Osteotomy Wedge System

K252254 · Sinaptic Surgical · PLF · Oct 17, 2025 · Orthopedic

Device Facts

Record IDK252254
Device NameOsteotomy Wedge System
ApplicantSinaptic Surgical
Product CodePLF · Orthopedic
Decision DateOct 17, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SiNAPTIC Osteotomy Wedge System is intended to be used for internal bone fixation for fractures or osteotomies in the ankle and foot, such as: 1. Opening wedge osteotomies of the bones of the foot, including osteotomies for Hallux Valgus 2. Opening wedge of the medial cuneiform or Cotton osteotomies. 3. Lateral Column Lengthening (Evans Lengthening Osteotomy or Calcaneal Z-Osteotomy) 4. Metatarsal Cuneiform osteotomies 5. Nonunion of arthrodesis of the Midfoot including Metatarsal Cuneiform osteotomies (TMT or Lapidus) 6. Hindfoot osteotomies such as Ankle fusion and Subtalar fusion The Osteotomy Wedge System MTP Wedges are intended to be used for internal fixation of metatarsal bones that have been surgically prepared (osteotomy) for correction of deformity. The Osteotomy Wedge System is indicated for use with supplemental fixation. The Osteotomy Wedge System is not indicated for use in the spine.

Device Story

Osteotomy Wedge System consists of sterile, single-use Silicon Nitride (Si3N4) implants; available in various shapes, footprints, and thicknesses. Designed for internal bone fixation during foot and ankle osteotomies or fracture repair. Implants feature insertion points and central voids for bone graft packing or supplemental fixation hardware. Used by orthopedic surgeons in clinical/surgical settings. Provides structural support to maintain osteotomy gap during bone healing; requires supplemental fixation (e.g., plates/screws) for stability. Benefits patient by facilitating bone alignment and fusion in corrective foot/ankle procedures.

Clinical Evidence

Bench testing only. Performance evaluated via static compressive strength, dynamic compressive strength, static compression shear strength, and expulsion testing. No clinical data provided.

Technological Characteristics

Material: Silicon Nitride (Si3N4). Form factor: Wedges with varying footprints/thicknesses. Features: Insertion feature, central open area for graft/fixation. Principle: Mechanical bone fixation/spacer. Sterilization: Sterile, single-use.

