AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252105 · Sep 26, 2025
Ligence Heart
Ligence Uab
Retrospective clinical echocardiography studies
A retrospective validation study was conducted using 524 previously acquired clinical echocardiograms to evaluate the performance and interchangeability of the device's automated measurements (LVEF, LVEDV, LVESV, and lateral e') against expert human sonographers.
Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi). Limitations: • Poor image capture will lead to poor annotations and subsequent measurements. • Multiple image quality algorithms are used to filter out images of poor quality. • Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. • If no gender is present, female referenced guideline values will be used for conclusions. • If Body Surface Area (BSA) is not present, indexed values cannot be provided. • During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software. • Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.
Device Story
Ligence Heart is an automated image post-processing software for adult transthoracic echocardiography. It ingests DICOM images; utilizes machine-learning-based view classification, cardiac cycle selection, and border detection to quantify left-ventricular function; and generates structured reports. Used in clinical settings by cardiologists or sonographers, the software provides automated measurements (LVEDV, LVESV, LVEF, lateral e') for review. Clinicians may accept, edit, or replace automated results. The device streamlines echocardiography workflows, supporting clinical decision-making by providing reproducible quantitative data.
Clinical Evidence
Retrospective non-interventional study of 524 adult echocardiographic studies. Performance compared against three expert sonographers. Primary endpoint: Individual Equivalence Coefficient (IEC) with non-inferiority margin < 0.25. All four parameters (LVEF, LVEDV, LVESV, lateral e') met non-inferiority criteria. Intraclass Correlation Coefficients (ICC) were ≥ 0.90. No device-related adverse events reported.
Technological Characteristics
Software-only platform operating on off-the-shelf PC hardware. Utilizes machine-learning-based algorithms for view classification, cardiac cycle selection, and border detection. Compliant with DICOM standards for image input/output. No specific hardware materials or energy sources; software-based processing.
Indications for Use
Indicated for adult patients undergoing transthoracic echocardiography to support cardiologists or sonographers in clinical decision-making. Not validated for congenital heart disease, valve disease, pericardial disease, intra-cardiac lesions, prosthetic/surgically altered valves, implantable intracardiac devices, or irregular (non-sinus) rhythms.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 26, 2025
Ligence, UAB
% Raymond Kelly
Consultant
Arazy Group Consultants Inc.
3422 Leonardo Lane
New Smyrna Beach, Florida 32168
Re: K252105
Trade/Device Name: Ligence Heart
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: September 5, 2025
Received: September 5, 2025
Dear Raymond Kelly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252105 - Raymond Kelly
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252105 - Raymond Kelly
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Ligence Heart
Page 10 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252105 | ? |
| Please provide the device trade name(s). | | ? |
| Ligence Heart | | |
| Please provide your Indications for Use below. | | ? |
| Ligence Heart is a fully automated software platform that processes, analyses and makes measurements
on acquired transthoracic cardiac ultrasound images, automatically producing a full report with
measurements of several key cardiac structural and functional parameters. The data produced by this
software is intended to be used to support qualified cardiologists or sonographers for clinical decision
making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the
assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions
(e.g., tumors, thrombi).
Limitations:
• Poor image capture will lead to poor annotations and subsequent measurements.
• Multiple image quality algorithms are used to filter out images of poor quality.
• Our software complements good patient care and does not exempt the user from the responsibility to
provide supervision, clinically review the patient, and make appropriate clinical decisions.
• If no gender is present, female referenced guideline values will be used for conclusions.
• If Body Surface Area (BSA) is not present, indexed values cannot be provided.
• During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes,
or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software.
• Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves,
implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated
measurement accuracy. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Page 1 of 5
510(k) Summary
21 CFR 807.92
K252105
I. SUBMITTER (21 CFR 807.92(a)(1))
Submitter: Ligence, UAB
Taikos pr. 54 Kaunas,
51305, Lithuania
Phone: +37061144345
Contact Person: Alvaro Perez-Moreno, Ph.D.
Date Prepared: July 2, 2025
II. DEVICE (21 CFR 807.92(a)(2))
Name of Device: Ligence Heart
Common or Usual Name: Automated Radiological Image Processing Software
Classification Name: Medical Image Management and Processing System (§892.2050)
Regulatory Class: II
Product Code: QIH
III. PREDICATE DEVICE (21 CFR 807.92(a)(3))
Us2.v1, K210791
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
IV. DEVICE DESCRIPTION (21 CFR 807.92(a)(4))
Ligence Heart is an image post-processing software used for viewing and quantifying adult cardiac ultrasound DICOM studies. The device is intended to generate structured measurement reports for echocardiography analysis and aid qualified sonographers and cardiologists in their decision-making process.
Ligence Heart automatically identifies standard transthoracic echo views with machine-learning-based view classification, cardiac cycle selection, and border detection, then
{5}
Page 2 of 5
generates reproducible quantitative left-ventricular volumetric and functional measurements. The results are inserted into a PACS-compatible report that the reviewing cardiologist or sonographer can accept, edit, supplement with additional manual measurements, or entirely replace with manual measurements. The software also organizes, displays, and compares each measurement with reference-guideline ranges. Completed reports are exported in PDF, streamlining routine echocardiography workflow while leaving final diagnostic responsibility with the clinician.
## V. INDICATIONS FOR USE (21 CFR 807.92(a)(5))
Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi).
## Limitations:
- Poor image capture will lead to poor annotations and subsequent measurements.
