← Product Code [POV](/productcode/POV) · K252228

# Seaman Pro/Seaman (K252228)

_Checkcells, Inc. · POV · Apr 9, 2026 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252228

## Device Facts

- **Applicant:** Checkcells, Inc.
- **Product Code:** [POV](/productcode/POV.md)
- **Decision Date:** Apr 9, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** AI/ML

## Indications for Use

The Seaman PRO Analyzer is an automated, point-of-care, in vitro use-only medical device for semen analysis performed by healthcare professionals. The desktop unit consists of a light source, optical elements, and an image sensor connected to a PC, which records a video of the semen sample and sends it to the cloud compute unit. The cloud compute unit runs software containing algorithms for assessing semen parameters. The product is a software-based medical device powered by ML that enables the user to further assess semen parameters, including concentration, motility, pH, morphology, and vitality. The product is intended for adult male patients to assess sperm quality. The product is not intended to process any of the PHI/PII. Seaman PRO for prescription use only. Seaman PRO used with Imaging Hardware is an optical device for human semen analysis, which provides direct and calculated quantitative measurements for: 1. Sperm concentration (10^6 per ml) 2. Total motility (PR+NP, %) 3. Progressive motility (%) 4. pH value 5. Sperm morphology (normal forms, %) 6. Vitality (%) The Seaman PRO device does not provide a comprehensive evaluation of a male's fertility status. Instead, it analyzes semen parameters which can be used by healthcare professionals as part of a broader assessment of male fertility.

## Device Story

Seaman PRO is an automated, point-of-care, in vitro diagnostic system for semen analysis. Input: video recordings of semen samples captured by desktop imaging hardware (light source, optics, USB camera). Processing: image data transferred to PC and processed in a secure cloud compute environment using AI/ML-based image analysis algorithms to quantify sperm parameters. Output: quantitative test report for concentration, motility (total/progressive), morphology, vitality, and pH. Used by healthcare professionals in clinical settings to support fertility assessments. Does not interpret results or diagnose fertility status. Benefits: provides objective, quantitative measurements of sperm quality to assist clinicians in broader fertility evaluations. Does not process PHI/PII.

## Clinical Evidence

Clinical method comparison study (N=300 adult male semen samples) compared Seaman PRO against established reference procedures. Results showed strong agreement with correlation coefficients >0.90 and regression slopes between 0.90 and 1.10. Supported by extensive non-clinical bench testing (precision, linearity, LoB/LoD/LoQ, interference, and environmental stability) per CLSI/WHO/ISO standards.

## Technological Characteristics

Modular system: desktop optical unit (light source, fixed optics, digital image sensor), PC, and cloud-based software. Uses GoldCyto semen analysis slides. Connectivity: PC to cloud. Software: ML-based image analysis. Sterilization: Not required (reusable hardware cleaned via standard lab procedures; single-use sample carriers).

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- LensHooke X1 Pro Semen Quality Analyzer ([K180343](/device/K180343.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number

K252228

B Applicant

CheckCells, Inc.

C Proprietary and Established Names

Seaman PRO

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  POV | Class II | 21 CFR 864.5220 - Automated Differential Cell Counter | HE - Hematology  |

## II Submission/Device Overview:

A Purpose for Submission:

Clearance of new device.

B Measurand:

Sperm concentration (M/mL), total motility (%), progressive motility (%), morphology (%), pH, and Vitality (%)

C Type of Test:

Analysis of Semen Parameters

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K252228 - Page 2 of 13

## III Intended Use/Indications for Use:

### A Intended Use(s):
See Indications for Use below.

### B Indication(s) for Use:

The Seaman PRO Analyzer is an automated, point-of-care, in vitro use-only medical device for semen analysis performed by healthcare professionals. The desktop unit consists of a light source, optical elements, and an image sensor connected to a PC, which records a video of the semen sample and sends it to the cloud compute unit. The cloud compute unit runs software containing algorithms for assessing semen parameters. The product is a software-based medical device powered by ML that enables the user to further assess semen parameters, including concentration, motility, pH, morphology, and vitality. The product is intended for adult male patients to assess sperm quality. The product is not intended to process any of the PHI/PII. Seaman PRO for prescription use only.

