EnSite™ X EP System

K252013 · ABBOTT MEDICAL · DQK · Sep 25, 2025 · Cardiovascular

Device Facts

Record IDK252013
Device NameEnSite™ X EP System
ApplicantABBOTT MEDICAL
Product CodeDQK · Cardiovascular
Decision DateSep 25, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2

Indications for Use

EnSite™ X EP System The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures. EnSite™ X EP System Contact Force Software License When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters. EnSite™ X EP System Surface Electrode Kit The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.

Device Story

Catheter navigation and mapping system; displays 3D position of conventional and Sensor Enabled™ (SE) electrophysiology catheters. Collects anatomic data via catheter movement within cardiac chambers to generate 3D models; displays cardiac electrical activity as waveform traces, isopotential, and isochronal maps. Used in clinical EP procedures by physicians. Updated version supports compatibility with Current™ PFA Generator and Volt™ PFA Catheter; enables integrated mapping and ablation workflows. Provides real-time visualization to assist clinicians in cardiac navigation and diagnostic procedures.

Clinical Evidence

Bench testing only. Included software verification at unit, software, and system levels; system-level verification; design validation; and human factors validation to ensure user needs were met and no new use errors were introduced.

Technological Characteristics

Programmable diagnostic computer; 3D catheter navigation and mapping. Supports Sensor Enabled™ (SE) catheters. Software-based mapping and model creation. Complies with IEC 62304 (software lifecycle), IEC 60601-1-2 (EMC), ISO 14971 (risk management), and IEC 62366-1 (usability).

Indications for Use

Indicated for patients undergoing electrophysiology studies requiring cardiac electrical activity monitoring and catheter navigation/mapping.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 25, 2025 Abbott Medical Alexandra Agre Principal Regulatory Affairs Specialist One St. Jude Medical Drive St. Paul, Minnesota 55117 Re: K252013 Trade/Device Name: EnSite™ X EP System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: June 27, 2025 Received: June 27, 2025 Dear Alexandra Agre: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252013 - Alexandra Agre Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K252013 - Alexandra Agre Page 3 Sincerely, MARCO CANNELLA -S for - Aneesh Deoras - Assistant Director - Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices - Office of Cardiovascular Devices - Office of Product Evaluation and Quality - Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K252013 Device Name EnSite™ X EP System Indications for Use (Describe) EnSite™ X EP System The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures. EnSite™ X EP System Contact Force Software License When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters. EnSite™ X EP System Surface Electrode Kit The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} EnSite X EP System Volt PFA Software Licenses Traditional 510(k) ^{}[] Abbott # 510(k) SUMMARY The 510(k) Summary was drafted in accordance with 21 CFR 807.92, and is included below. | 510(k) Information | | | --- | --- | | 510(k) Number | K252013 | | 510(k) Type | Traditional 510(k) | | Date Prepared | 27 June 2025 | | Submitter Information | | | Manufacturer Name & Address | Abbott Medical One St. Jude Medical Drive, St. Paul, Minnesota, 55117, USA Manufacturer of the EnSite X EP System | | | Abbott Medical Costa Rica Ltda. Edificio #44 Calle 0, Ave. 2 Zona Franca Coyol El Coyol Alajuela, Costa Rica 1897-4050 Manufacturer of the EnSite X EP System Surface Electrode Kit | | Contact Person | Alexandra Agre Principal Regulatory Affairs Specialist 612-322-3442 alexandra.agre@abbott.com | | EnSite™ X EP System Device Information | | | Trade Name | EnSite™ X EP System | | Common Name | Programmable Diagnostic Computer | | Class | II | | Classification Name | 870.1425, computer, diagnostic, programmable | | Product Code | DQK | | Predicate Device | EnSite™ X EP System (K242016) | | Device Description | The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber. The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points. | | Indications for Use | EnSite™ X EP System The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. | K252013 {5} EnSite X EP System Volt PFA Software Licenses Traditional 510(k) ^{}[] Abbott The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures. ## EnSite™ X EP System Contact Force Software License When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters. ## EnSite™ X EP System Surface Electrode Kit The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use. ## Predicate Comparison ### Comparison EnSite™ X EP System Version 4.0 with Volt™ PFA Software Licenses installed and the predicate EnSite™ X v3.1 have the same intended use and indications for use. They use the same fundamental scientific technology to facilitate catheter position and orientation, as well as cardiac mapping and model creation. The computer hardware has been updated to support the greater software capabilities and to maintain supply. The subject device software was revised to include the following updates: - Compatibility with Current™ PFA Generator - Compatibility with Volt™ PFA Catheter, Sensor Enabled™ - Software features to enable integrated mapping and ablation procedures with the newly compatible devices All risks were mitigated to acceptable levels. No new questions of safety or effectiveness were raised. ### Non-Clinical Testing Summary Design verification and validation activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are substantially equivalent to the predicate device. ### Testing The EnSite™ X EP System with Volt™ PFA Software Licenses in scope of this submission was developed and tested in accordance with the following industry guidance documents and standards: - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - IEC 62304: 2015-06 Edition 1.1, Medical Device Software - Software Life Cycle Processes - IEC 60601-1-2:2020-09 Edition 4.1, Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests - ISO 14971:2019 Medical Devices – Application of Risk Management to Medical Devices - IEC 62366-1:2015+A1:2020 Medical Devices – Part 1: Application of Usability Engineering to Medical Devices ### Types of Testing Performed – EnSite X EP System with Volt™ PFA Software License - Software Verification at unit, software and system level K252013 Page 2 of 3 {6} EnSite X EP System Volt PFA Software Licenses Traditional 510(k) ^{}[] | | • System Verification at the device and system level • Design Validation to ensure user needs are met • Human Factors Validation to ensure no new use errors introduced | | --- | --- | | Statement of Equivalence | All subject and predicate devices have the same intended use, and same indications for use. All devices operate using the same fundamental scientific technology to facilitate catheter position and orientation, as well as cardiac mapping and model creation. The testing completed and submitted in this Traditional 510(k) provides objective evidence the subject device is substantially equivalent to the predicate device. | K252013 Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...