Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
K252074 · Canon Medical Systems Corporation · IYN · Oct 31, 2025 · Radiology
Device Facts
Record ID
K252074
Device Name
Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
Applicant
Canon Medical Systems Corporation
Product Code
IYN · Radiology
Decision Date
Oct 31, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252074 · Oct 31, 2025
Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
Canon Medical Systems Corporation
Retrospective clinical ultrasound images from a U.S. clinical site
Retrospective clinical images from 30 patients were used to validate the performance of the new 3rd Harmonic Imaging (3-HI) feature by comparing it against predicate functionality (2nd harmonic imaging) via a blinded reader study.
3-HI Clinical Evaluation; Retrospective blinded observer study
Adult patients clinically scheduled for abdominal or pelvic imaging, or patients with a prior diagnosis of focal pathologies of the abdomen or pelvis.; Sample Size: 30 patients (minimum 294 images); Number of Sites: 1
Spatial resolution, contrast resolution, and artifact suppression using a 5-point ordinal scale.
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
3rd Harmonic Imaging (3-HI) Image Quality
Deep learning enabled filtering process
Median score > 3 on a 5-point ordinal scale
Scores for 3-HI were higher than the middle score of 3
—
—
Clinical evaluation: 30 patients from a U.S. clinical site
3 (US board certified radiologists)
Indications for Use
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs (thyroid, breast and testicle), trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural. This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler, Power Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or appropriately trained healthcare professionals. In addition to the aforementioned indications for use, when EUS transducer GF-UCT180 and BF-UC190F are connected, Aplio i800 Model TUS-AI800/E3 provides image information for diagnosis of the upper gastrointestinal tract and surrounding organs, airways, tracheobronchial tree and esophagus.
Device Story
Mobile diagnostic ultrasound system; uses various probes (2MHz-33MHz) to acquire B-mode, M-mode, Doppler, and elastography data. Operates in hospital/clinical settings; used by physicians/trained professionals. New V9.0 features include 3rd Harmonic Imaging (3-HI) using deep learning for abdominal image enhancement; improved SMI Angio Mode for blood flow separation; Precision+ Fine Processing for border depiction; and Auto Tune for shear wave elasticity/dispersion optimization. System outputs high-quality ultrasound images for diagnostic interpretation. Clinicians use these images to visualize anatomy and pathology, aiding clinical decision-making. Benefits include improved spatial/contrast resolution and artifact suppression, facilitating more accurate diagnosis.
Clinical Evidence
Clinical evaluation of 3-HI feature involved 30 patients (abdominal/pelvic imaging). Three board-certified radiologists performed a blinded comparison against predicate (2nd harmonic imaging) using a 5-point ordinal scale. Results showed statistically higher scores for 3-HI in spatial resolution, contrast resolution, and artifact suppression. Supplemented by phantom studies evaluating lateral/axial resolution, slice resolution, CNR, and reverberation suppression. SMI Angio and Auto Tune features validated via phantom data and volunteer imaging.
Technological Characteristics
Mobile ultrasound system; probes 2-33MHz. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, IEC 62359, ISO 10993-1. Connectivity: Windows 11 LTSC OS. Software: DL-based 3rd Harmonic Imaging (locked algorithm), rule-based image processing for SMI, Precision+, and Auto Tune. Sterilization: per transducer specifications.
Indications for Use
Indicated for visualization of structures and dynamic processes in fetal, abdominal, intra-operative, pediatric, small organ, trans-vaginal, trans-rectal, neonatal/adult cephalic, cardiac, peripheral vascular, transesophageal, musculo-skeletal, laparoscopic, and thoracic/pleural applications. Includes upper GI/airway/esophagus diagnosis when using specific EUS transducers.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 31, 2025
Canon Medical Systems Corporation
Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems USA, Inc.
2441 Michelle Dr
Tustin, California 92780
Re: K252074
Trade/Device Name: Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, QIH
Dated: October 3, 2025
Received: October 3, 2025
Dear Yoshiaki Cook:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252074 - Yoshiaki Cook
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252074 - Yoshiaki Cook
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252074
Device Name
Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
Indications for Use (Describe)
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs (thyroid, breast and testicle), trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural. This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler, Power Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or appropriately trained healthcare professionals.
In addition to the aforementioned indications for use, when EUS transducer GF-UCT180 and BF-UC190F are connected, Aplio i800 Model TUS-AI800/E3 provides image information for diagnosis of the upper gastrointestinal tract and surrounding organs, airways, tracheobronchial tree and esophagus.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
K252074
510(k) SUMMARY
1. SUBMITTER'S NAME
Junichiro Araoka
Sr. Manager, Quality Assurance Dept.