Seaman PRO used with Imaging Hardware is an optical device for human semen analysis, which provides direct and calculated quantitative measurements for:

1. Sperm concentration (10^6 per ml)
2. Total motility (PR+NP, %)
3. Progressive motility (%)
4. pH value
5. Sperm morphology (normal forms, %)
6. Vitality (%)

The Seaman PRO device does not provide a comprehensive evaluation of a male's fertility status. Instead, it analyzes semen parameters that healthcare professionals use as part of a broader assessment of male fertility.

### C Special Conditions for Use Statement(s):
Rx Only

### D Special Instrument Requirements:
Seaman PRO System

## IV Device/System Characteristics:

### A Device Description:

The Seaman PRO System is an automated in vitro diagnostic device for human semen analysis consisting of imaging hardware and software components designed for use in the point-of-care setting. The devices are intended for use in adult male individuals and are not to be used for post-vasectomy confirmation or in the diagnosis or treatment of disease. The device is intended for

{2}

multiple uses and does not require patient-contact reprocessing. Only single-use sample carriers come into direct contact with the biological material. The reusable imaging hardware does not require sterilization and is cleaned using standard laboratory procedures. The Seaman PRO kit includes the following components for analysis: Imaging Hardware (for precise camera positioning, sample illumination and image capturing), Power cable (for light source), Camera cable (for image recognition), Cartridge for pH strip, and scale (informational purposes only).

## B Principle of Operation:

For both the Seaman and Seaman PRO Systems, the imaging hardware includes a desktop optical unit composed of a light source, fixed optical elements, and a digital image sensor. The unit captures video recordings of semen samples prepared on sperm counting microscopic slides. The acquired image data is transferred to a connected PC and processed in a secure cloud compute environment. The software component is a Software as a Medical Device (SaMD) that uses image analysis algorithms, including machine learning (ML) techniques, to quantify semen parameters, including:

- Sperm concentration (10⁶/mL),
- Total Motility (%),
- Progressive Motility (%),
- Sperm morphology (% normal forms),
- pH value,
- Vitality (%).

Analysis of semen parameters require minimal sample manipulation. However, Vitality (%) is determined using WHO laboratory manual for the examination and processing of human semen, 6ᵗʰ edition Eosin-Nigrosin Stain procedure. A smear is prepared on a glass slide from fresh liquefied semen samples mixed with equal volume of eosin-nigrosin staining solution. Once dry, the stained slides are immediately analyzed on the Seaman PRO System.

The device provides quantitative outputs based on video analysis of sperm. The software does not interpret results or diagnose fertility status. It is intended to support healthcare professionals in assessing semen quality. The device does not process, store, or transmit any Protected Health Information (PHI) or Personally Identifiable Information (PII).

## C Instrument Description Information:

1. Instrument Name:
Seaman PRO System

2. Specimen Identification:
Manual Data Entry

K252228 - Page 3 of 13

{3}

K252228 - Page 4 of 13

3. Specimen Sampling and Handling:

The semen specimen should be collected via masturbation directly into a sterile, wide-mouthed, non-toxic, and appropriately labeled container made of polypropylene or polystyrene, in accordance with WHO laboratory manual recommendations. No lubricants or spermicidal agents should be used. The entire ejaculate should be collected, ensuring the first portion (rich in spermatozoa) is not lost. After collection, the sample must rest undisturbed at room temperature for approximately 15-30 minutes to allow for natural liquefaction. Incomplete liquefaction may affect analysis accuracy. The sample should be analyzed within 60 minutes of ejaculation for optimal accuracy. Delays in analysis beyond 60 minutes may result in decreased motility and alterations in morphology or pH. Samples should not be refrigerated or frozen prior to analysis.

4. Calibration:

User calibration is not required.

5. Quality Control:

The quality control of the device should be performed regularly and at least once a month. The quality can be controlled on a hardware level and on a software level.

Hamilton Thorne Accu-beads: three concentration levels (low, medium, high) of silicone beads, similar in size to human sperm, suspended in an aqueous solution.

Fertility Solutions Semen Test Recordings: control material represented by a set of microscopic recordings of human semen with three levels (low, medium, high) of motility, morphology, and vitality.

