The LIAISON® NES FLU A/B, RSV & COVID-19 assay is a real-time RT-PCR assay intended for use on the LIAISON® NES instrument for the simultaneous in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A (Flu A) virus, influenza B (Flu B) virus, and respiratory syncytial virus (RSV) in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The LIAISON® NES FLU A/B, RSV & COVID-19 assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections when used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the LIAISON® NES FLU A/B, RSV & COVID-19 assay may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infection and should not be used as the sole basis for patient management decisions.
Device Story
LIAISON NES FLU A/B, RSV & COVID-19 is a sample-to-answer, qualitative real-time RT-PCR assay performed on the LIAISON NES instrument. Input: anterior nasal swab specimen collected by healthcare professional or patient under supervision. Process: cartridge-based automated sample preparation (heating/pipetting), mixing with lyophilized PCR reagents, and multiplex RT-PCR amplification. Targets: Matrix gene (Flu A, Flu B, RSV); Spike and ORF7A genes (SARS-CoV-2). Output: automated interpretation of fluorescence intensity monitored during cycling; results displayed on instrument screen as Positive/Negative for each target. Used in point-of-care settings by untrained operators. Results aid differential diagnosis in conjunction with clinical/epidemiological data.
Clinical Evidence
Prospective clinical study (N=1591) across seven US sites. Compared LIAISON NES assay to FDA-cleared molecular assay. Results: Influenza A (PPA 96.5%, NPA 97.2%), Influenza B (PPA 97.1%, NPA 99.7%), RSV (PPA 90.3%, NPA 99.8%), COVID-19 (PPA 98.4%, NPA 99.4%). Reproducibility study (960 replicates) showed high agreement across sites/operators.
Technological Characteristics
Real-time RT-PCR; fluorescence detection. Single-use cartridge containing lyophilized reagents. Instrument: LIAISON NES (automated thermocycling, optical detection). Connectivity: Barcode reader for specimen ID. Sterilization: Not specified. Software: Embedded, automated result interpretation.
Indications for Use
Indicated for individuals with signs and symptoms of respiratory tract infection. Used for qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and RSV nucleic acids in anterior nasal swab specimens. Not intended for influenza C detection.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
Predicate Devices
Biofire® SPOTFIRE® Respiratory (R) Panel Mini (K230719)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K251978
B Applicant
Diasorin Molecular LLC
C Proprietary and Established Names
LIAISON NES FLU A/B, RSV & COVID-19
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QOF | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
The purpose of this submission is to demonstrate that the LIAISON NES FLU A/B, RSV & COVID-19 assay is substantially equivalent to the BIOFIRE SPOTFIRE Respiratory Panel Mini (SPOTFIRE R Panel Mini, K230719).
### B Measurand:
- Influenza A RNA
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
- Influenza B RNA
- Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) RNA
- Respiratory Syncytial Virus (RSV) RNA
## C Type of Test:
Qualitative RT-PCR
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The LIAISON NES FLU A/B, RSV & COVID-19 assay is a real-time RT-PCR assay intended for use on the LIAISON NES instrument for the simultaneous in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A (Flu A) virus, influenza B (Flu B) virus and respiratory syncytial virus (RSV) in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.
The LIAISON NES FLU A/B, RSV & COVID-19 assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections when used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the LIAISON NES FLU A/B, RSV & COVID-19 assay may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infection and should not be used as the sole basis for patient management decisions.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic Use Only
### D Special Instrument Requirements:
For use with LIAISON NES Instrument
## IV Device/System Characteristics:
### A Device Description:
The LIAISON NES Instrument is capable of analysis of a single cartridge containing a single specimen. A set of parameters specific to the assay is included in the instrument software to name target molecules, assign dyes to probes, specify cycling conditions, and to analyze data from runs. Fluorescence intensity is monitored at each PCR
K251978 - Page 2 of 30
{2}
cycle by detection modules within the instrument. The instrument software controls the thermocycling and, upon completion of the run, automatically interprets and displays results for the specimen.
The LIAISON NES instrument is designed to accept a cartridge, containing either a quality control (QC) or patient sample, to process and detect for the target nucleic acid.
The LIAISON NES FLU A/B, RSV & COVID-19 assay used on the LIAISON NES instrument is a real-time RT-PCR system that enables the direct amplification, detection, and differentiation of influenza A RNA, influenza B RNA, RSV RNA, and SARS-CoV-2 RNA directly from anterior nasal swabs.
## B Principle of Operation:
The anterior nasal swab specimen is collected by a healthcare professional or the patient under the supervision of the healthcare professional as indicated in section IV. C 3 Specimen Sampling and Handling.
The LIAISON NES instrument automates the LIAISON NES FLU A/B, RSV & COVID-19 assay sample analysis steps within a single cartridge. The sample analysis steps that occur within the cartridge are described below.
**Sample Preparation**: The specimen is processed inside the instrument with heating and pipetting to release the nucleic acids.
**PCR Reagents Preparation**: Nucleic acids released from the specimen are mixed with lyophilized PCR reagents (primers, nucleotides and enzymes needed for the amplification process) contained within the cartridge.
**Target Amplification**: Multiplex RT-PCR amplification of the target specific amplicons is performed using thermal cycling. Primers and fluorescent probes are used to specifically amplify conserved regions of influenza A RNA, influenza B RNA, RSV RNA, and SARS-CoV-2 RNA, while the internal control (IC) RNA is used to detect any PCR failures and/or inhibition. The genes targeted by the primers and probes include the Matrix gene for Influenza A, Influenza B, and RSV, and the Spike and ORF7A genes for SARS-CoV-2.
**Analysis**: The instrument's optical modules continuously monitor the fluorescence emitted during each PCR cycle. This real-time detection enables the system to determine the presence of specific pathogens. The instrument calculates the threshold cycle (Ct) value, which is the cycle number at which the fluorescence signal exceeds a predefined threshold.
Final results are displayed on the instrument's screen, indicating whether the sample is "Positive" or "Negative" for the targeted pathogens (e.g., influenza A, influenza B, RSV, SARS-CoV-2). These results are linked to the test and patient information entered at the beginning of each test to provide a results report.
## C Instrument Description Information:
1. **Instrument Name**:
K251978 - Page 3 of 30
{3}
LIAISON NES instrument
2. Specimen Identification:
The instrument incorporates a barcode reader to scan in specimen identification. Specimen identification may also be entered manually using an on-screen keyboard.
3. Specimen Sampling and Handling:
The NES Swab is utilized to collect the specimen from the patient by inserting the swab approximately 1 inch into one nostril, gently rotating the swab 5 times against the nasal wall for approximately 15 seconds, then repeating these steps in the other nostril. Specimen collection can be completed by the healthcare provider or by the patient under supervision of a healthcare provider.
After collection of the specimen from the patient, the operator inserts the nasal swab into the NES Sample Vial and squeezes the swab head while rotating the swab 10 times. Following rotation of the swab in the sample vial, the operator snaps the swab shaft off at a pre-defined mark on the swab shaft, leaves the swab head in the sample vial, and places the cap back on the sample vial. The operator then twists off the cap at the end of the sample vial and squeezes the body of the sample vial to deliver all of the liquid into the cartridge.
4. Calibration:
The instrument is factory calibrated. No user calibration is performed.
5. Quality Control:
Internal Controls
Each LIAISON NES FLU A/B, RSV & COVID-19 cartridge includes an internal control to ensure performance of sample preparation, amplification, and detection. The internal control must generate a signal above threshold in the Q705 channel for the system to report a valid test result when target RNA is not detected. Detection of the internal control is not required for a valid result when target RNA is detected.
External Controls
The LIAISON NES FLU A/B, RSV & COVID-19 Control Swab Kit contains positive and negative control swabs as external controls. Positive and negative external controls should be tested with each new lot or shipment of reagents, or monthly, (whichever occurs first), or in accordance with updated local, regional, state, and/or federal guidelines, per the instructions for use and facility procedures.
The NES FLU A/B, RSV & COVID-19 Positive Control Swab contains influenza A, influenza B, RSV A, and SARS-CoV-2 viral particles that have been inactivated using irradiation and chemical treatments. The viral particle solution is lyophilized into the swab. Viral particles included on the swab are influenza A (California 07/2009), Influenza B (Colorado 06/2017), RSV (A2) and SARS-CoV-2 (Washington 1/2020).
K251978 - Page 4 of 30
{4}
V Substantial Equivalence Information:
A Predicate Device Name(s):
BIOFIRE SPOTFIRE Respiratory (R) Panel Mini
B Predicate 510(k) Number(s):
K230719
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K251978 | K230719 |
| --- | --- | --- |
| Device Trade Name | LIAISON NES FLU A/B, RSV & COVID-19 | BIOFIRE SPOTFIRE Respiratory Panel Mini (SPOTFIRE R Panel Mini) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The LIAISON NES FLU A/B, RSV & COVID-19 assay is a real-time RT-PCR assay intended for use on the LIAISON NES instrument for the simultaneous in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A (Flu A) virus, influenza B (Flu B) virus and respiratory syncytial virus (RSV) in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The LIAISON NES FLU A/B, RSV & COVID-19 | The BIOFIRE SPOTFIRE Respiratory Panel Mini (SPOTFIRE R Panel Mini) is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE SPOTFIRE System for the simultaneous, qualitative detection and identification of multiple respiratory viral nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19. The following organism types and subtypes are identified and differentiated using the SPOTFIRE R Panel Mini:
• Coronavirus SARS-CoV-2
• Human rhinovirus
• Influenza A virus
• Influenza B virus
• Respiratory syncytial |
K251978 - Page 5 of 30
{5}
K251978 - Page 6 of 30
| | assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections when used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the LIAISON NES FLU A/B, RSV & COVID-19 assay may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infection and should not be used as the sole basis for patient management decisions. | virus
Nucleic acids from the viral organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection are indicative of the presence of the identified micro-organism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen.
Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R Panel Mini may not be the definite cause of disease.
Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible |
| --- | --- | --- |
{6}
| | | respiratory tract infection. |
| --- | --- | --- |
| Measurand | Same | Nucleic acid from Organisms detected |
| Patient Population | Individuals with signs and symptoms of respiratory tract infection | Individuals with signs and symptoms of respiratory tract infection, including COVID-19. |
| Organisms detected | Same, except no assay for Human rhinovirus | • Coronavirus SARS-CoV-2
• Human rhinovirus
• Influenza A virus
• Influenza B virus
• Respiratory syncytial virus |
| Automated System (Sample to Answer) | Same | Automated test interpretation and reporting. User cannot access raw data. |
| Reporting Features | Same | System software analyzes processed image data and provides test results for each panel target |
| Assay Software Functions | Same | Defines panel-specific parameters, instrument protocols and report requirements. |
| Time to Result | ~18 minutes | ~15 minutes |
| General Device Characteristic Differences | | |
| Sample Type | Anterior Nasal Swab (ANS) | Nasopharyngeal Swab (NPS) |
| Instrumentation | LIAISON NES Instrument | BIOFIRE SPOTFIRE System |
| Technological Principle | Highly multiplexed real-time RT-PCR with fluorescence detection and analysis | Highly multiplexed PCR with DNA melting analysis |
# VI Standards/Guidance Documents Referenced:
- Special controls listed under 21 CFR 866.3981—Respiratory Viral Panel Multiplex Nucleic Acid Assay – Class II Special Controls Guidance for Industry and FDA Staff (October 9, 2009)
- AAMI. Principles for medical device security - Risk Management. AAMI document
K251978 - Page 7 of 30
{7}
TIR57:2016. Association for the Advancement of Medical Instrumentation; 2016.
- AAMI. Principles for medical device security – Postmarket risk management for device manufacturers. AAMI document TIR97:2019. Association for the Advancement of Medical Instrumentation; 2019.
- CLSI. Information Technology Security of In Vitro Diagnostic Instruments and Software Systems; Approved Standard – Second Edition. CLSI document AUTO11-A2. Wayne, PA: Clinical Laboratory Standards Institute; 2014.
- CLSI. Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. CLSI document EP05-A3. Wayne, PA: Clinical Laboratory Standards Institute; 2019.
- CLSI. Interference Testing in Clinical Chemistry. 3rd Ed. CLSI Document EP07. Wayne, PA: Clinical Laboratory Standards Institute; 2018.
- CLSI. Evaluation of Qualitative, Binary Output Examination Performance; Approved Guideline – Third Edition. CLSI document EP12. Wayne, PA: Clinical Laboratory Standards Institute; 2023.
- CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. CLSI document EP17-A2. Wayne, PA: Clinical Laboratory Standards Institute; 2012.
- CLSI. Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline – Second Edition. CLSI document EP24-A2. Wayne, PA: Clinical Laboratory Standards Institute; 2011.
- CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI document EP25-A. Wayne, PA: Clinical Laboratory Standards Institute; 2009.
- CLSI. Collection Transport Preparation and Storage of Specimens for Molecular Methods. 2nd Edition. CLSI Document MM13. Wayne, PA: Clinical Laboratory Standards Institute; 2020.
- CLSI. Verification and Validation of Multiplex Nucleic Acid Assays. 2nd Edition. CLSI Document MM17. Wayne, PA: Clinical Laboratory Standards Institute; 2018.
- ISTA. Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb.) or Less. ISTA Document 3A. International Safe Transit Association. 2018.
- IEC 62366-1 Edition 1.1 2020-06 Consolidated Version; Medical devices – Part 1: Application of usability engineering to medical devices
- IEC 61010-1 Edition 3.1 2017-01 Consolidated Version; Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
- IEC 60601-1-2 Edition 4.1 2020-09 Consolidated Version; Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests
- IEC 61326-1 Edition 3.0 2020-10; Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 1: General requirements
- IEC 61326-2 Edition 3.0 2020-10; Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment
- IEC 62304 Edition 1.1 2015-06 Consolidated Version; Medical device software – Software life cycle processes
- IEC TR 60878 Ed. 4.0 2022-11; Graphical symbols for electrical equipment in
K251978 - Page 8 of 30
{8}
medical practice [Including: Corrigendum 1 (2023)]
- IEC TR 80001-2-2:2012. Application of risk management for IT Networks incorporating medical devices – Part 2-2: Guidance for the disclosure and communication of medical device security needs, risks and controls
- IEC TR 80001-2-8 Edition 1.0 206-05; Application of risk management for IT – networks incorporating medical devices – Part 2-8: Application guidance – Guidance on standards for establishing the security capabilities identified in IEC TR 80001-2-2
- ISO 14971:2019 Medical Devices – Application of risk management to medical devices
- ISO 15223-1: 2021-07 – Medical Devices- Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements
- UL ANSI 2900-1 First Edition 2017; Standard for Safety, Standard for Software Cybersecurity Network-Connectable Products, Part 1: General Requirements
- UL ANSI 2900-2-1 First Edition 2017; Standard for Safety, Software Cybersecurity for Network-Connectable Products, Part 2-1: Particular Requirements for Network Connectable Components of Healthcare and Wellness Systems
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
Analytical studies were performed using pooled negative human nasal matrix.
#### 1. Precision/Reproducibility:
##### a. Reproducibility
A reproducibility study was conducted assessing the total variability of the LIAISON NES FLU A/B, RSV & COVID-19 assay across operators, study sites, testing days and instruments.
Samples were prepared by a designated staff member of the Diasorin Molecular R&D study team. Each contrived sample panel member was randomized and labeled with a sample ID that did not contain any information regarding the sample content to ensure a minimum level of bias. Eight positive panel members consisted of contrived samples of influenza A virus, influenza B virus, RSV, and SARS-CoV-2 individually spiked onto dry nasal swabs at low positive (~2x LoD) and moderately positive (~5x LoD) concentrations. In addition, No Template Controls (NTC) which consisted of pooled negative human nasal matrix, and external positive controls, PC – LIAISON NES FLU A/B, RSV & COVID-19 Positive Control Swabs (PC), were included. The panel members are shown in Table 1. The reproducibility study was performed at three geographically distinct external sites, all under CLIA-waived status. There were two operators at each site and each operator tested two replicates of each panel member each day for eight non-consecutive days.
Table 1. Reproducibility Panel Members
K251978 - Page 9 of 30
{9}
| Panel Member | | Strain | Concentration |
| --- | --- | --- | --- |
| 1 | Influenza A | Darwin/9/21 | 2x LoD |
| 2 | Influenza A | | 5x LoD |
| 3 | Influenza B | Phuket 3073/2013 | 2x LoD |
| 4 | Influenza B | | 5x LoD |
| 5 | RSV | RSV B 12/2014 | 2x LoD |
| 6 | RSV | | 5x LoD |
| 7 | SARS-CoV-2 | USA/WA 1/2020 | 2x LoD |
| 8 | SARS-CoV-2 | | 5x LoD |
| 9 | NTC – Pooled Negative Human Nasal Matrix | NA | NA |
| 10 | PC - LIAISON NES FLU A/B, RSV & COVID-19 Positive Control Swab (PC) | NA | 5-10x LoD |
LoD = Limit of Detection, NTC = No Template Control, PC = Positive Control
Reproducibility was evaluated with reference to CLSI, EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures.
Table 2 shows the qualitative summary for influenza A virus, influenza B virus, RSV and SARS-CoV-2 during panel member testing. The LIAISON NES FLU A/B, RSV & COVID-19 assay demonstrated 100% agreement (96/96) for each target analyte with the moderate panel members (5x LoD). For low positive panel members (2x LoD), the assay yielded 100% agreement with the exception of influenza B, which yielded 99% (95/96) agreement. All negative (NTC) and positive control (PC) panel members also yielded 100% agreement with the expected results. This performance is acceptable and demonstrates acceptable assay reproducibility.
Table 2. Qualitative Summary of Results and Mean Ct Values +/- SD (%CV)
| | Site 1 | | Site 2 | | Site 3a | | All Sites | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Agreement with expected results | Mean Ct ±SD (%CV) | Agreement with expected results | Mean Ct ± SD (%CV) | Agreement with expected results | Mean Ct ± SD (%CV) | Agreement with expected results | Mean Ct ± SD (%CV) | 95% CI |
| Influenza A 2x | 100% (32/32) | 36.6 ± 0.77 (2.1%) | 100% (32/32) | 36.8 ± 1.28 (3.5%) | 100% (32/32) | 36.9 ± 1.18 (3.2%) | 100% (96/96) | 36.8 ± 1.09 (3.0%) | 96.2% - 100% |
| Influenza A 5x | 100% (32/32) | 34.6 ± 1.09 (3.2%) | 100% (32/32) | 34.7 ± 1.16 (3.4%) | 100% (32/32) | 35.7 ± 1.73 (4.8%) | 100% (96/96) | 35.0 ± 1.43 (4.1%) | 96.2% - 100% |
| Influenza B 2x | 100% (32/32) | 39.0 ± 1.89 (4.8%) | 100% (32/32) | 38.2 ± 0.88 (2.3%) | 96.9% (31/32a) | 38.2 ± 1.42 (3.7%) | 99% (95/96) | 38.5 ± 1.48 (3.9%) | 94.3% - 99.8% |
| Influenza B 5x | 100% (32/32) | 37.1 ± 1.40 (3.8%) | 100% (32/32) | 36.1 ± 0.78 (2.2%) | 100% (32/32) | 36.2 ± 1.47 (4.1%) | 100% (96/96) | 36.4 ± 1.32 (3.6%) | 96.2% - 100% |
K251978 - Page 10 of 30
{10}
The total Ct variability, as measured by the standard deviation, was less than or equal to 1.72 across all target viruses and concentrations. These results, shown in Table 3, indicate that the reproducibility of the LIAISON NES FLU A/B, RSV & COVID-19 assay is acceptable. Internal control reproducibility was also carried out and the result is acceptable.
