Sagami Original 002 Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
Device Story
Sagami Original 002 is a non-sterile, single-use, polyurethane male condom; features straight-walled, reservoir-ended design with smooth surface; lubricated with silicone oil (dimethylpolysiloxane). Device acts as a physical barrier to prevent exchange of bodily fluids during sexual activity. Used in home or private environments by individuals for contraception and STI risk reduction. Device is a passive barrier; no electronic or algorithmic components. Clinical benefit derived from mechanical integrity and barrier properties preventing pregnancy and STI transmission. Performance verified via biocompatibility, physical testing (air burst, tensile strength), and clinical in-use studies assessing slippage and breakage rates.
Clinical Evidence
Clinical in-use study (NCT04622306) conducted per ISO 29943-1:2017 and FDA guidance. Evaluated slippage and breakage rates of the subject device against a natural rubber latex condom comparator. Results demonstrated the subject device met non-inferiority requirements specified in ISO 23409:2011, showing low failure rates comparable to the control.
Technological Characteristics
Material: Polyurethane. Lubricant: Silicone oil (dimethylpolysiloxane). Form factor: Straight-walled, reservoir-ended sheath. Dimensions: Width 55-61mm, Length 170-190mm, Mid-body thickness 0.024mm. Non-sterile. Single-use. Shelf life: 3-5 years. Compliance: ISO 23409:2011, ISO 4074:2015.
Indications for Use
Indicated for contraception and prophylactic purposes to reduce the risk of pregnancy and STIs in individuals using male condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Okamoto 002 Lubricated Polyurethane Male Condom (K203541)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 30, 2026
Mayer Laboratories
David P. Mayer
President
1400 Shattuck Ave., Suite 9
Berkeley, CA 94709
Re: K252158
Trade/Device Name: Sagami Original 002 Polyurethane Male Condom
Regulation Number: 21 CFR§ 884.5300
Regulation Name: Condom
Regulatory Class: II
Product Code: MOL
Dated: July 8, 2025
Received: July 9, 2025
Dear David P. Mayer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252158 - David P. Mayer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252158 - David P. Mayer
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252158 | |
| Device Name Sagami Original 002 Polyurethane Male Condom | |
| Indications for Use (Describe) Sagami Original 002 Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs). | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252158
Page 1 of 4
510(k) Summary
K252158
Sagami Original 002 Polyurethane Male Condom
1. Submitter Information
Applicant: Mayer Laboratories
Address: 1400 Shattuck Ave Suite 9
Berkeley, CA 94709
Phone: (501) 229-5300
2. Correspondent Information
Company: Mayer Laboratories
Contact: David P. Mayer, President
Email: davidm@mayerlabs.com
3. Date prepared: March 30, 2026
4. Device Information
Device Name: Sagami Original 002 Polyurethane Male Condom
Common Name: Polyurethane Condom
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Product Code: MOL (Condom, Synthetic)
Regulatory Class: Class II
5. Predicate Device Information
Device Name: Okamoto 002 Lubricated Polyurethane Male Condom
510(k) Number: K203541
Sponsor: Okamoto U.S.A Inc.
The predicate device has not been subject to a design-related recall.
6. Device Description
The Sagami Original 002 Polyurethane Male Condom is made of a closely fitted sheath with a closed end, which completely covers the male penis with a non-latex, impermeable, non-sterile, fitted membrane made of a polyurethane material. Each device in the family is made of the same material: Polyurethane. All devices have a smooth-surface with a straight/parallel-walled shape, with a closed end that has a reservoir tip. Each device is lubricated with silicone oil ("dimethylpolysiloxane") and individually packaged. The Sagami Original 002 Condoms are a non-sterile, single use device. They are for home use and environments where persons have sexual activity. Based on currently available stability information, Sagami Original 002 Slim Fit and Sagami Original 002 Regular Mayer Labs have a 5 year shelf life and Sagami Original 002 Large has a 3 year shelf life.
{5}
K252158
Page 2 of 4
# 7. Indications for Use Statement
Sagami Original 002 Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
# 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below includes a comparison of the intended use and technological characteristics of the subject and predicate devices.
