K252037 · Curexo, Inc. · OLO · Mar 25, 2026 · Neurology
Device Facts
Record ID
K252037
Device Name
CUVIS-joint (CJ150)
Applicant
Curexo, Inc.
Product Code
OLO · Neurology
Decision Date
Mar 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252037 · Mar 25, 2026
CUVIS-joint (CJ150)
Curexo, Inc.
Published clinical literature; Clinical study cohorts
A systematic literature review and meta-analysis were conducted to evaluate clinical performance (HKA angle, patient-reported outcomes) and safety (adverse event rates) of the CUVIS-joint and its predicate device in real-world orthopedic surgical practice.
Systematic literature review; Meta-analysis; Clinical performance; Safety outcomes; Total knee arthroplasty
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic Literature Review and Meta-analysis; Systematic review and meta-analysis of clinical studies; Study Period: Literature search conducted September 2025
Patients undergoing total knee arthroplasty (TKA); Sample Size: >2,000 patients (total across review); 656 patients (meta-analysis subset)
Conventional total knee arthroplasty (TKA) procedures
The CUVIS-joint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with preoperative planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan. The preoperative planning software and robotic surgical tool are used as an alternative to manual planning and resecting techniques for the distal femur and proximal tibia preparation in primary total knee arthroplasty (TKA). The CUVIS-joint is indicated for orthopedic procedures in which resecting techniques used for the distal femur and proximal tibia may be considered to be safe and effective and where references to rigid anatomical structures may be made. The CUVIS-joint is also intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The CUVIS-joint facilitates accurate positioning of TKA implants, relative to these alignment axes. The CUVIS-joint is compatible with the following Knee Implant System: - United U2 Knee System
Device Story
CUVIS-joint is a robotic-assisted surgical system for primary total knee arthroplasty (TKA). It consists of a 3D preoperative planning workstation and an electromechanical robotic arm. Input: preoperative CT scans. Process: surgeon uses workstation to plan implant position/orientation relative to 3D bone models; robotic arm executes bone milling based on the plan. Intraoperative tracking: optical tracking arrays attached to bones allow real-time motion recovery relative to the robot base. Registration: pinless point-to-surface landmark collection. Operation: surgeon-directed; includes pendant control and optional foot switch. Output: precise bone preparation for TKA implant placement. Benefit: improved implant positioning accuracy and limb alignment compared to manual techniques. Used in clinical settings by orthopedic surgeons.
Clinical Evidence
Systematic literature review (PRISMA guidelines) of 22 studies, including a meta-analysis of 4 studies (656 patients). Primary endpoint: Hip-Knee-Ankle (HKA) angle. Results: mean HKA angle within clinically acceptable range of ±3°. Secondary outcomes: improved patient-reported scores (KSS, WOMAC, OKS, AKSS, FJS, SF-12). Safety: adverse event rate <1%, with rare device-related complications. Evidence supports equivalent safety and effectiveness to predicate and conventional TKA.
Technological Characteristics
System includes 3D planning workstation, electromechanical robotic arm, optical tracking system, display monitor, and pendant control. Materials have long history of orthopedic use. Registration: pinless point-to-surface. Tracking: optical arrays. Connectivity: networked/integrated system components. Sterilization: standard surgical protocols. Software: preoperative planning and robotic control algorithms.
Indications for Use
Indicated for patients requiring primary total knee arthroplasty (TKA) where distal femur and proximal tibia resection is required and reference to rigid anatomical structures is possible.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 25, 2026
Curexo, Inc.
Jay Koh
Regulatory Affairs Manager
3rd 4th FL. 480, Wiryesunhwan-ro, Songpa-gu
Seoul, 05814
Republic Of Korea
Re: K252037
Trade/Device Name: CUVIS-joint (CJ150)
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: February 22, 2026
Received: February 23, 2026
Dear Jay Koh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252037 - Jay Koh
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K252037 - Jay Koh
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K252037
Device Name
CUVIS-joint (CJ150)
Indications for Use (Describe)
## Indications for Use
The CUVIS-joint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with preoperative planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.
The preoperative planning software and robotic surgical tool are used as an alternative to manual planning and resecting techniques for the distal femur and proximal tibia preparation in primary total knee arthroplasty (TKA).
The CUVIS-joint is indicated for orthopedic procedures in which resecting techniques used for the distal femur and proximal tibia may be considered to be safe and effective and where references to rigid anatomical structures may be made.
The CUVIS-joint is also intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The CUVIS-joint facilitates accurate positioning of TKA implants, relative to these alignment axes.
The CUVIS-joint is compatible with the following Knee Implant System:
- United U2 Knee System
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1.