V Substantial Equivalence Information:

A Predicate Device Name(s):

LensHooke X1 Pro Semen Quality Analyzer

B Predicate 510(k) Number(s):

K180343

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K252228 | K180343  |
| --- | --- | --- |
|  Device Trade Name | Seaman PRO System | LensHooke X1 Pro Semen Quality Analyzer  |

{4}

K252228 - Page 5 of 13
|  General Device Characteristic Similarities |  |   |
| --- | --- | --- |
|  Intended Use/Indications For Use | The Seaman PRO Analyzer is an automated, point-of-care, in vitro use-only medical device for semen analysis performed by healthcare professionals. The desktop unit consists of a light source, optical elements, and an image sensor connected to a PC, which records a video of the semen sample and sends it to the cloud compute unit. The cloud compute unit runs software containing algorithms for assessing semen parameters. The product is a software-based medical device powered by ML that enables the user to further assess semen parameters, including concentration, motility, pH, morphology, and vitality. The product is intended for adult male patients to assess sperm quality. The product is not intended to process any of the PHI/PII. Seaman PRO for prescription use only.

Seaman PRO used with Imaging Hardware is an optical device for human semen analysis, which provides direct and calculated quantitative measurements for: | The LensHooke X1 PRO Semen Quality Analyzer used with LensHooke Semen Test Cassette is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
(1) Sperm concentration (10^{6} per ml)
(2) Total motility (PR+NP, %)
• Progressive motility (%)
• Non-Progressive motility (%)
(3) Sperm morphology (normal forms, %)
(4) pH value.

The LensHooke X1 PRO Semen Quality Analyzer does not provide a comprehensive evaluation of a male’s fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in healthcare professional setting to evaluate male fertility.  |

{5}

K252228 - Page 6 of 13
|   | 1. Sperm concentration (10^6 per ml)
2. Total motility (PR+NP, %)
3. Progressive motility (%)
4. pH value
5. Sperm morphology (normal forms, %)
6. Vitality (%).

The Seaman PRO device does not provide a comprehensive evaluation of a male's fertility status. Instead, it analyzes semen parameters that healthcare professionals use as part of a broader assessment of male fertility. |   |
| --- | --- | --- |
|  Regulation (product code) | 21 CFR 864.5220 (POV) | Same  |
|  Sample Type | Human Semen | Same  |
|  Parameters | Sperm concentration, motility (total, progressive), morphology, pH | Same  |
|  Test Principles | Optical design and Image processing method, Image analysis | Same  |
|  Software Algorithm | Custom image analysis with AI | Similar  |
|  Test Locale | POC | Same  |
|  **General Device Characteristic Differences** |  |   |
|  Standard/ Guidance Document | WHO 6th edition guidelines | WHO 5th edition guidelines  |
|  Consumables | GoldCyto semen analysis slides | Semen Test Cassette (CS0, CS1)  |
|  Hardware Configuration | Camera + optics + PC with AI software (modular) | Integrated desktop analyzer  |
|  Different Parameters | % Vitality, Semen | Nonprogressive  |

{6}

|   | Volume, Total Sperm Count | Motility  |
| --- | --- | --- |
|  Testing requiring additional reagents / slide from end-user (not provided) | % Vitality procedure | None  |
|  Control Material | Hamilton Thorne Accu-beads | X QC Beads, X QC Reticle  |
|   |  Fertility Solutions Semen Test Recordings | LensHooke X QC Video (For Semen)  |

# VI Standards/Guidance Documents Referenced:

WHO Laboratory Manual for the Examination and Processing of Human Semen 6th Edition (2021)
CLSI EP05-A3 (2014): Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
CLSI EP06-A2 (2020): Evaluation of the Linearity of Quantitative Measurement Procedures; Approved Guideline - Second Edition
CLSI EP07 (2018): Interference Testing in Clinical Chemistry; Approved Guideline - Third Edition
CLSI EP25-A2 (2023): Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline - Second edition
CLSI EP09-c (2018): Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Third Edition
CLSI EP17-A2 (2012): Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second edition
CLSI EP21 (2016): Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures; Approved Guideline - second edition
CLSI EP15-A3 (2019): Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition

