InferCare RECIST is a post-processing software application used to display, process, analyze, quantify and manipulate multi-time-point CT images. It is intended to be used by trained medical professionals in evaluating and managing tumors in various organs, tissues, and other anatomical structures based on RECIST criteria. InferCare RECIST offers functionalities for lesion measurement, registration, tracking, and RECIST report. It provides tools for interactive segmentation, and 3D reconstruction visualization. The software utilizes artificial intelligence algorithms for automated lesion segmentation and registration, and the results require confirmation by a medical professional. The software's artificial intelligence algorithms are intended for patients aged 21 years and older.
Device Story
Post-processing software for multi-time-point CT images; used by radiologists and oncologists in hospitals/clinics. Ingests DICOM CT images; performs automated lesion segmentation and registration using AI algorithms; provides interactive tools for 3D reconstruction, MPR, and MIP. Physician selects lesions; software tracks and compares across follow-up studies; generates RECIST-standard reports. Output assists clinicians in tumor management, lesion measurement, and diagnostic/treatment decision-making. Results require professional confirmation.
Clinical Evidence
Bench testing only. Performance evaluated on 212 target objects for registration (Match Rate 0.995; Centroid Error MAE 2.44mm). Segmentation accuracy tested on 102 cases (lung, liver, kidney, lymph node); mean Dice scores ranged from 0.782 to 0.929; Long/Short diameter MAE ranged from 3.2% to 5.3%.
Technological Characteristics
Software-based medical image processing system. Utilizes AI algorithms for automated segmentation and registration. Supports DICOM standard. Features include 2D viewing, 3D reconstruction, MPR, MIP, and interactive segmentation editing. Deployed in healthcare facilities or cloud-based platforms. Compliant with IEC 62304 and ISO 14971.
Indications for Use
Indicated for trained medical professionals to evaluate and manage tumors in various organs, tissues, and anatomical structures using RECIST criteria on multi-time-point CT images. AI algorithms are indicated for patients aged 21 years and older.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 13, 2026
Beijing Infervision Healthcare Medical Technology Co., Ltd.
% Matt Deng
Primary Contact
Room B403, 4th Floor, Building 1
No.12, Shangdi Information Road, Haidian District
BEIJING, 100085
CHINA
Re: K252261
Trade/Device Name: InferCare RECIST
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: February 12, 2026
Received: February 12, 2026
Dear Matt Deng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252261 - Matt Deng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252261 - Matt Deng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252261 | |
| Device Name InferCare RECIST | |
| Indications for Use (Describe) InferCare RECIST is a post-processing software application used to display, process, analyze, quantify and manipulate multi-time-point CT images. It is intended to be used by trained medical professionals in evaluating and managing tumors in various organs, tissues, and other anatomical structures based on RECIST criteria. InferCare RECIST offers functionalities for lesion measurement, registration, tracking, and RECIST report. It provides tools for interactive segmentation, and 3D reconstruction visualization. The software utilizes artificial intelligence algorithms for automated lesion segmentation and registration, and the results require confirmation by a medical professional. The software's artificial intelligence algorithms are intended for patients aged 21 years and older. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
InferCare RECIST
# 510(k) Summary-K252261
## InferCare RECIST
**Submitter:** Beijing Infervision Healthcare Medical Technology Co., Ltd.
**Address:** Room B403, 4th Floor, Building 1, No.12, Shangdi Information Road, Haidian District, Beijing 100085, China. Phone: +86 10-86462323
**Contact:** Mr. Matt Deng
Email: matt.deng@infervision.ai
Phone: 929-335-4879
**Date Prepared:** March 12, 2026
## Device Name and Classification
- **Trade Name:** InferCare RECIST
- **Classification:** Class II
- **Common Name:** Automated radiological image processing software
- **Regulation Name:** Medical image management and processing system
- **Regulation Number:** 21 CFR 892.2050
- **Classification Panel:** Radiology
- **Product Code:** QIH
## Predicate Device
- **Trade Name:** Multi-Modality Tumor Tracking (MMTT) application
- **Manufacturer:** Philips Medical Systems Nederland B.V.
