V series portable oxygen concentrator (V5, V5C, V6, V6C)
Applicant
Shenzhen Harveymed Technology Co., Ltd.
Product Code
CAW · Anesthesiology
Decision Date
Dec 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The V series Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis in a home, institutional, or travel environment. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
Device Story
Portable oxygen concentrator; utilizes molecular sieve/pressure swing adsorption (PSA) technology to extract oxygen from ambient air. Input: ambient air; Output: pulsed or continuous flow of high-concentration oxygen delivered via nasal cannula. Operated by patient or caregiver in home, institutional, or travel settings. Features breath detection technology to trigger pulse delivery. Device includes LCD display, buttons, and alarm system for monitoring system status (e.g., oxygen concentration, battery, flow). Healthcare providers prescribe the device to manage respiratory therapy needs; output affects patient oxygenation levels. Benefits include increased mobility and supplemental oxygen delivery for patients with respiratory conditions.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via oxygen pulse volume, continuous flow, trigger sensitivity, oxygen concentration, respiratory rate, outlet pressure, and alarm functionality testing. Electrical safety and EMC compliance verified per IEC 60601-1, 60601-1-2, 60601-1-8, 60601-1-11, and ISO 80601-2-67/69. Biocompatibility evaluated per ISO 10993 and ISO 18562 series.
Technological Characteristics
Molecular sieve/pressure swing adsorption (PSA) technology. Materials: PET (button film), PC+ABS (housing). Pulse and continuous flow modes. Powered by AC/DC adapter or Lithium-ion battery. Dimensions: 18.6 x 8.4 x 20.8 cm. Weight: 2.2-2.6 kg. Connectivity: None. Sterilization: Not applicable. Software: Firmware-controlled flow settings and alarm logic.
Indications for Use
Indicated for adult patients requiring supplemental oxygen therapy on a prescriptive basis in home, institutional, or travel environments.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 16, 2025
Shenzhen HarveyMed Technology Co., Ltd.
% James Tsai
Consultant
Shenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan
District
Shenzhen, Guangdong 518000
China
Re: K252268
Trade/Device Name: V series portable oxygen concentrator (V5, V5C, V6, V6C)
Regulation Number: 21 CFR 868.5440
Regulation Name: Portable Oxygen Generator
Regulatory Class: Class II
Product Code: CAW
Dated: November 17, 2025
Received: November 17, 2025
Dear James Tsai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252268 - James Tsai
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K252268 - James Tsai
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252268 | ? |
| Please provide the device trade name(s). | | ? |
| V series portable oxygen concentrator (V5, V5C, V6, V6C) | | |
| Please provide your Indications for Use below. | | ? |
| The V series Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis in a home, institutional, or travel environment. The device is not intended for life support, nor does it provide any patient monitoring capabilities. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
1. Contact details
1.1 Applicant information
| Applicant Name | Shenzhen HarveyMed Technology Co., Ltd. |
| --- | --- |
| Address | B203 Weike science and technology building, No. 20, Gaoxin West Road, Dongzhou community, Guangming Street
Guangming District, Shenzhen, Guangdong, China |
| Phone No. | +86-755-27900675 |
| Contact person | Huang Zhi'ang |
| E-mail | zhiang.huang@harveymed.com |
| Date Prepared | November 14th, 2025 |
1.2 Submission Correspondent
| Company | Shenzhen Joyantech Consulting Co., Ltd. |
| --- | --- |
| Address | 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong, 518000, China |
| Phone No. | +86-755-86069197 |
| Contact person | James Tsai |
| E-mail | james_tsai@cefda.com; zhiang.huang@harveymed.com |
| Website | http://www.cefda.com |
2. Device information
2.1 Subject device
| Trade name | V series portable oxygen concentrator |
| --- | --- |
| Model | V5, V5C, V6, V6C |
| Classification | II |
| Classification name | Portable Oxygen Generator |
| Product code | CAW |
| Regulation No. | 868.5440 |
2.2 Primary predicate device
| Trade Name | Inogen Rove 6 Portable Oxygen Concentrator |
| --- | --- |
| 510(k) Number | K230052 |
| Sponsor | Inogen, Inc. |
| Classification | II |
| Classification name | Portable Oxygen Generator |
| Product code | CAW |
| Regulation No. | 868.5440 |
2.3 Secondary predicate device
| Trade Name | Portable Oxygen Concentrator (Model: P2-TOC) |
| --- | --- |
| 510(k) Number | K242718 |
| Sponsor | Qingdao Kingon Medical Science and Technology Co., Ltd. |
| Classification | II |
| Classification name | Portable Oxygen Generator |
1 / 8
{5}
Product code CAW
Regulation No. 868.5440
# 3. Device Description
The V series Portable Oxygen Concentrator is intended to release oxygen for respiratory therapy by means of molecular sieve/pressure swing adsorption technology. The machine utilizes the physical principle of pressure swing adsorption with a molecular sieve to adsorb nitrogen and other gas components Device to increase oxygen concentration. It supplies a pulsed high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The V series devices are small, portable and may be used in home, institutional, or travel environment.