Indications for Use

Indicated for internal bone fixation for fractures or osteotomies in the ankle and foot, including Hallux Valgus, medial cuneiform/Cotton osteotomies, lateral column lengthening, metatarsal cuneiform osteotomies, midfoot nonunion/arthrodesis, and hindfoot osteotomies (ankle/subtalar fusion). MTP wedges indicated for metatarsal bone fixation post-osteotomy. Requires supplemental fixation. Contraindicated for use in the spine.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 17, 2025 SiNAPTIC Surgical Brian Hockett Vice President of Quality and Regulatory 1885 West 2100 South Slat Lake City, Utah 84119 Re: K252254 Trade/Device Name: Osteotomy Wedge System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PLF Dated: July 21, 2025 Received: July 21, 2025 Dear Brian Hockett: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252254 - Brian Hockett Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252254 - Brian Hockett Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S., Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3} K252254 - Page 1 of 1 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252254 | ? | | Please provide the device trade name(s). | | ? | | Osteotomy Wedge System | | | | Please provide your Indications for Use below. | | ? | | The SiNAPTIC Osteotomy Wedge System is intended to be used for internal bone fixation for fractures or osteotomies in the ankle and foot, such as: 1. Opening wedge osteotomies of the bones of the foot, including osteotomies for Hallux Valgus 2. Opening wedge of the medial cuneiform or Cotton osteotomies. 3. Lateral Column Lengthening (Evans Lengthening Osteotomy or Calcaneal Z-Osteotomy) 4. Metatarsal Cuneiform osteotomies 5. Nonunion of arthrodesis of the Midfoot including Metatarsal Cuneiform osteotomies (TMT or Lapidus) 6. Hindfoot osteotomies such as Ankle fusion and Subtalar fusion The Osteotomy Wedge System MTP Wedges are intended to be used for internal fixation of metatarsal bones that have been surgically prepared (osteotomy) for correction of deformity. The Osteotomy Wedge System is indicated for use with supplemental fixation. The Osteotomy Wedge System is not indicated for use in the spine. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} SiNAPTIC SURGICAL # 510(k) SUMMARY ## Submitter Information Submitter's Name: SiNAPTIC Surgical, LLC Address: 1885 West 2100 South Salt Lake City, UT 84119 Phone Number: (855) 839-3500 Prepared By: Brian Hockett Contact Person: Brian Hockett Date Prepared: 7/17/2025 ## Device Information Trade Name: Osteotomy Wedge System Common Name: Bone Wedge Classification: Class II per 21 CFR 888.3030 Panel: Orthopedic, Product Code: PLF ## Predicate Device Identification | 510(k) Number | Device Name | Predicate | | --- | --- | --- | | K243231 | Nvision Biomedical's Trigon HA Stand-Alone Wedge Fixation System | Primary | | K240461 | AMES's Medical OsteoSinter Evans and Cotton Wedges | Additional | | K201314 | Restor3d Utility Wedge | Additional | | K220513 | CTL Medical Nitro Interbody Fusion Cage System Family | Reference | ## Indications for Use The SiNAPTIC Osteotomy Wedge System is intended to be used for internal bone fixation for fractures or osteotomies in the ankle and foot, such as: 1. Opening wedge osteotomies of the bones of the foot, including osteotomies for Hallux Valgus 2. Opening wedge of the medial cuneiform or Cotton osteotomies. 3. Lateral Column Lengthening (Evans Lengthening Osteotomy or Calcaneal Z-Osteotomy) 4. Metatarsal Cuneiform osteotomies 5. Nonunion of arthrodesis of the Midfoot including Metatarsal Cuneiform osteotomies (TMT or Lapidus) 6. Hindfoot osteotomies such as Ankle fusion and Subtalar fusion The Osteotomy Wedge System MTP Wedges are intended to be used for internal fixation of metatarsal bones that have been surgically prepared (osteotomy) for correction of deformity. The Osteotomy Wedge System is indicated for use with supplemental fixation. The Osteotomy Wedge System is not indicated for use in the spine. ## Device Description The Osteotomy Wedge System devices are a sterile, single use Silicon Nitride (Si₃N₄) implants. The wedges are designed in a variety of shapes, each available in multiple footprints and thickness. The Osteotomy Wedge System is intended to be used for internal bone fixation for opening or lengthening osteotomies in the foot and ankle. The Page 1 of 2 {5} SiNAPTIC SURGICAL wedges contain an insertion feature as well as a central, open area for packing graft material or insertion of supplemental fixation. ## Summary of Non-Clinical Data No FDA performance standards have been established for the Osteotomy Wedge System. The device was subjected to the following bench performance tests to support the assertion of substantial equivalence and evidence that no new safety or effectiveness concerns were raised: - Static compressive strength - Dynamic compressive strength - Static compression shear strength - Expulsion testing The safety and effectiveness of the Osteotomy Wedge system is adequately supported by testing, substantial equivalence information, materials information and comparison of to the predicate device. ## Summary of Equivalence to Predicate Devices The Osteotomy Wedge system is substantially equivalent to the predicate devices. The subject implants have similar shapes, sizes and features as the previously cleared devices, including similar mechanism of fixation and principles of action. The devices also have the same indications for use. These technological characteristics have undergone a comparison of characteristics to ensure the device is as safe and effective as the predicates. ## Conclusion Although the Osteotomy Wedge System differs from the predicate in the device material, the devices have similar design / physical characteristics, intended use, and performance characteristics. The Osteotomy Wedge System does not raise different questions of safety or effectiveness. The Osteotomy Wedge System is substantially equivalent to the predicate device. Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...