{4}
CUREXO ROBOTICS BEYOND CORE
CUVIS-joint Traditional 510(k) Submission
# 510(k) SUMMARY K252037
## Applicant Information:
Owner Name: CUREXO, Inc.
Address: 3rd, 4th, 480 Wiryesunhwan-ro, Songpa-gu, Seoul, 05814, Republic of Korea
Phone number: +82-2-6202-8500
Fax number: +82-2-3446-4498
Establishment Registration Number: 3009522766
Contact Person: Jay Koh
Date Prepared: June 30, 2025
## Device Information:
Device Classification: Class II
Trade Name: CUVIS-joint
Common name: Orthopedic Stereotaxic Instrument
Classification name: Stereotaxic Instrument
Regulation number: 882.4560
Product Code: OLO
## Predicate Device:
The CUVIS-joint is substantially equivalent in intended use, materials, fundamental scientific technology and performance to the following legally marketed device in commercial distribution: THINK Surgical, Inc’s TSolution One Total Knee Application cleared via K203040 and cleared via K191369.
## Device Description:
The CUVIS-joint is a three-dimensional, graphical, preoperative planner and implementation tool for treatment of patients who require total joint arthroplasty. The device is intended as an alternative to manual template planning and preparation of the bone with patients requiring primary total knee arthroplasty (TKA).
The CUVIS-joint consists of two main components: 1) the Planner which is a three-dimensional (3D) preoperative planning workstation which uses preoperative CT scans of patient anatomy to
510(k) Summary
{5}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint Traditional 510(k) Submission
aid a surgeon in planning the position and orientation of the implant components relative to 3D models of the patient's anatomy, and 2) the Robotic Arm Tool which executes the surgical plan and includes a an electromechanical arm, an arm base including control electronics and computer, optical tracking system, a display monitor, operating software, pendant control, and tools and accessories. When used according to the Instructions for Use (IFU), the CUVIS-joint enables precise implant positioning possible before and during TKA surgical procedures.
# Intended Use / Indications for Use:
The CUVIS-joint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with preoperative planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.
The preoperative planning software and robotic surgical tool are used as an alternative to manual planning and resecting techniques for the distal femur and proximal tibia preparation in primary total knee arthroplasty (TKA).
The CUVIS-joint is indicated for orthopedic procedures in which resecting techniques used for the distal femur and proximal tibia may be considered to be safe and effective and where references to rigid anatomical structures may be made.
The CUVIS-joint is also intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The CUVIS-joint facilitates accurate positioning of TKA implants, relative to these alignment axes.
The CUVIS-joint is compatible with the following Knee Implant Systems:
- United U2 Knee System
# Substantial Equivalence:
Both the CUVIS-joint, and the predicate device, the TSolution One Total Knee Application, have the same intended use. Both are intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.
The Indications for Use of the CUVIS-joint and its predicate are identical, with the exception of the total joint implant systems with which each is compatible.
There are no significant differences in the patient contacting materials between the CUVIS-joint and the predicate, TSolution One Total Knee Application cleared via K203040. Both use materials that have had a long history of use in orthopaedic medical devices.
K252037, Page 2 of 7
510(k) Summary
{6}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint
Traditional 510(k) Submission
Both the CUVIS-joint and the predicate device have operating principles and technological characteristics which are substantially equivalent. Both devices use CT images of the patient’s anatomic structure and both use a three-dimensional (3D) preoperative planning workstation that aids a surgeon in planning the position and orientation of the implant components relative to 3D models of the patient's anatomy. Both devices use pinless point to surface registration of the femur and tibia by collecting landmarks on the bone surface to digitize the location of each bone.
Both devices use bone markers for intraoperative tracking of the position of the bones relative to the electromechanical arm and to recover from bone motion that may occur during the surgery. The CUVIS-joint uses tracking arrays that are optically tracked relative to the robot base, while the TSolution One Total Knee Application uses a mechanical motion detection arm attached from the robot base to each bone. Both use standard surgical exposures appropriate to the specific anatomic location and consistent with the exposures used in traditional manual surgical procedures.
Both devices have a display monitor, operating software, pendant control, and procedure specific tools and accessories for the implementation of the preoperative plan. Both devices use an electromechanical arm, an arm base including control electronics and computer which mills the bone to achieve the bone preparation required to accept the TKA implants. The CUVIS-joint allows the surgeon to make both pre-operative and intraoperative plan adjustments to the translation, rotation and size of the TKA components to support gap balancing.
Performance testing was performed on the CUVIS-joint to verify and validate that the performance of the system meets the user needs and that the device is substantially equivalent to the predicate. This testing included: Biocompatibility Testing, Full System Run Through Testing (Smoke Test), Cutting Accuracy Testing, Cadaver Validation Testing, Electrical Safety and EMC Testing, Software Testing and Usability Testing.
