The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.
Device Story
The Insufflator (OPTO-IFL1000) is a CO2 insufflation device used in clinical settings to create and maintain pneumoperitoneum for laparoscopic surgery. It connects to a CO2 gas source and delivers gas through specialized insufflation tubes (with or without heating) into the abdominal cavity. The device monitors intra-abdominal pressure and flow rates, automatically adjusting gas delivery to maintain a set pressure and compensate for leaks. It features multiple operational modes (Pediatric, Adult, Obese, Veress) with specific pressure and flow limits. The device provides visual and audible alarms for overpressure, gas supply depletion, overheating, and fluid contamination. Healthcare providers use the device interface to set parameters and monitor real-time pressure/flow data. By maintaining a stable surgical field of view and operating space, the device facilitates laparoscopic access for surgeons, potentially reducing procedural complexity and improving patient outcomes.
Clinical Evidence
No clinical data. Bench testing only. Verification included electromagnetic compatibility (IEC 60601-1-2:2020) and performance testing of gas supply, pressure/flow accuracy, leakage compensation, overpressure/overheating alarms, and heating function. Software verification and validation were performed per FDA guidance for 'Enhanced' documentation level.
Technological Characteristics
CO2 insufflator; 110-240V power; dimensions 370x410x166.5mm; weight 11kg. Features include heating function (OPTO-T1000H tubing), pressure range 1-30 mmHg, flow rate 0.1-50 L/min. Software-controlled with multiple modes (Pediatric, Adult, Obese, Veress). Sterilization: Steam sterilization for insufflation tubes. Connectivity: Standalone device.
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic laparoscopic procedures requiring pneumoperitoneum via abdominal gas distension.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2026
Guangdong OptoMedic Technologies, Inc.
Weijuan Guo
Regulatory Affairs Engineer
Suite 503, Building A, Golden Valley Intellicreation
Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai,
Foshan, Guangdong 528200
CHINA
Re: K252318
Trade/Device Name: Insufflator (OPTO-IFL1000)
Regulation Number: 21 CFR 884.1730
Regulation Name: Laparoscopic Insufflator
Regulatory Class: II
Product Code: HIF
Dated: July 23, 2025
Received: July 25, 2025
Dear Weijuan Guo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252318 - Weijuan Guo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252318 - Weijuan Guo
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252318 | ? |
| Please provide the device trade name(s). | | ? |
| Insufflator (OPTO-IFL1000) | | |
| Please provide your Indications for Use below. | | ? |
| The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
OptoMedic
Traditional 510(k) Premarket Notification
K252318
Page 1 of 5
# 510(k) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Date Prepared: April 15, 2026,
## I. General Information
510(k) Submitter/Owner: Guangdong OptoMedic Technologies, Inc.
Suite 503, Building A, Golden Valley Intellicreation Community,
No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan,
Guangdong, 528200, P.R. China
Establishment Registration Number: Not yet registered
Primary Contact Person: Weijuan Guo
Regulatory Affairs Engineer
Tel: +86 (757) 8670 2920
Email: guoweijuan@optomedic.com
Alternative Contact Person: Minghua Wu
Regulatory Affairs Engineer
Tel: +86 (757) 8670 2920
Email: Minghua Wu @optomedic.com
## II. Device Identification
Device Trade Name: Insufflator (OPTO-IFL1000)
Common or Usual Name: Insufflator
Model: OPTO-IFL1000
Regulation Name: Laparoscopic Insufflator
Regulation Number: 21 CFR 884.1730
Regulatory Class: Class II
Product Code: HIF
## III. Predicate Device
510(k) Number: K222812
Product Name: Insufflator (SL102)
{5}
OptoMedic
Traditional
510(k)
Premarket Notification
K252318
The predicate device has not been subject to a design-related recall.
# IV. Reference Device
510(k) Number: K231342
Product Name: OPTO-IFL-1000
Sponsor: Guangdong OptoMedic Technologies, Inc.
# V. Device Description
Insufflator (OPTO-IFL1000) is a $\mathrm{CO}_{2}$ insufflation device for creating and maintaining a pneumoperitoneum during laparoscopic examinations and operations. It is capable of establishing the surgical field of view and operating space. $\mathrm{CO}_{2}$ gas can be injected into abdominal cavity by the device, and the gas separates the abdominal wall from the internal organs of the abdominal cavity, forming a space for the operation and visual field. The device is to be used with the following insufflation tubes:
1. OPTO-T1000H (with heating function)
2. OPTO-T1000 (without heating function)
# VI. Indications for Use
The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.