The V series Portable Oxygen Concentrator has both pulse mode and continuous mode, there are 6 level/gear settings for pulse mode, and 1 level/gear setting for continuous mode, see the descriptions in the table below:
| Gear Settings | Level 1 (Gear 1) | Level 2 (Gear 2) | Level 3 (Gear 3) | Level 4 (Gear 4) | Level 5 (Gear 5) | Level 6 (Gear 6) | Mode C |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Rated Flow Rate | 210 mL/min | 420 mL/min | 630 mL/min | 840 mL/min | 1000 mL/min | 1200 mL/min | 1.2 L/min |
The 4 models have the same appearances, intended use, technical characteristics, and working principle, only a difference in gear setting, that is, the 4 models differ in the firmware/software, see the description in table below:
| Models | Level/Gear setting | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 | Level 6 | Mode C |
| V5 | ● | ● | ● | ● | ● | / | / |
| V5C | ● | ● | ● | ● | ● | / | ● |
| V6 | ● | ● | ● | ● | ● | ● | / |
| V6C | ● | ● | ● | ● | ● | ● | ● |
| “●” means “standard configuration”; “/” means “do not have the configuration” | | | | | | | |
# 4. Intended Use/Indication for Use
The V series Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis in a home, institutional, or travel environment. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
# 5. Substantial Equivalence Comparison
The subject device and predicate devices have the same intended use and the following similarities:
- Same Indications for Use.
- Same operating principle.
- Similar technological characteristics.
See the comparison information in details in the table below.
{6}
| Item | HarveyMed V series
(Proposed device) | Inogen Rove 6 Portable Oxygen Concentrator, K230052
(Primary predicate device) | Kingon, P2-TOC, K242718
(Secondary predicate device) | Comparisons |
| --- | --- | --- | --- | --- |
| Regulation number | 21 CFR 868.5440 | 21 CFR 868.5440 | 21 CFR 868.5440 | Substantial equivalence |
| Classification | Class II | Class II | Class II | Substantial equivalence |
| Product Code | CAW | CAW | CAW | Substantial equivalence |
| Intended use/Indications for use | The V series Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis in a home, institutional, or travel environment.
The device is not intended for life support, nor does it provide any patient monitoring capabilities. | The Inogen Rove 6 Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in the home, institution, and transport modalities.
This device is to be used as an oxygen supplement and is not intended to be life sustaining or life supporting. | The Portable Oxygen Concentrator P2-TOC is intended to provide supplemental low flow oxygen.
The device is not intended for life support, nor does it provide any patient monitoring capabilities. | Substantial equivalence |
| Environment of Use | Home, institutional, or travel environment | Home, institutional, or travel modalities | Home, outside the home | Substantial equivalence |
| Intended patient | 1. Adult
2. Patients requiring supplemental oxygen | 1. Adult
2. Patients requiring supplemental oxygen | 1. Adult
2. Patients requiring supplemental oxygen | Substantial equivalence |
| Prescriptive | Yes | Yes | Yes | Substantial equivalence |
| Scientific technology/Principle of operation | 1. Breath detection technology.
2. Molecular Sieve/pressure swing adsorption technology. | 1. Breath detection technology.
2. Molecular Sieve/pressure swing adsorption technology. | 1. Breath detection technology.
2. Pressure Swing Adsorption with molecular sieve. | Substantial equivalence |
3 / 8
{7}
| Item | HarveyMed V series
(Proposed device) | Inogen Rove 6 Portable Oxygen Concentrator, K230052
(Primary predicate device) | Kingon, P2-TOC, K242718
(Secondary predicate device) | Comparisons |
| --- | --- | --- | --- | --- |
| User/patient interface | -Buttons
-LCD Display
-Nasal cannula
-Sieve beds
-Particle filter
-Alarm Indicator
-Auditory Speaker | -LCD display
-Alarm Indicator
-Breath detect notification
-Auditory Speaker
-Battery release latch
-Sieve beds
-Particle filter | -Buttons
-LCD Display
-Nasal cannula
-Sieve beds
-Particle filter
-Alarm Indicator
-Auditory Speaker | Substantial equivalence |
| Components | -Oxygen concentrator
-AC adapter
-DC connector
-Nasal cannula
-Battery
-Shoulder bag | -Oxygen concentrator
-AC/DC power adapter
-DC power cable
-Cannula
-Battery
-Carry Bag, Backpack, Cart, External Battery Charger | -Oxygen concentrator
-AC Power Supply,
-Nasal cannula
-Battery
-Backpack, Cart | Substantial equivalence |
| Alarm function | -Respiratory detection failure.
