{6}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|  Trade/Device Name | Subject Device | Primary Predicate Device: DeepRESP | Additional Predicate Device: Nox Sleep System  |
| --- | --- | --- | --- |
|  Manufacturer | Nox Medical | Nox Medical | Nox Medical  |
|  510(k) Number | K252330 | K241960 | K192469  |
|  Regulation Number | 21 CFR 882.1400 | 21 CFR 882.1400 | 21 CFR 882.1400  |
|  Regulation Name | Electroencephalograph | Electroencephalograph | Electroencephalograph  |
|  Regulatory Class | Class II | Class II | Class II  |
|  Product Code(s) | OLZ | OLZ | Primary: OLZ
Subsequent: KZM  |
|  Intended Use | DeepRESP is a cloud-based artificial intelligence-enabled software application used for analysis (automatic scoring), retrieval, and summarization of data recorded with sleep monitoring devices. The device uses an algorithm to categorize sleep-related events that help aid in the diagnosis of sleep-related disorders. The results of the analyzed data are then transferred to another software for manual scoring, display and reporting | DeepRESP is a cloud-based artificial intelligence-enabled software application used for analysis (automatic scoring), retrieval, and summarization of data recorded with sleep monitoring devices. The device uses an algorithm to categorize sleep-related events that help aid in the diagnosis of sleep-related disorders. The results of the analyzed data are then transferred to another software for manual scoring, display and reporting | The Nox Sleep System is used as an aid in the diagnosis of different sleep disorders and for the assessment of sleep.
The Nox Sleep System is used to measure, record, display, organize, analyze, summarize, and retrieve physiological parameters during sleep and wake.
The Nox Sleep System allows the user to decide on the complexity of the study by varying the number and types of physiological signals measured.
The Nox Sleep System allows for generation of user/pre-defined reports based on subject’s data.
The user of the Nox Sleep System are medical professionals who have received training in the areas of hospital/clinical procedures, physiological monitoring of human subjects, or sleep disorder investigation.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient's home.  |
|  Indications for Use | DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, a patient's home, or an ambulatory setting. It is indicated for use with adults (18 years and above) in a clinical environment by | DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, patient home, or an ambulatory setting. It is indicated for use with adults (22 years and above) in a clinical environment by or on | The Nox Sleep System is used as an aid in the diagnosis of different sleep disorders and for the assessment of sleep.
The Nox Sleep System is used to measure, record, display, organize, analyze, summarize, and retrieve physiological parameters during sleep and wake.
The Nox Sleep System allows the user to decide on the complexity of the study by  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

Page 3 of 10

{7}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|   | or on the order of a medical professional.
DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotations of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); and infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional. | the order of a medical professional.
DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotation of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional. | varying the number and types of physiological signals measured.
The Nox Sleep System allows for generation of user/pre-defined reports based on subject’s data.
The user of the Nox Sleep System are medical professionals who have received training in the areas of hospital/clinical procedures, physiological monitoring of human subjects, or sleep disorder investigation.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient's home.  |
| --- | --- | --- | --- |
|  Use Environment | Physician office. No limitation on where data are acquired. | Physician office. No limitation on where data are acquired. | Physician office. No limitation on where data are acquired.  |
|  Patient Population | 18 years and older | 22 years and older | 22 years and older  |
|  Type of Use | Prescription Use only | Prescription Use only | Prescription Use only  |
|  Signals analyzed | EEG, Chin EMG, EOG, Abdomen and Thorax RIP, SpO2, Nasal pressure (Cannula flow), Body position | EEG, Chin EMG, EOG, Abdomen and Thorax RIP, SpO2, Nasal Pressure (Cannula flow), Body position | EEG, Chin EMG, EOG Abdomen and Thorax RIP, SpO2; Nasal pressure (Cannula flow); Body position  |
|  Sleep Scoring with EEG | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic ready for review and interpretation by a physician.
Manual  |
|  Sleep Scoring Without EEG | Automatic ready for review by medical professional | Automatic ready for review by medical professional | Not available  |
|  Score Sleep Disorder Respiratory Events | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic
Manual  |
|  Score Arousal Events with EEG | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Manual  |
|  Score Arousal Events without EEG | Automatic ready for review by medical professional | Automatic ready for review by medical professional | Not available  |
|  Score Central Sleep Apneas and Obstructive Sleep Apneas | Automatic ready for review by medical professional | Not available | Automatic ready for review and interpretation by a physician.  |
|  Score Body Position | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic ready for review by medical professional
Manual through a 3^{rd} party application. | Automatic
Manual  |
|  Manual Review of Automatic Scoring | Yes | Yes | Yes  |
|  Sleep Study Reporting | Yes | Yes | Yes  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

