Injector Force Max Single Use Injector (NM-400L); Injector Force Max Single Use Injector (NM-400U); Injector Force Max Single Use Injector (NM-400Y); Injector Force Max Single Use Injector (NM-401L)
K252247 · Olympus Medical Systems Corporation · FBK · Oct 16, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K252247
Device Name
Injector Force Max Single Use Injector (NM-400L); Injector Force Max Single Use Injector (NM-400U); Injector Force Max Single Use Injector (NM-400Y); Injector Force Max Single Use Injector (NM-401L)
Applicant
Olympus Medical Systems Corporation
Product Code
FBK · Gastroenterology, Urology
Decision Date
Oct 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Single Use Injector NM-400/401 Series is intended to be used to perform endoscopic vascular or submucosal injection in the digestive tract with an endoscope.
Device Story
The Olympus Single-Use Injector NM-400/NM-401 is a sterile, disposable needle device used with endoscopes for vascular or submucosal injections in the digestive tract. The device consists of a handle, needle, and sheath. A slider mechanism on the handle allows the clinician to extend or retract the needle. Fluids are delivered through an injection port connected to the needle. The device is used in clinical settings by physicians performing endoscopic procedures. By enabling precise delivery of therapeutic agents to targeted tissue, the device assists in managing various digestive tract conditions. The design includes ergonomic handle improvements and various needle gauges (21G-25G) and working lengths (up to 2700 mm) to accommodate different endoscope working channels and clinical requirements.
Clinical Evidence
Bench testing only. Performance verified through testing of insertion/withdrawal, needle advancement/retraction, puncture, repeated operation, fluid supply, and corrosion resistance. Biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility) performed per ISO 10993. Shelf life and package integrity validated via accelerated aging (ASTM F1980-21) and ISO 11607-1/2:2019 standards.
Technological Characteristics
Materials: Stainless steel (SUS304), polypropylene (PP), perfluoroalkoxy (PFA). Sterilization: Ethylene oxide (EtO). Mechanism: Slider-actuated needle extension/retraction. Dimensions: Needle gauges 21G-25G; working lengths 1650 mm-2700 mm; insertion diameter as small as 1.9 mm. Connectivity: None (mechanical accessory).
Indications for Use
Indicated for patients requiring endoscopic vascular or submucosal injection in the digestive tract using an endoscope.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 16, 2025
Olympus Medical Systems Corporation
% Jillian Connery
Program Manager Regulatory Affairs
Olympus Surgical Technologies of America
800 West Park Drive
Westborough, Massachusetts 01581
Re: K252247
Trade/Device Name: Injector Force Max Single Use Injector (NM-400L); Injector Force Max Single Use Injector (NM-400U); Injector Force Max Single Use Injector (NM-400Y); Injector Force Max Single Use Injector (NM-401L)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FBK
Dated: July 18, 2025
Received: July 18, 2025
Dear Jillian Connery:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252247 - Jillian Connery
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252247 - Jillian Connery
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Injector Force Max Single Use Injector
Page 11 of 37
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252247 | ? |
| Please provide the device trade name(s). | | ? |
| Injector Force Max Single Use Injector (NM-400L);
Injector Force Max Single Use Injector (NM-400U);
Injector Force Max Single Use Injector (NM-400Y);
Injector Force Max Single Use Injector (NM-401L) | | |
| Please provide your Indications for Use below. | | ? |
| The Single Use Injector NM-400/401 Series is intended to be used to perform endoscopic
vascular or submucosal injection in the digestive tract with an endoscope. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
OLYMPUS
Traditional 510(k)
Single Use Injector NM-400/NM-401
# SINGLE USE INJECTOR
K252247
Page 1 of 6
# 510(k) SUMMARY
510(k) Summary: K252247
# 1. COMPANY INFORMATION
- Applicant
OLYMPUS MEDICAL SYSTEMS CORPORATION
2951 Ishikawa-cho, Hachioji-shi
