The Stryker Robotic Precision System (RPS) Primary TKA Software application, when used with the Stryker Guidance Systems, is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to rigid anatomical structures during orthopedic procedures. The RPS Primary TKA Software application is indicated for use in total knee arthroplasty (TKA) procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified.
Device Story
RPS is a semi-active robotic bone preparation system for TKA; supports intraoperative planning, kinematic analysis, boundary control, and guided resection. System inputs include intraoperative localization data of patient anatomy. RPS Console processes data to control three actuators on a handheld robotic saw (pitch, elevation, roll) to maintain planned cut planes. Surgeon controls remaining degrees of freedom. If saw position exceeds actuator range, saw deactivates. Used in OR by trained medical professionals. Output is a guided bone resection for Triathlon CR/PS implants or 4-in-1 cut guides. Benefits include precise bone preparation and alignment. Healthcare providers use system to visualize anatomy and surgical plan, ensuring resections match planned spatial boundaries.
Clinical Evidence
No clinical data. Bench testing only. System accuracy evaluated: <2.0° limb alignment error, <2.0 mm gap attainment error, and cut error <1 mm and <1° (90% of values at 95% confidence).
Indicated for use in total knee arthroplasty (TKA) procedures where stereotactic surgery is appropriate and rigid anatomical bony structures can be identified.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 25, 2025
Stryker Instruments
Susanne Galin
Chief Regulatory Affairs Specialist
1941 Stryker Way
Portage, Michigan 49002
Re: K251971
Trade/Device Name: RPS Primary TKA Software; RPS Console; RPS Saw One; RPS Cable; RPS Saw Oscillating Tip Cartridge; RPS Console Cart Attachment; Ball Tip Pointer
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: June 26, 2025
Received: August 29, 2025
Dear Susanne Galin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251971 - Susanne Galin
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K251971 - Susanne Galin
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251971 | |
| Device Name RPS Primary TKA Software; RPS Console; RPS Saw One; RPS Cable; RPS Saw Oscillating Tip Cartridge; RPS Console Cart Attachment; Ball Tip Pointer | |
| Indications for Use (Describe) The Stryker Robotic Precision System (RPS) Primary TKA Software application, when used with the Stryker Guidance Systems, is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to rigid anatomical structures during orthopedic procedures. The RPS Primary TKA Software application is indicated for use in total knee arthroplasty (TKA) procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Stryker Instruments
510(k) Summary
# 510(k) Summary
| 807.92(a)(1) Submitter Information | |
| --- | --- |
| 510(k) Submitter | Stryker Instruments
1941 Stryker Way
Portage, MI 49002 USA |
| Contact Information | Primary:
Susanne Galin, RAC
Chief Regulatory Affairs Specialist
susanne.galin@stryker.com
Secondary:
Megan Guilbault
Senior Staff Regulatory Affairs Specialist
megan.guilbault@stryker.com |
| Date Summary Prepared | 26 June 2025 |
| 807.92(a)(2) Name of Device(s) | |
| --- | --- |
| Trade Name(s) and Manufacturer | • Stryker Robotic Power System (RPS) Primary TKA Software (Stryker Instruments)
• RPS Console, RPS Console Cart Attachment (Stryker Instruments)
• RPS Cable (Stryker Instruments)
• RPS Saw One (Stryker Instruments)
• RPS Saw Oscillating Tip Cartridge (Stryker Instruments)
• Ball Tip Pointer (Stryker Leibinger) |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Device Class | Class II Premarket Notification, 21 CFR 882.4560 |
| Product Code | OLO |
| 807.92(a)(3) Legally marketed device(s) to which equivalence is claimed | |
| --- | --- |
| Predicate: Real Intelligence™ CORI™ | K240139 |
| Reference: Mako Total Knee Application | K241011 |
| Reference: Ortho Guidance Precision Knee System | K233542 |
Traditional 510(k)
{5}
Stryker Instruments
510(k) Summary
## 807.92(a)(4) Device Description
**Stryker Robotic Power System (RPS) Overview**
RPS is a total joint bone preparation system to support total knee arthroplasty (TKA) consisting of functionality for intraoperative planning, kinematic analysis, boundary control, and guided resection. The Complete Workflow allows for the placement of Stryker Triathlon CR/PS Total Knee Implants (cemented or cementless), and plans/resects all 6 primary femoral/tibial resections. The Quick-Cut Workflow allows for placement of implants with implant-specific cut guides requiring primary distal femoral and proximal tibial cuts.