Table 3. Quantitative Summary of Reproducibility
| | | | Reproducibility | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Between Sites | | Between Operator (Site) | | Between Day (Operator Site) | | Between Instruments (Days Operator Site) | | Total Reproducibility |
| Sample | N | Mean Ct | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | |
| Influenza A 2x | 96 | 36.8 | 0.14 | 0.4 | 0.58 | 1.6 | 0.80 | 2.2 | 1.07 | 2.9 | 1.20 |
| Influenza A 5x | 96 | 35.0 | 0.61 | 1.8 | 0.97 | 2.8 | 1.20 | 3.4 | 1.03 | 3.0 | 1.60 |
| Influenza B 2x | 95 | 38.5 | 0.44 | 1.1 | 0.86 | 2.2 | 1.20 | 3.1 | 1.48 | 3.9 | 1.50 |
| Influenza B 5x | 96 | 36.4 | 0.54 | 1.5 | 0.82 | 2.3 | 1.06 | 2.9 | 1.22 | 3.3 | 1.54 |
| RSV 2x | 96 | 37.0 | 0.21 | 0.6 | 0.66 | 1.8 | 0.88 | 2.4 | 1.04 | 2.8 | 1.33 |
| RSV 5x | 96 | 35.2 | 0.09 | 0.3 | 0.82 | 2.3 | 1.02 | 2.9 | 0.81 | 2.3 | 1.33 |
| SARS-CoV-2 2x | 96 | 37.1 | 0.45 | 1.2 | 0.83 | 2.2 | 1.08 | 2.9 | 1.30 | 3.5 | 1.72 |
| SARS-CoV-2 5x | 96 | 35.2 | 0.26 | 0.7 | 0.91 | 2.6 | 1.12 | 3.2 | 0.76 | 2.2 | 1.52 |
| PC Influenza A | 96 | 28.5 | 0.46 | 1.6 | 0.57 | 2.0 | 0.75 | 2.7 | 0.45 | 1.6 | 0.92 |
| PC Influenza B | 96 | 29.1 | 0.45 | 1.5 | 0.53 | 1.8 | 0.58 | 2.0 | 0.32 | 1.1 | 0.69 |
| PC RSV | 96 | 30.0 | 0.33 | 1.1 | 0.49 | 1.6 | 0.59 | 2.0 | 0.42 | 1.4 | 0.80 |
| PC SARS-CoV-2 | 96 | 30.6 | 0.30 | 1.0 | 0.50 | 1.6 | 0.84 | 2.7 | 0.88 | 2.9 | 1.01 |
n = number of replicates, Ct = Cycle threshold, SD = Standard Deviation, %CV = Percent Coefficient of Variation.
NTC = No Template Control, PC = Positive Control
aOne NTC tested at site 3 failed to detect the IC and was repeated in accordance with product labeling.
The total Ct variability, as measured by the standard deviation, was less than or equal to 1.72 across all target viruses and concentrations. These results, shown in Table 3, indicate that the reproducibility of the LIAISON NES FLU A/B, RSV & COVID-19 assay is acceptable. Internal control reproducibility was also carried out and the result is acceptable.
K251978 - Page 11 of 30
{11}
b. Within-Laboratory Precision
Within-laboratory precision of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated with four different lots of LIAISON NES FLU A/B, RSV & COVID-19 Cartridges and four LIAISON NES instruments. The study was conducted by one operator over the course of twenty-four non-consecutive days. The qualitative study result is shown in Table 4. Target concentrations for Low Positive (LP) and Medium Positive (MP) were 2x LoD and 5x LoD, respectively.
There was 100% agreement between the expected results and the observed qualitative results. Each panel member showed 100% expected result (24/24). The results of the study demonstrate acceptable assay variability.
Table 4. Summary of Qualitative Performance for Reaction Mix Inter-Lot / Inter-Instrument Reproducibility
| Panel Member | Strain/ Concentration | Expected Qualitative Result | Observed Qualitative Result % Detection (# Detected/# Tested) |
| --- | --- | --- | --- |
| ILII Repro 1 | Influenza A Low Positive (LP) | 100% of the tested replicate results are Influenza A Positive | 100% (24/24) Influenza A Positive |
| ILII Repro 2 | Influenza A Medium Positive (MP) | 100% of the tested replicate results are Influenza A Positive | 100% (24/24) Influenza A Positive |
| ILII Repro 3 | Influenza B Low Positive (LP) | 100% of the tested replicate results are Influenza B Positive | 100% (24/24) Influenza B Positive |
| ILII Repro 4 | Influenza B Medium Positive (MP) | 100% of the tested replicate results are Influenza B Positive | 100% (24/24) Influenza B Positive |
| ILII Repro 5 | RSV Low Positive (LP) | 100% of the tested replicate results are RSV Positive | 100% (24/24) RSV Positive |
| ILII Repro 6 | RSV Medium Positive (MP) | 100% of the tested replicate results are RSV Positive | 100% (24/24) RSV Positive |
| ILII Repro 7 | SARS-CoV-2 Low Positive (LP) | 100% of the tested replicate results are SARS-CoV-2 Positive | 100% (24/24) SARS-CoV-2 Positive |
| ILII Repro 8 | SARS-CoV-2 Medium Positive (MP) | 100% of the tested replicate results are SARS-CoV-2 Positive | 100% (24/24) SARS-CoV-2 Positive |
| ILII Repro 9 | Negative Sample (No Template Control) | 100% of the tested replicate results are Negative | 0% (0/24) Negative |
| Positive Control (PC) | Positive Control (PC) | 100% of the tested replicate results are Positive for all targets | 100% (24/24) Positive |
2. Linearity:
Not applicable.
K251978 - Page 12 of 30
{12}
K251978 - Page 13 of 30
3. Analytical Specificity/Interference:
The inclusivity of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated using a combination of *in silico* analysis of publicly available sequence information and laboratory testing of contrived specimens containing viral isolates that were selected to represent phylogenetic, geographic, and temporal diversity.
Analytical Reactivity (Inclusivity)
a. *In silico*
The inclusivity of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated using *in silico* analysis of the forward primers, reverse primers, and probes for the SARS-CoV-2 target. Analysis was performed against SARS-CoV-2 sequences available in the GISAID EpiCoV database as of May 5, 2025. The sequence set encompassed 5,641,034 SARS-CoV-2 sequences, including available sequences from variants of concern or variants of interest defined as of May 5, 2025.
*In silico* analysis of the dual target SARS-CoV-2 primer and probe binding regions indicate ~100% detection of the 5,641,034 SARS-CoV-2 sequences, including sequences of all defined variants of concern or variants of interest, available in the GISAID EpiCoV database as of May 5, 2025.
b. Wet-Testing
The inclusivity study evaluates the ability of the test to detect clinically relevant strains, serotypes, or subtypes of influenza A virus, influenza B virus, RSV, and SARS-CoV-2. The reactivity/inclusivity of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated with 36 influenza A (15 H1N1, 1 H1N2, 14 H3N2, 4 H5N1, 1 H7N7, and 1 H9N2) isolates, 14 influenza B (7 Victoria, 6 Yamagata, and 1 subtype unknown) isolates, 9 RSV (4 RSVA and 5 RSVB) isolates, and 19 SARS-CoV-2 isolates. The inclusivity panel was prepared by spiking various target microorganisms into pooled negative ANS matrix. The concentration which showed a 100% hit rate is shown in Table 5 through Table 8.
Influenza A
The detected concentrations for the influenza A variants tested with the Diasorin LIAISON NES FLU A/B, RSV & COVID-19 assay are shown in Table 5.
All strains of influenza A (H1N1) have been tested at 1.8E4 cps/swab (3x LoD of influenza A Victoria/4897/2022 (H1N1)). All replicates were detected correctly at this concentration.
All strains of influenza A (H3N2) have been tested at 2.4E4 cps/swab (3x LoD of influenza A Darwin/9/21 (H3N2)). All replicates were detected correctly at this concentration with the exception of Flu A Hong Kong/8/68, which was detected at 5x LoD (4.0E4 cps/swab).
Other influenza A isolates were detected successfully as indicated.
