BRUSIO Trauma Toolbox System; TAMINA 3.5mm Proximal Humerus System
K252191 · Bonebridge AG · HRS · Dec 17, 2025 · Orthopedic
Device Facts
Record ID
K252191
Device Name
BRUSIO Trauma Toolbox System; TAMINA 3.5mm Proximal Humerus System
Applicant
Bonebridge AG
Product Code
HRS · Orthopedic
Decision Date
Dec 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BRUSIO Trauma Toolbox System is indicated for internal fixation of fractures, osteotomies and nonunions of bones and bone fragments of the appendicular skeleton that are appropriate for the implant size in adult patients. When used in the clavicle, scapula, humerus, olecranon, pelvis, femur, tibia or fibula, the BRUSIO Trauma Toolbox System is intended only for non-load-bearing stabilization and reduction. The TAMINA, TAMINA Long and TAMINA TF Proximal Humerus System is indicated for • Dislocated two-, three-, and four fragment fractures of the proximal humerus including fractures involving osteopenic bone • Pseudarthroses (non-unions) of the proximal humerus • Osteotomies of the proximal humerus
Device Story
Bone fixation system comprising metallic plates, locking/non-locking screws, and surgical instruments; used for internal fixation of appendicular skeleton fractures, osteotomies, and nonunions. Plates feature titanium inlay clips; screws and plates manufactured from stainless steel. System variants include BRUSIO Trauma Toolbox and TAMINA 3.5mm Proximal Humerus System. Operated by surgeons in clinical settings; provides mechanical stabilization of bone fragments. Implants are MR Conditional. Benefits include fracture stabilization and support for bone healing.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance validated via static and dynamic testing (four-point bending, construct stiffness, fatigue testing) compared to predicate devices. Biocompatibility assessed per ISO 10993. Sterilization and packaging validated per ISO 11137-2, ISO 17664, ISO 17665-1, and ISO 11607.
Technological Characteristics
Plates: Stainless steel (ISO 5832-1, ASTM F138/F139) with pure titanium inlay clips (ASTM F67/ISO 5832-2). Screws/Instruments: Stainless steel (ASTM F899, ISO 7153-1, ISO 5832-1, ASTM F138/139), PEEK, EPDM, silicone. Mechanical fixation via locking/non-locking screws. Sterilization: Gamma irradiation (plates) or steam sterilization (screws/instruments). MR Conditional.
Indications for Use
Indicated for adult patients requiring internal fixation of fractures, osteotomies, and nonunions of appendicular skeleton bones. Specific indications for TAMINA system include dislocated 2-, 3-, and 4-fragment proximal humerus fractures (including osteopenic bone), pseudarthroses, and osteotomies of the proximal humerus. Clavicle, scapula, humerus, olecranon, pelvis, femur, tibia, or fibula use restricted to non-load-bearing stabilization and reduction.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Bonebridge Osteosynthesis Plating System (K250933)
Synthes VOLT Mini Fragment Plating System (K233665)
Synthes (USA) Modular Mini Fragment LCP System (K063049)
Synthes (USA) LCP® Proximal Humerus Plates, Long (K041860)
Synthes (USA) 3.5 mm Locking Compression Plate (LCP) System with Expanded Indications (K082807)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 17, 2025
Bonebridge AG
Michelle Gumpelmayer
Head of QM&RA
Bahnhofstrasse 11
Zug, Zug 6300
Switzerland
Re: K252191
Trade/Device Name: BRUSIO Trauma Toolbox System; TAMINA 3.5mm Proximal Humerus System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS
Dated: November 18, 2025
Received: November 18, 2025
Dear Michelle Gumpelmayer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252191 - Michelle Gumpelmayer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252191 - Michelle Gumpelmayer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
BRUSIO Trauma Toolbox System
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252191 | ? |
| Please provide the device trade name(s). | | ? |
| BRUSIO Trauma Toolbox System;
TAMINA 3.5mm Proximal Humerus System | | |
| Please provide your Indications for Use below. | | ? |
| The BRUSIO Trauma Toolbox System is indicated for internal fixation of fractures, osteotomies and
nonunions of bones and bone fragments of the appendicular skeleton that are appropriate for the implant
size in adult patients. When used in the clavicle, scapula, humerus, olecranon, pelvis, femur, tibia or fibula,
the BRUSIO Trauma Toolbox System is intended only for non-load-bearing stabilization and reduction.