Page 4 of 10

{8}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|  Hardware Components | Not included | Not included | Included  |
| --- | --- | --- | --- |
|  Physical Characteristics | API-based software operates in the cloud. | API-based software operates in the cloud. | Operates on any PC with Windows 10.  |
|  Performance Testing | IEC 62304:2006/ A1:2015
ISO 14971:2019
ANSI/AAMI SW96:2023 | IEC 62304:2006/ A1:2015
ISO 14971:2019
ANSI/AAMI SW96:2023 | IEC 62304:2006
ISO 14971:2007  |

Clinical Performance

|  Trade/Device Name | Subject Device: DeepRESP v2.0 | Primary Predicate Device: DeepRESP v1.0 (K241960) | Additional Predicate Device: Nox Sleep System (K192469) | Observed paired differences (ΔPPA, ΔNPA, ΔOPA) between subject device and predicate devices. Bootstrapped 95% CI  |
| --- | --- | --- | --- | --- |
|  Clinical performance | Severity classification | Severity classification | Severity classification | Severity classification DeepRESP (K241960)  |
|  Type I/II studies | AHI ≥ 5
PPA%: 87.7
NPA%: 76.5
OPA%: 87.3 | AHI ≥ 5
PPA%: 87.5
NPA%: 91.9
OPA%: 87.9 | AHI ≥ 5
Not used in comparison. | AHI ≥ 5
ΔPPA%: -4.7 [-5.6, -3.9]
ΔNPA%: 27.0 [19.2, 35.3]
ΔOPA%: -3.5 [-4.4, -2.6]  |
|   |  AHI ≥ 15
PPA%: 71.9
NPA%: 94.8
OPA%: 78.2 | AHI ≥ 15
PPA%: 74.1
NPA%: 94.7
OPA%: 81.5 | AHI ≥ 15
Not used in comparison. | AHI ≥ 15
ΔPPA%: -3.2 [-4.3, -2.1]
ΔNPA%: 4.9 [3.3, 6.7]
ΔOPA%: -1.0 [-1.9, -0.1]  |
|   |  CAHI ≥ 5
PPA%: 80.0
NPA%: 98.0
OPA%: 97.2 | CAHI ≥ 5
N/A | CAHI ≥ 5
Not reported in K192469.
Non-inferiority is demonstrated in next column. | Nox Sleep System (K192469)
CAHI ≥ 5
ΔPPA%: 26.6 [18.4, 34.6]
ΔNPA%: -1.0 [-1.5, -0.5]
ΔOPA%: 0.3 [-0.3, 1.0]  |
|   |  Sleep stages | Sleep stages | Sleep stages | Sleep stages  |
|   |  Wake
PPA%: 92.8
NPA%: 95.8
OPA%: 95.1 | Wake
PPA%: 95.4
NPA%: 94.6
OPA%: 94.8 | Wake
Not used in comparison. | Sleep stages
DeepRESP (K241960)
Wake
ΔPPA%: -2.8 [-3.0, -2.6]
ΔNPA%: 1.5 [1.4, 1.7]
ΔOPA%: 0.6 [0.5, 0.7]  |
|   |  REM
PPA%: 82.5
NPA%: 98.8
OPA%: 96.5 | REM
PPA%: 84.3
NPA%: 98.3
OPA%: 96.3 | REM
Not used in comparison. | REM
ΔPPA%: -2.4 [-2.7, -2.1]
ΔNPA%: 0.3 [0.2, 0.3]
ΔOPA%: -0.1 [-0.2, -0.1]  |
|   |  NREM1
PPA%: 43.1
NPA%: 94.5
OPA%: 91.7 | NREM1
PPA%: 42.8
NPA%: 89.7
OPA%: 87.1 | NREM1
Not used in comparison. | NREM1
ΔPPA%: 1.7 [1.0, 2.3]
ΔNPA%: 4.6 [4.4, 4.9]
ΔOPA%: 4.5 [4.3, 4.7]  |
|   |  NREM2
PPA%: 78.1
NPA%: 91.5
OPA%: 85.3 | NREM2
PPA%: 74.2
NPA%: 82.8
OPA%: 78.7 | NREM2
Not used in comparison. | NREM2
ΔPPA%: 4.4 [3.9, 5.0]
ΔNPA%: 8.0 [7.7, 8.3]
ΔOPA%: 6.4 [6.1, 6.6]  |
|   |  NREM3
PPA%: 87.5
NPA%: 94.6
OPA%: 93.7 | NREM3
PPA%: 43.1
NPA%: 98.5
OPA%: 91.6 | NREM3
Not used in comparison. | NREM3
ΔPPA%: 45.6 [44.4, 46.8]
ΔNPA%: -4.0 [-4.1, -3.8]
ΔOPA%: 2.0 [1.8, 2.2]  |
|   |  Respiratory events
PPA%: 71.2
NPA%: 93.2 | Respiratory events
PPA%: 72.0
NPA%: 94.2 | Respiratory events
Not used in comparison. | Respiratory events
DeepRESP (K241960)
ΔPPA%: -1.0 [-1.3, -0.6]
ΔNPA%: 2.3 [2.1, 2.5]  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{9}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|   | OPA%: 85.6