-Oxygen output detection failure or low concentration.
-Battery/system temperature too High or battery temperature too low.
-Adapter voltage too low/high.
-Battery empty or not connected or life depleted.
-Airway transmission blockage
-Molecular sieve replacement needed.
-Compressor or fan stopped.
-Gas tank pressure failure or overpressure. | -Alarm Indicator - yellow LED on UIP above “Alarm/Warning” triangle symbol that illuminates to indicate abnormal operating conditions in compliance with ISO 60601-1-8.
-Auditory Speaker - Audible beeps are emitted to indicate alarm or status change conditions in compliance with ISO 60601-1-8. | -Loss of power.
-Low Battery.
-Low O₂ Concentration.
-Gas Delivery Fail.
-Absence of Breath (Only in -Pulse Mode).
-System Startup Fail. | Substantial equivalence (Note 1) |
| Status indicator | -Tool icon
-Battery status gauge
-External Power Indicator
-Flow Control Setting
-Technical Alarm
-Buzzer | -Tool icon
-Battery status gauge
-External Power Indicator
-Flow Control Setting
-Technical Alarm
-Auditory buzzer | -Tool icon
-Battery status gauge
-External Power Indicator
-Flow Control Setting
-Low Priority Technical Alarm
-Buzzer | Substantial equivalence |
{8}
| Item | HarveyMed V series
(Proposed device) | Inogen Rove 6 Portable Oxygen Concentrator, K230052
(Primary predicate device) | Kingon, P2-TOC, K242718
(Secondary predicate device) | Comparisons |
| --- | --- | --- | --- | --- |
| Power supply | AC Adaptor: a.c.100-240V 50/60Hz, 2.0-1.0A
DC Adaptor: input-d.c.13.5-15.0V, output-d.c.13.5-15.0V,10A
Battery type: Lithium Ion | AC power: 100 to 240 VAC, 50 to 60 Hz
Autosensing 2.0 - 1.0A
DC power: 13.5-15.0, 24 VDC, 120W Max voltage: 12.0 to 16.8 VDC (+ 0.5)
Battery type: Lithium Ion | AC Adaptor (Input: 100-240 VAC, 50-60 Hz)
DC Adaptor (Input: 11-16 VDC
Output: 19V)
Battery (Lithium Ion) | Substantial equivalence
(Note 2) |
| Filter | Air Inlet Filter | Air Inlet Filter | Air Inlet Filter | Substantial equivalence |
| Weight | 2.2kg/4.85lbs (with standard battery)
2.6kg/5.73lbs (with long-lasting battery) | 4.8 lbs (2.18kg, with standard battery)
5.8 lbs (2.63kg, with extended batter) | 18.8 lbs/8.5kg (with battery) | Substantial equivalence
(Note 3) |
| Size | 18.6 * 8.4 * 20.8 cm (with standard battery version)
(i.e., 8.2" * 3.3" * 7.3" inch) | 8.1" * 3.3" * 7.2" (with standard 8-cell battery)
(i.e., 20.6 * 8.4 * 18.3cm) | 11.1 L x 6.7W x 15H (inch)
28.2 L x 17.1 W x 38.2 H (cm) | Substantial equivalence
(Note 3) |
| Operating Environment | 41°F to 104°F (5°C to 40°C)
0% to 90%, non-condensing
0 to 10,000 ft. (0 to 3048 meters, 70.0kPa to 101.3kPa) | 41°F to 104°F (5°C to 40°C)
15% to 90%, non-condensing
0 to 10,000 ft (0 to 3048 meters) | 41°F to 104°F (5°C to 40°C)
10% to 90%, non-condensing
0 to 10,000 ft. (0 to 3048 meters, 70kPa to 106kPa) | Substantial equivalence |
| Shipping Storage | -13°F to 158°F (-25°C to 70°C);
0% to 95%, non-condensing
Store in a dry environment
0 to 10,000 ft (0 to 3048 meters, 70.0kPa to 101.3kPa) | -13°F to 158°F (-25°C to 70°C);
Up to 90%, non-condensing
Store in a dry environment.
