K252296 · Oncopatch, Inc. · KXK · Dec 5, 2025 · Radiology
Device Facts
Record ID
K252296
Device Name
OncoPatch
Applicant
Oncopatch, Inc.
Product Code
KXK · Radiology
Decision Date
Dec 5, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OncoPatch is indicated for the surface treatment of localized tumors. It is designed to deliver radiation therapy for the treatment of superficial lesions. It can be used either as primary treatment or as treatment for residual disease after excision of primary or recurrent tumors.
Device Story
OncoPatch is a surface brachytherapy system for treating localized skin cancers (e.g., basal/squamous cell carcinoma) and keloids. The device utilizes yttrium-90 (Y-90) as a pure beta-emitter source, contained within thin, polymer-coated metal foils. These foils are integrated into a shield assembly that provides positional stability and absorbs backward-directed radiation. The assembly is coupled with a skin collimator to deliver low-energy beta radiation to the epidermis and upper dermis. The device is applied temporarily to the patient's skin surface for single or fractionated treatment sessions (typically < 60 minutes). It is operated by healthcare providers in a clinical setting. By delivering concentrated surface radiation while sparing deeper healthy tissue, the device provides a non-invasive therapeutic option for managing superficial lesions.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical bench testing, including shielding/bremsstrahlung attenuation, dose distribution accuracy, simulated-use testing on anatomical phantoms, package integrity testing per ASTM D7386-16, and biocompatibility evaluation per ISO 10993-18. All tests met predefined acceptance criteria.
Technological Characteristics
Surface brachytherapy system using Yttrium-90 beta-emitter. Components: polymer-coated metal foil source, composite tungsten collimator, photopolymer adapter/shield, neodymium magnets. Manual operation; non-sterile. Biocompatibility per ISO 10993-18; risk management per ISO 14971:2019. Energy: 2.28 MeV endpoint energy.
Indications for Use
Indicated for surface treatment of localized tumors and superficial lesions in patients; can be used as primary treatment or for residual disease after excision of primary or recurrent tumors.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
RIC Conformal Source Model 100 Brachytherapy Source (K090321)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 5, 2025
OncoPatch, Inc.
% Sam Murray
Regulatory Consultant
TAMM Net, Inc.
3101 Cobb Pkwy
Atlanta, Georgia 30339
Re: K252296
Trade/Device Name: OncoPatch
Regulation Number: 21 CFR 892.5730
Regulation Name: Radionuclide brachytherapy source
Regulatory Class: Class II
Product Code: KXK
Dated: July 22, 2025
Received: July 23, 2025
Dear Sam Murray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252296 - Sam Murray
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252296 - Sam Murray
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2025.12.05 14:49:04 -05'00" For
Lora Weidner
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
| Submission Number (if known) |
| --- |
| K252296 |
| Device Name |
| OncoPatch |
| Indications for Use (Describe) |
| OncoPatch is indicated for the surface treatment of localized tumors. It is designed to deliver radiation therapy for the treatment of superficial lesions. It can be used either as primary treatment or as treatment for residual disease after excision of primary or recurrent tumors. |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
OncoPatch
OncoPatch K252296
# 1. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
# 1.1. Submitter's Information
Name: OncoPatch, Inc.
Address: 12635 East Monteview Blvd., Suite 100 Aurora, CO 80045
Phone: (720) 778-1317
Contact Person: David Westerly Co-Founder & CEO
Preparation Date: 5 December 2025
# 1.2. Device Name
Trade Name: OncoPatch
Common Name: Radionuclide Brachytherapy Source
Classification Name: Radionuclide Brachytherapy Source
Regulation 21 CFR § 892.5730
Regulatory Class: Class II
Product Code KXK
# 1.3. Legally Marketed Predicate Device
The legally marked predicate is the RIC Conformal Source Model 100 Brachytherapy Source (RIC-100) (K090321).
# 1.4. Device Description
The OncoPatch device is engineered for the surface treatment of skin cancers, such as basal and squamous cell carcinoma, as well as for other skin conditions, including keloids. Using yttrium-90 (Y-90) as the radioactive source, this device delivers low-energy beta radiation directly to the skin surface, targeting cancerous or abnormal tissue while sparing underlying healthy tissue. The OncoPatch device makes use of radioactive patches constructed from thin, polymer-coated foils containing Y-90. These foils are adhered to the underside of a plastic or metal composite shield, which ensures safety for both the patient and healthcare providers during treatment. The patch and shield are coupled with an applicator. The applicator uses collimators with predetermined shapes and sizes that assume a flat patient geometry. The OncoPatch is secured temporarily to the patient surface for single or multiple treatment fractions, typically with short (< 60 min) treatment durations.
{5}
OncoPatch
OncoPatch K252296
# 1.4.1. Brief Written Description of the Device
The OncoPatch device is a surface brachytherapy system that applies the principles of low-energy beta radiation to treat localized skin lesions. Its therapeutic function is based on the physical properties of yttrium-90 (Y-90), a pure beta emitter with short penetration range and high surface dose deposition. This enables concentrated radiation delivery to the epidermis and upper dermis, while minimizing exposure to deeper tissues.
The core technology centers on a foil-based Y-90 source created by neutron activation of yttrium-89. These thin metal foils are polymer-coated to contain the radioactive material and minimize contamination risk. The foil source is mechanically supported by a rigid shield that not only ensures positional stability but also serves to absorb backward-directed radiation. Magnetic coupling elements integrated into the shield facilitate precise alignment with the treatment applicator.
The applicator includes Y-90 patch and shield assembly which is coupled with a skin collimator to deliver the treatment. These components are available in fixed geometries and are fabricated from rigid polymers or composites and metal compatible with medical use.