- **Classification:** Class II
- **510(k) Number:** K162955
- **Common Name:** System, image processing, radiological
- **Regulation Name:** Medical image management and processing system
- **Regulation Number:** 21 CFR 892.2050
- **Classification Panel:** Radiology
- **Product Code:** LLZ
{5}
InferCare RECIST
# Device Description
InferCare RECIST is based on advanced image processing technologies and temporal sequence analysis methods. The software can be installed in a healthcare facility or a cloud-based platform. It ingests medical imaging data compliant with the DICOM standard and supports multi-timepoint image registration. It allows automatic navigation to match lesion locations across different image series based on physician-selected lesions, ensuring accurate tracking and comparison of the same lesion across follow-up studies.
InferCare RECIST integrates machine learning based algorithms to provide interactive segmentation tools, enabling users to generate 3D segmentation and measurement results simply by clicking on the target lesion.
In addition, the system offers a variety of image processing tools, such as window width/level adjustment, multi-planar reconstruction (MPR), and maximum intensity projection (MIP), to meet the needs of different clinical viewing scenarios. Through the integrated use of these technologies, InferCare RECIST delivers accurate lesion measurements, follow-up trend analyses, and RECIST-standard evaluation reports, effectively supporting physicians in making precise diagnostic and treatment decisions.
# Indications for Use
InferCare RECIST is a post-processing software application used to display, process, analyze, quantify and manipulate multi-time-point CT images. It is intended to be used by trained medical professionals in evaluating and managing tumors in various organs, tissues, and other anatomical structures based on RECIST criteria.
InferCare RECIST offers functionalities for lesion measurement, registration, tracking, and RECIST report. It provides tools for interactive segmentation, and 3D reconstruction visualization.
The software utilizes artificial intelligence algorithms for automated lesion segmentation and registration, and the results require confirmation by a medical professional. The software's artificial intelligence algorithms are intended for patients aged 21 years and older.
# Substantial Equivalence
The subject device is substantially equivalent to the predicate device in the following ways:
| Item | Subject Device: InferCare RECIST | Predicate Device: Multi-Modality Tumor Tracking (MMTT) application | Comparison |
| --- | --- | --- | --- |
| Application Number | K252261 | K162955 | - |
| Product Code | QIH | LLZ | - |
| Class | Class II | Class II | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
{6}
| Item | Subject Device: InferCare RECIST | Predicate Device: Multi-Modality Tumor Tracking (MMTT) application | Comparison |
| --- | --- | --- | --- |
| Indications for Use/ Intended use | InferCare RECIST is a post-processing software application used to display, process, analyze, quantify and manipulate multi-time-point CT images. It is intended to be used by trained medical professionals in evaluating and managing tumors in various organs, tissues, and other anatomical structures based on RECIST criteria. InferCare RECIST offers functionalities for lesion measurement, registration, tracking, and RECIST report. It provides tools for interactive segmentation, and 3D reconstruction visualization. The software utilizes artificial intelligence algorithms for automated lesion segmentation and registration, and the results require confirmation by a medical professional. The software's artificial intelligence algorithms are intended for patients aged 21 years and older. | Multi-Modality Tumor Tracking (MMTT) application is a post processing software application used to display, process, analyze, quantify and manipulate anatomical and functional images, for CT, MR PET/CT and SPECT/CT images and/or multiple time-points. The MMTT application is intended for use on tumors which are known/confirmed to be pathologically diagnosed cancer. The results obtained may be used as a tool by clinicians in determining the diagnosis of patient disease conditions in various organs, tissues, and other anatomical structures. | Same. Both devices are used for medical image processing to display, process, analyze, quantify, manipulate multi-time-point images and enable RECIST 1.1 categorization. |
| Intended Users | Radiologists, oncologists, and others trained in reading medical images and applying RECIST criteria | Radiologists, Technologist | Same |
| Where used | Healthcare facilities such as hospitals and clinics | Healthcare facilities such as hospitals and clinics | Same |
| Image Input | CT DICOM images | CT, MR, PET/CT and SPECT/CT DICOM images | Similar. The predicate device includes a wider variety |
InferCare RECIST
{7}
| Item | Subject Device: InferCare RECIST | Predicate Device: Multi-Modality Tumor Tracking (MMTT) application | Comparison |
| --- | --- | --- | --- |
| | | | of imaging modalities. |
| 2D DICOM viewing | Yes | Yes | Same |
| 3D Reconstruction | Automatic | Automatic | Same |
| Multi-Planar Reconstructions (MPR) | Yes | Yes | Same |
| Registration | Automatic and manual | Automatic and manual | Same |
| Lesion segmentation | Semi-automatic and manual | Semi-automatic and manual | Same |
| Segmentation editing tools | Add, Delete, Edit, Undo | Add, Delete, Edit, Undo | Same |
| Automatic software measurement of segmented lesion | Long axis, short axis, volume, key slice. | Long axis, short axis, volume, key slice, density, etc. | Same |
| RECIST Report | Yes | Yes | Same |
| Algorithm characteristics | Artificial Intelligence (AI) algorithm | Traditional algorithm | The algorithm performance of the subject device has been validated through performance testing. |
The subject device and the predicate device are substantially equivalent in indications for use, input/output, image registration, and interactive lesion segmentation. Regarding the technical differences between the two devices, standalone performance testing has demonstrated that the subject device meets the predetermined target values.