**All apnea**
PPA%: 83.7
NPA%: 98.2
OPA%: 97.1 | OPA%: 87.2

**All apnea**
PPA%: 85.1
NPA%: 97.7
OPA%: 96.7 | **All apnea**
Not used in comparison. | ΔOPA%: 1.2 [1.0, 1.3]

**All apnea (Bootstrapped paired median differences)**
DeepRESP (K241960)
ΔPPA%: 0.0 [-0.7, 0.8]
ΔNPA%: 0.9 [0.8, 1.0]
ΔOPA%: 0.8 [0.7, 0.9]  |
| --- | --- | --- | --- | --- |
|   |  **Central apnea**
PPA%: 79.3
NPA%: 99.2
OPA%: 99.0 | **Central apnea**
N/A | **Central apnea**
PPA, NPA, and OPA not reported in K192469.
Non-inferiority is demonstrated in next column | **Central apnea**
Nox Sleep System (K192469)
ΔPPA%: 21.4 [16.2, 27.0]
ΔNPA%: -0.4 [-0.5, -0.4]
ΔOPA%: -0.2 [-0.3, -0.2]  |
|   |  **Obstructive apnea**
PPA%: 76.2
NPA%: 98.4
OPA%: 97.0 | **Obstructive apnea**
N/A | **Obstructive apnea**
PPA, NPA, and OPA not reported in K192469.
Non-inferiority is demonstrated in next column | **Obstructive apnea (Bootstrapped paired median differences)**
Nox Sleep System (K192469)
ΔPPA%: 9.8 [7.6, 11.9]
ΔNPA%: -0.5 [-0.6, -0.4]
ΔOPA%: 0.2 [0.0, 0.3]  |
|   |  **Hypopnea**
PPA%: 60.1
NPA%: 92.9
OPA%: 83.5 | **Hypopnea**
PPA%: 59.6
NPA%: 94.2
OPA%: 85.4 | **Hypopnea**
Not used in comparison. | **Hypopnea (Bootstrapped paired median differences)**
DeepRESP (K241960)
ΔPPA%: -0.2 [-0.7, 0.2]
ΔNPA%: 2.0 [1.8, 2.2]
ΔOPA%: 1.3 [1.2, 1.5]  |
|   |  **Desaturation**
PPA%: 98.5
NPA%: 95.5
OPA%: 96.1 | **Desaturation**
PPA%: 96.4
NPA%: 99.6
OPA%: 98.9 | **Desaturation**
Not used in comparison. | **Desaturation (Bootstrapped paired median differences)**
DeepRESP (K241960)
ΔPPA%: 0.5 [0.5, 0.6]
ΔNPA%: -0.9 [-0.9, -0.9]
ΔOPA%: -0.6 [-0.6, -0.6]  |
|   |  **Arousal events**
PPA%: 62.1
NPA%: 89.1
OPA%: 81.5 | **Arousal events**
PPA%: 62.2
NPA%: 89.3
OPA%: 81.4 | **Arousal events**
N/A | **Arousal events**
DeepRESP (K241960)
ΔPPA%: -0.8 [-0.8, -0.7]
ΔNPA%: 0.2 [0.2, 0.2]
ΔOPA%: -0.1 [-0.1, -0.1]  |
|  Trade/Device Name | Subject Device: DeepRESP v2.0 | Primary Predicate Device: DeepRESP v1.0 (K241960) | Additional Predicate Device: Nox Sleep System (K192469) | Observed paired differences (ΔPPA, ΔNPA, ΔOPA) between subject device and predicate devices. Bootstrapped 95% CI  |
