K252130 · Icu Medical, Inc. · PHC · Apr 2, 2026 · General Hospital
Device Facts
Record ID
K252130
Device Name
LifeShield Infusion Safety Software Suite
Applicant
Icu Medical, Inc.
Product Code
PHC · General Hospital
Decision Date
Apr 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The LifeShield™ Infusion Safety Software Suite is a collection of software products that facilitates networked communication between compatible systems. The Infusion Safety Software Suite provides trained healthcare professionals the ability to manage data for compatible infusion pumps. All data entry and validation of infusion parameters on compatible infusion pumps is performed by a trained healthcare professional. LifeShield™ Infusion Safety Software Suite is indicated for use in patients including adult, pediatric and neonate undergoing infusion therapy with connected compatible infusion pumps (as per the indications for use specified for the compatible infusion pump). The LifeShield™ Drug Library Management (DLM) software product is intended to be used by pharmacists to create, configure, edit, and manage drug library data, including infusion pump settings, for use with compatible infusion pumps. Drug library contents are constructed based on the healthcare provider's defined best practices. The LifeShield™ Clinical Dashboards & Reports (CDR) software product provides trained healthcare professionals with the capability to view and manage infusion information collected from compatible infusion pumps. Healthcare professionals may choose to use the collected infusion information to support continuous quality improvement programs, or to analyze and trend various aspects of the infusion pumps and therapies used. It is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways. The LifeShield™ Data Flow Management (DFM) software product is intended to facilitate bi-directional communication with compatible infusion pumps, information technology systems, and other LifeShield™ Infusion Safety Software Suite products. LifeShield™ DFM provides a way to automate the programming of infusion parameters, thereby decreasing the number of manual steps necessary to enter infusion data. LifeShield™ DFM forwards infusion-related information received from the infusion pump to compatible information technology systems.
Device Story
LifeShield Infusion Safety Software Suite is a cloud-based SaaS platform hosted on AWS; facilitates networked communication and data management for compatible Plum Duo and Plum Solo infusion pumps. Comprises three modules: Drug Library Management (DLM) for pharmacists to configure pump settings/drug libraries; Clinical Dashboards & Reports (CDR) for clinicians to view infusion data/trends and integrate third-party location system data; and Data Flow Management (DFM) for bi-directional communication between pumps and hospital IT systems. DFM automates infusion parameter programming to reduce manual entry errors. Operated by pharmacists and clinical administrators in hospital settings. Does not provide real-time pump control, alarm management, or clinical decision support. Benefits include streamlined fleet management, standardized drug library deployment, and improved infusion documentation efficiency.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including software verification (IEC 62304:2015), design validation, cybersecurity evaluation, risk management (ISO 14971:2019), and human factors studies (IEC 62366-1:2020).
Technological Characteristics
Cloud-based SaaS platform (AWS); web-based application compatible with Google Chrome and Microsoft Edge. Supports bi-directional communication with Plum Duo/Solo pumps. Features include multi-factor authentication (TOTP), digitally-signed drug libraries, and integration with third-party location systems. Software development follows IEC 62304:2015; risk management per ISO 14971:2019; human factors per IEC 62366-1:2020.
Indications for Use
Indicated for adult, pediatric, and neonate patients undergoing infusion therapy with connected compatible infusion pumps.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
LifeShield™ Infusion Safety Software Suite (K242117)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 2, 2026
ICU Medical, Inc.