Tokyo 192-8507, Japan
FDA Establishment Registration #: 8010047
- Official Correspondent
Jillian Connery
c/o Olympus Surgical Technologies of the Americas
800 West Park Drive
Westborough, MA 01581
Cell: 404-542-5854
Email: jillian.connery@olympus.com
- Manufacturing Site
Olympus Vietnam Company Limited
8 Street
Long Thanh Industrial Zone, Tam An Commune
Long Thanh District Dong Nai, VN 810000
FDA Establishment Registration #: 3008040402
- Date Prepared: 04-Sep-2025
# 2. PRODUCT INFORMATION
- Trade Name: Single Use Injector
- Models: NM-400L-0421/0423/0425/0523/0525/0621/0623/0625
NM-400U-0323/0423/0425/0523/0525/0623/0625
NM-400Y-0423, NM-401L-0423/0425/0523/0525/0623/0625
- Common Name: Injector
- Classification Name: Endoscope and Accessories
- Product Code: FBK
510k Summary
{5}
OLYMPUS
Traditional 510(k) Single Use Injector NM-400/NM-401
K252247
Page 2 of 6
- Regulation Number: 21 CFR 876.1500
- Regulation Name: Endoscope and accessories
- Device Class: II
## 3. PREDICATE DEVICE
- Trade Name: Single Use Injector
- Model: NM-600/610
- 510(k) Number: K153625
## 4. REFERENCE DEVICES
- Trade Name: Injection Needles
- Model: NM-10L/11L/12L/13L/14L/15L/16L/17L
- 510(k) Number: K902736
- Trade Name: Olympus Injector
- Model: NM-4L-1/5L-1/6L-1/7L-1/8L-1/9L-1
- 510(k) Number: K011484
## 5. DEVICE DESCRIPTION
The Olympus Single-Use Injector NM-400/NM-401 is a sterile, disposable injection needle designed for use with Olympus endoscopes to perform endoscopic vascular or submucosal injections in the digestive tract. The device comprises a handle section, needle section, and sheath section, and operates via a slider mechanism that extends and retracts the needle. The needle is connected to an injection port through which fluids can be delivered to targeted tissue. The NM-400/NM-401 series includes various models with needle gauges ranging from 21G to 25G and working lengths from 1650 mm to 2700 mm, accommodating different clinical needs and endoscope compatibilities. The device is constructed from biocompatible materials including stainless steel (SUS304), polypropylene (PP), and perfluoroalkoxy (PFA), and is sterilized using ethylene oxide (EtO). It is intended for single use only.
## 6. INDICATIONS FOR USE
The Single Use Injector NM-400/401 Series is intended to be used with an Olympus endoscope to perform endoscopic vascular or submucosal injection in the digestive tract.
Page 2 of 6
510k Summary
{6}
OLYMPUS
K252247
Page 3 of 6
Traditional 510(k)
Single Use Injector NM-400/NM-401
# 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Olympus NM-400/NM-401 and NM-600/610 injection needles are technologically similar in many core aspects.
Both are intended for endoscopic vascular or submucosal injection in the digestive tract and are classified as Class II medical devices under 21 CFR 876.1500 with the FDA product code FBK. They are designed for single use and are sterilized using ethylene oxide (EtO). Both devices are compatible with endoscopes and support a range of needle gauges and working lengths suitable for various clinical applications.
Despite these similarities, the NM-400/NM-401 introduces several enhancements and modifications. It offers both lancet bevel and flat bevel needle shapes, whereas the NM-600/610 only features lancet bevel needles. The NM-400/401 also includes models with smaller maximum insertion diameters (as small as 1.9 mm) compared to the 2.6 mm diameter of the NM-600/610, allowing compatibility with endoscopes that have smaller working channels. Additionally, the NM-400/401 includes a longer working length option of 2700 mm, which is not available in the predicate device. Material differences are also present. The NM-400/401 uses polypropylene (PP) for the outer tube and perfluoroalkoxy (PFA) for the inner tube, while the NM-600/610 uses polytetrafluoroethylene (PTFE) for both. These changes have been evaluated through biocompatibility testing to ensure safety. The handle design of the NM-400/NM-401 is more ergonomic and less bumpy than that of the predicate, although the operational mechanism remains the same. Furthermore, the NM-400/NM-401 uses a different packaging film material, which has been tested to confirm seal integrity and sterility maintenance.