A handheld robotic saw is guided via camera visualization in relation to patient anatomy and the surgical plan established intraoperatively. The robotic saw movement is partially controlled by the RPS Console which determines, based on surgical plan and localization data, how the saw actuators need to move within three degrees of freedom (elevation, pitch, roll) to maintain the planned cut plane. The surgeon controls the remaining degrees of freedom.
## 807.92(a)(5) Intended Use/Indications of Use of the Device
The Stryker Robotic Power System (RPS) Primary TKA Software application, when used with the Stryker Guidance Systems, is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to rigid anatomical structures during orthopedic procedures.
The RPS Primary TKA Software application is indicated for use in total knee arthroplasty (TKA) procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified.
## 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to the Predicate
| Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| Intended Use | The Stryker Robotic Power System (RPS) Primary TKA Software application, when used with the Stryker Guidance Systems, is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to rigid anatomical structures during orthopedic procedures. | Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. |
| Indications for Use | The RPS Primary TKA Software application is indicated for use in total knee arthroplasty (TKA) | Indicated for use in surgical procedures in which the use of stereotactic surgery may be |
Traditional 510(k)
2 of 6
{6}
Stryker Instruments
510(k) Summary
| | procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified. | appropriate, and where reference to rigid anatomical body structures can be determined. These procedures include:
• Unicondylar knee replacement (UKR)
• Total knee arthroplasty (TKA)
• Revision knee arthroplasty
• Total hip arthroplasty (THA) |
| --- | --- | --- |
| **System Elements** | • Cart mounted guidance systems
• Anatomy and device trackers
• Software application
• Pointer
• Console/Cart Attachment
• Saw/Cable
• Cutting Accessory – cartridge-style oscillating blade | • Guidance cart
• Anatomy and device trackers
• Software application
• Pointer
• Console
• Drill/Cable
• Foot pedal
• Cutting Accessory - bur |
| **Modes of Operation** | • System set-up
• Calibration
• Anatomy registration
• Initial limb evaluation
• Implant planning
• Navigated bone resection
• Freehand bone resection
• Trial/final implant evaluation | • System set-up
• Calibration
• Anatomy registration
• Initial limb evaluation
• Implant planning
• Navigated bone resection
• Freehand bone resection
• Trial/final implant evaluation |
| **Operating Principle** | RPS uses established methods and technologies to prepare bone for attachment of identified TKA implants. The bone surface may also be prepared to receive 4-in-1 cutting guides for other implants.
RPS uses intraoperative data collection to create a model of the patient’s femur and tibia in the RPS tracking space and allows the surgeon to prepare and modify the surgical plan.
RPS uses a tracked saw whose head sits atop three actuators that adjust the positioning/alignment of the cutting accessory based on its location as it approaches the | For knee applications, CORI uses established technologies to prepare bone for attachment of TKA implant components. The bone surface may also be prepared to receive 4-in-1 cut guides.
CORI uses either pre-operative data (for image-based cases) or intraoperative data collection (for image-free or non-CT data generation cases) to create a model of the patient’s femur and tibia in CORI tracking space and allows the surgeon to prepare/modify the surgical plan.
The system uses predefined boundaries generated during the planning process to control the motion |
Traditional 510(k)
3 of 6
{7}
Stryker Instruments
510(k) Summary
| | patient bone to be resected per the surgical plan. When actuators are unable maintain the cut plane due to gross positioning of the saw, the saw will not activate or will deactivate. When gross positioning of the saw is adequate for actuators to maintain the cut plane, the saw can be activated. | of the surgical bur and limit the amount of bone removed to shape the operative condyle or tibial plateau in preparation for placement of the surgical implant. |
| --- | --- | --- |
| Preoperative Imaging Requirements | None – Uses image-free, intraoperative planning only | Preoperative/image-based or intraoperative/image-free planning available |
| Intended Users/Use Environment | RPS is intended to be used by trained medical professionals in an operating room setting of a hospital or ambulatory surgery center | CORI is intended to be used by trained medical professionals in a hospital or clinical setting equivalent to an orthopedic surgery suite. |
| Robotic Type | Semi-active | Semi-active |
| Compatible Implants | Triathlon CR/PS Total Knee Implants, cemented and cementless (Complete Workflow)
Implants with implant-specific cut guide requiring primary distal femoral and proximal tibial cuts (Quick-Cut Workflow) | Smith+Nephew knee implants as defined in product labeling, cemented and cementless models |
| Robotic Handpiece Type | Saw | Drill |
| Robotic Handpiece Motion | Three degrees of freedom for pitch, elevation, and roll | Single degree of freedom – bur retracts to limit exposure |
| Cutting Accessory Type | Oscillating blade (cartridge-style) | Bur |
| 807.92(b)(1) Nonclinical Testing to Support Submission |
| --- |
| The function and performance of the subject devices have been evaluated through non-clinical design verification and validation testing. Additional testing was performed on the subject devices to ensure that they meet their design requirements. The results of the tests demonstrate that the subject devices meet the requirements for their intended use.