The TAMINA, TAMINA Long and TAMINA TF Proximal Humerus System is indicated for
• Dislocated two-, three-, and four fragment fractures of the proximal humerus including fractures involving
osteopenic bone
• Pseudarthroses (non-unions) of the proximal humerus
• Osteotomies of the proximal humerus | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K252191 - Page 1 of 5
K252191
510(k) Summary
In accordance with 21 CFR 807.92 the following information is provided for the Bonebridge Osteosynthesis Plating System.
ADMINISTRATIVE INFORMATION
Date prepared July 14, 2025
Submitter Bonebridge AG
Bahnhofstrasse 11
6300 Zug
Switzerland
Official Contact Michelle Gumpelmayer
Head of QM&RA, Bonebridge AG
Phone: +41 76 731 07 32
Email michelle.gumpelmayer@bonebridge.ch
Alternative Contact Alexander Häusler
COO, Bonebridge AG
Phone: +41 79 660 81 42
Email alexander.haeusler@bonebridge.ch
DEVICE NAME AND CLASSIFICATION
Trade name: BRUSIO Trauma Toolbox System
TAMINA 3.5mm Proximal Humerus System
Common name: Plate, Fixation, Bone
Regulation number: 21 CFR 888.3030
Classification name: Single/multiple component metallic bone fixation appliances and accessories
Regulatory class: Class II
Product Code: HRS
Page 1 of 5
{5}
K252191 - Page 2 of 5
# PREDICATE DEVICES
Primary predicate device: Bonebridge Osteosynthesis Plating System K250933 TAMINA 3.5mm Proximal Humerus System
| Subject Device: | Additional predicate devices: |
| --- | --- |
| BRUSIO Trauma Toolbox System | K233665 Synthes VOLT Mini Fragment Plating System |
| | K063049 Synthes (USA) Modular Mini Fragment LCP System |
| TAMINA 3.5mm Proximal Humerus System | K041860 Synthes (USA) LCP® Proximal Humerus Plates, Long |
| | K011815 Synthes LCP Proximal Humerus Plates |
| | K082807 Synthes (USA) 3.5 mm Locking Compression Plate (LCP) System with Expanded Indications |
Page 2 of 5
{6}
K252191 - Page 3 of 5
# INDICATIONS FOR USE
The BRUSIO Trauma Toolbox System is indicated for internal fixation of fractures, osteotomies and nonunions of bones and bone fragments of the appendicular skeleton that are appropriate for the implant size in adult patients. When used in the clavicle, scapula, humerus, olecranon, pelvis, femur, tibia or fibula, the BRUSIO Trauma Toolbox System is intended only for non-load-bearing stabilization and reduction.
The TAMINA, TAMINA Long and TAMINA-TF Proximal Humerus System is indicated for:
- Dislocated two-, three-, and four-fragment fractures of the proximal humerus including fractures involving osteopenic bone
- Pseudarthroses (non-unions) of the proximal humerus
- Osteotomies of the proximal humerus
# DEVICE DESCRIPTION
The Bonebridge Osteosynthesis Plating System is intended for treating fractures of various bones. It consists of plates, locking and non-locking screws for fixation and corresponding instruments. The plating system is further subdivided into variants/types based on the anatomical location of the fracture. The subject devices BRUSIO Trauma Toolbox System and TAMINA 3.5mm Proximal Humerus System are variants of the Bonebridge Osteosynthesis Plating System.
