K252031 · Adin Dental Implant Systems , Ltd. · DZE · Mar 23, 2026 · Dental
Device Facts
Record ID
K252031
Device Name
Adin Long Dental Implant System
Applicant
Adin Dental Implant Systems , Ltd.
Product Code
DZE · Dental
Decision Date
Mar 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252031 · Mar 23, 2026
Adin Long Dental Implant System
Adin Dental Implant Systems , Ltd.
Retrospective clinical records
Retrospective analysis of 37 patients to evaluate clinical performance and success rates of the Adin Long Dental Implant System (including angled multi-unit abutments) in a real-world clinical setting.
37 patients (10 females, 27 males; age range 45–75 years, mean 58.1 ± 8.2) with severe atrophic maxilla; Sample Size: 37 patients (36 analyzed)
Not applicable for this study
Implant and angled abutment (45°–60°) success at ≥ 6 months post-loading
Indications for Use
The Adin Long Dental Implant System is intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to restore the patient's chewing function in fully or partially edentulous patients. The implants are intended for multi-unit, full-arch, screw-retained prosthetic restorations and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Adin Touareg-OS Zygomatic Dental Implants are intended for placement in the zygomatic bone in patients with severe atrophic maxilla; they may be used with implant-level framework designs with 45° to 60° angulation.
Device Story
Adin Long Dental Implant System consists of threaded, root-form titanium alloy implants and screw-retained abutments (TMA). Implants are surgically placed through the maxillary arch into the zygomatic bone; used for extra-maxillary, extra-sinus procedures. System supports multi-unit, full-arch, screw-retained prosthetic restorations. Designed for one-stage or two-stage surgical operations; immediate loading is applicable if primary stability is achieved. Used by clinicians in dental surgery settings. Output is a stable foundation for artificial teeth, restoring chewing function. Benefits include enabling prosthetic restoration in patients with severe maxillary atrophy where standard implants may be insufficient.
Clinical Evidence
Retrospective real-world data (RWD) study of 36 patients (92 abutments, 72 with 60° angulation) with 6-22 months follow-up. Primary endpoint: implant and abutment success at ≥ 6 months. Results: 100% success rate; no device-related adverse events, infection, inflammation, or fistula reported. Subgroup analysis (≥ 12 months) also showed 100% success.
Indicated for fully or partially edentulous patients requiring prosthetic support in the upper jaw; specifically for patients with severe atrophic maxilla requiring zygomatic bone anchorage.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Noris Medical Zygomatic Dental Implants System (K151909)
Reference Devices
Noris Medical Zygomatic Dental Implants System's multi-unit abutment (K210356)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 23, 2026
Adin Dental Implant Systems Ltd.
% Tal Bresler Stramer
Consultant
OrbitRA
325 Betty Ann Drive
Toronto, M2R 1B4
CANADA
Re: K252031
Trade/Device Name: Adin Long Dental Implant System
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: February 18, 2026
Received: February 19, 2026
Dear Tal Bresler Stramer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252031 – Tal Bresler Stramer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252031 – Tal Bresler Stramer
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252031 | |
| Device Name Adin Long Dental Implant System | |
| Indications for Use (Describe) The Adin Long Dental Implant System is intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to restore the patient's chewing function in fully or partially edentulous patients. The implants are intended for multi-unit, full-arch, screw-retained prosthetic restorations and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Adin Touareg-OS Zygomatic Dental Implants are intended for placement in the zygomatic bone in patients with severe atrophic maxilla; they may be used with implant-level framework designs with 45° to 60° angulation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) SUMMARY FOR ADIN'S LONG DENTAL IMPLANT SYSTEM
Date Prepared: March 23, 2026
# 510(k) OWNER NAME
Adin Dental Implants Systems Ltd.
Alon Tavor Industrial Zone P.O. Box 1128, Afula 1811101, Israel
Phone: +972-4-642-6732, Fax: +972-4-642-6733
E mail: Dimitry@adin-implants.com
Contact person name: Tal Bresler-Stramer, RA consultant
OrbitRA Consulting
Phone: +1-647-544-2911
Email: tal@orbitra.ca
# DEVICE NAME
Common/Usual Name: Adin Long Dental Implant System
Proprietary/Trade name: Adin Long Dental Implant System
Classification: Adin Long Dental Implant System has been classified as Class II devices under the following classification names:
| Classification Name | Product Code | 21 CFR Ref. | Panel |
| --- | --- | --- | --- |
| Implant, Endosseous, Root-Form | DZE (Primary) | 872.3640 | Dental |
| Abutment, Implant, Dental, Endosseous | NHA (Secondary) | | |
# PREDICATE DEVICES
Adin's Long Dental Implant System is substantially equivalent to the following Predicate Devices:
- Primary Predicate Device: Noris Medical Zygomatic Dental Implants System, cleared under 510(k) number K151909 on April 8, 2016.