| --- | --- | --- | --- | --- |
|  Clinical performance | Severity classification | Severity classification | Severity classification | Severity classification DeepRESP (K241960)  |
|  Type III studies | AHI ≥ 5
PPA%: 91.0
NPA%: 78.0
OPA%: 90.6 | AHI ≥ 5
PPA%: 93.1
NPA%: 81.1
OPA%: 92.5 | AHI ≥ 5
Not used in comparison | AHI ≥ 5
ΔPPA%: -4.1 [-4.7, -3.5]
ΔNPA%: 33.9 [26.7, 42.0]
ΔOPA%: -3.0 [-3.6, -2.3]  |
|  HSAT-Flow | AHI ≥ 15
PPA%: 78.1
NPA%: 93.9
OPA%: 81.7 | AHI ≥ 15
PPA%: 82.1
NPA%: 92.3
OPA%: 84.7 | AHI ≥ 15
Not used in comparison | AHI ≥ 15
ΔPPA%: -3.6 [-4.5, -2.8]
ΔNPA%: 6.7 [5.1, 8.3]
ΔOPA%: -1.2 [-1.9, -0.4]  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{10}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|  CAHI ≥ 5
PPA%: 80.7
NPA%: 98.0
OPA%: 97.2 | CAHI ≥ 5
N/A | CAHI ≥ 5
Not reported in K192469.
Non-inferiority is demonstrated in next column | Nox Sleep System (K192469)
CAHI ≥ 5
ΔPPA%: 24.3 [17.0, 31.0]
ΔNPA%: -0.6 [-1.0, -0.2]
ΔOPA%: 0.5 [-0.1, 1.0]  |
| --- | --- | --- | --- |
|  Sleep states | Sleep states | Sleep states | Sleep states
DeepRESP (K241960)  |
|  Wake
PPA%: 79.7
NPA%: 96.6
OPA%: 92.9 | Wake
PPA%: 76.2
NPA%: 96.8
OPA%: 92.7 | Wake
Not used in comparison. | Wake
ΔPPA%: 2.3 [2.1, 2.4]
ΔNPA%: 0.6 [0.6, 0.7]
ΔOPA%: 1.0 [1.0, 1.0]  |
|  REM
PPA%: 77.0
NPA%: 98.1
OPA%: 95.2 | REM
PPA%: 79.1
NPA%: 98.1
OPA%: 95.4 | REM
Not used in comparison. | REM
ΔPPA%: 0.4 [0.3, 0.5]
ΔNPA%: 0.1 [0.1, 0.1]
ΔOPA%: 0.1 [0.1, 0.2]  |
|  NREM
PPA%: 94.2
NPA%: 80.1
OPA%: 89.1 | NREM
PPA%: 94.5
NPA%: 79.0
OPA%: 89.2 | NREM
Not used in comparison. | NREM
ΔPPA%: 0.7 [0.7, 0.7]
ΔNPA%: 1.2 [1.1, 1.3]
ΔOPA%: 0.9 [0.8, 0.9]  |
|  Respiratory events
PPA%: 74.4
NPA%: 92.0
OPA%: 85.5 | Respiratory events
PPA%: 75.4
NPA%: 87.8
OPA%: 83.7 | Respiratory events
Not used in comparison. | Respiratory events
DeepRESP (K241960)
ΔPPA%: -0.5 [-0.8, -0.3]
ΔNPA%: 2.7 [2.5, 2.8]
ΔOPA%: 1.5 [1.4, 1.6]  |
|  All apnea
PPA%: 84.5
NPA%: 98.2
OPA%: 97.0 | All apnea
PPA%: 84.9
NPA%: 96.7
OPA%: 95.8 | All apnea
Not used in comparison. | All apnea (Bootstrapped paired median differences)