## Performance data
Software verification and validation activities were conducted in accordance with IEC 62304:2006+A1:2015 – Medical device software – Software lifecycle processes and ISO 14971:2019 Medical devices – Application of risk management to medical devices, and in accordance with relevant FDA guidance documents, Guidance for the Content of Premarket Submissions for Device Software Functions (issued June 14, 2023), and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (issued September 27, 2023).
InferCare RECIST
{8}
Cybersecurity and vulnerability analyses were conducted, and it has been determined that InferCare RECIST conforms to the cybersecurity requirements.
The following processes were followed and applied during the design and development of InferCare RECIST:
- Risk Analysis
- Unit Testing
- Integration Testing
- System Testing
Performance Testing
- Software Verification & Validation
- Cybersecurity Testing and Analysis
# Performance of AI algorithm
# 1. Image Registration Performance Test
(1) For the primary endpoint: Match Rate, a total of 212 target objects including 184 paired, 21 new and 7 disappeared target objects were annotated as ground truth. Among all 212 pairs, 183 paired, 21 new and 7 disappeared target objects were successfully matched across prior and primary scans by InferCare RECIST. Patient-level bootstrapping: the Match Rate was 0.995 (95% CI: 0.986-1.000), with the lower limit of the 95% C.I. being 0.986.
(2) For the secondary endpoint: Centroid Error Distance: Patient-level bootstrapping: A total of 183 successfully matched target object pairs in 86 patient-scans were included in the analysis. The MAE of centroid error distance was $2.44\mathrm{mm}$ (95% C.I.: $2.22\mathrm{mm} - 2.68\mathrm{mm}$ ), with the higher limit of the $95\%$ C.I. being $2.68\mathrm{mm}$ .
# 2. Lesion Segmentation Performance Test
(1) Segmentation accuracy was set as the primary endpoint, and Long/Short diameter measurement was set as the secondary endpoint. We used the same batch of data to test these two endpoints. The dataset comprises a total of 102 cases, including 42 cases of lung nodules, 23 cases of liver lesions, 23 cases of kidney lesions, and 14 cases of lymph node lesions. The result is listed below:
| | | Primary endpoint Segmentation Accuracy | | Secondary endpoint Long/Short Diameter | |
| --- | --- | --- | --- | --- | --- |
| Item | N | Mean (Dice) | 95% CI | MAE (%) | 95% CI |
| Lung nodule | 42 | 0.913 | 0.895-0.932 | 3.2 | 2.2-4.2 |
| Liver lesion | 23 | 0.929 | 0.913-0.944 | 4.3 | 3.1-5.5 |
| Kidney lesion | 23 | 0.904 | 0.889-0.918 | 4.5 | 3.3-5.6 |
| Lymph node | 14 | 0.782 | 0.750-0.814 | 5.3 | 3.0-7.6 |
InferCare RECIST
{9}
# Conclusion
The InferCare RECIST is substantially equivalent in intended use, main functions, operating principles, and safety features to the predicate device. The minor differences in technological characteristics do not raise different questions of safety or effectiveness. The performance testing reports and the submitted documentations demonstrate that the subject device and the predicate device are substantially equivalent.
InferCare RECIST
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.