- Reference Device: Noris Medical Zygomatic Dental Implants System's multi-unit abutment, cleared under 510(k) number K210356 on January 2, 2022.
- Reference Device: Adin Short Implants, cleared under 510(k) number K212775 on June 8, 2022.
- Reference Device: NobelZygoma 0° abutments, cleared under 510(k) number K161598 on February 15, 2017.
- Reference Device: UniFit Dental Implant System, cleared under 510(k) number K223714 on December 29, 2023.
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
{5}
0 7 2
# DEVICE DESCRIPTION
Adin Long Dental Implant System is an extension of previously cleared dental implant devices, designed to expand the product offering and provide additional options for dental restoration procedures. Specifically, this submission seeks marketing clearance for the following implants and abutments:
| (1) Adin Touareg™-OS Zygomatic Dental Implants: | • Outer diameter of 4.2 mm
• Available Lengths: 35.0mm, 37.5mm, 40.0mm, 42.5mm, 47.5mm, 50.0mm, 52.5mm and 55.0mm. |
| --- | --- |
| (2) Adin Long Dental Implant System -RS TMA™: | • Connection: RS
• Diameter: 4.9 mm
• Available Configurations (Angle / Gingival Height):
• 45° / 4 mm
• 52° / 5 mm
• 60° / 5 mm |
# ADIN TOUAREG™-OS ZYGOMATIC DENTAL IMPLANTS:
Adin Touareg™-OS Zygomatic dental implants is a threaded, root-form dental implants made of biocompatible Ti-6Al-4V ELI alloy complying with ASTM F136-13(2021)e1 – “Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications” and ISO 5832-3:2021 “Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy”.
Adin Touareg™-OS Zygomatic Dental Implants are used for surgical placement through the maxillary arch and anchored in the zygomatic bone to provide support for prosthetic devices such as artificial teeth to restore the patient's chewing function in edentulous or partially edentulous patients only with severe atrophic maxilla.
Adin Touareg™-OS Zygomatic Dental Implants are intended for extra-maxillary extra-sinus procedures, and their use is limited to ZAGA-4 extra-maxillary implant placement.
The device is intended to be used under the following clinical conditions and surgical parameters:
- Multi-unit reconstruction (TMA), with a rigid splinting of minimum two implants.
- For an edentulous/full mouth restoration, should be used together with at least two standard implants.
- It should be used only in patients with severe atrophic maxilla.
- Extra-maxillary extra-sinus procedures intended to be anchored in the zygoma bone
- One stage or two stage surgical operation.
- Immediate loading (function) is applicable provided that stability requirements are satisfied.
All Touareg™-OS Zygomatic implants have OsseoFix™ surface treatment which was previously cleared under Adin’s 510(k) K212775 for Adin's legally marketed Adin Short Implants and remained unchanged since these clearances.
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
{6}
0 7 3
# ADIN TOUAREG™-OS ZYGOMATIC IMPLANTS – RS TMA™
The TMA™ system is indicated for multiple-unit, screw-retained restorations, and may be used in combination with an implant level framework design.
Adin Touareg™-OS Zygomatic implants are intended for TMA in combination with a rigid splinting of a minimum of 2 conventional dental implants.
Screw-retained abutments are made of titanium alloy 6Al-4V ELI complying with ASTM F136-13 and ISO 5832-3:2021 - Implants for surgery — Metallic materials — Part 3: Wrought titanium 6-aluminium 4-vanadium alloy.
The TMA system is used to elevate seating platform of restoration when restoration at implant level is not indicated or practical due to the depth or emergence angle of the implant.
The Adin Touareg™-OS Zygomatic Implants - RS Angled TMA™ (45°, 52°, 60°) are only intended for use with the proposed Adin Touareg™-OS Zygomatic Dental Implants.
