Images from multiple facilities; included model pre-training, model training to the disease target, architecture optimization, threshold determination, and validation.
—
—
—
Indications for Use
Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD. Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The results of Fibresolve are intended to be used only by clinicians qualified in the care of lung disease, specifically in caring for patients with ILD, in conjunction with the patient’s clinical history, symptoms, and other diagnostic tests, as well as the clinician’s professional judgment. The input to Fibresolve must include a DICOM-compliant lung CT scan. Clinical case eligibility includes the following criteria: - Age > 22 years old. - Pulmonary symptoms suggestive of possible ILD including IPF.
Device Story
Fibresolve is a cloud-based software-only device for qualitative assessment of DICOM-compliant chest CT scans. It functions as an adjunct tool for clinicians specializing in ILD to assist in diagnosing IPF. The system consists of an Image Receiver API, an Ingestion Pipeline, and an Output API. The Ingestion Pipeline validates and prepares CT series; the Analysis System uses an ensemble 3D deep learning model to generate a binary classification result regarding the target disease state. The device does not perform image segmentation or provide visual annotations. Results are transmitted to the clinician via EHR integration or human-readable report. The output serves as adjunctive information for multidisciplinary discussions, potentially reducing the need for invasive testing. The device includes a Predetermined Change Control Plan (PCCP) allowing for future updates to model architecture, training data, and input types, provided performance remains non-inferior.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by software verification and validation performed per IEC 62304 and the inclusion of a PCCP for algorithm lifecycle management.
Technological Characteristics
Software-only device; cloud-hosted; utilizes HTTP/1.1 REST APIs for data transmission. Core technology is an ensemble 3D deep learning model. Does not perform image segmentation or visual annotation. Compliant with DICOM standards for image input. Includes a PCCP for controlled algorithm updates.
Indications for Use
Indicated for patients > 22 years old with pulmonary symptoms suggestive of interstitial lung disease (ILD), including idiopathic pulmonary fibrosis (IPF), to provide diagnostic subtype classification as an adjunct to clinical assessment and multidisciplinary discussion.
Regulatory Classification
Identification
Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD. Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing.
Special Controls
In combination with the general controls of the FD&C Act, radiology software for referral of findings related to fibrotic lung disease is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 7, 2025
Imvaria, Inc.
% Dulciana Chan
Principal Consultant
RQM+
2790 Mosside Blvd
Monroeville, Pennsylvania 15146
Re: K252041
Trade/Device Name: Fibresolve (with PCCP)
Regulation Number: 21 CFR 892.2085
Regulation Name: Radiology Software For Referral Of Findings Related To Fibrotic Lung Disease
Regulatory Class: Class II
Product Code: QWO
Dated: October 10, 2025
Received: October 14, 2025
Dear Dulciana Chan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252041 - Dulciana Chan
Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K252041 - Dulciana Chan
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252041
Device Name
Fibresolve (with PCCP)
Indications for Use (Describe)
Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD. Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The results of Fibresolve are intended to be used only by clinicians qualified in the care of lung disease, specifically in caring for patients with ILD, in conjunction with the patient’s clinical history, symptoms, and other diagnostic tests, as well as the clinician’s professional judgment.
The input to Fibresolve must include a DICOM-compliant lung CT scan. Clinical case eligibility includes the following criteria:
- Age > 22 years old.
- Pulmonary symptoms suggestive of possible ILD including IPF.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
IMVARIA
510(k) Summary
{5}
IMVARIA
# K252041 510(k) Summary
## DATE PREPARED
November 7, 2025
## MANUFACTURER AND 510(k) OWNER
IMVARIA, Inc.
2930 Domingo Ave #1496
Berkeley, CA 94705
Telephone: (650) 683-9800
Official Contact: Joshua Reicher, MD
CEO
## REPRESENTATIVE/CONSULTANT
Dulciana D. Chan, MSE
RQM+
2790 Mosside Blvd
Monroeville, PA 15146
Telephone: +1 (412) 816-8253
Email: dchan@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: Fibresolve (with PCCP)
Common Name: Radiology software for referral of findings related to fibrotic lung disease
Regulation Number: 21 CFR 892.2085
Class: II
Product Code: QWO
Review Panel: Radiology
## PREDICATE DEVICE IDENTIFICATION
Fibresolve (with PCCP) is substantially equivalent to the following predicate:
| 510(k) Number | Device/Manufacturer | Predicate/Reference |
| --- | --- | --- |
| DEN220040 | Fibresolve / Imvaria, Inc | Predicate |
## DEVICE DESCRIPTION
Fibresolve is a software system developed for qualitative disease assessment of DICOM-compliant chest computed tomography (CT) imaging. The software system is based on a machine learning model component and a [Docker](https://docs.docker.com/get-docker/) based HTTP/1.1 Representational State Transfer (REST) software application programming interfaces (APIs) to enable image transfer, analysis, and output of results.