The system is designed for single-fraction or fractionated therapy, allowing flexibility in clinical implementation while ensuring consistent and directional surface radiation.
# 1.5. Intended Use
OncoPatch is intended for the treatment of cancer and other lesions by temporary surface irradiation.
# 1.6. Indications for Use
OncoPatch is indicated for the surface treatment of localized tumors. It is designed to deliver radiation therapy for the treatment of superficial lesions. It can be used either as primary treatment or as treatment for residual disease after excision of primary or recurrent tumors.
# 1.7. Substantial Equivalence Discussion
The OncoPatch is substantially equivalent to the legally marketed predicate device RIC-100 (K090321) with respect to intended use, mechanism of action, technological characteristics, performance, and safety. A comparison of the similarities and differences of the OncoPatch and cleared predicate are provided in Table 1.
{6}
onco patch
OncoPatch
K252296
Table 1: Predicate Comparison
| Specification/Characteristic | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| | OncoPatch | RIC Conformal Source Model 100 Brachytherapy Source (K090321) | |
| Classification | 21 CFR § 892.5730, Class II | 21 CFR 872.5570, Class II | Identical |
| Product Code | KXK | KXK, ONL | Substantially Equivalent |
| Intended Use | Treatment of cancer by temporary surface irradiation. | Treatment of cancer by surface irradiation. | Identical |
| Indications for Use | OncoPatch is indicated for the surface treatment of localized tumors. It is designed to deliver radiation therapy for the treatment of superficial lesions. It can be used either as primary treatment or as treatment for residual disease after excision of primary or recurrent tumors. | RIC Conformal Source Model 100 Brachytherapy Source, containing Phosphorus-32 with activity up to 200 mCi is indicated for treatment of temporary intraoperative, interstitial, intracavitary or surface application to treat selected localized tumors. It can be used either as primary treatment or as treatment for residual disease after excision of primary or recurrent tumors. This brachytherapy source may be used concurrently with or following treatment with other interventions, such as external beam therapy. Chordomas, chondrosarcomas, soft tissue sarcomas, skin cancers and other accessible tumors could be commonly treated by the RIC Conformal Source Model 100 Brachytherapy Source, containing Phosphorus-32. | Substantially Equivalent |
| Mode of Use | Rx Only | Rx Only | Identical |
| Radionuclide | Yttrium-90 (Y-90) | Phosphorus-32 (P-32) | Substantially Equivalent |
| Radionuclide Half life | 64.1 hours | 14.3 Days | Substantially Equivalent |
| Type of Radiation | Beta particles | Beta particles | Identical |
| Source Configuration | Foil array | Planar array | Substantially Equivalent |
| Radioactivity | 200 mCi | 200 mCi | Identical |
{7}
onco patch
OncoPatch
K252296
| Specification/Characteristic | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| | OncoPatch | RIC Conformal Source Model 100 Brachytherapy Source (K090321) | |
| End Point Energy | 2.28 MeV | 1.71 MeV | Substantially Equivalent |
| Non-radioactive Materials | Composite tungsten (collimator), Photopolymer (adapter/shield), Neodymium (magnets) | Polymeric Film, silicone | Substantially Equivalent |
| Provided Sterile | No | No | Identical |
{8}
OncoPatch
OncoPatch K252296
# 1.7.1. Statement on Substantial Equivalence
The proposed device and predicate device are both surface brachytherapy systems intended to deliver localized radiation for the treatment of superficial tumors. Both devices operate on the same principle of beta-emitting radionuclide therapy and are designed for temporary, non-invasive application to the skin. The proposed and predicate devices share core technological characteristics including intended use, energy source (beta radiation), manual operation, and customizable treatment geometry. Differences in radionuclide (Yttrium-90 vs. Phosphorus-32), source configuration, and shielding materials have been thoroughly evaluated through bench testing and dosimetric analysis. These differences do not alter the intended therapeutic function and do not raise new questions of safety or effectiveness.
# 1.8. Performance Data
Performance requirements were established through a comprehensive risk management process in accordance with ISO 14971:2019. Non-clinical bench testing included shielding and bremsstrahlung attenuation, relative and absolute dose distribution accuracy, simulated-use, end-to-end testing of fit, stability, and dose delivery on an anatomically relevant phantom, simulated distribution and package integrity testing, and material characterization / biocompatibility evaluation of the patient-contacting components.
Key acceptance criteria required that surface registration errors remain within clinically acceptable ranges for accurate targeting, that dose and depth-dose profiles remain within predefined dosimetric agreement limits, and that patch activity assays remain within an acceptable tolerance around the nominal activity. Criteria also required that ambient dose rates remain within institutional and regulatory radiation safety limits for handling and transport, that device form and function are preserved following a simulated distribution cycle per ASTM D7386-16, and that material characterization per ISO 10993-18 identifies no new or unexpected chemical constituents of concern. All tests met their respective acceptance criteria and confirmed that the OncoPatch device performs as intended and is suitable for its proposed use.
Collectively, these non-clinical data demonstrate that the OncoPatch device delivers dose, provides shielding, maintains mechanical stability, and withstands shipping stresses in a manner that is comparable to the legally marketed predicate device, and that the materials are suitably biocompatible for their intended use. Therefore, the OncoPatch device is as safe and effective as, and performs as well as the predicate device.
# 1.8.1. Clinical Studies
Clinical testing was not necessary for the demonstration of substantial equivalence.
# 1.9. Conclusions
The OncoPatch device employs the same technological principles, intended use, and core therapeutic mechanism as the identified predicate device. Differences in radionuclide selection, source configuration, and shielding design have been evaluated through comprehensive non-clinical testing. These differences do not raise new questions of safety or effectiveness and do not alter the clinical application of the device. Therefore, the OncoPatch is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.