Substantial equivalence in technological characteristics and performance of the CUVIS-joint compared to the predicate device is outlined in the table below:
K252037, Page 3 of 7
510(k) Summary
{7}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint Traditional 510(k) Submission
| Product | CUVIS-joint | TSolution One Total Knee Application (Predicate Device) | TSolution One Total Knee Application (Predicate Device) | Conclusion |
| --- | --- | --- | --- | --- |
| 510(k) number | Subject Device | K203040 | K191369 | |
| Manufacturer | CUREXO, Inc. | THINK Surgical Inc | THINK Surgical Inc | |
| Materials | | | | |
| -Material Used | Uses materials with a long history of use in orthopedic procedures. | Uses materials with a long history of use in orthopedic procedures. | Uses materials with a long history of use in orthopedic procedures. | SAME. |
| Technological Characteristics | | | | |
| -Patient Imaging | CT Scan | CT Scan | CT Scan | SAME |
| -User Controls | Keyboard, mouse, Pendant with mechanically latched Stop Button, foot switch | Keyboard, mouse, Pendant with mechanically latched Stop Button | Keyboard, mouse, Pendant with mechanically latched Stop Button | Substantially Equivalent
(Both systems have similar user controls – CUVIS-joint adds an optional foot switch for use in collecting registration points) |
| -Preoperative planning workstation | Three-dimensional preoperative planning workstation | Three-dimensional preoperative planning workstation | Three-dimensional preoperative planning workstation | SAME |
| -Presurgical Plan | CT images used to create a 3D model of the bone, library of FDA cleared components used to develop implant size and location | CT images used to create a 3D model of the bone, library of FDA cleared components used to develop implant size and location | CT images used to create a 3D model of the bone, library of FDA cleared components used to develop implant size and location | SAME |
| -Surgical Plan Data | High level operative plan based on preoperative plan with predetermined control file developed to control the robotic arm. System allows pre-operative and intraoperative plan adjustment to the translation, rotation and size of the TKA components to support gap balancing. | High level operative plan based on preoperative plan with predetermined control file developed to control the robotic arm. System allows only pre-operative plan adjustment to the translation, rotation and size of the TKA components to support gap balancing. | High level operative plan based on preoperative plan with predetermined control file developed to control the robotic arm. System allows only pre-operative plan adjustment to the translation, rotation and size of the TKA components to support gap balancing. | Substantially Equivalent
(Both systems use a high level preoperative plan to control the robotic arm – CUVIS-joint allows for intraoperative plan adjustments) |
K252037, Page 4 of 7
510(k) Summary
{8}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint Traditional 510(k) Submission
| Product | CUVIS-joint | TSolution One Total Knee Application (Predicate Device) | TSolution One Total Knee Application (Predicate Device) | Conclusion |
| --- | --- | --- | --- | --- |
| 510(k) number | Subject Device | K203040 | K191369 | |
| -Surgical Exposure | Similar to traditional surgical exposure for the anatomic site | Similar to traditional surgical exposure for the anatomic site | Similar to traditional surgical exposure for the anatomic site | SAME |
| -Preparation of bones to accept TKA implants | -Electromechanical arm mills the bone to implement pre-surgical plan | -Electromechanical arm mills the bone to implement pre-surgical plan | -Electromechanical arm mills the bone to implement pre-surgical plan | SAME |
| -Intraoperative tracking of the position of the bones relative to the electromechanical arm | Uses an optical tracking array to track the position of the bones relative to the electromechanical arm | Uses a mechanical motion detection arm to track the position of the bones relative to the electromechanical arm | Uses a mechanical motion detection arm to track the position of the bones relative to the electromechanical arm | Substantially Equivalent
(Both systems use an intraoperative tracking system – CUVIS-joint uses optical tracking while TS1 uses a mechanical motion arm) |
| -Patient/Robot Registration | Pinless point to surface registration of the femur and tibia by collecting landmarks on the bone surface matched to the 3D bone model created from the CT scan | Pinless point to surface registration of the femur and tibia by collecting landmarks on the bone surface matched to the 3D bone model created from the CT scan | Pinless point to surface registration of the femur and tibia by collecting landmarks on the bone surface matched to the 3D bone model created from the CT scan | SAME |
| -Use of Markers Attached to the Bone for Motion Recovery | Uses Tracking Arrays attached to the bone which are optically tracked to detect motion. | Uses Bone Motion Recovery Markers attached to the bone which are tracked using the digitizing arm. | Uses Bone Motion Recovery Markers attached to the bone which are tracked using the digitizing arm. | Substantially Equivalent