0 to 10,000 ft (0 to 3048 meters, 70kPa to 106 kPa) | -4°F to 158°F (-20°C to 70°C)
5% to 90%, non-condensing
Store in a dry environment
0 to 10,000 ft (0 to 3048 meters, 70kPa to 106 kPa) | Substantial equivalence |
| Oxygen delivery mode | Pulse dose and continuous mode | Pulse dose | Pulse dose and continuous mode | Substantial equivalence |
{9}
| Item | HarveyMed V series (Proposed device) | | | | | | | Inogen Rove 6 Portable Oxygen Concentrator, K230052 (Primary predicate device) | | | | | | | Kingon, P2-TOC, K242718 (Secondary predicate device) | Comparisons |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Output flow (pulse mode) | BPM | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 | Level 6 | BPM | Setting 1 | Setting 2 | Setting 3 | Setting 4 | Setting 5 | Setting 6 | Pulse Dose Mode: 5-133 ml Pulse Volumes, breathing rate is 15 times a minute Pulse Dose Mode: 5-200 ml Pulse Volumes, breathing rate is 10 times a minute | Substantial equivalence (Note 4) |
| | 10 | 21 | 42 | 63 | 84 | 100 | 120 | 10 | 21.0 | 42.0 | 63.0 | 84.0 | 105.0 | 126.0 | | |
| | 15 | 14 | 28 | 42 | 56 | 66.7 | 80 | 15 | 14.0 | 28.0 | 42.0 | 56.0 | 70.0 | 84.0 | | |
| | 20 | 10.5 | 21 | 31.5 | 42 | 50 | 60 | 20 | 10.5 | 21.0 | 31.5 | 42.0 | 52.5 | 63.0 | | |
| | 25 | 8.4 | 16.8 | 25.2 | 33.6 | 40 | 48 | 25 | 8.4 | 16.8 | 25.2 | 33.6 | 42.0 | 50.4 | | |
| | 30 | 7 | 14 | 21 | 28 | 33.3 | 40 | 30 | 7.0 | 14.0 | 21.0 | 28.0 | 35.0 | 42.0 | | |
| | 35 | 6 | 12 | 18 | 24 | 28.6 | 34 | 35 | 6.0 | 12.0 | 18.0 | 24.0 | 30.0 | 36.0 | | |
| | 40 | 5.25 | 10.5 | 15.8 | 21 | 25 | 30 | 40 | 5.25 | 10.5 | 15.75 | 21.0 | 26.3 | 31.5 | | |
| | Total (ml/min) | 210 | 420 | 630 | 840 | 1000 | 1200 | Total (ml/min) | 210 | 420 | 630 | 840 | 1050 | 1260 | | |
| Output flow (continuous mode) | V5, V6: Not applicable V5C, V6C: 1 gear, continuous Flow Mode (Mode C): 1.2 lpm | | | | | | | Not applicable | | | | | | | 1-5 gear, continuous Flow Mode: 1.0-3.0 lpm | Substantial equivalence (Note 5) |
| Oxygen purity | ≥90% at all settings | | | | | | | 90%-3%/+6% at all settings (87%-96%) | | | | | | | 90%-3%/+6% at all settings (87%-96%) | Substantial equivalence |
| Inspiratory trigger sensitivity | ≤0.12 cmH2O | | | | | | | <0.12cm/H2O | | | | | | | Not publicly available | Substantial equivalence |
| Maximum outlet pressure | < 199.3kPa (28.9 PSI) | | | | | | | < 28.9 PSI (199.3kPa) | | | | | | | Not publicly available | Substantial equivalence |
| Performance standard | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-8 IEC 60601-1-11 ISO 80601-2-69 ISO 80601-2-67 | | | | | | | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-8 IEC 60601-1-11 ISO 80601-2-69 ISO 80601-2-67 | | | | | | | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-8 IEC 60601-1-11 ISO 80601-2-69 ISO 80601-2-67 | |
| Material of Patient contact components | Button film: PET Main housing: PC+ABS | | | | | | | Not publicly available | | | | | | | Button panel: PET Main housing: PC+ABS | Substantial equivalence (Note 6) |
| Biocompatibility | ISO 10993-1 ISO 18562-1 | | | | | | | ISO 10993-1 ISO 18562-1 | | | | | | | ISO 10993-1 ISO 18562-1 | Substantial equivalence |
{10}
Note 1: Although the descriptions of the alarms are different, they are all related to battery status, oxygen flow and concentration, and device status. And the subject device is tested the standard of IEC60601-1-8 and ISO80601-2-69. Therefore, the difference will not raise any different questions of safety and effectiveness for the subject device.