A comparison of the subject and predicate device technological characteristics is provided below.
| Feature | NM-400/NM-401 | NM-600/610 |
| --- | --- | --- |
| Intended Use | Endoscopic vascular or submucosal injection in the digestive tract | Same |
| FDA Classification | Class II, 21 CFR 876.1500, Product Code FBK | Same |
| Sterilization | Ethylene oxide (EtO), single-use | Same |
| Endoscope Compatibility | Compatible with Olympus endoscopes | Same |
Page 3 of 6
510k Summary
{7}
OLYMPUS
K252247
Page 4 of 6
Traditional 510(k)
Single Use Injector NM-400/NM-401
| Feature | NM-400/NM-401 | NM-600/610 |
| --- | --- | --- |
| Needle Options | Lancet bevel and flat bevel | Lancet bevel only |
| Insertion Diameter | As small as 1.9 mm | 2.6 mm |
| Working Length | Up to 2700 mm | Shorter length only |
| Tube Materials | Outer: Polypropylene (PP); Inner: Perfluoroalkoxy (PFA) | Both Inner and Outer: Polytetrafluoroethylene (PTFE) |
| Biocompatibility | Tested and confirmed | Tested and confirmed |
| Handle Design | More ergonomic and streamlined | Bulkier, less ergonomic |
| Packaging Material | Different film, tested for seal integrity and sterility | Different material (not specified) |
| Operational Mechanism | Same basic mechanism | Same |
In summary, while the NM-400/NM-401 maintains the same fundamental design and intended use as the NM-600/610, it incorporates improvements in needle design, material composition, dimensional flexibility, and ergonomic features, all supported by appropriate performance and safety testing.
# 8. PERFORMANCE DATA
The following performance bench tests were conducted on the Olympus Single-use Injection Needle NM-400/NM-401 to demonstrate substantial equivalence to the predicate device and verify design integrity, including simulated distribution and accelerated aging:
1. Insertion into and Withdrawal from an Endoscope
2. Needle Advancement from the Tube
3. Needle Retraction into the Tube
4. Needle Puncture
5. Needle Return during Needle Puncture
6. Repeated Device Operations
7. Fluid Supply
Page 4 of 6
510k Summary
{8}
OLYMPUS
K252247
Page 5 of 6
8. Resistance to Corrosion (performed only on non-aged samples per ISO 9626)
These tests were performed on final, sterilized devices and included both baseline (T=0) and accelerated aged (T=3 years) samples. All tests met their predefined acceptance criteria, confirming the device's performance and stability over its intended shelf life.
9. BIOCOMPATIBILITY
The biocompatibility testing for the Olympus Single Use Injector (NM-400/NM-401 series) included evaluations for cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity, and hemocompatibility, all of which passed according to ISO 10993 standards. Testing confirmed the device is non-cytotoxic, non-sensitizing, non-irritating, non-pyrogenic, non-hemolytic, and not acutely toxic. Additional chemical characterization and toxicological risk assessments support the conclusion that the device is biologically safe and stable over its intended shelf life.
10. STERILIZATION AND SHELF LIFE
The shelf life of the Olympus Single-use Injection Needles NM-400/401 series has been validated for three years through accelerated aging and package integrity testing in accordance with ASTM F1980-21 and ISO 11607-1/2:2019. Accelerated aging was conducted at 60°C for 162 days to simulate three years of real-time storage, with a real-time aging study also initiated. Post-aging performance testing confirmed that the devices maintained functionality across key parameters including insertion and withdrawal, needle advancement and retraction, puncture, repeated use, and fluid delivery. Additionally, package integrity was verified through visual inspection, peel strength, dye penetration, and splitting resistance tests using a worst-case representative model. All test samples met predefined acceptance criteria, supporting the device's stability and performance over its intended shelf life.
11. SUBSTANTIAL EQUIVALENCE
Based on the intended use, technological characteristics, performance testing, and comparison to the predicate device, the Single Use Injector NM-400/NM-401 needles raise no new issues of safety and effectiveness and are substantially equivalent to the predicate device Single Use Injector NM-600/610.
510k Summary
{9}
OLYMPUS
Traditional 510(k)
Single Use Injector NM-400/NM-401
K252247
Page 6 of 6
## 12. CONCLUSION
Based on the results of the comparison of the indications for use, technological characteristics, and performance testing of the Subject and Predicate devices, the Subject Single Use Injector NM-400/NM-401 needle raises no new issues of safety and effectiveness and the device is substantially equivalent to the Predicate device.
Page 6 of 6
510k Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.