# VII Performance Characteristics (if/when applicable):

# A Analytical Performance:

K252228 - Page 7 of 13

{7}

# 1. Precision/Reproducibility:

# Repeatability

A precision study was conducted evaluating sperm concentration, motility, progressive motility, morphology, pH, and vitality using fourteen native semen samples covering low, medium, and high ranges for all parameters. Some samples were only utilized for certain parameters (i.e. pH and vitality) due to limited sample stability, but at least seven samples were used for each parameter evaluation. The analysis was carried out at one POC site over the course of one day by two POC operators using three lots, and three analyzers. Due to the limited stability of semen samples, each "day" in the statistical analysis represents different times of the one day. The study included two replicates per run with 2 runs per 20 "days" (all over one actual day) (20 "days" × two replicates × two runs = 80 data points per level). Data analyses were performed to provide within-run, between-run, between-day, within-laboratory, between-operator, between-instrument, and between-lot repeatability. All results met the predefined acceptance criteria.

|  Concentration [M/mL] |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot  |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  1 | 131.3 | 80 | 6.5 | 1.1 | 5.8 | 1.0 | 3.9 | 0.7 | 8.4 | 1.4 | 1.3 | 0.2 | 1.8 | 0.3 | 1.3 | 1.7  |
|  2 | 12.9 | 80 | 7.2 | 5.9 | 5.0 | 4.1 | 2.9 | 2.4 | 8.2 | 6.7 | 1.8 | 1.4 | 1.4 | 1.2 | 1.2 | 0.2  |
|  3 | 51.8 | 80 | 3.9 | 2.5 | 5.2 | 3.4 | 3.4 | 2.2 | 6.2 | 4.0 | 1.6 | 1.1 | 2.0 | 1.3 | 0.5 | 0.3  |
|  4 | 7.0 | 80 | 5.6 | 3.9 | 4.8 | 3.4 | 3.8 | 2.7 | 6.9 | 4.9 | 0.7 | 0.5 | 0.6 | 0.5 | 1.3 | 0.1  |
|  5 | 36.9 | 80 | 6.0 | 5.3 | 5.4 | 4.8 | 3.4 | 3.0 | 7.4 | 6.5 | 1.6 | 1.4 | 1.1 | 1.0 | 3.3 | 1.2  |