- Multiple image quality algorithms are used to filter out images of poor quality.
- Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions.
- If no gender is present, female referenced guideline values will be used for conclusions.
- If Body Surface Area (BSA) is not present, indexed values cannot be provided.
- During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software.
- Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.
{6}
Page 3 of 5
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE (21 CFR 807.92(a)(5) & 21 CFR 807.92(a)(6))
| Attribute | Predicate Device: K210791 | Subject Device |
| --- | --- | --- |
| Product Code | QIH | QIH |
| Regulation | 892.2050 | 892.2050 |
| Classification | II | II |
| Rx/OTC | Rx | Rx |
| Physical Characteristics | Software package that operates utilizing off-the-shelf hardware | Software package that operates utilizing off-the-shelf hardware |
| User interface | The software is designed for use on a personal computer that has received images from a compatible PACS | The software is designed for use on a personal computer that has received images from a compatible PACS |
| Automation level | Fully automated, including clip selection | Fully automated, including clip selection |
| User confirmation / rejection of result | Yes | Yes |
| Manual editing of automated result by user | Yes | Yes |
| Input Data | Echocardiographic images | Echocardiographic images |
| Communication protocol | DICOM Standard Compliance | DICOM Standard Compliance |
| Patient Populations | Adults | Adults |
Both the subject and predicate devices share the following technological characteristics:
Variations in automated-measurement implementation:
Ligence Heart automatically generates four core left-ventricular measurements essential for efficient and accurate routine cardiac assessments: lateral mitral-annular early-diastolic velocity (lateral e', Le), left-ventricular end-diastolic volume by the biplane Method of Disks (LVEDV MOD), left-ventricular ejection fraction by the biplane Method of Disks (LVEF MOD), and left-ventricular end-systolic volume by the biplane Method of Disks (LVESV MOD).
{7}
Page 4 of 5
# VII. PERFORMANCE DATA (21 CFR 807.92(b))
Ligence Heart was developed and tested in accordance with Ligence Heart's life cycle development processes. The device has been subject to extensive safety and performance testing. Verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at the unit, integration, and system levels. Risk management analysis generated multiple risk mitigation measures and verification activities. A cybersecurity analysis and testing were conducted to verify that security measures are properly implemented in Ligence Heart to protect Ligence Heart's data, including patient health information (PHI). A summative human factors study was performed to validate that Ligence Heart can be used safely and effectively by the intended user population.
A formal retrospective non-interventional validation study was conducted at a U.S. based independent Echocardiography Core Laboratory using 600 previously acquired echocardiograms to evaluate the performance of the device for each claimed measurement. Test datasets were strictly segregated from algorithm training datasets, as they are from completely separate cohorts.
A cumulative total of 76 studies were excluded because of missing views or corrupt metadata. Ultimately, 524 echocardiographic studies were assessed by all three human readers, satisfied the predefined quality criteria, and were included in the final analysis. Baseline demographic characteristics of the cohort are summarized in the table below.
| Participants characteristics | |
| --- | --- |
| Age, years | 49.48 ± 13.56 |
| Male, n (%) | 319 (60.88%) |
| Female, n (%) | 205 (39.12%) |
| Height, cm | 173.36 ± 10.52 |
| Weight, kg | 88.92 ± 21.11 |
| LVEF, % | 34.45% ± 15.78% |
| Normal LVEF (>50%), n (%) | 120 (22.90%) |
| Reduced LVEF (<50%), n (%) | 404 (77.10%) |
{8}
Page 5 of 5
| Participants characteristics | |
| --- | --- |
| Mildly reduced LVEF (40-50%), n (%) | 64 (12.21%) |
| Moderately reduced LVEF (30-40%), n (%) | 85 (16.22%) |
| Severely reduced LVEF (<30%), n (%) | 255 (48.66%) |
Primary Clinical Performance Endpoint was Individual Equivalence Coefficient (IEC). Agreement between Ligence Heart and three independent expert sonographers was quantified using the reference-scaled IEC. Automated measurements are deemed interchangeable when the upper level of the IEC 95% confidence interval was < 0.25, a non-inferiority margin derived from individual-bioequivalence principles.
All four parameters met the non-inferiority criterion, confirming that Ligence Heart measurements are interchangeable with expert human measurements. Secondary agreement metrics were consistent with the IEC result. Intraclass Correlation Coefficients (ICC) were ≥ 0.90 for all four parameters. No device-related adverse events, malfunctions, or unmediated issues occurred during the study.
These combined non-clinical and clinical data demonstrate that Ligence Heart satisfies its specified performance criteria and support determination of substantial equivalence.
# VIII. CONCLUSIONS (21 CFR 807.92(b))
Ligence Heart is substantially equivalent to the predicate device Us2.v1 (K210791). The subject and predicate share the same intended use, core technological characteristics, and fundamental principles of operation. No distinctions between the two systems do not introduce new questions of safety or effectiveness.
Comprehensive non-clinical testing, including software verification and validation, cybersecurity assessment, and summative human-factors validation, demonstrated that the software platforms as intended and can be used safely under the specified conditions. Complementary clinical validation confirmed that automated measurements of LVEF, LVEDV, LVESV, and lateral e' are interchangeable with expert manual measurements and remain within normal inter-observer variability.
Taken together, the non-clinical and clinical data show that Ligence Heart is at least as safe and effective as the predicative device for post processing analysis of adult transthoracic echocardiograms.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.