Table 5 Influenza A Results
{13}
| Strain | Concentration (cps/swab) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) |
| --- | --- | --- | --- |
| Influenza A Michigan/45/2015 (H1N1) | 1.8E4 | 100% (3/3) | 35.3 ± 1.1 (3.0%) |
| Influenza A Brisbane/59/07 (H1N1) | 1.8E4 | 100% (3/3) | 34.9 ± 0.1 (0.4%) |
| Influenza A NY/02/09 (H1N1) | 1.8E4 | 100% (3/3) | 34.0 ± 0.3 (1.0%) |
| Influenza A Mexico/4108/09 (H1N1)* | 1.8E4 | 100% (3/3) | 36.2 ± 0.8 (2.3%) |
| Influenza A New York/18/09 (H1N1) | 1.8E4 | 100% (3/3) | 37.0 ± 0.3 (0.9%) |
| Influenza A Taiwan/42/06 (H1N1) | 1.8E4 | 100% (3/3) | 37.1 ± 1.2 (3.1%) |
| Influenza A Brisbane/02/2018 (H1N1) (CDC panel) | 1.8E4 | 100% (3/3) | 35.8 ± 0.6 (3.1%) |
| Influenza A Wisconsin/588/2019 (H1N1) | 1.8E4 | 100% (3/3) | 37.0 ± 1.1 (3.0%) |
| Influenza A California/07/2009 (H1N1) | 1.8E4 | 100% (3/3) | 33.3 ± 0.5 (1.5%) |
| Influenza A New Caledonia/20/1999 (H1N1) | 1.8E4 | 100% (3/3) | 38.6 ± 2.5 (6.5%) |
| Influenza A Guangdong- Moanan/SWL1536/2019 (H1N1) | 1.8E4 | 100% (3/3) | 38.5 ± 0.4 (1.0%) |
| Influenza A Puerto Rico/8/34 (H1N1) | 1.8E4 | 100% (3/3) | 31.9 ± 0.3 (1.1%) |
| Influenza A Victoria/2570/2019 (H1N1) | 1.8E4 | 100% (3/3) | 37.9 ± 0.4 (1.0%) |
| Influenza A Nebraska/14/2018 (H1N1) | 1.8E4 | 100% (3/3) | 34.3 ± 0.3 (0.7%) |
| Influenza A Sydney/05/2021 (H1N1) | 1.8E4 | 100% (3/3) | 33.5 ± 1.4 (4.2%) |
| Influenza A swine/Ohio/09SW1484E/2009 (Extracted RNA) (H1N2) | 1.8E4 | 100% (3/3) | 30.7 ± 0.5 (1.7%) |
| Influenza A Switzerland/9715293/13 (H3N2) | 2.4E4 | 100% (3/3) | 33.7 ± 0.6 (1.8%) |
| Influenza A Hong Kong/4801/2014 (H3N2) | 2.4E4 | 100% (3/3) | 36.0 ± 0.5 (1.3%) |
| Influenza A Singapore/INFIMH-16-0019/2016 (H3N2) | 2.4E4 | 100% (3/3) | 33.0 ± 0.3 (0.8%) |
| Influenza A Perth/16/2009 (H3N2) (CDC Panel) | 2.4E4 | 100% (3/3) | 35.0 ± 0.5 (1.3%) |
| Influenza A Hong Kong/2671/2019 (H3N2) (CDC Panel) | 2.4E4 | 100% (3/3) | 36.4 ± 0.2 (0.7%) |
| Influenza A Hong Kong/8/68 (H3N2) | 2.4E4 (3x LoD) | 67% (2/3) | 38.1 ± 0.1 (0.2%) |
| | 3.2E4 (4x LoD) | 67% (2/3) | 38.9 ± 1.0 (2.6%) |
| | 4E4 (5x LoD) | 100% (3/3) | 37.5 ± 0.3 (0.9%) |
| Influenza A Cambodia/E0826360/2020 (H3N2) | 2.4E4 | 100% (3/3) | 35.4 ± 0.4 (1.3%) |
| Influenza A Kansas/14/2017 (H3N2) (CDC Panel) | 2.4E4 | 100% (3/3) | 34.0 ± 0.6 (1.7%) |
| Influenza A Port Chalmers/1/73 (H3N2) | 2.4E4 | 100% (3/3) | 34.8 ± 0.6 (1.7%) |
| Influenza A Darwin/6/2021 (H3N2) | 2.4E4 | 100% (3/3) | 35.1 ± 0.4 (1.1%) |
| Influenza A Tasmania/503/2020 (H3N2) | 2.4E4 | 100% (3/3) | 35.1 ± 0.5 (1.4%) |
| Influenza A Thailand/08/2022 (H3N2) | 2.4E4 | 100% (3/3) | 33.4 ± 0.5 (1.6%) |
| Influenza A Texas/50/2012 (H3N2) | 2.4E4 | 100% (3/3) | 33.3 ± 0.1 (0.2%) |
| Influenza A Massachusetts/18/2022 (H3N2) | 2.4E4 | 100% (3/3) | 36.9 ± 0.5 (1.2%) |
| Influenza A/Anhui/01/2005 (H5N1) | 1.8E4 | 100% (3/3) | 30.2 ± 0.2 (0.7%) |
| Influenza A/Egypt/N03072/2010 (H5N1) | 1.8E4 | 100% (3/3) | 29.4 ± 0.1 (0.5%) |
| Influenza A/Hubei/1/2010 (H5N1) | 1.8E4 | 100% (3/3) | 30.2 ± 0.2 (0.7%) |
| Influenza A A/bovine/Ohio.B24OSU-439/2024 gRNA (H5N1) | 1.8E4 | 100% (3/3) | 33.0 ± 0.2 (0.5%) |
| Influenza A/Mallard/Netherlands/12/2000 (H7N7) | 1.8E4 | 100% (3/3) | 30.7 ± 0.3 (1.0%) |
| Influenza A/Hong Kong/33982/2009 (H9N2) | 1.8E4 | 100% (3/3) | 32.8 ± 0.5 (1.6%) |
Ct = Cycle threshold, SD = Standard Deviation, %CV = Percent Coefficient of Variation
*One replicate was not completed due to instrumental error and was repeated as per protocol
## Influenza B
All strains of influenza B (Victoria) were tested at 1.2E4 cps/swab (3x LoD of influenza B Austria/1359417/2021 (Victoria)). All replicates were detected correctly at this concentration.
All strains of influenza B (Yamagata) were tested at 1.5E4 cps/swab (3x LoD of influenza B Phuket/3073/2013 (Yamagata)). All replicates were detected correctly at this concentration.
K251978 - Page 14 of 30
{14}
Flu B Maryland/1/1959 (subtype unknown) was tested at 1.2E4 cps/swab (3x LoD of influenza B Austria/1359417/2021 (Victoria)). All three replicates were detected correctly at this concentration.
The results for influenza B strains/variants are presented in Table 6.
Table 6. Influenza B Results
| Strain | Concentration (cps/swab) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) |
| --- | --- | --- | --- |
| Influenza B Brisbane/60/2008 (Victoria) | 1.2E4 | 100% (3/3) | 35.4 ± 0.2 (0.7%) |
| Influenza B Brisbane/33/2008 (Victoria) | 1.2E4 | 100% (3/3) | 35.4 ± 0.2 (0.6%) |
| Influenza B Washington/02/2019 (Victoria) | 1.2E4 | 100% (3/3) | 37.0 ± 0.9 (2.3%) |
| Influenza B Colorado/06/2017 (Victoria) (CDC Panel) | 1.2E4 | 100% (3/3) | 35.8 ± 0.3 (0.7%) |
| Influenza B Florida/02/06 (Victoria) | 1.2E4 | 100% (3/3) | 37.2 ± 0.7 (1.9%) |
| Influenza B Malaysia/2506/04 (Victoria) | 1.2E4 | 100% (3/3) | 35.8 ± 0.9 (2.5%) |
| Influenza B Texas/2/13 (Victoria) (CDC Panel) | 1.2E4 | 100% (3/3) | 35.0 ± 0.3 (1.0%) |
| Influenza B Maryland/1/1959 (subtype unknown) | 1.2E4 | 100% (3/3) | 43.3 ± 0.7 (1.5%) |
| Influenza B Massachusetts/02/2012 (Yamagata) | 1.5E4 | 100% (3/3) | 38.6 ± 0.2 (0.4%) |
| Influenza B Wisconsin/1/10 (Yamagata) | 1.5E4 | 100% (3/3) | 39.7 ± 0.5 (1.4%) |
| Influenza B Florida/07/04 (Yamagata) | 1.5E4 | 100% (3/3) | 38.9 ± 0.3 (0.7%) |
| Influenza B Florida/04/06 (Yamagata) | 1.5E4 | 100% (3/3) | 36.9 ± 0.6 (1.5%) |
| Influenza B Panama/45/90 (Yamagata) | 1.5E4 | 100% (3/3) | 36.5 ± 0.5 (1.4%) |
| Influenza B Utah/9/14 (Yamagata) | 1.5E4 | 100% (3/3) | 37.0 ± 0.2 (0.5%) |
Ct = Cycle threshold, SD = Standard Deviation, %CV = Percent Coefficient of Variation
**RSV**
All strains of RSV were tested at 2.4E4 cps/swab (3x LoD of RSV A Isolate 2006 and RSV B Isolate 12/2014). All replicates were detected correctly at this concentration.
The results for RSV are presented in Table 7.
Table 7. RSV Results
| Strain | Concentration (cps/swab) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) |
| --- | --- | --- | --- |
| RSV-A 1/2015 Isolate #1 | 2.4E4 | 100% (3/3) | 39.3 ± 0.5 (1.4%) |
| RSV-A2 | 2.4E4 | 100% (3/3) | 41.2 ± 0.1 (0.3%) |
| RSV-A 3/2015 Isolate #3 | 2.4E4 | 100% (3/3) | 41.2 ± 1.2 (2.9%) |
| RSV-A 4/2015 Isolate 1 | 2.4E4 | 100% (3/3) | 41.9 ± 0.7 (1.7%) |
| RSV-B 9320 | 2.4E4 | 100% (3/3) | 40.7 ± 0.1 (0.2%) |
| RSV-B CH93(18)-18 | 2.4E4 | 100% (3/3) | 37.8 ± 0.6 (1.5%) |
| RSV-B 3/2015 Isolate #1 | 2.4E4 | 100% (3/3) | 41.3 ± 1.3 (3.0%) |
| RSV-B1 | 2.4E4 | 100% (3/3) | 38.6 ± 0.6 (1.5%) |
| RSV-B 3/2015 Isolate #2 | 2.4E4 | 100% (3/3) | 40.8 ± 1.3 (3.3%) |
Ct = Cycle threshold, SD = Standard Deviation, %CV = Percent Coefficient of Variation
**SARS-CoV-2**
All strains of SARS-CoV-2 were tested at 3.0E3 cps/swab (3x LoD of SARS-CoV-2 USA/WA 1/2020 and SARS-CoV-2 USA/MDHP20874/2021 (B.1.1.529 Omicron)). All replicates were detected correctly at this concentration, except for SARS-CoV-2 Isolate
K251978 - Page 15 of 30
{15}
hCoV-19/USA/CA-Stanford -109_S21/2022 PANGO Lineage B.1.1.529, (XBB Omicron), which was detected at 4x LoD (4.0E3 cps/swab).