Pernell Abrantes
Director of Regulatory Affairs
600 North Field Drive
Lake Forest, Illinois 60045
Re: K252130
Trade/Device Name: LifeShield Infusion Safety Software Suite
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: PHC
Dated: February 25, 2025
Received: February 25, 2025
Dear Pernell Abrantes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252130 - Pernell Abrantes
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252130 - Pernell Abrantes
Page 3
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director, Infusion Devices
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252130 | |
| Device Name LifeShield Infusion Safety Software Suite | |
| Indications for Use (Describe) The LifeShield™ Infusion Safety Software Suite is a collection of software products that facilitates networked communication between compatible systems. The Infusion Safety Software Suite provides trained healthcare professionals the ability to manage data for compatible infusion pumps. All data entry and validation of infusion parameters on compatible infusion pumps is performed by a trained healthcare professional. LifeShield™ Infusion Safety Software Suite is indicated for use in patients including adult, pediatric and neonate undergoing infusion therapy with connected compatible infusion pumps (as per the indications for use specified for the compatible infusion pump). The LifeShield™ Drug Library Management (DLM) software product is intended to be used by pharmacists to create, configure, edit, and manage drug library data, including infusion pump settings, for use with compatible infusion pumps. Drug library contents are constructed based on the healthcare provider's defined best practices. The LifeShield™ Clinical Dashboards & Reports (CDR) software product provides trained healthcare professionals with the capability to view and manage infusion information collected from compatible infusion pumps. Healthcare professionals may choose to use the collected infusion information to support continuous quality improvement programs, or to analyze and trend various aspects of the infusion pumps and therapies used. It is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways. The LifeShield™ Data Flow Management (DFM) software product is intended to facilitate bi-directional communication with compatible infusion pumps, information technology systems, and other LifeShield™ Infusion Safety Software Suite products. LifeShield™ DFM provides a way to automate the programming of infusion parameters, thereby decreasing the number of manual steps necessary to enter infusion data. LifeShield™ DFM forwards infusion-related information received from the infusion pump to compatible information technology systems. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
# 510(k) Summary
The 510(k) summary is prepared in accordance with the requirements of 21 CFR Part 807.92.
| Submitter Information | |
| --- | --- |
| Name | ICU Medical, Inc. |
| Address | 600 North Field Drive
Lake Forest, IL 60045 |
| Establishment Registration Number | 3013319212 |
| Name of contact person | Pernell Abrantes |
| Phone number | (224) 706-2229 |
| Date Prepared | March 26, 2026 |
| Name of device | |
| Trade or proprietary name | LifeShield™ Infusion Safety Software Suite |
| Common or usual name | Infusion Safety Management Software |
| Classification | Class II |
| Classification Reason | 21 CFR 880.5725 |
| Panel | General Hospital |
| Product Code(s) | PHC |
| Predicate Device | K242117, LifeShield™ Infusion Safety Software Suite |
| Reason for 510(k) submission | The submission is a traditional pre-market notification for software modifications made to LifeShield™ Infusion Safety Software Suite. |
| Device description | The LifeShield™ Infusion Safety Software Suite is a cloud-based platform provided as a software-as-a-service (SaaS) and is hosted by Amazon Web Services (AWS). LifeShield™ Infusion Safety Software Suite is designed to be compatible with ICU Medical large volumetric pumps (Plum Duo™ Precision IV infusion pump, Plum Solo™ Precision IV infusion pump).
LifeShield™ Infusion Safety Software Suite consists of a collection of software services which, when used together, provide a comprehensive set of data management capabilities for trained healthcare professionals when working with infusion pumps. LifeShield™ Infusion Safety Software Suite does not remotely control or program the infusion pump or provide the ability to remotely manage pump alarms such as real-time monitoring, clearing and silencing alarms.
The LifeShield™ Drug Library Management (DLM) software is used by pharmacists to create approved drug libraries that can be downloaded by the |
Page 1 of 8
{5}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| | compatible infusion pumps. The latest software version implements general workflow enhancements.
The LifeShield™ Clinical Dashboards & Reports (CDR) software is used by clinical administrators to view infusion or device-related information received from the infusion pumps via the LifeShield™ DFM. The information presented by the software does not create decisions or treatment pathways for patients. The latest version of the software displays infusion and device information including reporting of pump location from the third-party location system.