{6}
IMVARIA
The device consists of the following 3 components: (1) Image Receiver API for image acquisition in the cloud; (2) Ingestion Pipeline and Analysis System for image processing and analysis; and (3) Output API for device output transmission.
(1) The Image Receiver API is accessed via any DICOM-compliant system (e.g. PACS). The hospital or clinic either accesses the API directly via secure software integration and submits the images electronically; or the images are transmitted manually (e.g. by mail) to the device manufacturer and the case is submitted to the device through the API directly by the manufacturer. The API passes the images to the Ingestion Pipeline and Analysis System (2).
(2) (a) The Ingestion Pipeline and (b) Analysis System accept the images, select cases appropriate for processing, process the images for analysis, and analyze the images. This Analysis System includes the Fibresolve Model Inference Graph that generates the assessment for the case. The final device output report data including identifying information and technical details about the case data and a binary result stating whether the data are determined to be suggestive for the target disease state.
- The Ingestion Pipeline identifies applicable CT imaging series from the case and verifies that the series is valid, completes quality checks, and confirms adequacy for analysis.
- The Fibresolve Model Inference Graph, the core component of the Analysis System, is an ensemble 3D deep learning model developed and trained using images from multiple facilities. Analysis System algorithm development phases included model pre-training, model training to the disease target, architecture optimization, threshold determination, and validation. No segmentation is performed as part of the Analysis System.
(3) The Output API transmits the Report data for the clinician to review. The Output API is either integrated into the hospital or clinic notification software (e.g. electronic health record) for electronic transmission or the device manufacturer transmits the Report in human-readable format directly (e.g. via fax). The clinician then incorporates the device Report as part of diagnostic decision-making.
The system does not include an image viewer or visual output for diagnostic use. The source images are reviewed for subjective assessment prior to submission to the device using the facility's standard diagnostic viewer as part of routine standard-of-care and the source images can be reassessed by the clinical team at any time before or after submission of the case to the device. The system only assesses for the target disease described in the Intended Use and does not replace imaging interpretation generally.
{7}
IMVARIA
# INDICATIONS FOR USE
Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD.
Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The results of Fibresolve are intended to be used only by clinicians qualified in the care of lung disease, specifically in caring for patients with ILD, in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.
The input to Fibresolve must include a DICOM-compliant lung CT scan. Clinical case eligibility includes the following criteria:
- Age > 22 years old.
- Pulmonary symptoms suggestive of possible ILD including IPF.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Fibresolve (with PCCP) includes an update to the original Fibresolve device to include a Predetermined Change Control Plan (PCCP) for the Fibresolve device, including an Algorithm Change Protocol (ACP) for the underlying Fibresolve Analysis Algorithm. The PCCP enables updates to the underlying Analysis Algorithm.
{8}
IMVARIA
A table comparing the key features of the subject and predicate device is provided below.
| | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| | Fibresolve (with PCCP)K252041 | FibresolveDEN220040 | - |
| Indications for Use | Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD. Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The results of Fibresolve are intended to be used only by clinicians qualified in the care of lung disease, specifically in caring for patients with ILD, in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.The input to Fibresolve must include a DICOM-compliant lung CT scan. Clinical case eligibility includes the following criteria: | Fibresolve is a software-only device that receives and analyzes lung computed tomography (CT) imaging data in order to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). The device supplements the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD) or as part of an MDD. Specifically, the tool is used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The results of Fibresolve are intended to be used only by clinicians qualified in the care of lung disease, specifically in caring for patients with ILD, in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.The input to Fibresolve is a DICOM-compliant lung CT scan. Clinical case eligibility includes the following criteria: | Same |
{9}
#
IMVARIA
| | Age > 22 years old.
Pulmonary symptoms suggestive of possible ILD including IPF. | Age > 22 years old.