|  6 | 179.3 | 80 | 6.5 | 2.4 | 5.8 | 2.1 | 3.0 | 1.1 | 8.0 | 3.0 | 1.1 | 0.4 | 0.3 | 0.1 | 0.2 | 0.4  |
|  7 | 95.2 | 80 | 5.3 | 3.0 | 4.9 | 2.8 | 3.3 | 1.9 | 7.0 | 4.0 | 2.1 | 1.2 | 2.0 | 1.2 | 2.4 | 2.3  |
|  Total Motility [%] |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot  |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  1 | 53.9 | 80 | 6.1 | 3.3 | 5.2 | 2.8 | 3.5 | 1.9 | 7.5 | 4.1 | 0.9 | 0.5 | 0.9 | 0.5 | 1.4 | 0.8  |
|  2 | 30.5 | 80 | 6.4 | 2.0 | 5.1 | 1.5 | 3.9 | 1.2 | 7.6 | 2.3 | 0.1 | 0.0 | 1.4 | 0.4 | 0.6 | 0.2  |
|  3 | 73.1 | 80 | 5.9 | 4.3 | 5.8 | 4.3 | 4.3 | 3.1 | 7.7 | 5.7 | 1.4 | 1.0 | 3.2 | 2.4 | 1.7 | 1.3  |
|  4 | 61.1 | 80 | 5.6 | 3.4 | 6.1 | 3.7 | 4.7 | 2.8 | 7.7 | 4.7 | 0.5 | 0.3 | 1.0 | 0.6 | 1.7 | 1.1  |
|  5 | 24.1 | 80 | 6.6 | 1.6 | 6.6 | 1.6 | 5.7 | 1.4 | 8.6 | 2.1 | 0.9 | 0.2 | 1.7 | 0.4 | 2.8 | 0.7  |
|  6 | 84.5 | 80 | 4.9 | 4.1 | 4.3 | 3.6 | 2.8 | 2.4 | 6.3 | 5.3 | 0.6 | 0.5 | 0.2 | 0.2 | 2.5 | 2.1  |
|  7 | 47.4 | 80 | 5.9 | 2.8 | 5.2 | 2.5 | 3.1 | 1.5 | 7.5 | 3.5 | 1.6 | 0.7 | 1.8 | 0.9 | 1.2 | 0.6  |
|  Progressive Motility [%] |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot  |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  1 | 48.4 | 80 | 7.3 | 3.6 | 6.2 | 3.0 | 3.7 | 1.8 | 8.8 | 4.2 | 0.1 | 0.1 | 1.6 | 0.8 | 1.1 | 0.6  |
|  2 | 24.6 | 80 | 8.1 | 2.0 | 6.5 | 1.6 | 4.9 | 1.2 | 9.3 | 2.3 | 0.1 | 0.0 | 1.6 | 0.4 | 0.4 | 0.1  |
|  3 | 62.3 | 80 | 6.9 | 4.3 | 6.6 | 4.1 | 4.6 | 2.9 | 9.2 | 5.7 | 1.4 | 0.9 | 3.5 | 2.2 | 1.2 | 0.7  |
|  4 | 52.1 | 80 | 7.0 | 3.7 | 6.0 | 3.1 | 4.8 | 2.5 | 8.5 | 4.4 | 0.5 | 0.3 | 1.2 | 0.6 | 2.1 | 1.1  |
|  5 | 20.5 | 80 | 8.4 | 1.7 | 6.4 | 1.3 | 5.2 | 1.1 | 9.4 | 1.9 | 1.3 | 0.3 | 2.2 | 0.4 | 2.3 | 0.5  |
|  6 | 71.3 | 80 | 5.8 | 4.2 | 5.9 | 4.2 | 4.1 | 2.9 | 7.9 | 5.6 | 0.5 | 0.4 | 0.8 | 0.5 | 2.8 | 2.0  |
|  7 | 40.8 | 80 | 6.4 | 2.6 | 5.9 | 2.4 | 3.6 | 1.5 | 8.1 | 3.3 | 0.8 | 0.3 | 2.0 | 0.8 | 1.7 | 0.7  |