# INTENDED USE
The Adin Long Dental Implant System is intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to restore the patient's chewing function in fully or partially edentulous patients. The implants are intended for multi-unit, full-arch, screw-retained prosthetic restorations and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Adin Touareg-OS Zygomatic Dental Implants are intended for placement in the zygomatic bone in patients with severe atrophic maxilla; they may be used with implant-level framework designs with 45° to 60° angulation.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
Adin Long Dental Implant System has the same intended use as the cleared predicate Noris Medical Zygomatic Dental Implants System, cleared under 510(k) number K151909. Both systems are intended to provide endosseous support for prosthetic devices to restore patient's chewing function in the upper jaw of edentulous or partially edentulous patients. Both devices are intended for extra-maxillary, extra-sinus surgical procedures and enable a one-stage or two-stage surgical operation. In addition, both systems support multi-unit, screw-retained restoration with 45° to 60° angulated abutments. The predicate device was cleared with a 45° abutment. The 52° and 60° angled abutment options were subsequently cleared under the Noris Medical Zygomatic Dental Implant System (K210356) reference device.
Both the subject and predicate devices utilize a Resorbable Blast Media (RBM) surface treatment applied to titanium implants to create a moderately rough surface intended to support osseointegration.
The subject device incorporates Adin’s proprietary OsseoFix™ calcium phosphate RBM surface treatment, which uses calcium phosphate-based resorbable media. This surface treatment has been previously cleared in other Adin implant systems (K212775).
# PERFORMANCE DATA (NON-CLINICAL)
Non-clinical data provided in this submission in support of a determination of substantial equivalence includes:
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Page 3
Revision 3.0
Adin Long Dental Implant System
{7}
0 7 2
# BIOCOMPATIBILITY:
Assessments regarding biological compatibility were performed according to ISO 10993-1 and to the FDA Guidance document “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process’, Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016”.
Representative samples of the worst-case devices were subjected to the following testing:
- Cytotoxicity: AAMI ANSI ISO 10993-5: 2009 (R) 2014 - Biological evaluation of medical devices
Part 5: Tests for in vitro cytotoxicity
- USP Rabbit Pyrogen Study: 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
The results of the testing demonstrated that the device materials are biocompatible for the intended use.
# STERILIZATION VALIDATION:
Sterilization: Gamma sterilization validation was conducted in accordance with ISO 11137-2:2013 for Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose in conjunction with ISO/TS 13004, in order to assure SAL (Sterility Assurance Level) of 10-6 using VDmax 20kGy.
# BACTERIAL ENDOTOXINS
Bacterial endotoxin testing was performed using the Kinetic Turbidimetric Limulus Amebocyte Lysate (LAL) method in accordance with USP <85> by a qualified third-party laboratory. The acceptance criterion was ≤ 20 EU/device, consistent with USP <161> and FDA guidance “Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.”
Periodic LAL endotoxin testing is performed as part of routine quality control to ensure continued compliance with the specified endotoxin limits.
# SHELF LIFE
Accelerated shelf-life testing was conducted on the worst-case packaged device configuration in accordance with ISO 11607-1 and ISO 11607-2 to support a five (5) year shelf life. Packaging integrity and sterility evaluations following accelerated aging demonstrated that the sterile barrier system maintains integrity and sterility throughout the labeled shelf life.
# CLEANING AND REPROCESSING VALIDATION:
Adin Trans Mucosal Abutments (TMAs) are provided non-sterile for single use and must be cleaned and steam sterilized prior to clinical use according to the instructions provided in the device’s Instructions for Use (IFU). Cleaning and steam sterilization procedures were validated in accordance with applicable standards and FDA guidance, including ISO 17664, AAMI TIR12, AAMI TIR30, and the FDA guidance document “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.” The validated cleaning process demonstrated effective removal of organic residues and microbial contamination.
Cleaning and steam sterilization validation previously reviewed by FDA under K223714 was leveraged, as the subject devices utilize the same materials, manufacturing processes, and reprocessing methods.
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
{8}
0 7 2
# FATIGUE TESTING
Comparative fatigue testing was performed on the subject device and the predicate device (K151909) to evaluate mechanical performance under worst-case conditions. Testing was conducted using a modified ISO 14801 setup consistent with FDA guidance for endosseous dental implants and adapted to simulate the ZAGA-4 extra-maxillary clinical configuration.