LifeShield™ Data Flow Management (DFM) software facilitates bi-directional communications between the infusion pump and hospital information systems (HIS); it routes pharmacy-validated orders to the connected pumps and infusion-related information and alarms to the HIS. The latest software version allows connectivity with third-party location systems. |
| --- | --- |
| Intended Use of Device/Indication for use | The LifeShield™ Infusion Safety Software Suite is a collection of software products that facilitates networked communication between compatible systems. The Infusion Safety Software Suite provides trained healthcare professionals the ability to manage data for compatible infusion pumps. All data entry and validation of infusion parameters on compatible infusion pumps is performed by a trained healthcare professional. LifeShield™ Infusion Safety Software Suite is indicated for use in patients including adults, pediatrics and neonates undergoing infusion therapy with connected compatible infusion pumps (as per the indications for use specified for the compatible infusion pump).
The LifeShield™ Drug Library Management (DLM) software product is intended to be used by pharmacists to create, configure, edit, and manage drug library data, including infusion pump settings, for use with compatible infusion pumps. Drug library contents are constructed based on the healthcare provider's defined best practices.
The LifeShield™ Clinical Dashboards & Reports (CDR) software product provides trained healthcare professionals with the capability to view and manage infusion information collected from compatible infusion pumps. Healthcare professionals may choose to use the collected infusion information to support continuous quality improvement programs, or to analyze and trend various aspects of the infusion pumps and therapies used. It is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.
The LifeShield™ Data Flow Management (DFM) software product is intended to facilitate bi-directional communication with compatible infusion pumps, information technology systems, and other LifeShield™ Infusion Safety Software Suite products. LifeShield™ DFM provides a way to automate the programming |
Page 2 of 8
{6}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| | of infusion parameters, thereby decreasing the number of manual steps necessary to enter infusion data. LifeShield™ DFM forwards infusion-related information received from the infusion pump to compatible information technology systems. |
| --- | --- |
Page 3 of 8
{7}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| Characteristic | Subject Device | Predicate K242117 | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The LifeShield™ Infusion Safety Software Suite is a collection of software products that facilitates networked communication between compatible systems. The Infusion Safety Software Suite provides trained healthcare professionals the ability to manage data for compatible infusion pumps. All data entry and validation of infusion parameters on compatible infusion pumps is performed by a trained healthcare professional. LifeShield™ Infusion Safety Software Suite is indicated for use in patients including adult, pediatric and neonate undergoing infusion therapy with connected compatible infusion pumps (as per the indications for use specified for the compatible infusion pump).
The LifeShield™ Drug Library Management (DLM) software product is intended to be used by pharmacists to create, configure, edit, and manage drug library data, including infusion pump settings, for use with compatible infusion pumps. Drug library contents are constructed based on the healthcare provider's defined best practices.
The LifeShield™ Clinical Dashboards & Reports (CDR) software product provides trained healthcare professionals with the capability to view and manage infusion information collected from compatible infusion pumps. Healthcare professionals may choose to use the collected infusion information to support continuous quality improvement programs, or to analyze and trend various aspects of the infusion pumps and therapies used. It is not intended to be a substitute for good clinical management practices, nor does its operation create decision or treatment pathways.