K252228 - Page 8 of 13

{8}

|  Morphology [%]  |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot  |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  1 | 2.7 | 80 | 6.4 | 0.2 | 6.4 | 0.2 | 3.6 | 0.1 | 8.6 | 0.2 | 3.5 | 0.09 | 1.0 | 0.03 | 1.3 | 0.03  |
|  2 | 4.3 | 80 | 6.6 | 0.3 | 5.1 | 0.2 | 3.8 | 0.2 | 7.6 | 0.3 | 0.0 | 0.00 | 1.4 | 0.06 | 2.2 | 0.09  |
|  3 | 8.6 | 80 | 4.9 | 0.4 | 4.5 | 0.4 | 2.7 | 0.2 | 6.3 | 0.5 | 0.6 | 0.05 | 1.7 | 0.14 | 1.2 | 0.11  |
|  4 | 2.5 | 80 | 7.3 | 0.2 | 5.3 | 0.1 | 4.2 | 0.1 | 8.1 | 0.2 | 1.1 | 0.03 | 2.1 | 0.05 | 1.2 | 0.03  |
|  5 | 14.2 | 80 | 4.2 | 0.6 | 4.2 | 0.6 | 2.9 | 0.4 | 5.6 | 0.8 | 1.1 | 0.15 | 1.1 | 0.15 | 0.6 | 0.09  |
|  6 | 3.5 | 80 | 8.2 | 0.3 | 4.9 | 0.2 | 3.2 | 0.1 | 8.2 | 0.3 | 1.6 | 0.06 | 0.7 | 0.03 | 1.5 | 0.05  |
|  7 | 4.6 | 80 | 6.4 | 0.3 | 5.6 | 0.3 | 4.3 | 0.2 | 7.9 | 0.4 | 0.0 | 0.00 | 0.7 | 0.03 | 2.3 | 0.11  |
|  Vitality [%]  |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot  |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  8 | 17.2 | 80 | 6.5 | 1.1 | 5.8 | 1.0 | 3.9 | 0.7 | 8.4 | 1.4 | 1.3 | 0.2 | 1.8 | 0.3 | 0.6 | 0.1  |
|  9 | 81.8 | 80 | 7.2 | 5.9 | 5.0 | 4.1 | 2.9 | 2.4 | 8.2 | 6.7 | 1.8 | 1.4 | 1.4 | 1.2 | 0.5 | 0.4  |
|  10 | 65.6 | 80 | 3.9 | 2.5 | 5.2 | 3.4 | 3.4 | 2.2 | 6.2 | 4.0 | 1.6 | 1.1 | 2.0 | 1.3 | 1.8 | 1.2  |
|  11 | 70.9 | 80 | 5.6 | 3.9 | 4.8 | 3.4 | 3.8 | 2.7 | 6.9 | 4.9 | 0.7 | 0.5 | 0.6 | 0.5 | 0.7 | 0.5  |
|  12 | 88.7 | 80 | 6.0 | 5.3 | 5.4 | 4.8 | 3.4 | 3.0 | 7.4 | 6.5 | 1.6 | 1.4 | 1.1 | 1.0 | 0.4 | 0.4  |
|  13 | 37.0 | 80 | 6.5 | 2.4 | 5.8 | 2.1 | 3.0 | 1.1 | 8.0 | 3.0 | 1.1 | 0.4 | 0.3 | 0.1 | 1.0 | 0.4  |
|  14 | 57.4 | 80 | 5.3 | 3.0 | 4.9 | 2.8 | 3.3 | 1.9 | 7.0 | 4.0 | 2.1 | 1.2 | 2.0 | 1.2 | 0.9 | 0.5  |
|  pH |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  ID | Mean Value | n | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory |   | Between Operator |   | Between Instrument |   | Between Lot |   |
|   |   |   |  CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD  |
|  8 | 7.8 | 80 | 7.6 | 0.6 | 5.8 | 0.5 | 3.6 | 0.3 | 8.9 | 0.7 | 0.4 | 0.0 | 1.4 | 0.1 | 1.2 | 0.1  |
|  9 | 8.4 | 80 | 4.8 | 0.4 | 4.2 | 0.4 | 3.4 | 0.3 | 6.2 | 0.5 | 1.4 | 0.1 | 0.6 | 0.1 | 1.7 | 0.1  |
|  10 | 7.8 | 80 | 5.1 | 0.4 | 4.3 | 0.3 | 2.8 | 0.2 | 6.4 | 0.5 | 0.7 | 0.1 | 1.2 | 0.1 | 0.8 | 0.1  |
|  11 | 6.8 | 80 | 5.6 | 0.4 | 6.3 | 0.4 | 5.1 | 0.3 | 7.9 | 0.5 | 0.9 | 0.1 | 0.9 | 0.1 | 1.1 | 0.1  |
|  12 | 8.3 | 80 | 4.5 | 0.4 | 4.9 | 0.4 | 3.6 | 0.3 | 6.3 | 0.5 | 1.2 | 0.1 | 1.1 | 0.1 | 1.4 | 0.1  |
|  13 | 7.4 | 80 | 5.9 | 0.4 | 5.6 | 0.4 | 4.1 | 0.3 | 7.7 | 0.6 | 1.8 | 0.1 | 1.7 | 0.1 | 2.3 | 0.2  |
|  14 | 8.6 | 80 | 8.0 | 0.7 | 4.7 | 0.4 | 3.6 | 0.3 | 8.2 | 0.7 | 0.5 | 0.0 | 1.7 | 0.1 | 0.3 | 0.0  |

## Reproducibility

To evaluate the reproducibility performance of the Seaman PRO System, a precision study was conducted at three sites (two US, one OUS) with six POC operators. For sperm concentration and vitality, three levels of QC control solution or slides (low negative, low positive, high positive) were evaluated. For the pH evaluation, three potassium/sodium phosphate buffer levels were utilized. The studies included three sites × five replicates × five days, totaling 75 data points per level. Data analyses were performed for each site and overall to provide within-day repeatability, between-day, between-Operator, between-site, and total reproducibility. All results met the predefined acceptance criteria.