The evaluation assessed the fatigue performance of the zygomatic implant when used with a 60° angled abutment with 5 mm gingival height under worst-case loading conditions, including 10° uncorrected angulation and simulated bone resorption. A minimum of twelve specimens per system were tested using identical methods and loading configurations to enable direct comparison. Results demonstrated comparable fatigue performance between the subject and predicate devices.
# MRI COMPATIBILITY
Non-clinical worst-case MRI review was performed to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental abutments, and fixation screws) and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque.
# PERFORMANCE DATA (CLINICAL)
The company retrospectively collected real-world data (RWD) to evaluate the clinical performance of the subject system, including the use of angled multi-unit abutments greater than 30°, and compared outcomes to publicly available data for the predicate device (Noris Medical Multi-Unit systems; K151909 and K210356). Records from 37 patients (10 females, 27 males; age range 45–75 years, mean 58.1 ± 8.2) were reviewed. Follow-up ranged from 6 to 22 months (mean 10.4 ± 3.8 months).
One patient (2.7%) was excluded due to loss to follow-up before 6 months. The final analysis population included 36 patients with 92 multi-unit abutments (72 with 60° angulation). All cases involved extra-sinus placement.
Primary effectiveness was defined as implant and angled abutment (45°–60°) success at ≥ 6 months post-loading. In the 36-patient cohort (92 abutments), a 100% success rate was observed, with stable implants, functional abutments, favorable soft tissue healing, and no reported device-related adverse events. Available clinical evaluations did not indicate evidence of evidence of infection, inflammation or fistula. A post hoc subgroup analysis of patients with ≥ 12 months follow-up (16 patients; 40 abutments) also demonstrated a 100% success rate with no reported adverse events or prosthetic complications.
CONFIDENTIAL
March 23, 2026
Revision 3.0
Adin Dental Implant Systems Ltd. 510(k)
Adin Long Dental Implant System
{9}
#
# SUBSTANTIAL EQUIVALENCE OF ADIN'S ADIN LONG DENTAL IMPLANT SYSTEM WITH PREDICATE DEVICE
Table 1 - Adin Touareg™-OS Zygomatic Implants comparison
| Feature | Adin Long Dental Implant System – Adin Touareg™-OS Zygomatic Implants - Subject Device - | Noris Medical Zygomatic Dental Implants System - Primary Predicate Device - | Adin Short Implants (Touareg™-OS) - Reference Device - | Equivalence Discussion |
| --- | --- | --- | --- | --- |
| 510(k) number | K252031 | K151909 | K212775 | NA |
| Indications For Use | The Adin Long Dental Implant System is intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to restore the patient's chewing function in fully or partially edentulous patients. The implants are intended for multi-unit, full-arch, screw-retained prosthetic restorations and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Adin Touareg-OS Zygomatic Dental Implants, intended for placement in the zygomatic bone in patients with severe atrophic maxilla; may be used with implant-level framework designs with 45° to 60° angulation. | Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae. | Adin Dental Implants are intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in fully or partially edentulous patients in order to restore a patient's chewing function. Adin Dental Implants may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading. Adin short implants are to be used only with straight abutments. | Similar, for subject device and primary predicate device. (K151909) |
| Patient population | Edentulous or partially edentulous patients | Edentulous or partially edentulous patients | Edentulous or partially edentulous patients | Same |
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
{10}
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Page 7
Revision 3.0
Adin Long Dental Implant System
| Feature | Adin Long Dental Implant System – Adin Touareg™-OS Zygomatic Implants - Subject Device - | Noris Medical Zygomatic Denta Implants System - Primary Predicate Device - | Adin Short Implants (Touareg™-OS) - Reference Device - | Equivalence Discussion |
| --- | --- | --- | --- | --- |
| Sterility | Sterile using Gamma Irradiation | Sterile using Gamma Irradiation | Sterile using Gamma Irradiation | Same |
| Nature of body contact | bone/tissue for long term duration (>30d) | bone/tissue contact for long term duration (>30d) | bone/tissue contact for long term duration (>30d) | Same |
| Prescription or Over-the-Counter (OTC) | Prescription | Prescription | Prescription | Same |