DeepRESP (K241960)
ΔPPA%: 0.0 [-0.5, 0.5]
ΔNPA%: 1.0 [0.9, 1.1]
ΔOPA%: 0.9 [0.8, 1.0]  |
|  Central apnea
PPA%: 77.5
NPA%: 99.2
OPA%: 99.0 | Central apnea
N/A | Central apnea
PPA, NPA, and OPA not reported in K192469.
Non-inferiority is demonstrated in next column | Central apnea
Nox Sleep System (K192469)
ΔPPA%: 15.2 [11.4, 19.1]
ΔNPA%: -0.4 [-0.5, -0.4]
ΔOPA%: -0.3 [-0.3, -0.3]  |
|  Obstructive apnea
PPA%: 78.4
NPA%: 98.4
OPA%: 97.0 | Obstructive apnea
N/A | Obstructive apnea
PPA, NPA, and OPA not reported in K192469.
Non-inferiority is demonstrated in next column | Obstructive apnea (Bootstrapped paired median differences)
Nox Sleep System (K192469)
ΔPPA%: 2.4 [1.1, 3.7]
ΔNPA%: -0.0 [-0.1, 0.1]
ΔOPA%: 0.2 [0.1, 0.3]  |
|  Hypopnea
PPA%: 63.9
NPA%: 91.7
OPA%: 83.3 | Hypopnea
PPA%: 64.0
NPA%: 89.1
OPA%: 82.6 | Hypopnea
Not used in comparison. | Hypopnea (Bootstrapped paired median differences)
DeepRESP (K241960)
ΔPPA%: 0.3 [0.0, 0.6]
ΔNPA%: 2.3 [2.2, 2.4]
ΔOPA%: 1.7 [1.6, 1.8]  |
|  Desaturation
PPA%: 98.8
NPA%: 96.3
OPA%: 96.9 | Desaturation
PPA%: 96.9
NPA%: 99.5
OPA%: 98.8 | Desaturation
Not used in comparison. | Desaturation (Bootstrapped paired median differences)
DeepRESP (K241960)
ΔPPA%: 0.6 [0.6, 0.6]
ΔNPA%: -1.0 [-1.0, -0.9]
ΔOPA%: -0.6 [-0.6, -0.5]  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{11}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|   | Arousal events
PPA%: 64.0
NPA%: 90.5
OPA%: 83.1 | Arousal events
PPA%: 66.8
NPA%: 86.8
OPA%: 81.8 | Arousal events
N/A | Arousal events
DeepRESP (K241960)
ΔPPA%: -0.8 [-0.9, -0.8]
ΔNPA%: 0.5 [0.5, 0.5]
ΔOPA%: 0.1 [0.1, 0.2]  |
| --- | --- | --- | --- | --- |
|  Trade/Device Name | Subject Device: DeepRESP v2.0 | Primary Predicate Device: DeepRESP v1.0 (K241960) | Additional Predicate Device: Nox Sleep System (K192469) | Observed paired differences (ΔPPA, ΔNPA, ΔOPA) between subject device and predicate devices. Bootstrapped 95% CI  |
| --- | --- | --- | --- | --- |
|  Clinical performance
Type III studies
HSAT-RIP | Severity classification
AHI ≥ 5
PPA%: 93.7
NPA%: 63.5
OPA%: 92.8