| | Subject Device
Sagami Original 002
Polyurethane Male
Condom
K252158 | Subject Device
Okamoto 002
Lubricated
Polyurethane Male
Condom
K203541 | Comparison |
| --- | --- | --- | --- |
| Device & Predicate
Device | Sagami Original 002
Polyurethane Male
Condom | Okamoto 002 Lubricated
Polyurethane Male
Condom | -- |
| 510(K) Number | K252158 | K203541 | -- |
| Product Code | MOL | MOL | Same |
| Regulation Number | 21 CFR 884.5300 | 21 CFR 884.5300 | Same |
| Regulation Name | Condom | Condom | Same |
| Indications for Use | Sagami Original 002
Polyurethane Male
Condom is used for
contraception and for
prophylactic purposes
(to help reduce the risk
of pregnancy and the
transmission of sexually
transmitted infections,
STIs). | The Okamoto 002
Lubricated Polyurethane
Male Condom is used
for contraception and for
prophylactic purposes
(to help reduce the risk
of pregnancy and the
transmission of sexually
transmitted infections,
STIs). | Same |
| Prescription or Over-
The-Counter Use | Over-The-Counter | Over-The-Counter | Same |
| Condom Material | Polyurethane | Polyurethane | Same |
| Nominal Width | Slim fit: 55 ± 2 mm
Regular: 58 ± 2 mm
Large: 61 ± 2 mm | 57 ± 2 mm | Different |
| Nominal Length | Slim fit: 170 ± 10 mm
Regular: 190 ± 10 mm
Large: 190 ± 10 mm | 180 ± 10 mm | Different |
| Nominal Thickness | Mid-body: 0.024 ± 0.008
mm | 0.028 ± 0.008 mm | Different |
| Lubricant | Silicone Oil | Silicone Oil | Same |
| Lubricant Quantity | 400 ± 100 mg | 400 ± 40 mg | Different |
| Air Burst Pressure | Not less than 2.0 kPa | Not less than 1.8 kPa | Different |
| Air Burst Volume | Not less than 3.6 dm³ | Not less than 4.0 dm³ | Different |
| Sterilization | Non-Sterile | Non-Sterile | Same |
{6}
K252158
Page 3 of 4
| Shape | Straight-walled & Reservoir-ended | Straight-walled & Reservoir-ended | Same |
| --- | --- | --- | --- |
| Texture | Smooth Surface | Smooth Surface | Same |
| Shelf Life | Slim fit, regular: 5 Years
Large: 3 Years | 5 Years | Different |
| Color Additives | N/A | N/A | Same |
| Flavor Additives | N/A | N/A | Same |
The subject and predicate device have similar indications for use and have the same intended use. The technological characteristics of the subject and predicate devices are similar in that they are polyurethane-based, are lubricated with silicone oil, and have the same shelf-life duration. The subject and predicate condoms also have identical formulations. There are differences in dimensions and specifications; the subject device has reduced thickness and length compared to the predicate device. These differences do not raise different questions of safety and effectiveness.
## 9. Summary of Non-Clinical Performance Testing
### Biocompatibility:
Biocompatibility studies, including Acute Systemic Toxicity, Vaginal Irritation Testing, Cytotoxicity and Sensitization testing were performed in accordance with a surface medical device with limited contact duration (<24 hr contact with skin and mucosal membranes) per the 2023 FDA guidance document *Use of International Standard ISO 10993-1*, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as follows:
- Cytotoxicity (ISO 10993-5:2009/R 2014)
- Sensitization (ISO 10993-10:2010/R 2014)
- Vaginal Irritation (ISO 10993-10:2010/R 2014)
- Acute Systemic Toxicity (ISO 10993-11:2017)
The results of testing demonstrate that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and not acutely, systemically toxic.
### Physical Performance Testing:
The subject condom was tested for compliance with the ISO 23409:2011, "Male Condoms – Requirements and test methods for condoms made from synthetic materials," "Guidance for Male Condoms Made from New Material (Non-Latex), issued by the FDA in June 1995, and meets requirements of ISO 4074:2015," "Natural rubber latex male condoms – Requirements and test methods" for dimensional, tensile strength, force at break, lubricant quantity, visible defects, tear resistance and propagation, elongation, air burst volume and air burst pressure requirements.
### Package Integrity:
An evaluation of seal integrity was performed on three lots (3) of the subject device according to ISO 23409: 2011 with satisfactory results.
### Barrier Properties/Permeability:
A viral penetration study was performed on three (3) test lots of the subject polyurethane condom, one lot of the commercially available predicate device, and a natural rubber latex condom comparator per
{7}
K252158
Page 4 of 4
ISO 23409: 2011. The cumulative results of the studies demonstrate the barrier effectiveness of the subject device as compared to the predicate device and natural rubber latex control condom for viral penetration under conditions of the *in vitro* study.
## Shelf Life:
The Sagami Original 002 Polyurethane Male Condom has a shelf life based on the results of accelerated stability evaluations conducted as required in 21 CFR 801.435 and ISO 23409:2011. Based on the currently available, stability information, Mayer Labs is claiming a 5 year shelf life on Sagami Original 002 Slim Fit and Sagami Original 002 Regular, and a 3 year shelf life on Sagami Original 002 Large.
All samples met predefined acceptance criteria.
## Clinical In-Use Slip/Break:
Clinical testing was performed. The Study design complied with the recommendations of the FDA guidelines for clinical studies on synthetic condoms and ISO 29943 -1:2017. The Study is registered on the clinicaltrials.gov website with the identifier NCT04622306. The slippage and breakage study, which conforms to ISO 23409: 2011 and follows a protocol prepared to meet the FDA guidance "Testing Guidance for Male Condoms Made From New Material (Non-Latex)" June 1995, was conducted. Condom Polyurethane A, representative of the Sagami Original 002 Polyurethane Male condom, met the noninferiority requirements specified in ISO 23409:2011 and demonstrated low failure rates relative to a standard natural rubber latex condom as a comparator in the full study population.
## 10. Conclusion
The results of the performance testing described above demonstrate that the Sagami Original 002 Polyurethane Male Condom is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.