The results for SARS-CoV-2 are presented in Table 8.
Table 8. SARS-CoV-2 Results
| Strain | Concentration (cps/swab) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) |
| --- | --- | --- | --- |
| SARS-CoV-2 Isolate hCoV-1 9/USA/PHC658/2021 B.1.617.2 (Delta) | 3.0E3 | 100% (3/3) | 38.1 ± 0.7 (1.8%) |
| SARS-CoV-2 Isolate Hong Kong/VM20001061/2020 | 3.0E3 | 100% (3/3) | 36.1 ± 0.5 (1.5%) |
| SARS-CoV-2 Isolate hCoV- 19/Japan/TY7-503/2021 (Gamma Variant) | 3.0E3 | 100% (3/3) | 35.5 ± 0.3 (0.8%) |
| SARS-CoV-2 Isolate USA/MD- HP24556/2022 BA.2.3 (Omicron) | 3.0E3 | 100% (3/3) | 32.8 ± 0.3 (1.1%) |
| SARS-CoV-2 Isolate South Africa/KRISP-EC-K005325/2020 (B.1.351 South Africa variant) | 3.0E3 | 100% (3/3) | 34.4 ± 0.2 (0.5%) |
| SARS-CoV-2 Isolate England/204820464/2020 B.1.1.7 (UK variant) | 3.0E3 | 100% (3/3) | 35.7 ± 0.6 (1.6%) |
| SARS-CoV-2 USA/CA-Stanford- 15_S02/2021 B.1.617.1 (Kappa Variant) | 3.0E3 | 100% (3/3) | 37.1 ± 0.7 (2.0%) |
| SARS-CoV-2 USA/NY-Wadsworth- 21025952-01/2021 B.1.526_2021 (Iota Variant) | 3.0E3 | 100% (3/3) | 37.7 ± 1.0 (2.7%) |
| SARS-CoV-2 USA-WI 1/2020 (Lineage B) | 3.0E3 | 100% (3/3) | 36.5 ± 1.2 (3.4%) |
| SARS-CoV-2 NY-Wadsworth- 21006055-01/2021 Lineage P2_2021 (Zeta Variant) | 3.0E3 | 100% (3/3) | 38.3 ± 0.5 (1.4%) |
| SARS-Related Coronavirus 2 Isolate hCoV-1 9/USA/New York/PV96109/2023 (Omicron Variant JN.1) | 3.0E3 | 100% (3/3) | 34.5 ± 0.2 (0.6%) |
| SARS-CoV-2 Lineage JN.1.4 Omicron Variant USA/NY-Wadsworth-23068107-01/2023 | 3.0E3 | 100% (3/3) | 36.2 ± 0.2 (0.5%) |
| SARS-CoV-2 Isolate USA-CA3/2020 | 3.0E3 | 100% (3/3) | 38.4 ± 0.7 (1.9%) |
| SARS-CoV-2 Isolate Germany/BavPat1/2020 (Lineage B) | 3.0E3 | 100% (3/3) | 36.3 ± 0.7 (1.8%) |
| SARS-CoV-2 Isolate USA/CA/VRLC014/2021 PANGO Lineage B.1.429 (Epsilon) | 3.0E3 | 100% (3/3) | 35.5 ± 0.9 (2.6%) |
| SARS-CoV-2 Isolate hCoV-1 9/USA/VA-FBCH_675/2021 PANGO Lineage AY4.2 | 3.0E3 | 100% (3/3) | 36.0 ± 1.0 (2.8%) |
| SARS-CoV-2 Isolate SARS-CoV-2 Peru/un-CDC-2-4069945/2021 PANGO Lineage C.37 (Lambda) | 3.0E3 | 100% (3/3) | 35.3 ± 0.9 (2.4%) |
| SARS-CoV-2 isolate hCoV-191 USA/MD-HP38861/2022 PANGO Lineage B.1.1.529, BQ.1.1 (Omicron) | 3.0E3 | 100% (3/3) | 34.9 ± 0.3 (0.9%) |
| SARS-CoV-2 Isolate hCoV- 19/USA/CA-Stanford - 109_S21/2022 PANGO Lineage B.1.1.529, (XBB Omicron) | 3.0E3 (3x LoD) | 67% (2/3) | 39.2 ± 0.6 (1.5%) |
| | 4.0E3 (4x LoD) | 100% (3/3) | 37.6 ± 0.4 (1.1%) |
$\mathrm{Ct} =$ Cycle threshold, $\mathrm{SD} =$ Standard Deviation, $\% \mathrm{CV} =$ Percent Coefficient of Variation
This study demonstrates that the LIAISON NES FLU A/B, RSV & COVID-19 assay is able to detect multiple strains/variants for all four (4) viral targets.
# Cross-Reactivity / Microbial Interference
Cross-reactivity (analytical specificity) of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated. The purpose of this study was to determine the ability of the assay to not react with microorganisms that have a similar genome to influenza A virus, influenza B
K251978 - Page 16 of 30
{16}
virus, RSV and SARS-CoV-2 or normal and pathogenic flora that are reasonably likely to be present in the clinical sample.
For this study, forty-eight potential cross-reactants have been tested. Samples containing whole organisms or purified nucleic acid from bacteria, fungi, parasites and viruses quantified and titered in CFU/mL, TCID50/mL, cells/mL, cps/mL or other industry-accepted units were diluted in Pooled Negative Human Nasal Matrix and spiked on a dry swab. Each potential cross-reactant was tested in triplicate.
The results of the study are shown in Table 9. No cross-reactivity of LIAISON NES FLU A/B, RSV & COVID-19 was observed with any of the potential cross-reactants tested.
Table 9. Cross-Reactivity Results for Non-Target Microorganisms.
| Organism | Concentration tested | Influenza A | Influenza B | RSV | SARS-CoV-2 | IC |
| --- | --- | --- | --- | --- | --- | --- |
| | | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) |
| Bordetella parapertussis | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Bordetella pertussis | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Candida albicans | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Corynebacterium diphtheriae | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Escherichia coli | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Fusobacterium necrophorum | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Haemophilus influenzae | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Lactobacillus casei | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Legionella pneumophila | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Moraxella catarrhalis | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Neisseria elongata | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Neisseria gonorrhoeae | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Neisseria meningitidis | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Pneumocystis jirovecii | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Pseudomonas aeruginosa | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Staphylococcus epidermidis BAA-3171 | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Streptococcus pneumoniae | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Streptococcus pyogenes M1 | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Streptococcus salivarius | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Staphylococcus aureus | 1E6 CFU/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Aspergillus fumigatus | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Chlamydia pneumoniae | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Mycoplasma genitalium | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
K251978 - Page 17 of 30
{17}
| Mycoplasma pneumonia | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| --- | --- | --- | --- | --- | --- | --- |
| Mycobacterium tuberculosis (genomic DNA) | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Adenovirus 7A | 1E6 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Adenovirus Type 31 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| CMV AD-169 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Coronavirus 229E | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Coronavirus NL63 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Coronavirus OC43 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Enterovirus Type 68 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| hMPV 9 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Influenza C SendailTU2IO8 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Measles | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Mumps Virus | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Parainfluenza Virus Type 2 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Parainfluenza Virus Type 3 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Parainfluenza Virus Type 4A | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Parechovirus Type 1 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Rhinovirus Type 1A | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| EBV | 1E5 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| MERS Coronavirus (Florida/USA-2_Saudi Arabia 2014) | 1E5 cps/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Parainfluenza Virus Type 1 | 1E5 TCID50/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Coronavirus HKU1 (synthetic RNA) | 1E5 genome copies/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Leptospira (interrogans) | Quantification not provideda | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| SARS Coronavirus | Quantification not provideda | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Pooled human nasal wash | 10% | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
a Quantification not provided by the manufacturer. Maximum testable volume used.
Microbial interference of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated. The purpose of this study was to determine the ability of the assay to detect influenza A virus, influenza B virus, RSV and SARS-CoV-2 in nasal swab specimens that contain other bacteria, fungi, and viruses commonly found in the respiratory tract, using contrived dry nasal swab samples. Forty-seven microorganisms were tested for cross-reactivity and interference along with pooled human nasal wash.
The test samples consisted of contrived dry nasal swabs spiked with low concentrations (3x LoD) of influenza A virus, influenza B virus, RSV and SARS-CoV-2 respectively, prepared together with dilutions of bacteria, fungi, parasites and viruses from stocks that have been quantified and titered in CFU/mL, TCID50/mL, cells/mL or other industry-accepted units. Test samples were prepared in pooled negative human nasal matrix that was confirmed influenza A virus, influenza B virus, RSV and SARS-CoV-2-negative through screening.
K251978 - Page 18 of 30
{18}
Results of the microbial interference study are presented in Table 10.