Quality, Safety and Regulation Center
Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-shi, Tochigi-ken, Japan 324-8550
2. ESTABLISHMENT REGISTRATION
9614698
3. OFFICIAL CORRESPONDENT/CONTACT PERSON
Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems USA, Inc.
2441 Michelle Drive
Tustin, CA 92780
ycook@us.medical.canon
+1 (657) 270-5595
4. DATE PREPARED
July 01, 2025
5. DEVICE NAME/TRADE NAME
Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System
6. COMMON NAME
System, Diagnostic Ultrasound
7. DEVICE CLASSIFICATION
Class II
Ultrasonic Pulsed Doppler Imaging System – Product Code: IYN [per 21 CFR 892.1550]
Ultrasonic Pulsed Echo Imaging System – Product Code: IYO [per 21 CFR 892.1560]
Diagnostic Ultrasonic Transducer – Product Code: ITX [per 21 CFR 892.1570]
Medical Image Management and Processing System – Product Code: QIH [per 21 CFR 892.2050]
Page 1 of 5
{5}
Page 2 of 5
# 8. PREDICATE DEVICE
| Product | Marketed by | 510(k) Number | Clearance Date |
| --- | --- | --- | --- |
| Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V8.5 | Canon Medical Systems USA, Inc. | K242808 | May 13, 2025 |
# 9. REASON FOR SUBMISSION
Modification of a cleared device.
# 10. DEVICE DESCRIPTION
The Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700, V9.0 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex, and sector array with frequency ranges between approximately 2MHz to 33MHz.
# 11. INDICATIONS FOR USE
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs (thyroid, breast and testicle), trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural. This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler, Power Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or appropriately trained healthcare professionals.
In addition to the aforementioned indications for use, when EUS transducer GF-UCT180 and BF-UC190F are connected, Aplio i800 Model TUS-AI800/E3 provides image information for diagnosis of the upper gastrointestinal tract and surrounding organs, airways, tracheobronchial tree and esophagus.
# 12. SUBSTANTIAL EQUIVALENCE
The Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700, V9.0 are substantially equivalent to the Aplio i900/i800/i700, Diagnostic Ultrasound System, V8.5 (K242808). The subject devices employ the same fundamental scientific technology as the predicate devices and function in a manner similar to, and are intended for the same use as the predicate devices. The subject devices include new features, as well as improvements to existing functionality. These differences between the subject devices and cleared predicate devices do not raise any new questions about the safety and effectiveness of the subject devices. This submission includes details which demonstrate the substantial equivalence of the improved features, to those currently cleared with the predicate device.
- The subject devices have the same clinical intended use and use the same imaging modes as the predicate devices
{6}
- The transducers supported by the subject and predicate devices are identical
- The software features and functionality supported in the subject and predicate devices are identical except for the following
- 3rd Harmonic Imaging (3-HI), a new deep learning (DL) enabled filtering process for images of the abdominal region, which enables higher spatial resolution, contrast resolution, and reduced artifacts
- SMI Angio Mode, an improvement to the SMI functionality available with the predicate device, which emphasizes separation of adjacent blood flows
- Precision+ Fine Processing Mode, an improvement to existing Precision+ functionality, to enable improved depiction of borders
- Auto Tune, and improvement to existing Shear wave elasticity(SWE) and Shear wave dispersion(SWD) functionality, which introduces a pre-scan which optimizes scan conditions in advance of the main scan for more accurate SWE results
- The operating system incorporated in the subject devices will be changed to Windows 11 LTSC due to the forthcoming termination of support of the Windows 10 LTSC incorporated in the predicate devices
## 14. SAFETY
The subject devices are designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. These devices are in conformance with the applicable parts of the ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012(Cons. Text) [Incl. AMD2:2021], IEC 60601-1-2 (2020), IEC 60601-2-37 (2015), IEC 62304 (2015), IEC 62359 (2017) and ISO 10993-1 (2018) standards.
## 15. TESTING
Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met. Additional performance testing included in the submission was conducted in order to demonstrate that the requirements for the new and improved features and functionality were met. The results of all these studies demonstrate that the subject devices meet established specifications and perform as intended and in accordance with labeling.
FDA guidance document “Marketing Clearance of Diagnostic Ultrasound Systems and Transducers”, issued February 21, 2023, was referenced for this submission, and software documentation appropriate for the Basic Documentation Level, per the FDA guidance document, “Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023, was included in this submission.
Additionally, cybersecurity documentation, per the FDA guidance document “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”, issued on June 27, 2025, was included in this submission.
Testing of this device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and Ultrasound systems.
