← Product Code [HBE](/productcode/HBE) · K252060

# GREEN (K252060)

_Phasor Health, LLC · HBE · Oct 31, 2025 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252060

## Device Facts

- **Applicant:** Phasor Health, LLC
- **Product Code:** [HBE](/productcode/HBE.md)
- **Decision Date:** Oct 31, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4310
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Phasor GREEN™ Drill is a re-sterilizable drill driver (for use on one or more patients, up to 10 total holes), connected to a separately-packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial bone.

## Device Story

Phasor GREEN Drill is a battery-operated, re-sterilizable drill driver used with a single-use, gamma-sterilized drill bit assembly. Operated by qualified users in a clinical setting, the device drills cranial bone. The driver is powered by non-rechargeable lithium batteries and features forward/reverse buttons. The driver is provided non-sterile and must be sterilized by the user facility using Vaporized Hydrogen Peroxide (VHP) before each use, up to a maximum of 10 uses. The drill bit assembly is single-use. The device provides mechanical drilling power to assist surgeons in neurosurgical procedures. Clinical benefit is derived from the device's ability to perform cranial bone drilling safely and effectively, comparable to existing predicate technology.

## Clinical Evidence

Bench testing only. Performance testing confirmed the device successfully drilled 10 holes per driver/bit assembly, consistent with the predicate device's performance. Sterilization validation was conducted according to ISO 22441 for VHP and ISO 11607-1 for the drill bit assembly.

## Technological Characteristics

Powered by non-rechargeable lithium batteries (18V). Drill bits are 440C steel (6.35-mm or less). Driver housing is plastic. Sterilization: VHP (ISO 22441) for the driver; Gamma irradiation (ISO 11607-1) for the drill bit assembly. Mechanical activation via forward/reverse buttons.

## Regulatory Identification

Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.

## Predicate Devices

- Phasor Drill ([K161704](/device/K161704.md))

## Reference Devices

- Stryker Reprocessed Pulse Oximeter Sensor ([K211138](/device/K211138.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 31, 2025

Phasor Health, LLC
Ray King
Chief Executive Officer
8944 Kirby Drive
Houston, Texas 77054

Re: K252060
Trade/Device Name: Green
Regulation Number: 21 CFR 882.4310
Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories
Regulatory Class: Class II
Product Code: HBE
Dated: October 2, 2025
Received: October 2, 2025

Dear Ray King:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}

K252060 - Ray King
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252060 - Ray King
Page 3

Sincerely,

Adam D. Pierce -S
Digitally signed by Adam D. Pierce
Date: 2025.10.31
07:20:38 -04'00'

Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252060  |   |
|  Device Name GREEN  |   |
|  Indications for Use (Describe) Phasor GREEN Drill is a re-sterilizable drill driver (for use on one ore more patients, up to 10 total holes), with separately packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial bone.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K252060, Page 1 of 4

# Phasor GREEN™

A. Device Description

|  Category | Comments  |
| --- | --- |
|  Sponsor | Phasor Health, LLC, Ray King, 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234  |
|  Correspondent Contact Information | Ray King 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234  |
|  Device Common Name | Powered simple cranial drills, burrs, trephines, and their accessories  |
|  Device Regulation & Name | 21 CFR 882.4310  |
|  Classification & Product Code | Class II, HBE  |
|  Device Proprietary Name | PHASOR GREEN  |

# Predicate Device(s) Information:

Legally Marketed reference device used to support substantial equivalence in terms of cleaning and sterilizing the equipment using Vaporized Hydrogen Peroxide (VHP).

|  Predicate Device(s)/ Reference Devices(s) | Phasor Drill / Stryker Reprocessed Pulse Oximeter Sensor  |
| --- | --- |
|  Predicate Device/ Reference device Manufacturer(s) | (1) Phasor Health, LLC, (2) Stryker Sustainability Solutions  |
|  Predicate Device Common Name / Reference Device Common Name | (1) Powered simple cranial drills, burrs, trephines, and their accessories; (2) Oximeter  |
|  Predicate Device Notification # / Reference Device Notification # | (1) K161704; (2) K211138  |
|  Predicate Device Classification & Name; Reference Device Classification & Name | (1) 21 CFR 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories; (2) 21 CFR 870.2700 Oximeter.  |
|  Predicate Device Classification & Product Code / Reference Device Classification & Name | Class II for each of (1) HBE & (2) NLF, respectively  |

B. Date Summary Prepared: October 2, 2025

# C. Description of Device

The Phasor™ GREEN Drill is composed of 2 items: (1) a re-sterilizable, non-rechargeable battery-operated Phasor Green Driver (made of plastic housing, capable of drilling up to 10 total holes in one or more patients) in conjunction with (2) a separately packaged, single-use Phasor Green Drill Bit Assembly comprised of a steel bit (of 6.35-mm or less), plastic adapter, and polybag secured using latex-free bands. The device is for drilling cranial or orthopedic bone, by prescription only and used by qualified users, with (1) Drill Driver (provided non-sterile) for

{5}

K252060, Page 2 of 4

sterilization using vaporized hydrogen peroxide (VHP) prior to use at user facility and (2) Drill Bit Assembly provided gamma-sterilized for single-use respectively.