AHI ≥ 15
PPA%: 81.0
NPA%: 91.1
OPA%: 83.4

CAHI ≥ 5
PPA%: 79.5
NPA%: 97.6
OPA%: 96.9

Sleep states
Wake
PPA%: 79.7
NPA%: 96.6
OPA%: 92.9

REM
PPA%: 77.0
NPA%: 98.1
OPA%: 95.2

NREM
PPA%: 94.2
NPA%: 80.1
OPA%: 89.1

Respiratory events
PPA%: 75.0
NPA%: 90.7
OPA%: 84.8

All apnea
PPA%: 81.1
NPA%: 95.7
OPA%: 94.5

Central apnea
PPA%: 78.8
NPA%: 99.2
OPA%: 99.0 | Severity classification
AHI ≥ 5
PPA%: 93.1
NPA%: 81.1
OPA%: 92.5

AHI ≥ 15
PPA%: 82.1
NPA%: 92.3
OPA%: 84.7

CAHI ≥ 5
N/A

Sleep states
Wake
PPA%: 76.2
NPA%: 96.8
OPA%: 92.7

REM
PPA%: 79.1
NPA%: 98.1
OPA%: 95.4

NREM
PPA%: 94.5
NPA%: 79.0
OPA%: 89.2

Respiratory events
PPA%: 75.4
NPA%: 87.8
OPA%: 83.7

All apnea
PPA%: 84.9
NPA%: 96.7
OPA%: 95.8

Central apnea
N/A | Severity classification
AHI ≥ 5
Not used in comparison

AHI ≥ 15
Not used in comparison

CAHI ≥ 5
Not reported in K192469.
Non-inferiority is demonstrated in next column

Sleep states
Wake
Not used in comparison.

REM
Not used in comparison.

NREM
Not used in comparison.

Respiratory events
Not used in comparison.

All apnea
Not used in comparison.

Central apnea
PPA, NPA, and OPA not reported in K192469. | Severity classification
DeepRESP (K241960)
AHI ≥ 5
ΔPPA%: -1.4 [-1.9, -0.8]
ΔNPA%: 19.3 [12.0, 26.9]
ΔOPA%: -0.8 [-1.4, -0.2]

AHI ≥ 15
ΔPPA%: -0.6 [-1.5, 0.3]
ΔNPA%: 3.9 [2.3, 5.6]
ΔOPA%: 0.5 [-0.3, 1.2]

Nox Sleep System (K192469)
CAHI ≥ 5
ΔPPA%: 23.0 [15.3, 31.1]
ΔNPA%: -0.9 [-1.3, -0.5]
ΔOPA%: 0.1 [-0.4, 0.6]

Sleep states
DeepRESP (K241960)
Wake
ΔPPA%: 2.3 [2.1, 2.4]
ΔNPA%: 0.6 [0.6, 0.7]
ΔOPA%: 1.0 [1.0, 1.0]

