The sponsor utilized a review of clinical literature and real-world evidence to support the expansion of the device's indications for use from pediatric/adolescent patients to include adults (8 years and older).
Clinical literature; Label expansion; Pediatric to adult extrapolation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Review of clinical literature; Literature review
Patients 8 years and older with functional abdominal pain associated with IBS and FD
Not applicable for this study
Safety and effectiveness of the device
Indications for Use
The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8 years and older with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms.
Device Story
Battery-operated, disposable, 5-gram micro-stimulation appliance; placed behind patient's ear. System includes PENFS stimulator, sterile multi-pin wire harness array (1-1-1-4 configuration), bandages, adhesive, alcohol swab, dressing, tweezers, surgical marker, and transilluminator. Transilluminator aids clinician in identifying auricular nerve bundles (Cranial Nerves V, VII, IX, X, and occipital nerves) for needle placement. Device delivers low-frequency electrical pulses for 120 hours/week over 4 consecutive weeks. Used as an aid in pain reduction for IBS and FD when combined with other therapies; also reduces nausea in FD patients. Operated by clinicians in a clinical setting. Output is electrical nerve field stimulation; helps manage chronic abdominal pain and nausea symptoms.
Clinical Evidence
Evidence includes randomized clinical studies (DEN180057), real-world evidence, and clinical literature. Data supports safety and effectiveness in pediatric and adolescent populations; clinical literature and data extrapolation support extension to adult patients 8 years and older.
Technological Characteristics
Battery-operated percutaneous electrical nerve field stimulator (PENFS). Delivers low-frequency electrical pulses. Components include sterile multi-pin wire harness array (1-1-1-4 configuration) and transilluminator for nerve identification. Disposable, single-use, 5-gram form factor. No changes to design or materials from predicate.
Indications for Use
Indicated for patients 8 years and older with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD).
Regulatory Classification
Identification
A nonimplanted nerve stimulator for functional abdominal pain relief is a device that stimulates nerves remotely from the source of pain with the intent to relieve functional abdominal pain. This generic type of device does not include devices designed to relieve pelvic pain.
Special Controls
In combination with the general controls of the FD&C Act, the non-implanted nerve stimulator for functional abdominal pain relief is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the labeled shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 16, 2025
NeurAxis
% Dawn Norman
Partner
MRC Global, LLC
9085 E. Mineral Cir., Suite 110
Centennial, Colorado 80112
Re: K252024
Trade/Device Name: NeurAxis IB-Stim (01-1020)
Regulation Number: 21 CFR 876.5340
Regulation Name: Nonimplanted Nerve Stimulator For Functional Abdominal Pain Relief
Regulatory Class: Class II
Product Code: QHH
Dated: July 17, 2025
Received: July 18, 2025
Dear Dawn Norman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K252024 - Dawn Norman
Page 2 of 3
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252024 - Dawn Norman
Page 3 of 3
Sincerely,
SIVAKAMI VENKATAChalam -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
NeurAxis IB-Stim
Page 8 of 29
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252024 | ? |
| Please provide the device trade name(s). | | ? |
| NeurAxis IB-Stim (01-1020) | | |
| Please provide your Indications for Use below. | | ? |
| The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8 years and older with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252024
Page 1 of 2
510(k) Summary
IB-Stim
June 26, 2025
Company: NeurAxis, Inc.
11611 N. Meridian St.
Suite 330
Carmel, IN 46032
Primary/Secondary Contact: Dawn Norman – Partner
MRC Global
9085 E. Mineral Cir., Suite 110
Centennial, CO 80112
Phone: (618) 604-3064
Email: dawn.norman@askmrcglobal.com
Kathy Remsen – Regulatory Consultant
(901) 606-4856
Email: kathy.remsen@askmrcglobal.com
Company Contact: Dr. Thomas J. Carrico, CRO
Trade Name: NeurAxis IB-Stim (01-1020)
Common Name: Non-implanted nerve stimulator for pain associated with irritable bowel syndrome (ibs)
Classification: Class II
Regulation: 21 CFR 876.5340
Regulation Name: Non-implanted nerve stimulator for functional abdominal pain relief
Panel: Neurology
Product Code: QHH
Primary Predicate: NeurAxis IB-Stim (K250451)
Device Description:
IB-Stim stimulator is a battery-operated micro-stimulation appliance weighing 5 grams designed as a disposable product for a single use. IB-Stim stimulator is placed behind the patient's ear and connected to stimulation needles on the auricle. IB-Stim stimulator offers regular therapy over several days. The appliance transmits low-frequency electric pulses.
IB-Stim is a device system that consists of a (1) percutaneous electrical nerve field stimulator (PENFS); (2) sterile multi-pin wire harness array; (3) round bandages to fasten the appliance leads, stimulator skin adhesive, alcohol
{5}
K252024
Page 2 of 2
swab, transparent dressing, biohazard bag, and Patient ID card; (4) tweezers; (5) surgical marker; and (6) transilluminator for use in transillumination technique that aids in needle implantation. The wire harness array consists of 4 leads. The 1-1-1-4 configuration consists of three single-needle leads, and one 4-needle array. The transilluminator emits a light for use in the trans-illumination technique to aid the clinician in identifying the nerve bundles in the ear.
## Indications for Use:
The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8 years and older with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms.
## Substantial Equivalence:
The subject NeurAxis IB-Stim is substantially equivalent to the predicate device, NeurAxis, Inc. IB-Stim – K250451.
The design, materials, packaging, sterilization, and package label of the subject device are identical to those of the predicate device. The indication for use is being expanded for the desired age group (from "8-21 years of age" to "8 years and older"). The extrapolation of clinical data from pediatric to adult indications to support that this change does not raise new questions about safety and effectiveness.
## Performance Testing:
There are no changes to the design. Therefore, previously conducted testing remains applicable and no additional performance testing is required.
## Clinical Data:
The randomized clinical studies sponsored and conducted by the sponsor in the previous DEN180057, real world evidence including clinical literature support the safety and effectiveness of the device in pediatrics and adolescents. Review of clinical literature and extrapolation of that data to the adult population supports the use of the device in patients 8 years and older.
## Conclusion:
The subject device is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.