Note 2: Although the specifications of the AC adapter and DC connector between the subject device and predicate devices are different, the power supply of the subject device complies with the requirements of IEC60601-1, IEC60601-1-2, ISO80601-2-67, ISO80601-2-69 and IEC62133-2. Therefore, the difference will not raise any different questions of safety and effectiveness for the subject device.
Note 3: The weight and size of the subject device is similar to the Inogen Rove 6, and different from Kingon P2-TOC due to their different designs. The risk is mitigated by testing the standards of IEC60601-1, ISO80601-2-67, and ISO80601-2-69. Therefore, the difference will not raise any different questions of safety and effectiveness for the subject device.
Note 4: The pulse mode settings of the subject device are the same as the Inogen Rove 6: 6 levels/ gears, 10-40 breaths per minutes. Only the pulse volumes in gear 5 & gear 6 of the subject device is a little smaller than that of Inogen Rove 6. The risks are mitigated by testing the standards of IEC60601-1, ISO80601-2-67, ISO80601-2-69, and ISO10993-1 & ISO18562-1, as well as verification & validation by FDA's software guidance. Therefore, the difference will not raise any different questions of safety and effectiveness for the subject device.
Note 5: V5, V6 of the subject device do not have the continuous mode, which is the same as Inogen Rove 6. The flowrate range of Kingon P2-TOC is 1.0-3.0L/min with 5 gears under continuous mode, the flowrate of V5C, V6C is 1.2L/min under continuous mode, which is in the flowrate range of Kingon P2-TOC. The continuous mode is controlled by the firmware, V5C, V6C of the subject device have been designed, verified and validated the software according to FDA's software guidance to mitigate the risks of software. The risks of safety and biocompatibility are mitigated by testing the standards of IEC60601-1, ISO80601-2-67, ISO80601-2-69, and ISO10993-1 & ISO18562-1. Therefore, the difference will not raise any different questions of safety and effectiveness for the subject device.
Note 6: Materials of Inogen Rove 6 are not publicly available, but the biocompatibility evaluations including the biocompatibility tests of the subject device are carried out according to FDA biological guidance, and the results indicate that the subject device has a good biocompatibility performance. Therefore, the differences will not raise any different questions of safety and effectiveness issues for the subject device.
The subject device and the predicate devices are similar/same in indication for use, operating principle, and technology characteristics. There is no significant difference between subject device and predicate devices that raise any different questions of safety and effectiveness for the subject device.
## 6. Non-clinical Testing
### 6.1 Electrical Safety and EMC
Tests have been conducted on the subject device. The system complies with the following standards for electrical safety and EMC:
- IEC60601-1:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - requirements and tests
- IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC60601-1-6:2010+A2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
7 / 8
{11}
- IEC 60601-1-8:2020, Medical electrical device - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests, and guidance for alarm systems in medical electrical device and medical electrical systems
- IEC 60601-1-11:2020 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in The Home Healthcare Environment
- ISO 80601-2-67:2020, Medical electrical device - Part 2-67: Particular requirements for basic safety and essential performance of oxygen conserving device
- ISO 80601-2-69:2020, Medical electrical device - Part 2-69: Particular requirements for basic safety and essential performance of oxygen concentrator device
- IEC 62133-2:2017 + A1:2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
## 6.2 Biocompatibility
The subject device is categorized as both surface contact medical device and external communication device. Evaluation and tests have been conducted in accordance with the following standards:
- ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices - Part 10: Tests for irritation
- ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds (VOCs)
## 6.3 Performance
The following nonclinical tests were conducted with the subject device:
- Oxygen pulse volume
- Continuous flow
- Trigger sensitivity
- Oxygen concentration
- Respiratory rate
- Outlet pressure
- Low oxygen concentration alarm
## 6.4 Software
Software verification and validation were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff:
- Guidance for Industry and Food and Drug Administration Staff - Content of Premarket Submissions for Device Software Functions.
## 7. Clinical testing
Not applicable.
## 8. Conclusions
Based on the results of substantial equivalence assessment, and the safety and performance testing data, it concludes that the subject device is substantially equivalent to the predicate devices.
8 / 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.