REM
ΔPPA%: 0.4 [0.3, 0.5]
ΔNPA%: 0.1 [0.1, 0.1]
ΔOPA%: 0.1 [0.1, 0.1]

NREM
ΔPPA%: 0.7 [0.7, 0.7]
ΔNPA%: 1.2 [1.1, 1.3]
ΔOPA%: 0.9 [0.8, 0.9]

Respiratory events
DeepRESP (K241960)
ΔPPA%: 0.0 [-0.2, 0.3]
ΔNPA%: 1.4 [1.2, 1.5]
ΔOPA%: 0.9 [0.8, 1.0]

All apnea (Bootstrapped paired median differences)
DeepRESP (K241960)
ΔPPA%: -3.3 [-4.3, -2.5]
ΔNPA%: -1.5 [-1.7, -1.3]
ΔOPA%: -1.6 [-1.8, -1.5]

Central apnea
NX Sleep System (K192469)
ΔPPA%: 16.6 [12.9, 20.2]
ΔNPA%: -0.5 [-0.5, -0.5]
ΔOPA%: -0.3 [-0.4, -0.3]  |

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{12}

nox medical

510(k) Summary

DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040

Revision Number: 04

|   | Obstructive
apnea
PPA%: 74.3
NPA%: 96.0
OPA%: 94.4 | Obstructive
apnea
N/A | Non-inferiority is
demonstrated in next
column | Obstructive apnea (Bootstrapped
paired median differences)
Nox Sleep System (K192469)
ΔPPA%: -1.7 [-3.4, -0.2]
ΔNPA%: -2.5 [-2.6, -2.3]
ΔOPA%: -2.4 [-2.6, -2.2]  |
| --- | --- | --- | --- | --- |
|   | Hypopnea
PPA%: 58.9
NPA%: 90.7
OPA%: 81.0 | Hypopnea
PPA%: 64.0
NPA%: 89.1
OPA%: 82.6 | Not used in
comparison. | Hypopnea (Bootstrapped paired
median differences)
DeepRESP (K241960)
ΔPPA%: -4.7 [-5.1, -4.2]
ΔNPA%: 1.2 [1.1, 1.4]
ΔOPA%: -0.6 [-0.7, -0.4]  |
|   | Desaturation
PPA%: 98.8
NPA%: 96.3
OPA%: 96.9 | Desaturation
PPA%: 96.9
NPA%: 99.5
OPA%: 98.8 | Not used in
comparison. | Desaturation (Bootstrapped paired
median differences)
DeepRESP (K241960)
ΔPPA%: 0.6 [0.6, 0.6]
ΔNPA%: -1.0 [-1.0, -0.9]
ΔOPA%: -0.6 [-0.6, -0.5]  |
|   | Arousal events
PPA%: 64.0
NPA%: 90.5
OPA%: 83.0 | Arousal events
PPA%: 66.8
NPA%: 86.8
OPA%: 81.8 | Arousal events
N/A | Arousal events
DeepRESP (K241960)
ΔPPA%: -0.8 [-0.9, -0.8]
ΔNPA%: 0.5 [0.5, 0.5]
ΔOPA%: 0.1 [0.1, 0.2]  |

# SUMMARY OF NON-CLINICAL TESTING

Software verification and validation testing were performed per IEC 62304:2006/A1:2015 Medical device software – Software life cycle processes, IEC 82304-1:2017 Health Software – general requirements for product safety, and the FDA guidance General Principles of Software Validation. Nox Medical demonstrated safety and effectiveness based on current industry standards. All documentation was prepared as recommended in the FDA Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

# SUMMARY OF CLINICAL TESTING

A retrospective study validated the clinical performance of the subject device DeepRESP. It used manually scored sleep recordings, originating from sleep clinics in the United States, performed as part of routine clinical work, matching the indicated use of the study device and patient population.