The plates are primarily manufactured from stainless steel (ISO 5832-1, ASTM F138, or ASTM F139) and include Titanium Inlay Clips (TICs) made from pure titanium (ASTM F67 or ISO 5832-2). The screws are manufactured entirely from stainless steel (ISO 5832-1, ASTM F138, or ASTM F139).
All materials used are biocompatible, corrosion-resistant and nontoxic in a biological environment. Surgical instruments are made of stainless steel (ASTM F899, ISO 7153-1, ISO 5832-1, and ASTM F138/139), medical grade PEEK, medical grade EPDM terpolymer, or medical grade silicone. All plates are sterilized with gamma irradiation and delivered sterile. Screws and instruments are delivered non-sterile. Devices supplied in a non-sterile condition must be cleaned and steam sterilized prior to surgical use. Non-clinical testing has demonstrated the implants are MR Conditional.
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICE
The subject devices and the predicate devices have the same intended use, similar indications for use, and technological characteristics. The subject and predicate devices are all fabricated from the same materials and share the same design characteristics, including plate screw holes to accommodate locking and non-locking screws. The subject and predicate devices encompass the same range of physical dimensions, and the subject device is compatible with screws from the primary predicate device. The subject and predicate devices are sterilized by the standard methods.
{7}
K252191 - Page 4 of 5
# SUMMARY OF NON-CLINICAL TESTS
Sterilization validation:
Gamma irradiation: The minimal dose of 25kGy is validated using VDmax25 method as described in ISO 11137-2 and confirmed a Sterility Assurance Level SAL of 10⁻⁶.
Steam sterilization: Cleaning and sterilization procedures have been successfully validated in accordance with ISO 17664 and ISO 17665-1 at 132°C (270F) for 4 minutes and 20 min drying time.
Packaging validation:
Validation of the sterile packaging has been successfully performed in accordance with ISO 11607 1/2 and ASTM F1980. Furthermore, a transport simulation was conducted according ISTA 2A followed by these packaging verification tests:
- Dye-Penetration, ASTM F1929
- Visual inspection, ASTM F1886/1886M
- Seal strength, ASTM F88/F88M
- Microbial barrier testing, DIN 58953-6, Chapter 2.14
Biocompatibility:
Biological Evaluation and toxicological risk assessment to evaluate the device's biological safety for the intended use, in accordance with ISO 10993-series and FDA guidance.
Mechanical testing:
Static and/or dynamic comparative testing was performed for each subject plate variant, including statistical analysis and comparison to the respective predicate devices. Substantial equivalence of each Bonebridge subject plate variant to the corresponding predicate devices was demonstrated with respect to maximum force, yield strength and construct stiffness (static test) and maximum force for a given number of cycles (dynamic test) or with respect to static four-point bending strength and bending structural stiffness. All predefined acceptance criteria were successfully met, demonstrating that the mechanical performance of each subject plate variant is substantially equivalent to the corresponding predicate devices.
MRI safety:
The Bonebridge Osteosynthesis Plating System is MR conditional considering local SAR based on the following tests
- Assessment of displacement force and torque effects in the main static magnetic field at 3Tesla. Additionally, the expected magnetic force in a stronger magnetic field gradient of 30T/m was extrapolated. (According to ASTM F2052-21 and ASTM F2213-17)
- Assessment of heating effects due to the RF-field during MR scans at 1.5Tesla and 3Tesla according to ASTM F2182-19
- Assessment of image artifacts at 3Tesla according to ASTM F2119-24
The tested implant and associated product family can be claimed as MR conditional.
Page 4 of 5
{8}
K252191 - Page 5 of 5
# CONCLUSION
The subject TAMINA 3.5mm Proximal Humerus System and BRUSIO Trauma Toolbox System, have the same intended use and similar indications for use, technological characteristics, and materials as their respective predicate devices. Non-clinical performance testing demonstrated that the subject devices are substantially equivalent to the predicate devices.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.