Two instances of microbial interference with the detection of influenza B by the LIAISON NES FLU A/B, RSV & COVID-19 assay were observed. At 1E6 CFU/mL, Staphylococcus aureus completely inhibited the detection of influenza B (0/3). The assay was repeated with Staphylococcus aureus at 3.13E5 CFU/mL. Inhibition of Influenza B detection was no longer observed with $100\%$ of the influenza B samples detected.
At 1E5 $\mathrm{TCID}_{50} / \mathrm{mL}$ , Parainfluenza Virus Type 1 resulted in detection of $33\%$ (1/3) of influenza B samples. Repeating the assay after reducing the concentration of Parainfluenza to 3.31E4 $\mathrm{TCID}_{50} / \mathrm{mL}$ resulted in the detection of $100\%$ (3/3) of influenza B samples.
No microbial inhibition was observed with any of the other microorganisms tested.
Table 10. Microbial Interference Results for Non-Target Microorganisms.
| Organism | Concentration tested | Influenza A | Influenza B | RSV | SARS-CoV-2 | IC |
| --- | --- | --- | --- | --- | --- | --- |
| | | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) | % Detection (#Detected /#Tested) |
| Bordetella parapertussis | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Bordetella pertussis | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Candida albicans | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Corynebacterium diphtheriae | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Escherichia coli | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Fusobacterium necrophorum | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Haemophilus influenzae | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Lactobacillus casei | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Legionella pneumophila | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Moraxella catarrhalis | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Neisseria elongata | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Neisseria gonorrhoeae | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Neisseria meningitidis | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Pneumocystis jirovecii | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Pseudomonas aeruginosa | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Staphylococcus epidermidis BAA-3171 | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Streptococcus pneumoniae | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Streptococcus pyogenes M1 | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Streptococcus salivarius | 1E6 CFU/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Staphylococcus aureus* | 1E6 CFU/mL | 100% (3/3) | 0% (0/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| | 3.13E5 CFU/mL | N/A | 100% (3/3) | N/A | N/A | N/A |
K251978 - Page 19 of 30
{19}
K251978 - Page 20 of 30
| Aspergillus fumigatus | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| --- | --- | --- | --- | --- | --- | --- |
| Chlamydia pneumoniae | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Mycoplasma genitalium | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Mycoplasma pneumoniae | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Mycobacterium tuberculosis (genomic DNA) | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Adenovirus 7A | 1E6 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Adenovirus Type 31 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| CMV AD-169 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Coronavirus 229E | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Coronavirus NL63 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Coronavirus OC43 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Enterovirus Type 68 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| hMPV 9 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Influenza C SendaiITU2IO8 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Measles | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Mumps Virus | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Parainfluenza Virus Type 2 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Parainfluenza Virus Type 3 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Parainfluenza Virus Type 4A | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Parechovirus Type 1 | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Rhinovirus Type 1A | 1E5 TCID50/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| EBV | 1E5 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| MERS Coronavirus (Florida/USA-2 Saudi Arabia 2014) | 1E5 cps/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Parainfluenza Virus Type 1** | 1E5 TCID50/mL | 100% (3/3) | 33% (1/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| | 3.31E4 TCID50/mL | N/A | 100% (3/3) | N/A | N/A | N/A |
| Coronavirus HKU1 (synthetic RNA) | 1E5 genome copies/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Leptospira (interrogans) | Quantification not provided^{a} | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| SARS Coronavirus | Quantification not provided^{a} | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Pooled human nasal wash | 10% | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
{20}
* Inhibition of Influenza B detection was observed with Staphylococcus aureus tested at 1E6 CFU/mL. No inhibition was observed when a lower concentration of 3.13E5 CFU/mL was tested.
** Inhibition of Influenza B detection was observed with Parainfluenza Virus Type 1 tested at 1E5 TCID50/mL. No inhibition was observed when a lower concentration of 3.31E4 TCID50/mL was tested.
a Quantification not provided by the manufacturer. Maximum testable volume used.
# Interfering Substances
An analytical study was performed to assess the potential inhibitory effects of endogenous and exogenous substances that may be commonly found in ANS specimens. Low positive samples were prepared by spiking low doses (3x LoD) of influenza A virus, influenza B virus, RSV and SARS-CoV-2 in nasal swab specimens. For negative samples, pooled negative human nasal matrix was used. Each interfering substance was tested in triplicate, at the indicated concentration, for each target. Results for the negative sample study are presented in Table 11.
Table 11: LIAISON NES FLU A/B, RSV & COVID-19 Assay Interference for Negative Sample
| Potentially Interfering Substances | Active Ingredient | Interferent Concentration | Influenza A | Influenza B | RSV | SARS-CoV-2 | IC |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) |
| Afrin Nasal Spray | Oxymetazoline | 15% (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Antibacterial, systemic | Tobramycin | 0.004 mg/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Antibiotic, nasal ointment | Mupirocin | 6.6 mg/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Blood | N/A | 2% (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Purified Mucin Protein | Bovine submaxillary gland mucin, type I-S | 5 mg/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Nasal corticosteroid (Beconase AQ) | Beclomethasone | 5% (w/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| | Dexamethasone | 5% (w/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| | Mometasone | 5% (w/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Nasal corticosteroid (Fluticasone) | Fluticasone | 5% (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Tamiflu Antiviral drug | Oseltamivir | 0.001 mM | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Zicam Nasal Gel | Luffa operculata, Galphimia glauca, histaminum hydrochloricum | 5 % (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Budesonide | Budesonide | 5% (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Flunisolide | Flunisolide | 5% (v/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Relenza Antiviral Drug | Zanamivir | 3.3 mg/mL | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
| Throat lozenges | Avena sativa, Zinc gluconate, Sambucus nigra, Echinacea, Rose hips, Licorice root | 1.25% (w/v) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 0% (0/3) | 100% (3/3) |
{21}
The results for the positive sample study are presented in Table 12. One replicate of the RSV target was not detected when tested with Remdesivir at a concentration of $0.1\mathrm{mg / mL}$ . The test was repeated in triplicate at a reduced concentration (Remdesivir, $0.05\mathrm{mg / mL}$ ). No interference in the detection of the RSV target was observed at the lower concentration.
Table 12. LIAISON NES FLU A/B, RSV & COVID-19 assay Interference for Positive Sample
| Potentially Interfering Substances | Active Ingredient | Interferent Concentration | Influenza A | Influenza B | RSV | SARS-CoV-2 |
| --- | --- | --- | --- | --- | --- | --- |
| | | | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) | % Detection (#Detected/#Tested) |
| Afrin Nasal Spray | Oxymetazoline | 15% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Antibacterial, systemic | Tobramycin | 0.004 mg/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Antibiotic, nasal ointment | Mupirocin | 6.6 mg/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Blood | N/A | 2% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Purified Mucin Protein | Bovine submaxillary gland mucin, type I-S | 5 mg/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Nasal corticosteroid (Beconase AQ) | Beclomethasone | 5% (w/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| | Dexamethasone | 5% (w/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| | Mometasone | 5% (w/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Nasal corticosteroid (Fluticasone) | Fluticasone | 5% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Tamiflu Antiviral drug | Oseltamivir | 0.001 mM | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Zicam Nasal Gel | Luffa operculata, Galphimia glauca, histaminum hydrochloricum | 5% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Budesonide | Budesonide | 5% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Flunisolide | Flunisolide | 5% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Relenza Antiviral Drug | Zanamivir | 3.3 mg/mL | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Throat lozenges | Avena sativa, Zinc gluconate, Sambucus nigra, Echinacea, Rose hips, Licorice root | 1.25% (w/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
| Remdesivir | Remdesivir | 0.1 mg/mL | 100% (3/3) | 100% (3/3) | 67% (2/3)a | 100% (3/3) |
| | | 0.05 mg/mL | N/A | N/A | 100% (3/3) | N/A |
| Neo-Synephrine nasal spray | Phenylephrine hydrochloride | 15% (v/v) | 100% (3/3) | 100% (3/3) | 100% (3/3) | 100% (3/3) |
$\mathrm{mg / mL} =$ milligram/milliliter, $\mathrm{v / v} =$ volume per volume, $\mathrm{w / v} =$ weight per volume, $\mathrm{mM} =$ millimolar
K251978 - Page 22 of 30
{22}
0.1 mg/mL Remdesivir resulted in one out of three replicates 'Negative' for RSV. The concentration was reduced to
0.05\mathrm{mg / mL}
and retested and no interference was observed.
# Competitive Inhibition
A competitive interference study was performed to assess the ability of the LIAISON NES FLU A/B, RSV & COVID-19 assay to detect low concentration (3x LoD) of one target analyte in the presence of high concentration (1000x LoD) of another. Twelve target combinations were generated and each was tested in triplicate. A total of 38 tests (36 tests plus 2 repeats) were carried out by one operator in one day.
The four strains/isolates tested were:
Influenza A Victoria/4897/2022
Influenza B Austria/1359417/2021
- Respiratory Syncytial Virus A 2006 Isolate
SARS-CoV-2 USA/MDHP20874/2021
Results of the competitive inhibition assay are presented in Table 13. No evidence of competitive inhibition was observed.