## D. Indications for Use

Phasor GREEN™ Drill is a re-sterilizable drill driver (for use on one or more patients, up to 10 total holes), connected to a separately-packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial bone.

## E. Comparison of the Technological Characteristics

|   | Application Device | Predicate Device | Reference Device | Impact on Substantial Equivalence  |
| --- | --- | --- | --- | --- |
|  Company | Phasor Health, LLC | Phasor Health, LLC | Stryker | N/A  |
|  Regulation Number | 21 CFR 882.4310 | 21 CFR 882.4310 | 21 CFR 870.2700 | No impact  |
|  Product Code | HBE (Class II) | HBE (Class II) | NLF (Class II) | No impact  |
|  Intended Use & Indications for Use | Phasor GREEN™ Drill is a re-sterilizable drill driver (for use on one or more patients, up to 10 total holes), with separately packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial or orthopedic bone. | Phasor GREEN™ Drill is a re-sterilizable drill driver (for use on one or more patients, up to 10 total holes), with separately packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial or orthopedic bone. | The sensor is indicated for single patient use for continuous noninvasive arterial oxygen saturation and pulse rate monitoring. | Comparable to predicate device.  |
|  Technology  |   |   |   |   |
|  Driver Power Source | Single-use Lithium non-rechargeable Batteries | Single-use Lithium non-rechargeable Batteries; | No onboard battery or power source; contains sensors. | Same as predicate device  |
|  Performance Characteristics | Drills bone | Drills bone | optical sensor transmits light to photodetector that detects signal transmission. The signal transmission is processed by the Pulse Oximeter to provide SpO2 and pulse rate. | comparable performance to predicate device  |
|  Drill bit | 440C | 440C | N/A | Same as predicate device  |
|  Activation buttons | Forward/Reverse button on housing | Forward/Reverse button for driver | N/A no activation required | comparable to reference device  |

{6}

K252060, Page 3 of 4

|  Biocompatibility | Yes | Yes | N/A | Same as predicate device  |
| --- | --- | --- | --- | --- |
|  Electrical components including connections, wiring, switches, etc. | Motor, wiring, on/off switch, forward/reverse switch, batteries 18V | Motor, wiring, on/off switch, forward/reverse switch, batteries 18V | N/A | Identical  |
|  Sterility Status at Shipping | (1) Drill Driver shipped non-sterile to user's facility, to be sterilized using VHP by user prior to initial and each subsequent use (2) Drill Bit Assembly shipped sterile (using gamma sterilization) to user | Sterile | Sterile | comparable to reference device  |
|  Sterilization Method | (1)VHP for the Drill Driver at user facility prior to each use; (2) Gamma irradiation for Drill Bit Assembly | Gamma irradiation | Vaporous Hydrogen Peroxide | comparable to reference device  |
|  Cleaning and Sterilization Testing Standard(s) Met | (1) ISO 22441 for VHP for Drill driver (2) ISO 11607-1 for Drill Bit Assembly | ISO 11607-1 | ISO 17665-1 (utilizing VHP, but noting K211138 preceded the ISO 22441 release date of August 26, 2022) | Met testing to noted standard for Application Device  |

## F. Summary of Supporting Data

Phasor GREEN™ has the same intended use (while being multi-use vs. single-use), material composition, manufacturing process, manufacturer, patient population, classification, and initial packaging as the predicate. Moreover, it has the same sterilization method as the reference device. Phasor Drill had already been utilized as a "quick-connect" per justification based upon documentation, compatible with bits of the same range. Otherwise, the subject and predicate devices are based upon the following technological elements, when the predicate and reference device are taken together:

* Drilling function into cranial bone in the same manner, with same drill bit diameter size ranges as supported for the predicate device;
* Both the GREEN DRILL BIT DRIVER and predicate Phasor Drill are sterilized in the same manner (gamma sterilization);
* Both the GREEN DRILL BIT DRIVER and predicate Phasor Drill are intended for single-use, and are disposable after the single use.

The following technical characteristics of the subject device differ from the predicate device:

* Phasor GREEN DRIVER is a multi-use (&lt;= 10 uses), compared to the single-use Phasor Drill;
* Phasor GREEN DRIVER is vaporized hydrogen peroxide (VHP) sterilized vs. Gamma sterilized for the Phasor Drill.

The proposed device does not differ from the predicate and reference device taken together in any manner which would negatively impact the safety or effectiveness of the device.

## G. Discussion of Performance Testing

{7}

K252060, Page 4 of 4

Post-respective sterilization using VHP and gamma for the GREEN Driver and Drill Bit Assembly, respectively, performance testing of 10 holes with each Driver and bit were accomplished within the time period needed, similar to the predicate Phasor Drill.

## H. Conclusion

The performance and design validation testing conducted on Phasor GREEN™ device on the bench demonstrated that it performs comparably to the stated predicate and reference device that are currently marketed individually for the similar intended use. The proposed device, comprised of the elements noted above, demonstrates that it should perform as safely and as effectively as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K252060](https://fda-staging.innolitics.com/device/K252060)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