# VII. Comparison of Technological Characteristics with the Predicate Device
Table 1 Technological Characteristics Comparison
| Description | Subject Device (K252318) | Predicate Device (K222812) |
| --- | --- | --- |
| Product | Insufflator (OPTO-IFL1000) | Insufflator (SL102) |
| Regulation Number | 21 C.F.R. § 884.1730 | 21 C.F.R. § 884.1730 |
| Product Code | HIF | HIF |
| Device class | Class II | Class II |
| Indication for use | The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures. | The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it. |
| Prescription/Over-the-counter use | Prescription | Prescription |
{6}
OptoMedic
Traditional 510(k) Premarket Notification
K2523181
Page 3 of 5
| Distension Medium | CO₂ | CO₂ |
| --- | --- | --- |
| Configuration | Insufflator, heatable insufflation tube set insufflation tube, central intake pipe (optional), gas filter, power supply cord, fuse (F3.15AH250V), cart | Insufflator (housing, power supply, pressure reducers, venting system, fluid sensor, gas heater, various setting keys and display elements), high pressure tube and pneumoperitoneum tube |
| Insufflation Mode Selection | Pediatric Mode: 1-15 mmHg, 0.1-15 L/min
Adult Mode: 1-20 mmHg, 1-50 L/min
Obese Mode: 1-30 mmHg, 1-50 L/min
Veress Mode: 1-15 mmHg, 1-5 L/min | Pediatric mode: 1-10mmHg, 1-20 L/min
Adult mode: 1-30mmHg, 1-50 L/min
Bariatric mode: 1-30mmHg, 1-50 L/min |
| Pressure range | 1-30 mmHg | 1-30 mmHg |
| Accuracy of the measured pressure | ±2mmHg | ±2mmHg |
| Flow Rate Range | 0.1~50L/min | 1-50L/min |
| Accuracy of the measured flow | the measured flow<2L/min: ±0.5L/min;
2L/min≤the measured flow≤10L/min: ±2L/min;
the measured flow>10L/min: ±20% | the measured flow≤10L/min:±2L/min;
the measured flow>10L/min: ±20% |
| Overpressure alarm | When the nominal pressure is exceeded by more than 4 mmHg, visual and audible alarms will be issued.
When the measured pressure exceeds 30 mmHg for 0.5 seconds, a medium priority overpressure alarm will be triggered | When the actual pressure is > 3 mmHg above the nominal pressure, the overpressure warning is initiated. |
| Gas supply warning function | When gas cylinder pressure falls below 15 bar or central gas supply pressure falls below 3 bar, injection will be interrupted, device warning initiated. | When gas cylinder pressure falls below 15 bar or central gas supply pressure falls below 2 bar, injection will be interrupted, device warning initiated. |
| Gas Overheating alarm | At >41°C, visual and acoustic alarm; gas injection interrupted | At>41°C, visual and acoustic alarm; gas injection interrupted |
| Heating Function | Support heating function (with OPTO-T1000H tubing) | Does not support heating function (SL102) |
| Contamination warning | If liquid enters the device through the insufflation connection port, the device warning will be initiated and device will be allowed to complete the current operation. Insufflation cannot be restarted after device is | When the fluid penetrates into the device through the pneumoperitoneum tube joint, the gas injection will be interrupted, device warning initiated. |
{7}
OptoMedic
Traditional 510(k) Premarket Notification
K252318
Page 4 of 5
| | turned off. | |
| --- | --- | --- |
| Reprocessing for insufflation tube | Yes, Steam sterilization | Yes, Steam sterilization |
| Dimensions (W*H*D) | 370mm* 410mm *166.5mm | 267mm x 395mm x134mm |
| Weight | 11Kg (N.W.) | About 8.2kg |
| Power supply | 110-240 V~ | 110-240 V~ |
The differences between the subject and predicate device include the mode, flow rate and pressure accuracy, overpressure alarm indications and actions, gas supply warning indication, heating functions, and dimensions/weight. The different technological characteristics do not raise different questions of safety and effectiveness.
## VIII. Performance data
Non-clinical tests were conducted to verify that the subject device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- IEC 60601-1-2:2020 Edition 4.1 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
The reference device is the previous version of the subject device, without modes. The operation of the subject hardware is unchanged from the reference device. Therefore, electrical safety information from the reference device is leveraged to support the electrical safety of the subject device.
Per the FDA guidance document *Content of Premarket Submissions for Device Software Functions*, issued June 2024, the software verification and validation testing were conducted and the test results demonstrated the software function met the requirements. The software for this device was considered a “Enhanced” software documentation level.
Performance testing was also conducted and demonstrated that the proposed system performs according to specifications and functions as intended and substantially equivalent to the predicate device. And the test result shows that the preset acceptance criteria are met and are substantially equivalent to the predicate device performance:
1. Gas Supply Indication
2. Accuracy of the Pressure
3. Accuracy of the Pressure-Under Leak Condition (Continuous leakage compensation testing)
4. Overpressure Alarm
5. Overpressure Reduction
6. Under-pressure Replenishment (Transient leakage compensation testing)
{8}
OptoMedic
Traditional 510(k) Premarket Notification
K252318
Page 5 of 5
7. Accuracy of the Flow
8. Heating Function
9. Overheating Alarm
10. Accuracy of Gas Consumption Display
## IX. Conclusions
The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.