The LifeShield™ Data Flow Management (DFM) software product is intended to facilitate bi-directional communication with compatible infusion pumps, information technology systems, and other LifeShield™ Infusion Safety Software Suite products. LifeShield™ DFM provides a way to automate the | Same | |
Page 4 of 8
{8}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate K242117 | Comparison |
| | programming of infusion parameters, thereby decreasing the number of manual steps necessary to enter infusion data. LifeShield™ DFM forwards infusion-related information received from the infusion pump to compatible information technology systems. | | |
| Prescription Device | Yes | | Same |
| Stand-alone software or embedded software | Stand-alone | | Same |
| Application Type | Web-based | | Same |
| Compatible Browser | Google Chrome, Microsoft Edge | | Same |
| Database provider | Amazon Web Services Cloud | | Same |
| Compatible Infusion Pump | Plum Duo
Plum Solo | Plum Duo
Plum Solo | Same |
| Drug library Capacity | 40 Clinical Care Areas / 40,000 total rulesets | | Same |
| No Drug Selected | Available for all drug libraries | | Same |
| Drug Library Report | Single printable report | | Same |
| Facility support | Multiple facilities for drug library and data reporting | | Same |
| Available Drug Library | Plum LVP (Plum Duo™, Plum Solo™) | Plum LVP (Plum Duo™, Plum Solo™) | Same |
| Drug Library Workflow | -Enhanced copy ruleset feature
-Drug library content review and approval | -Ability to copy existing ruleset
-Drug library content review and approval | Similar
Updates included workflow and user interface refinements to better |
Page 5 of 8
{9}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate K242117 | Comparison |
| | -UI improvements for managing a larger number of drug library and pump software updates | -Ability to view status of drug library and pump software updates. | facilitate creating and scheduling updates of drug libraries. Identified risks are effectively mitigated, confirming the changes did not raise different questions of safety and effectiveness. |
| Drug Library Settings | Plum LVP (Plum Duo™, Plum Solo™) settings | | Same |
| Infusion and Device Status Display | Near-real time | | Same |
| Historical Reports | Historical reports can be configured for the customer account and are displayed for drug library management, infusion data management and asset management. | | Same |
| Auto-programming | Supported | | Same |
| Alarm Forwarding | Supported | | Same |
| Infusion Documentation | Supported | | Same |
| Asset Management | -Drug Library Update
-Software Update of Pump
-Device status view with customizable access point description or with locations service data
-Display of Pump Certificate Status
-Pump Log Download | -Drug Library Update
-Software Update of Pumps
-Device status view with customizable access point description
-Display of Pump Certificate Status
-Pump Log Download | Similar
Both subject and predicate devices offer the same set of features to manage the pump fleet, with the subject device introducing the ability to archive pumps to help streamline the list of pumps to be managed and reporting pump locations from third-party location systems. Minor improvement to display of device |
Page 6 of 8
{10}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate K242117 | Comparison |
| | | | status does not raise different questions of safety and effectiveness. |
| Multi-factor Authentication | Time-based one-time password with ability to enforce for all users. | Time-based one-time password | Similar
The subject device adds an additional layer of security by providing the option to enforce multi-factor authentication on all users. |
| Digitally-signed Drug Library | Supported | | Same |
Page 7 of 8
{11}
LifeShield™ Infusion Safety Software Suite
K252130
eSTAR: 510(k) Summary
icumedical human connections
# Summary of Non-Clinical Testing
Non-clinical testing of LifeShield™ Infusion Safety Software Suite has been conducted to evaluate the safety, performance and functionality. The results of these tests have demonstrated its overall safety and effectiveness, and support a substantial equivalence determination to the predicate device. A summary of the testing conducted is presented below:
- Design verification tests pass established acceptance criteria, confirming the subject device meets design requirements. Verification activities also include software verification, performance, reliability, compatibility, and interoperability tests. Software verification follows the software development process outlined in IEC 62304:2015.
- Design validation tests pass established acceptance criteria, confirming the subject device meets all intended users' needs for the device's intended use and intended use environment.
- Cybersecurity evaluation and testing demonstrate that the software is reasonably secure.
- Risk management activities are performed in accordance with ISO 14971:2019, and risk mitigations have been incorporated into the design and have been tested for correct implementation and effectiveness as part of design verification and validation. Safety Assurance Case conducted in accordance with AAMI TIR38:2018 and FDA Guidance: Total Product Lifecycle of Infusion Pumps issued in 2014 concludes that the subject device is reasonably safe.
- Human Factors studies demonstrate the effectiveness of the user interface design for additional features and their associated critical tasks. The human factors studies are conducted in accordance with IEC 62366-1:2020 and ANSI/AAMI HE75:2009/(R)2018.
# Summary of Clinical Testing
Not applicable. A clinical study is not required for this submission to support substantial equivalence.
# Conclusion
The subject device, LifeShield™ Infusion Safety Software Suite, is substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.