| Single use | Yes | Yes | Yes | Same |
| Operation Principle | Single or Two-Stages procedure (immediate or delayed loading) | Single or Two-Stages procedure (immediate or delayed loading) | Single or Two-Stages procedure (immediate or delayed loading) | Same |
| Placement Method | Placing the implant in the upper jawbone immediately after drilling | Placing the implant in the upper jawbone immediately after drilling | Placing the implant in the jawbone immediately after drilling | Same |
| Implant Material | Titanium Alloy 6Al-4V-ELI | Titanium Alloy 6Al-4V-ELI | Titanium Alloy 6Al-4V-ELI | Same |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Same |
| Shape | Screw type | Screw type | Screw type | Same |
| Connection | Internal hex | Internal hex | Internal hex | Same as primary predicate device |
| Surface Treatment | OsseoFix™ Calcium Phosphate | RBM (Resorbable blast media) | OsseoFix™ Calcium Phosphate | Identical as reference device |
| Length | 35-55mm | 35-57.5 mm | 6.25 mm | Similar, the subject device is within the length range of the primary predicate device |
| Outer Diameter (OD) | 4.2mm | 4.2mm | 4.2, 5, 6mm | Same |
| Shelf-Life | 5 years | 5 years | 5 years | Same |
{11}
C
#
#
Table 2 - Adin RS TMA ${}^{\mathrm{{TM}}}$ comparison:
| Feature | Adin TouaregTM-OS Zygomatic implants - RS TMA™ - Subject Device - | Noris Medical Zygomatic Dental Implants System - Multi unit 45° - Primary Predicate Device - | Noris Medical Zygomatic Dental Implants System Multi unit - Reference Device - | Nobel Biocare's NobelZygoma 0° Abutments - Reference Device - | Equivalence Discussion |
| --- | --- | --- | --- | --- | --- |
| 510(k) number | K252031 | K151909 | K210356 | K161598 | --- |
| Indications For Use | The Adin Long Dental Implant System is intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to restore the patient's chewing function in fully or partially edentulous patients. The implants are intended for multi-unit, full-arch, screw-retained prosthetic restorations and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Adin Touareg-OS Zygomatic Dental Implants are intended for placement in the zygomatic bone in patients with severe atrophic maxilla; they may be used with | Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae. | Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae. | NobelZygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arch to provide support for prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function. The NobelZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. | Same. Although primary device doesn't specifically refer to the Multi Unit- it is included as part of the system of K151909. Additionally, the primary device includes a Multi unit of 45°, this is bridges by reference device K210356 which include Multi units 52°-60° . |
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
{12}
7
#
| Feature | Adin Touareg™-OS Zygomatic implants - RS TMA™ - Subject Device - | Noris Medical Zygomatic Dental Implants System - Multi unit 45° - Primary Predicate Device - | Noris Medical Zygomatic Dental Implants System Multi unit - Reference Device - | Nobel Biocare's NobelZygoma 0° Abutments - Reference Device - | Equivalence Discussion |
| --- | --- | --- | --- | --- | --- |
| | implant-level framework designs with 45° to 60° angulation. | | | | |
| Patient population | Edentulous or partially edentulous patients | Edentulous or partially edentulous patients | Edentulous or partially edentulous patients | Edentulous or partially edentulous patients | Same |
| Abutment and it's screw materials | Titanium Alloy 6Al-4V-ELI | Titanium Alloy 6Al-4V-ELI | Titanium Alloy 6Al-4V-ELI | Titanium Alloy 6Al-4V-ELI | Same |
| Angulated Abutment's Gingival Height (GH) | 45° abutment: 4 mm 52°, 60° abutment: 5 mm | 45° abutment: 2 mm | 45° abutment: 3, 4, 5mm 52°, 60° abutment: 2mm | 45° abutment: 6, 8, 10mm 60° abutment: 6, 8mm | Similar, within range of the primary predicate and reference devices K210356 and K161598 |
| Compatible Implant-abutment | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Same as primary predicate and reference devices |
| Sterility | Supplied non-sterile. Steam sterilized before use | Supplied non-sterile. Steam sterilized before use | Supplied non-sterile. Steam sterilized before use | Supplied sterile | Same as primary predicate and reference devices (excluding K161598) |
| Single use | Yes | Yes | Yes | Yes | Same |
CONFIDENTIAL
March 23, 2026
Adin Dental Implant Systems Ltd. 510(k)
Revision 3.0
Adin Long Dental Implant System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.