SOL-M Luer Lock Syringe with Exchangeable Needle Tray
K252184 · Sol-Millennium Medical, Inc. · FMF · Aug 25, 2026 · General Hospital
Device Facts
Record ID
K252184
Device Name
SOL-M Luer Lock Syringe with Exchangeable Needle Tray
Applicant
Sol-Millennium Medical, Inc.
Product Code
FMF · General Hospital
Decision Date
Aug 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Luer Lock Syringe with Exchangeable Needle Tray is used to inject fluids into, or withdraw fluids from, the body.
Device Story
Sterile, single-use 3-piece piston syringe with detachable pre-attached needle; used for manual injection or withdrawal of fluids. System comprises syringe barrel, plunger, Isoprene rubber piston, and exchangeable needle; components made of polypropylene. Available in 3, 5, and 10mL volumes with 20-23G needle options. Operated manually by medical professionals in clinical settings. Packaged in bulk thermal-formed plastic trays with Tyvek sterile-barrier material. Output is the delivery or aspiration of fluids; device facilitates medication administration or fluid sampling. Benefits include sterile, single-use delivery mechanism for clinical procedures.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-1), package integrity (visual, seal strength, dye penetration), sterility (USP <71>), endotoxin (EP 10.2.6.14), and simulated transportation (ISTA-3A:2018).
Technological Characteristics
Materials: Polypropylene (barrel/plunger/hub/cap), Isoprene rubber (piston), SUS 304 (needle). Manual operation. Luer lock connector (ISO 80369-7). Sterilization: EtO (SAL 10^-6). Packaging: Bulk plastic tray with Tyvek sterile barrier. Standards: ISO 7864, ISO 9626, ISO 11607-1, ASTM F88/F88-15, ASTM F1929-15.
Indications for Use
Indicated for medical professionals for the injection of fluids into, or withdrawal of fluids from, the body. Prescription use only.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Luer Lock Syringe with Exchangeable Needle (K241821)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 25, 2026
Sol-Millennium Medical, Inc.
Mike Xie
VP of QA/RA
311 S Wacker Dr.
Suite 4100
Chicago, Illinois 60606
Re: K252184
Trade/Device Name: SOL-M Luer Lock Syringe with Exchangeable Needle Tray
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF, FMI
Dated: July 20, 2026
Received: August 3, 2026
Dear Mike Xie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252184 - Mike Xie
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252184 - Mike Xie
Page 3
Sincerely,
**SHRUTI N. MISTRY -S**
Shruti Mistry, M.S.
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252184 | ? |
| Please provide the device trade name(s). | | ? |
| SOL-M Luer Lock Syringe with Exchangeable Needle Tray | | |
| Please provide your Indications for Use below. | | ? |
| The Luer Lock Syringe with Exchangeable Needle Tray is used to inject fluids into, or withdraw fluids from, the body. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
**SolM**Building^{}[] ^{}[] a Healthier^{}[] ^{}[] Tomorrow
# 510(k) Summary
### 1. Preparation date: August 25, 2026
### 2. Submitter
Manufacturer: Sol-Millennium Medical Inc.
Address: 311 S. Wacker Dr, Suite 4100 Chicago, IL 60606, United States
Contact person: Mike Xie, VP of QA/RA, +1 847-400-4821, yxie@solm.com
### 3. Device
Trading name: SOL-M Luer Lock Syringe with Exchangeable Needle Tray
Regulation No.: 21 CFR § 880.5860; 21 CFR § 880.5570
Classification name: Syringe, Piston; Needle, Hypodermic, Single Lumen
Classification: Class II
Product code: FMF; FMI
Review panel: General Hospital
### 4. Predicate device
510(k) Number: K241821
Trading name: Luer Lock Syringe with Exchangeable Needle
Regulation No.: 21 CFR § 880.5860; 21CFR 880.5570
Classification name: Syringe, Piston, Antistick, Needle, Hypodermic, Single Lumen
Classification: Class II
Product code: FMF; MEG; FMI
Review panel: General Hospital
### 5. Device description
This sterile, single-use 3-piece piston syringe features a detachable pre-attached needle for manual use. The system includes multiple syringes with exchangeable needles in a bulk tray. Each unit consists of a needle cap, tube, hub, barrel, plunger, and Isoprene rubber piston, with polypropylene barrel and plunger components. Available in 3, 5, and 10mL volumes with 20-23G needle options and Luer lock connector. The device is packaged in a thermal-formed plastic tray protected with a Tyvek sterile-barrier material, sterilized with EO gas, and offers a 5-year shelf life.