**Accuracy**
Overall system accuracy was evaluated and determined to be <2.0° limb alignment error, <2.0 mm gap attainment error, and cut error <1 mm and <1° (90% of values at 95% confidence). |
Traditional 510(k)
4 of 6
{8}
Stryker Instruments
510(k) Summary
## Software
Software verification and validation testing was conducted as required by IEC 62304 and FDA Guidance “Guidance for the Content of Premarket Submissions for Device Software Functions.” All requirements were met and no new questions of safety or effectiveness were identified.
## Cybersecurity
Cybersecurity testing was conducted in accordance with FDA Guidance “Cybersecurity of Networked Medical Devices Containing Off-the-Shelf Software (Jan 2025),” FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (Sep 2023),” AAMI TIR57, and NIST SP 800-30 Rev.1. All requirements were met and no new questions of safety or effectiveness were identified.
## Biocompatibility
Biocompatibility testing of patient-contacting components was conducted in accordance with ISO 10993-1. All requirements were met and no new questions of safety or effectiveness were identified.
## Reprocessing/Sterilization
Reprocessing and sterilization validation (moist heat) covering critical devices was successfully performed in accordance with AAMI ST98, AAMI TIR12, ISO 17664-1, AAMI TIR39, AAMI ST70, ISO 17665-1, and FDA Guidance “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (2015).” All requirements were met and no new questions of safety or effectiveness were identified.
Cleaning validation was performed for non-sterile devices was successfully performed in accordance with AAMI TIR12, AAMI ST98, ISO 17664-2, and FDA Guidance “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (2015).” All requirements were met and no new questions of safety or effectiveness were identified.
Sterility Testing for terminally sterile RPS Saw Oscillating Tip Cartridge was successfully performed in accordance with BS EN 556-1, BS EN ISO 11137-1, BS EN ISO 11137-2, BS EN ISO 11137-3. All requirements were met and no new questions of safety or effectiveness were identified.
Shelf-Life Evaluation for Terminally Sterile RPS Saw Oscillating Tip Cartridge was successfully performed in accordance with ASTM F1980-21, ISO 11607-1. All requirements were met and no new questions of safety or effectiveness were identified.
## Electrical Safety and EMC
Electrical Safety and EMC testing was successfully performed in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-4-2, and IEC 80601-2-77. All requirements were met and no new questions of safety or effectiveness were identified.
Traditional 510(k)
5 of 6
{9}
Stryker Instruments
510(k) Summary
## Human Factors Evaluation
Human factors were evaluated in accordance with IEC 62366-1 and FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices (2016).” All requirements were met and no new questions of safety or effectiveness were identified.
## 807.92(b)(2) Clinical Testing
No clinical testing was required to support this submission.
## 807.92(b)(3) Conclusions Drawn from Testing Performed
The results of verification and validation testing were determined to be successful for all protocols and demonstrate the functionality, integrity, and safety and effectiveness of the subject devices are sufficient for their intended use. Test results for the subject devices support a determination of substantial equivalence to the predicate devices.
## Substantial Equivalence Rationale
The subject devices, in comparison with the legally marketed predicates with supporting similarities in comparison to the two reference systems, have the same or equivalent intended use, indications for use, operating principle, technical characteristics, and functional outputs. Risk analysis and verification/validation testing support that any differences between the subject and predicate don’t raise different types of questions of safety or effectiveness, and that the subject devices are at least as safe and effective as the predicate. A determination of substantial equivalence is supported.
Traditional 510(k)
6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.