(Both systems use bone markers for motion recovery – CUVIS-joint tracks them optically while TS1 tracks them with the digitizer arm) |
| -System is Compatible with specified Knee Implant Systems | Verification and Validation Testing of Individual Implant Systems is Provided to Support the use of specified Knee Implant Systems | Verification and Validation Testing of Individual Implant Systems is Provided to Support the use of specified Knee Implant Systems | Verification and Validation Testing of Individual Implant Systems is Provided to Support the use of specified Knee Implant Systems | SAME |
| Performance Testing | | | | |
| -Full System Run Through Testing (Smoke Test) | Passed | Passed | Passed | SAME |
K252037, Page 5 of 7
510(k) Summary
{9}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint Traditional 510(k) Submission
| Product | CUVIS-joint | TSolution One Total Knee Application (Predicate Device) | TSolution One Total Knee Application (Predicate Device) | Conclusion |
| --- | --- | --- | --- | --- |
| 510(k) number | Subject Device | K203040 | K191369 | |
| -Cutting Accuracy Verification | Passed | Passed | Passed | SAME |
| -Cadaver Lab Validation Testing | Passed | Passed | Passed | SAME |
| -Software Testing | Passed | Passed | Passed | SAME |
| -Electromagnetic Compatibility and Electrical Safety | Passed | Passed | Passed | SAME |
| -Biocompatibility Testing ISO 10993-1:2018 | Passed | Passed | Passed | SAME |
| -Usability Testing | Passed | Passed | Passed | SAME |
# Clinical Data:
Following PRISMA guidelines, a structured literature search was conducted in September 2025 to identify studies reporting key clinical performance and safety endpoints related to CUVIS-joint and its equivalent device, TSolution One. The review included data on implant positioning, mechanical axis correction, functional recovery, and adverse event profiles, with particular attention to adverse events rates and device-related complications. The collected evidence was analyzed to determine whether the devices meet established clinical safety criteria and demonstrate acceptable performance in orthopedic surgical practice.
The search identified 1,571 papers. These papers were evaluated sequentially based on predefined exclusion criteria. Studies were excluded if they were not relevant to the total knee arthroplasty (TKA) or if they did not meet methodological or clinical relevance standards. Only publications meeting the predefined criteria for study design, relevance, and contribution to the evaluation of clinical safety and performance were included in the final analysis.
Of the 22 clinical studies on the performance of CUVIS-joint and its equivalent device, a meta-analysis was conducted on 4 studies that reported the mean Hip-Knee-Ankle (HKA) angle as a quantitative surgical outcome. The combined analysis of 656 patients yielded a mean HKA angle within the clinically acceptable range of $\pm 3^{\circ}$ , which met the acceptance criteria of the analysis and suggested that the CUVIS-joint system achieves clinically acceptable alignment accuracy. Further analysis of selected studies demonstrated that both systems contributed to accurate implant positioning and limb alignment, improved postoperative patient reported outcomes (including: KSS, WOMAC, OKS, AKSS, FJS and SF-12), and consistent surgical outcomes.
K252037, Page 6 of 7
510(k) Summary
{10}
CUREXO ROBOTICS BEYOND CURE
CUVIS-joint Traditional 510(k) Submission
These findings support equivalent clinical efficacy of robotic-assisted TKA using CUVIS-joint to other comparable technologies and to those commonly reported for conventional total knee arthroplasty (TKA) procedures.
In terms of safety, the overall adverse event rate across the reviewed studies remained below 1%, with device-related adverse events being rare and typically resolved without further complications. Therefore, both CUVIS-joint and TSolution One can be considered comparable in terms of clinical safety and similar to results reported for conventional TKA procedures. Taken together, the evidence from more than 2,000 clinical cases reported in this systematic review supports the use of the CUVIS-joint system and its predicate device as equivalently safe and effective options for robotic-assisted TKA. The CUVIS-joint system demonstrates clinically meaningful benefits in both performance and safety in total knee arthroplasty (TKA).
## Conclusion
The CUVIS-joint is equivalent to the predicate, TSolution One Total Knee Application, cleared via K203040 in the following ways: it has the same intended use and Indications for Use and substantially equivalent technological characteristics and operating principles and incorporates the same basic design and similar materials. The results of performance testing demonstrate that the CUVIS-joint performed substantially equivalent to the predicate and did not raise any new questions of safety and effectiveness. Clinical data from the published literature further supports equivalent safety and effectiveness of the CUVIS-joint and demonstrated a positive risk/benefit for the indication for total knee arthroplasty. The data presented supports a determination of Substantial Equivalence.
K252037, Page 7 of 7
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.