## Validation of improvements to existing features
### SMI Angio Mode
SMI Angio Mode is introduced as an improvement to the existing SMI feature, to emphasize the separation of adjacent blood flows. This improved functionality was tested using phantom data
{7}
and clinical images acquired from three volunteers. All prespecified performance criteria were achieved, confirming that the improved SMI Angio Mode performs as intended and emphasizes the separation of adjacent blood flow compared with the predicate function.
## Auto Tune
Auto Tune is introduced as an improvement to the existing Shear wave elasticity (SWE) and Shear wave dispersion (SWD) features, to enable a pre-scan which optimizes scan conditions in advance of the subsequent SWE or SWD main scan to achieve more accurate SWE results. This study utilized phantom data as well as images of the thyroid and liver acquired from three volunteers. All prespecified performance criteria were achieved, confirming that Auto Tune performs as intended.
## Precision+ Fine Processing Mode
Precision+ Fine Processing Mode is an improvement in the functionality of the existing Precision+ feature, to improve the depiction of borders in an image. This test compared clinical images acquired using a standard preset, to which both the Precision+ Fine Processing Mode and predicate functionality, Precision+ were applied. The images were acquired from anatomy representative of the intended use of Precision+ Fine Processing Mode and in all cases, improved visibility of tissue structure was demonstrated when using Precision+ Fine Processing Mode, confirming it achieves its purpose as an improvement of the existing Precision+ feature.
## Validation of new DL based feature, 3rd Harmonic Imaging (3-HI)
This clinical evaluation was conducted in order to demonstrate improvement relative to predicate functionality (conventional 2nd harmonic imaging), with respect to spatial resolution, contrast resolution, and artifact suppression. The evaluation was conducted using abdominal images acquired from 30 patients from a U.S. clinical site. The validation data set was entirely independent of the data set used to train the algorithm during its development. The algorithm was locked upon completion of development, prior to implementation into the 3-HI feature and no post-market, continuous learning capability is deployed on the subject device. Inclusion criteria specified adult patients clinically scheduled for abdominal or pelvic imaging, or patients with a prior diagnosis of focal pathologies of the abdomen or pelvis. Images of representative organs and anatomical structures and representative focal pathologies were included in the study to simulate expected clinical variation encountered by this feature during clinical use. Patients were selected to ensure diverse demographic characteristics representative of the intended U.S. patient population, including a wide range of body mass indices. Three (3) U.S. board certified radiologists participated in a blinded observer study, in which images with 3-HI were compared to images without 3-HI (predicate functionality) using a 5-point ordinal scale. The median score was compared with the middle score of 3. A statistical analysis of the results demonstrated that scores for 3-HI were higher than the middle score of 3 with respect to spatial resolution, contrast resolution, and artifact suppression.
A sub-group analysis was conducted in order to demonstrate the generalizability of the evaluated performance for the intended U.S. patient population, confirming improvement of 3-HI compared with predicate functionality.
Page 4 of 5
{8}
Validation details:
| Demographic distribution | This study included representative images from 30 patients selected from among previously acquired data • Gender: roughly equivalent number of males and females • Age: Ranging from 23-89 years old • Ethnicity (Country): USA • BMI: ranging from 18.5-36.3 kg/m2, equally distributed across underweight or healthy, overweight, and obese categories • Images were collected of various representative abdominal organs and anatomical structures, as well as various representative focal pathologies |
| --- | --- |
| Data collection | Minimum 294 images from demographically diverse patients acquired at a U.S. clinical site |
| Evaluation Method | Three (3) board certified U.S. radiologists participate in a blind scoring and visibility evaluation in accordance with Canon's study design |
This clinical evaluation was supplemented by phantom studies which examined the lateral and axial resolution, slice resolution, contrast-to-noise ratio (CNR), reverberation artifact suppression and frequency spectra aspects of image quality. Five abdominal phantoms with various physical properties including targets at depths ranging from 3cm to 11.5cm, various target sizes from pin targets to 10mm targets and targets with high and low contrast were used for the evaluation of lateral and axial resolution, slice resolution and CNR improvements by 3-HI. In passing all prespecified performance criteria, these phantom studies demonstrated the capability of 3-HI to visualize abdominal images better than conventional 2nd harmonic imaging with respect to lateral and axial resolution, slice resolution, CNR, reverberation artifact suppression and frequency spectra aspects of image quality, as well as support the overall finding of improved performance by 3-HI compared to predicate for its intended use.
# 16. CONCLUSION
The Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700, V9.0 are substantially equivalent to the Aplio i900/i800/i700, Diagnostic Ultrasound System, V8.5, K242808. The subject devices function in a manner similar to, and are intended for the same use as the predicate devices, as described in labeling. The evidence provided in this submission demonstrate that Aplio i900/i800/i700, Diagnostic Ultrasound System, V9.0 are safe and effective for their intended use and perform with substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.