The device is for medical professionals use only and for prescription use only.
1
{5}
SolM Building a Healthier Tomorrow
Table 1. Device specifications
| 3ml Luer Lock Syringe w/Exch Needle Tray 21G*1" |
| --- |
| 3ml Luer Lock Syringe w/Exch Needle Tray 22G*1" |
| 3ml Luer Lock Syringe w/Exch Needle Tray 23G*1" |
| 5ml Luer Lock Syringe w/Exch Needle Tray 20G*1" |
| 5ml Luer Lock Syringe w/Exch Needle Tray 21G*1" |
| 10ml Luer Lock Syringe w/Exch Needle Tray 20G*1" |
| 10ml Luer Lock Syringe w/Exch Needle Tray 21G*1" |
## 6. Indications for use/Intended use
The Luer Lock Syringe with Exchangeable Needle Tray is used to inject fluids into, or withdraw fluids from, the body.
## 7. Comparison of technological characters between proposed and predicate devices
Table 2. Characters comparison
| Element of Comparison | Predicate Device Luer Lock Syringe with exchangeable needle (K241821) | Subject Device SOL-M Luer Lock Syringe with Exchangeable Needle Tray |
| --- | --- | --- |
| **Product Code** | FMF, FMI | FMF, FMI |
| **Intended Use/ Indications for Use** | The Syringe with exchangeable needle is used to inject medicines and vaccines into, or withdraw fluids from, the body. | The Luer Lock Syringe with Exchangeable Needle Tray is used to inject fluids into, or withdraw fluids from, the body. |
2
{6}
**SolM**Building a Healthier Tomorrow
| Element of Comparison | Predicate Device Luer Lock Syringe with exchangeable needle (K241821) | Subject Device SOL-M Luer Lock Syringe with Exchangeable Needle Tray |
| --- | --- | --- |
| | | Same, Different phrases that convey the same meaning. |
| **Principle of Operation** | Manual | Same |
| **Source of Energy** | Manual | Same |
| **Needle/Syringe Connection** | ISO 80369-7 | Same |
| **Needle physical performance** | ISO 7864 ISO 9626 | Same |
| **Device Specifications** | Syringe volume: 1ml, 3ml, 5ml, 10ml Needle gauge: 20G, 21G, 22G, 23G, 25G Needle length: 1', 5/8', 11/2' | Syringe volume: 3ml, 5ml and 10ml Needle gauge: 20G, 21G, 22G, 23G Needle length: 1' Same, within scope. |
| **Device Configuration** | Syringe - Barrel: PP - Plunger: PP - Gasket: IR rubber - Lubricant: silicone oil | Same |
3
{7}
**SolM**Building^{}[] ^{}[] a Healthier^{}[] ^{}[] Tomorrow
| Element of Comparison | Predicate Device Luer Lock Syringe with exchangeable needle (K241821) | Subject Device SOL-M Luer Lock Syringe with Exchangeable Needle Tray |
| --- | --- | --- |
| | Needle - Needle hub: PP - Needle cap: PP - Needle: SUS 304 - Adhesive: Glue | |
| **Packaging Method** | Individual packaging, Tyvek pouch like | Bulk packaging, plastic tray sealed with a Tyvek sterile-barrier material. Different, new packaging types were applied, packaging integrity and sterility testing results verified its effectiveness. So, difference didn't raise any risk, demonstrating the substantial equivalence. |
| **Sterilization** | EtO SAL 10^{-6} | Same |
| **Biocompatibility** | Meet ISO 10993 | Same |
## 8. Non-clinical testing
Non-clinical tests, and sterilization validation were conducted under K241821 with no changes except to the packaging type.
Biocompatibility testing was re-conducted according to the guidance of Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
4
{8}
SolM Building a Healthier Tomorrow
Package integrity testing (including visual inspection, Seal strength, Dye penetration, and Clean peel) had been conducted on simulated transportation treated products. In addition, package testing had been conducted after aging.
Tyvek is used as the sterile barrier material for the package, and the package integrity was tested by the following methods:
- Visual Check: Internal standard
- Clean peel: ISO 11607-1:2019
- Seal strength: ASTM F88/F88-15
- Dye penetration: ASTM F1929-15
- Sterility: USP <71>
- Endotoxin: European Pharmacopoeia 10.2.6.14.
- Simulated transportation: ISTA-3A:2018
Additionally, Bacterial Aerosol Challenge Testing was conducted to validate the suitability of the syringe inside the tray to maintain a sterile barrier after tray opening.
## 9. Clinical testing
Not applicable for this submission.
## 10. Conclusion
The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.