The performance of DeepRESP with regard to scoring sleep recordings with electroencephalography (EEG) (i.e., Type I and II), and sleep recordings without EEG (i.e., Type III) was validated. Two separate studies were conducted, one to validate DeepRESP with scoring Type I-II recordings and a second one with scoring Type III recordings. The performance of DeepRESP on Type III recordings was validated in two configurations: one where thorax and abdomen RIP, nasal pressure, and SpO2 are available (HSAT-Flow), the second where thorax and abdomen RIP and SpO2 are available (HSAT-RIP). The studies were done by evaluating the agreement in scoring and clinical indices resulting from the automatic scoring by DeepRESP compared to manual scoring. The performance of DeepRESP was compared to the agreement between the automatic scoring of the predicate device and manual scoring. The same collection of sleep recordings and the same manual scoring were used when

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

{13}

nox medical

510(k) Summary DeepRESP v2.0 – Traditional 510(k)

Document Number: REG-0040 Revision Number: 04

comparing the automatic scoring of DeepRESP and the predicate device. The study method was a retrospective data study comparing paired differences.

In both studies, the subject device DeepRESP's performance was validated against the primary predicate DeepRESP (K241960). To evaluate the performance scoring of central sleep apnea (CSA) and obstructive sleep apnea (OSA), Nox Sleep System (K192469) was used as a comparator.

To assess the performance of the subject device DeepRESP for use in patients aged 18 years and older, an evaluation of the performance of the primary endpoints was validated in the subgroup aged between 18 and 21 years old compared to the other sub-groups ages. The performance of classifying sleep recordings was validated for AHI ≥ 5, AHI ≥ 15, and CAHI ≥ 5. The PPA, NPA, and OPA for these classifications were evaluated for this age group as well as age groups of older patients. Further analysis was done by plotting histograms of the AHI and CAHI values of the patients in each age group, and plotting histograms of the AHI and CAHI values of the misclassified sleep recordings in each age group. This analysis demonstrated that misclassifications occurred most frequently for AHI and CAHI values close to the decision boundary. This behavior was consistent across all age groups. Therefore, it was concluded that the patient's age did not impact the performance of the subject device.

For validating scoring performance for Type I/II PSG recordings, a total of 4,030 PSG recordings were used. For validating scoring performance for Type III recordings, a total of 5,771 sleep recordings were used, including 4,037 Type I recordings and 1,734 Type II recordings. The Type I and II recordings were processed as Type III recordings by utilizing only the subset of signals from them that are common to Type III recordings.

The recording collection used for validation of Type I/II scoring consisted of 44% Females, had individuals in all age groups (18-21, 22-35, 36-45, 46-55, 56-65, &gt;65), and all BMI groups (&lt;25, 25-30, &lt;30). The recording collection used for validation of Type III scoring consisted of 35% females, had individuals in all age groups (18-21, 22-35, 36-45, 46-55, 56-65, &gt;65), and all BMI groups (&lt;25, 25-30, &lt;30). Type I and II sleep recordings were collected as part of standard clinical care for patients suspected of suffering from sleep disorders. The patients came from urban, suburban, and rural areas with a high level of race/ethnicity diversity (Caucasian or White, Black or African American, Other). The patient population was therefore considered representative of patients seeking medical services for sleep disorders in the United States. Medical professionals should assess whether DeepRESP's outputs are reliable for rare disease presentations, as such cases may be underrepresented in their training and validation data.

## CONCLUSION

Based on the verification and validation testing performed in accordance with IEC 62304:2006/A1:2015, it can be concluded that DeepRESP (K252330) does not raise new issues of safety or effectiveness compared to the primary predicate device. The subject device DeepRESP has been demonstrated to be substantially equivalent to the predicate devices.

TEMP-0482 Controlled Document Template 01

Copyright © 2025 Nox Medical ehf

---

**Source:** [https://fda-staging.innolitics.com/device/K252330](https://fda-staging.innolitics.com/device/K252330)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com