Pulmonary symptoms suggestive of possible ILD including IPF. | |
| --- | --- | --- | --- |
| User population | Clinicians qualified in the care of lung disease, specifically in caring for patients with ILD | Clinicians qualified in the care of lung disease, specifically in caring for patients with ILD | Same |
| Target Population | Age > 22 years old. | Age > 22 years old. | Same |
| Anatomical region of interest | Chest | Chest | Same |
| Scan type and protocol | DICOM-compliant lung CT scan | DICOM-compliant lung CT scan | Same |
| Software Inputs | CT scan
Potentially age, sex, and pulmonary function tests (PFTs) | CT scan | Similar |
| Segmentation of region of interest | No; device does not mark, annotate, or direct users’ attention to a specific location in the original image | No; device does not mark, annotate, or direct users’ attention to a specific location in the original image | Same |
| Algorithm | Machine learning pattern recognition | Machine learning pattern recognition | Same |
| Alteration of original image | No | No | Same |
| Data Displayed | Qualitative classification output of imaging findings
Inputs that contribute to the diagnostic result | Qualitative classification output of imaging findings | Similar |
{10}
IMVARIA
# SUMMARY OF NON-CLINICAL TESTING
Software Verification and Validation (per IEC 62304) were performed to demonstrate safety based on current industry standards. The results of these tests indicate that the subject device is equivalent to the predicate device.
# PREDETERMINED CHANGE CONTROL PLAN
Fibresolve includes a predetermined change control plan (PCCP), detailing the specific modifications (SaMD Pre-Specifications (SPS)) that may be made to the device and the specific methods in place to achieve and appropriately control the risks of the anticipated types of modifications (Algorithm Change Protocol (ACP)). The ACP outlines the process for data management, model re-training, performance evaluation, and update procedures associated with the change. The plan allows for modifications and updates to the underlying Analysis Algorithm within a limited scope of changes, specifically model architecture modification, introduction of new training data, and incorporation of ancillary inputs (while maintaining original inputs). Changes are evaluated via pre-specified statistical analyses in-line with those as part of the original device testing, to ensure, at minimum, non-inferior absolute performance, and potential improvements in performance, training data, or generalizability.
{11}
IMVARIA
| Modification | Rationale | Testing Methods | Impact Assessment |
| --- | --- | --- | --- |
| Model architecture modification | With advancements in underlying model architecture through new computer science developments, new model training allows for improvements in performance in terms of overall metrics as well as consistency and speed. | Substantial equivalence as compared to the prior version.
Statistical assessments following same standards used in original device clearance. | Revised generalizability or discriminatory capacity metrics for the system.
**Benefit-Risk Analysis:**
Benefit: Enhanced performance; generalizability.
Risk: Reduction in clinical performance or generalizability.
**Risk Mitigation:**
Evaluate device model on Test dataset metrics.
Execute unit and integration tests for the product code. |
| Introduction of new training data | With new training data, new model training allows for improvements in generalizability, robustness, and reductions in biases, which provides greater | Substantial equivalence as compared to the prior version.
Statistical assessments following same standards used in original device | Revised sensitivity and specificity metrics for the system.
**Benefit-Risk Analysis:**
Benefit: Enhanced performance; generalizability. |
{12}
IMVARIA
| | clinical value. Performance improvements may also be achieved. | clearance. | Risk: Reduction in clinical performance or generalizability.
Risk Mitigation: Evaluate device model on Test dataset metrics. Execute unit and integration tests for the product code. |
| --- | --- | --- | --- |
| Incorporation of ancillary inputs | Addition of non-imaging ancillary inputs into the model to improve overall performance with model component contributions displayed with the final result. improvements enhance performance and overall clinical value. | Substantial equivalence as compared to the prior version. Statistical assessments following same standards used in original device clearance | Revised sensitivity and specificity metrics for the system.
Benefit-Risk Analysis: Benefit: Enhanced performance; generalizability; explainability. Risk: Reduction in clinical performance or generalizability.
Risk Mitigation: Evaluate device model on Test dataset metrics. Execute unit and integration tests for the product code. |
{13}
IMVARIA
# CONCLUSION
The subject device and predicate devices are both intended to receive and analyze lung computed tomography (CT) imaging data to provide a diagnostic subtype classification in suspected cases of interstitial lung disease (ILD). Both devices supplement the standard-of-care workflow by providing a qualitative, diagnostic classification output of imaging findings based on machine learning pattern recognition, in order to provide adjunctive information as part of a referral pathway to an appropriate Multidisciplinary Discussion (MDD). Both devices are used to serve as an adjunct in the diagnosis of idiopathic pulmonary fibrosis (IPF) prior to invasive testing. The subject device incorporates a predetermined change control plan (PCCP) where changes are evaluated via pre-specified analyses to ensure non-inferior absolute performance ensuring the device performs as intended in the specified use conditions and does not present any new issues of safety or effectiveness. Thus, Fibresolve (with PCCP) is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.