K252228 - Page 9 of 13

{9}

|  Reproducibility  |   |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Parameter [units] | Level | N | Mean | Within-Day |   | Between-Day |   | Between Site |   | Between Operator |   | Reproducibility  |   |
|   |   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  Conc [M/mL] | 3 | 75 | 3.2 | 0.2 | 6.4 | 0.1 | 4.5 | 0.1 | 2.1 | 0.1 | 2.4 | 0.3 | 7.9  |
|   |  18 | 75 | 17.8 | 1.1 | 6.0 | 0.5 | 3.0 | 0.2 | 1.3 | 0.2 | 1.4 | 1.1 | 6.4  |
|   |  35 | 75 | 36.1 | 1.7 | 4.6 | 0.8 | 2.3 | 0.5 | 1.5 | 0.7 | 1.8 | 1.8 | 4.9  |
|  pH | 7.0 | 75 | 7.1 | 0.6 | 8.3 | 0.3 | 4.0 | 0.1 | 1.4 | 0.2 | 2.6 | 0.6 | 8.8  |
|   |  8.0 | 75 | 8.0 | 0.5 | 6.1 | 0.2 | 2.7 | 0.1 | 1.5 | 0.1 | 1.6 | 0.5 | 6.7  |
|   |  9.0 | 75 | 8.9 | 0.5 | 5.5 | 0.3 | 3.9 | 0.1 | 1.3 | 0.2 | 1.9 | 0.6 | 6.4  |
|  Vitality [%] | 16 | 75 | 16.3 | 1.5 | 9.2 | 0.7 | 4.5 | 0.3 | 1.9 | 0.6 | 3.4 | 1.6 | 9.8  |
|   |  70 | 75 | 71.6 | 4.3 | 6.0 | 1.9 | 2.7 | 0.7 | 1.0 | 1.1 | 1.5 | 4.6 | 6.4  |
|   |  81 | 75 | 79.3 | 5.4 | 6.8 | 2.6 | 3.3 | 0.3 | 0.4 | 1.1 | 1.4 | 5.7 | 7.1  |

2. Linearity:

Linearity for sperm concentration, pH, and vitality were evaluated using one analyzer to measure semen samples prepared at nine levels. Sperm concentrations, prepared via concentration or dilution of fresh semen samples, ranged from 2.1–387.1 M/mL (n=3), verified using manual microscope. Sperm pH ranged from 6.0–9.0 (n=3), verified using manual pH strip. The different levels of pH were created by mixing different levels of semen with phosphate buffer (pH 5.0 and pH 9.0). For Vitality, commercial Sperm Viability Quality Control stained slides, with known vitality levels were utilized. For all parameters, the mean and SD of results were calculated; and Weighted Least Squares Regression analysis was used to verify the linear range. Sperm concentration was demonstrated to be linear from 2.1–387.1 M/mL, Sperm pH was demonstrated to be linear from 6.0–9.0, and sperm vitality was demonstrated to be linear from 5–81 %.

3. Analytical Specificity/Interference:

The potential interference of various substances on Seaman PRO results were evaluated by using two sperm concentration levels (&lt; 50 M/mL and &gt; 50 M/mL). The following 11 interfering substances were tested in the study: vitamin B, testosterone, yeast, E. Coli, RBC, WBC, urine, saliva, agglutination of semen sample, D-norgestrel, and β-estradiol. Samples were tested in five replicates on one analyzer using one lot of test slides. Results of test group were compared to the control group. Study results showed that all tested interfering substances met the acceptance criteria, and no significant interference was caused by the tested substances.

4. Detection Limit and Assay Reportable Range:

Detection Limit:

A study was conducted to determine the detection limits of the Seaman PRO System. Semen samples were taken from volunteers. For the limit of blank (LoB), semen was centrifuged to obtain sperm-free seminal plasma to a concentration of ~0 M/mL (blank sample) as verified by manual microscope. For assessing limit of detection (LoD), samples were prepared by diluting semen with seminal plasma to four samples with concentration of ~2-8 M/mL. Blank and low-level samples were divided into four aliquots and tested in five replicates once a day for three days, using one analyzer. For assessing LoQ, samples were prepared by diluting semen with seminal plasma to four samples with concentration of approximately ~2 M/mL.

K252228 - Page 10 of 13

{10}

The four LoQ low-level samples were tested in three replicates once a day for three days, using one analyzer and two lots of sperm counting slides.

Results were calculated, and the detection limits were determined to be:

Limit of Blank (LoB) = 0.0 M/mL
Limit of Detection (LoD) = 1.9 M/mL
Limit of Quantitation (LoQ) = 2.1 M/mL.

**Assay Reportable Range:**
Sperm concentration: 2.1–380 M/mL
Vitality: 1–100%
pH: 6.0–9.0

The assay reportable range was determined by the linearity and detection limit studies and verified in the method comparison study.