Table 13. Results for Competitive Inhibition Target Combinations
| Target combinations | Influenza A | | Influenza B | | RSV | | SARS-CoV-2 | | Internal Control | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) | % Detection (#Detected/#Tested) | Mean Ct ± SD (%CV) |
| Influenza B HP/ Influenza A LP | 100% (3/3) | 32.2 ± 0.3 (1.1%) | 100% (3/3) | 25.4 ± 0.1 (0.6%) | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 24.6 ± 0.1 (0.2%) |
| RSVA HP/ Influenza A LP | 100% (3/3) | 33.0 ± 0.6 (1.7%) | 0% (0/3) | N/A | 100% (3/3) | 23.6 ± 0.2 (0.9%) | 0% (0/3) | N/A | 100% (3/3) | 24.5 ± 0.3 (1.2%) |
| SARS-CoV-2 HP/ Influenza A LP | 100% (3/3) | 32.3 ± 0.8 (2.5%) | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 25.2 ± 0.5 (1.8%) | 100% (3/3) | 24.8 ± 0.2 (0.9%) |
| Influenza A HP/ Influenza B LP | 100% (3/3) | 23.2 ± 0.9 (4.0%) | 100% (3/3) | 35.4 ± 0.6 (1.6%) | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 24.5 ± 0.2 (1.0%) |
| RSVA HP/ Influenza B LP* | 0% (0/3) | N/A | 100% (3/3) | 36.9 ± 1.8 (5.0%) | 100% (3/3) | 24.4 ± 0.1 (0.6%) | 0% (0/3) | N/A | 100% (3/3) | 24.4 ± 0.2 (0.9%) |
| SARS-CoV-2 HP/ Influenza B LP | 0% (0/3) | N/A | 100% (3/3) | 36.8 ± 1.9 (5.1%) | 0% (0/3) | N/A | 100% (3/3) | 25.4 ± 0.1 (0.2%) | 100% (3/3) | 25.0 ± 0.3 (1.2%) |
| Influenza A HP/ RSVA LP | 100% (3/3) | 23.7 ± 0.7 (2.9%) | 0% (0/3) | N/A | 100% (3/3) | 36.6 ± 0.6 (1.8%) | 0% (0/3) | N/A | 100% (3/3) | 25.3 ± 0.2 (1.0%) |
| Influenza B HP/ RSVA LP | 0% (0/3) | N/A | 100% (3/3) | 25.3 ± 0.3 (1.0%) | 100% (3/3) | 35.7 ± 0.5 (1.3%) | 0% (0/3) | N/A | 100% (3/3) | 24.9 ± 0.3 (1.2%) |
| SARS-CoV-2 HP/ RSVA LP* | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 36.9 ± 1.1 (3.0%) | 100% (3/3) | 25.7 ± 0.4 (1.5%) | 100% (3/3) | 25.3 ± 0.1 (0.6%) |
| Influenza A HP/ SARS-CoV-2 LP | 100% (3/3) | 23.2 ± 0.1 (0.5%) | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 35.7 ± 0.4 (1.2%) | 100% (3/3) | 25.0 ± 0.1 (0.6%) |
K251978 - Page 23 of 30
{23}
| Influenza B HP/ SARS-CoV-2 LP | 0% (0/3) | N/A | 100% (3/3) | 25.3 ± 0.2 (0.7%) | 0% (0/3) | N/A | 100% (3/3) | 37.3 ± 0.4 (1.0%) | 100% (3/3) | 25.0 ± 0.2 (0.7%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| RSVA HP/ SARS-CoV-2 LP | 0% (0/3) | N/A | 0% (0/3) | N/A | 100% (3/3) | 25.2 ± 0.3 (1.4%) | 100% (3/3) | 37.3 ± 1.4 (3.7%) | 100% (3/3) | 25.0 ± 0.1 (0.4%) |
Ct = Cycle threshold, SD = Standard Deviation, %CV = Percent Coefficient of Variation, N/A = Not Applicable
* One (1) run was repeated due to an instrumental error (a total of 4 runs were carried out)
4. Assay Reportable Range:
Not applicable; this is a qualitative assay.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
a. Controls
The assay contains an Internal Control (IC) added to each test specimen and external positive and negative controls. For more information, see Section IV.C.5. Quality Control, above.
b. Sample Stability
Sample storage and handling were evaluated for LIAISON NES FLU A/V, RSV & COVID-19 to support the following:
- After sample collection, the anterior nasal swab should be stored in its tube and processed within 120 minutes at 15-30°C ambient temperature if immediate testing is not possible.
c. Device Stability (Open pouch)
- The LIAISON NES FLU A/B, RSV & COVID-19 Cartridge is stable for up to 2 hours after pouch opening at 15-30°C and 80% RH.
- The LIAISON NES FLU A/B, RSV & COVID-19 Positive and negative control swabs must be used immediately after opening the sleeved pouch.
6. Detection Limit:
Limit of detection (LoD) studies were conducted to determine the lowest detectable concentration of influenza A virus, influenza B virus, RSV and SARS-CoV-2 at which ≥95% of all replicates test positive.
The LoD was established with serial dilutions of two quantified viral stocks for each target (influenza A virus, influenza B virus, RSV and SARS-CoV-2) in pooled negative human nasal matrix. Subsequently, 10 μL of the dilution was spiked onto dry swabs for testing.
Confirmation of the LoD was performed using a minimum of twenty replicates for each of three cartridge lots. The strains evaluated, as well as their corresponding LoD values are shown in Table 14.
K251978 - Page 24 of 30
{24}
Table 14. LoD Determination for Influenza A, Influenza B, RSV and SARS-CoV-2
| Target | Strain/Isolate | LoD Concentration (Cps/swab) | Positivity |
| --- | --- | --- | --- |
| Influenza A | Victoria/4897/2022 (H1N1) | 6E3 | 98% (59/60) |
| | Darwin/9/21 (H3N2) | 8E3 | 98% (59/60) |
| Influenza B | Austria/1359417/2021 (Victoria) | 4E3 | 98% (59/60) |
| | Phuket/3073/2013 | 5E3 | 98% (59/60) |
| Respiratory Syncytial Virus A | Isolate 2006 | 8E3 | 98% (59/60) |
| Respiratory Syncytial Virus B | Isolate 12/2014 | 8E3 | 98% (59/60) |
| SARS-CoV-2 | USA/WA 1/2020 | 1E3 | 98% (59/60) |
| | USA/MDHP20874/2021 (B.1.1.529 Omicron) | 1E3 | 98% (59/60) |
7. Assay Cut-Off:
For each channel, the respective target is detected if (1) the fluorescence signal crosses the fluorescence threshold, and (2) the cycle at which it crosses is at/before the last cycle (the Ct threshold, 45 cycles).
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over/Cross-Contamination:
The carry-over and cross-contamination rate for the LIAISON NES FLU A/B, RSV & COVID-19 assay was assessed through testing of alternating high positive samples and negative samples between successive runs. The high positive sample was prepared by spiking SARS-CoV-2 USA/MDHP20874/2021 into pooled negative human nasal matrix at a final concentration of 1E8 copies/mL (1000x LoD). The negative sample consisted of pooled negative human nasal matrix spiked onto the swab. A total of 107 experimental runs were performed by one operator over the course of four non-consecutive days.
Results of the study adequately demonstrate that alternating between highly positive and negative samples does not result in cross-contamination.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not Applicable
2. Matrix Comparison:
Not Applicable
C Clinical Studies:
1. Clinical Sensitivity:
K251978 - Page 25 of 30
{25}
# Prospective Clinical Study
The clinical performance of the LIAISON NES FLU A/B, RSV & COVID-19 assay was evaluated using clinical specimens prospectively collected between October 2024 and March 2025 from seven geographically diverse clinical sites located within the United States. The clinical study utilized prospective specimens collected from pediatric and adult patients exhibiting clinical signs and symptoms of respiratory tract infection. Anterior nasal swabs (ANS) were professionally collected by a Healthcare Provider or self-collected under the Healthcare Provider's supervision. All collection and assay testing were performed at CLIA-waived facilities, representative of the intended use, point-of-care environment.
## Subject/Specimen Enrollment
A total of 1692 unique prospectively collected specimens that met the pre-determined inclusion criteria, were enrolled in the study. 101 specimens were excluded from performance evaluation. Of these 101 (6.0%) specimens,
- 55 specimens did not have a reference method result due to fresh specimen testing time window violations,
- 19 specimens were excluded due to investigational device testing occurring outside of the allowed time window,
- 19 specimens were excluded due to improper sample collection,
- 5 specimens were excluded for sample handling issues,
- 1 specimen was excluded due to operator error,
- 1 specimen was excluded due to patient withdrawing consent, and
- 1 patient for whom the reference method testing swab could not be collected.
This resulted in 1591 specimens that were eligible for performance evaluation. The demographic data for the study population are shown in Table 15. Clinical sample testing using the LIAISON NES FLU A/B, RSV & COVID-19 assay was performed using the LIAISON NES device by untrained operators at each of the seven collection sites.
Table 15: Demographic Data for the Prospective Study Population (N=1591)
| | Site 01 | Site 02 | Site 03 | Site 04 | Site 05 | Site 06 | Site 13 | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sex | | | | | | | | |
| Male | 191 (53.1%) | 129 (42.7%) | 104 (53.3%) | 31 (40.3%) | 99 (46.3%) | 107 (51.4%) | 100 (42.6%) | 761 (47.8%) |
| Female | 169 (46.9%) | 173 (57.3%) | 91 (46.7%) | 46 (59.7%) | 115 (53.7%) | 101 (48.6%) | 135 (57.4%) | 830 (52.2%) |
| Total | 360 (100.0%) | 302 (100.0%) | 195 (100.0%) | 77 (100.0%) | 214 (100.0%) | 208 (100.0%) | 235 (100.0%) | 1591 (100.0%) |
| Age (years) | | | | | | | | |
| 0-1 | 47 (13.1%) | 33 (10.9%) | 42 (21.5%) | 0 (0.0%) | 9 (4.2%) | 10 (4.8%) | 5 (2.1%) | 146 (9.2%) |
K251978 - Page 26 of 30
{26}
After a single test of each specimen, 1571 specimens generated valid LIAISON NES results for a success rate of $98.7\%$ (1571/1591). There were 20 specimens $(1.3\%)$ with invalid results. Due to sample type limitations, repeat testing of specimens with invalid results was not possible. Table 16 presents the number of positives identified stratified by clinical site and operator.