5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**

**Device Stability**

A stability study was conducted to determine the shelf-life of the Seaman PRO System at room temperature. Three semen samples with varying sperm concentrations (low/near cutoff, medium, high) were tested in 80 replicates using three analyzers before and after storage. The evaluation met the acceptance criteria when comparing results from the stressed devices to control analyzers. Study results demonstrated analyzer performance for three years at room temperature conditions.

**Sample Stability**

Stability of semen samples was determined with 42 samples at different concentration levels (half low and half high concentration samples). The semen samples were stored at room temperature. At time zero (0 hour), the samples were tested to establish the baseline. These samples were tested again at various timepoints and for each time point, the results were compared to the baseline results. The results support a 1-hour semen sample stability duration.

6. **Assay Cut-Off:**

See Reference Range study, section VII.D.

**B Comparison Studies:**

1. **Method Comparison with Predicate Device:**

A method comparison study was conducted to evaluate the performance of the Seaman PRO System when used by intended users (i.e. healthcare professionals). The study was performed

K252228 - Page 11 of 13

{11}

at three sites (two US; one OUS site), with a total of three POC operators for semen analysis. Following collection of the specimen, the operators analyzed the sample on the Seaman PRO System. After recording the results, a different operator at the study site performed semen analysis on the sample using either the predicate (i.e., the LensHooke X1 PRO Semen Quality Analyzer for concentration, motility, and morphology) or the reference method (i.e., manual microscope for % vitality analysis as recommended in the WHO 6th edition; and manual pH strip reading for pH analysis). A total of 300 semen specimens spanning the analytical measuring range were analyzed by using the Passing-Bablok regression. All results met the predefined acceptance criteria.

|  Seaman PRO vs. LensHooke X1 Pro  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Parameter [Units] | N | Result Range | Slope (95% CI) | Intercept (95% CI) | R² (95% CI)  |
|  Concentration [10^6/mL] | 300 | 3.1 - 391.3 | 1.04 (1.00, 1.07) | 1.52 (-1.11, 4.15) | 0.92 (0.91, 0.94)  |
|  Total Motility [%] | 300 | 0 - 84 | 1.02 (0.99, 1.05) | -0.01 (-0.02, 0.00) | 0.93 (0.92, 0.95)  |
|  Progressive Motility [%] | 300 | 0 - 84 | 0.95 (0.92, 0.99) | 0.02 (0.01, 0.04) | 0.91 (0.89, 0.93)  |
|  Morphology [%] | 300 | 2.7 - 7.2 | 1.10 (1.06, 1.14) | 0.00 (-0.00, 0.00) | 0.90 (0.88, 0.92)  |
|  pH | 300 | 6.0 - 10.0 | 0.99 (0.96, 1.03) | 0.05 (-0.25, 0.34) | 0.90 (0.88, 0.92)  |
|  Vitality [%] | 300 | 14 - 100 | 1.01 (0.97, 1.05) | -0.01 (-0.04, 0.02) | 0.90 (0.87, 0.92)  |

2. Matrix Comparison:

Not applicable.

C Clinical Studies:

1. Clinical Sensitivity:

Not applicable.

2. Clinical Specificity:

Not applicable.

3. Clinical Cut-Off

Not applicable.

4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable.

K252228 - Page 12 of 13

{12}

K252228 - Page 13 of 13

# D Expected Values/Reference Range:

The lower limit of the reference range for parameters Sperm Concentration (M/mL), Total Motility (%), Progressive Motility (%), Morphology (%), and Vitality (%) were directly reported as the 5th percentile in the WHO Laboratory Manual for the Examination and Processing of Human Semen, 6th Edition (Table 8.3). No reference range has been established for pH.

|  Semen Parameter | Reference Range | Source  |
| --- | --- | --- |
|  Sperm Concentration (M/mL) | ≥ 16 | WHO 6th Edition  |
|  Total Motility (%) | ≥ 42  |   |
|  Progressive Motility (%) | ≥ 30  |   |
|  Morphology (%) | ≥ 4  |   |
|  Vitality (%) | ≥ 54  |   |

# E Other Supportive Instrument Performance Characteristics Data:

Electrical safety and electromagnetic compatibility (EMC) testing were performed, and the system was found to be acceptable.

Software and cybersecurity documentation was reviewed and found to be acceptable.

# VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

# IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K252228](https://fda-staging.innolitics.com/device/K252228)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