Table 16. Operator Testing Summary
| Clinical Site # | Operator # | Samples Tested | # of Positives | % Positive | # of Invalids | % Invalid |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 21 | 358 | 122 | 34.1% | 3 | 0.8% |
| | Unknown2 | 2 | 0 | 0.0% | 2 | 100% |
| | Subtotal | 360 | 122 | 33.9% | 5 | 1.4% |
| 2 | 1 | 244 | 80 | 32.8% | 1 | 0.4% |
| | 2 | 58 | 22 | 37.9% | 0 | 0.0% |
| | Subtotal | 302 | 102 | 33.8% | 1 | 0.3% |
| 3 | 1 | 21 | 9 | 42.9 % | 2 | 9.5% |
| | 2 | 30 | 15 | 50.0 % | 0 | 0.0% |
| | 3 | 16 | 10 | 62.5 % | 1 | 6.3% |
| | 4 | 73 | 38 | 52.1 % | 0 | 0.0% |
| | 5 | 24 | 17 | 70.8 % | 0 | 0.0% |
| | 6 | 29 | 15 | 51.7 % | 1 | 3.4% |
| | Unknown3 | 2 | 0 | 0.0% | 2 | 100% |
| Subtotal | 195 | 104 | 53.3% | 6 | 3.1% | |
K251978 - Page 27 of 30
{27}
For each target in the assay panel, the diagnostic performance of the LIAISON NES FLU A/B, RSV & COVID-19 assay was compared to the reference method to determine the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA), along with the associated 95% confidence intervals. PPA was calculated as 100% × (TP / (TP + FN)). True positive (TP) indicates that both the assay and the comparator method had a positive result for the specific analyte, and false negative (FN) indicates that the assay was negative while the comparator result was positive. NPA was calculated as 100% × (TN / (TN + FP)). True negative (TN) indicates that both the assay and the comparator method produced negative results, and false positive (FP) indicates that the assay was positive while the comparator result was negative. The results for the sample analysis are summarized in Table 17.
The Clinical Study results are acceptable.
Table 17. Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) (N=1571)
| Pathogen Target | Positive Percent Agreement | | | Negative Percent Agreement | | |
| --- | --- | --- | --- | --- | --- | --- |
| | TP / (TP+FN) | PPA (%) | 95% CI | TN / (TN+FP) | NPA (%) | 95% CI |
| Influenza A | 384/398a | 96.5% | 94.2%-97.9% | 1140/1173b | 97.2% | 96.1%-98.0% |
| Influenza B | 34/35c | 97.1% | 85.5%-99.5% | 1531/1536d | 99.7% | 99.2%-99.9% |
| Respiratory Syncytial Virus | 102/113e | 90.3% | 83.4%-94.5% | 1455/1458f | 99.8% | 99.4%-99.9% |
| SARS-CoV-2 | 63/64g | 98.4% | 91.7%-99.7% | 1498/1507h | 99.4% | 98.9%-99.7% |
{28}
Eight (8) of the 14 Influenza A False Negative specimens were negative by Standard of Care.
bTwelve (12) of the 33 Influenza A False Positive specimens were positive by bi-directional sequencing (BDS).
c The one (1) Influenza B False Negative specimen was negative by BDS.
Two (2) of the five (5) Influenza B False Positive specimens were positive by Standard of Care.
One (1) of the 11 RSV False Negative specimens was negative by BDS, and another two (2) specimens were negative by Standard of Care.
Two (2) of the three (3) RSV False Positive specimens were positive by another FDA-cleared molecular assay.
The one (1) SARS-CoV-2 False Negative specimen was negative by Standard of Care.
${}^{\mathrm{b}}$ Five (5) of the nine (9) SARS-CoV-2 False Positive specimens were positive by Standard of Care.
Observed co-infection combinations are shown in Table 18. Of the four specimens in which both influenza A and COVID-19 were detected by the LIAISON NES FLU A/B, RSV & COVID-19 assay, influenza A was detected in 4/4 specimens by the comparator and SARS-CoV-2 was detected in 1/4 samples by the comparator. Of the 6 specimens in which both influenza A and RSV were detected by the assay, influenza A was detected in 5/6 specimens by the comparator and RSV was detected in 5/6 specimens by the comparator.
Table 18. Observed Co-infection Combinations by the LIAISON NES FLU A/B, RSV & COVID-19 Assay Stratified by Age (N=1571)
| Age (years) | 0-1 Years (N=145) | 2-5 Years (N=246) | 6-21 Years (N=679) | 22-65 Years (N=439) | >65 Year (N=62) | Overall (N=1571) |
| --- | --- | --- | --- | --- | --- | --- |
| Co-infection | #POS | % | #POS | % | #POS | % | #POS | % | #POS | % | #POS | % |
| Influenza A | SARS-CoV-2 | 0 | 0.00% (0/145) | 0 | 0.00% (0/246) | 3 | 0.44% (3/679) | 1 | 0.23% (1/439) | 0 | 0.00% (0/62) | 4 | 0.25% (4/1571) |
| Influenza A | RSV | 1 | 0.68% (1/145) | 2 | 0.79% (2/246) | 1 | 0.15% (1/679) | 1 | 0.23% (1/439) | 1 | 1.61% (1/62) | 6 | 0.38% (6/1571) |
2. Clinical Specificity: See Section "Clinical Sensitivity" above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.
# D Clinical Cut-Off:
Not applicable.
# E Expected Values/Reference Range:
The LIAISON NES FLU A/B, RSV & COVID-19 assay prospective clinical study included a total of 1,591 prospectively collected ANS specimens, of which 1,571 specimens were evaluable. The number and percentage of cases positive for influenza A virus, influenza B virus, RSV, and SARS-CoV-2, as determined by the test, are presented in Table 19, stratified by collection site, and by age in Table 20.
Table 19. Expected Values for Prospective Specimens by Site $\left( {\mathrm{N} = {1571}}\right) {}^{ \dagger }$
| | Site 1 (N=355) | Site 2 (N=301) | Site 3 (N=189) | Site 4 (N=75) | Site 5 (N=213) | Site 6 (N=205) | Site 13 (N=233) | Total (N=1571) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
K251978 - Page 29 of 30
{29}
| Target | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Influenza A | 78 | 22.0% (78/355) | 58 | 19.3% (58/301) | 81 | 42.9% (81/189) | 10 | 13.3% (10/75) | 76 | 35.7% (76/213) | 57 | 27.8% (57/205) | 57 | 24.5% (57/233) | 417 | 26.5% (417/1571) |
| Influenza B | 7 | 2.0% (7/355) | 18 | 6.0% (18/301) | 2 | 1.1% (2/189) | 0 | 0.0% (0/75) | 3 | 1.4% (3/213) | 5 | 2.4% (5/205) | 4 | 1.7% (4/233) | 39 | 2.5% (39/1571) |
| RSV | 35 | 9.9% (35/355) | 14 | 4.7% (14/301) | 12 | 6.3% (12/189) | 2 | 2.7% (2/75) | 6 | 2.8% (6/213) | 16 | 7.8% (16/205) | 20 | 8.6% (20/233) | 105 | 6.7% (105/1571) |
| SARS-CoV-2 | 4 | 1.1% (4/355) | 14 | 4.7% (14/301) | 10 | 5.3% (10/189) | 9 | 12.0% (9/75) | 9 | 4.2% (9/213) | 15 | 7.3% (15/205) | 11 | 4.7% (11/233) | 72 | 4.6% (72/1571) |
1Twenty (20) specimens out of 1591 returned invalid results on the LIAISON NES FLU A/B, RSV & COVID-19 Assay and were excluded from this analysis.
Table 20. Expected Values for Prospective Specimens by Age (N=1571)¹
| | 0-1 Years (N=145) | | 2-5 Years (N=246) | | 6-21 Years (N=679) | | 22-65 Years (N=439) | | >65 Year (N=62) | | Overall (N=1571) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Target | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) | #Pos | (%) |
| Influenza A | 31 | 21.4% (31/145) | 81 | 32.9% (81/246) | 184 | 27.1% (184/679) | 106 | 24.1% (106/439) | 15 | 24.2% (15/62) | 417 | 26.5% (417/1571) |
| Influenza B | 0 | 0.0% (0/145) | 4 | 1.6% (4/246) | 29 | 4.3% (29/679) | 6 | 1.4% (6/439) | 0 | 0.0% (0/62) | 39 | 2.5% (39/1571) |
| RSV | 23 | 15.9% (23/145) | 41 | 16.7% (41/246) | 29 | 4.3% (29/679) | 6 | 1.4% (6/439) | 6 | 9.7% (6/62) | 105 | 6.7% (105/1571) |
| SARS-CoV-2 | 8 | 5.5% (8/145) | 4 | 1.6% (4/246) | 23 | 3.4% (23/679) | 30 | 6.8% (30/439) | 7 | 11.3% (7/62) | 72 | 4.6% (72/1571) |
¹Twenty (20) specimens out of 1591 returned invalid results on the LIAISON NES FLU A/B, RSV & COVID-19 Assay and were excluded from this analysis.
## F Other Supportive Instrument Performance Characteristics Data:
Not applicable
## VIII Proposed Labeling:
The labeling supports or the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
The submitted information in this CLIA waiver application supports a CLIA waiver approval decision.
K251978 - Page 30 of 30
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.