← Product Code [PGM](/productcode/PGM) · K251961

# SDS Growing Rod (K251961)

_BAAT Medical Products B.V. · PGM · Jan 15, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251961

## Device Facts

- **Applicant:** BAAT Medical Products B.V.
- **Product Code:** [PGM](/productcode/PGM.md)
- **Decision Date:** Jan 15, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K251961 · Jan 15, 2026 | SDS Growing Rod | BAAT Medical Products B.V. | Retrospective clinical data | Retrospective clinical data were used to support the safety and effectiveness profile of the SDS Growing Rod by comparing outcomes to predicate devices in the same patient population. | Retrospective clinical data; Clinical outcomes; Safety profile |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical study | Skeletally immature patients less than 10 years of age with severe, progressive, life-threatening early-onset deformities | Predicate pedicle screw-based growing rod systems | Safety and effectiveness profile; mitigation of risks associated with metallic components (tissue ingrowth, wear/corrosion byproducts, irritation) |

## Indications for Use

The SDS Growing Rod is indicated for skeletally immature patients less than 10 years of age with the potential for additional spinal growth who require surgical treatment for correction and maintenance of the correction of severe, progressive, life-threatening early-onset deformities, including early-onset scoliosis, which are associated with or at risk of thoracic insufficiency syndrome.

## Device Story

The SDS Growing Rod is a spinal implant system designed for surgical correction and active lengthening of the spine in children with early-onset spinal deformities. The device consists of 5.5 mm rods, titanium alloy springs, sliding bearings (axial and polyaxial), and UHMWPE bushings. It is implanted by a surgeon to provide a distraction force via the spring mechanism to correct spinal curvature, while the sliding bearing allows for longitudinal displacement to accommodate natural spinal growth. The system is used in conjunction with pedicle screws (e.g., RESPONSE system) and dedicated instrumentation. The device remains in the patient for up to 5 years, after which it is removed or locked. By enabling growth without requiring repeated surgical lengthening procedures, the device aims to maintain spinal correction and mitigate the risks associated with thoracic insufficiency syndrome.

## Clinical Evidence

Retrospective clinical data provided. Analysis demonstrated a safety and effectiveness profile substantially equivalent to the predicate. Evaluated risks included tissue ingrowth/encapsulation, local tissue reactions from wear/corrosion, and irritation from mechanical elements. Non-clinical testing included mechanical integrity verification per ASTM F1717 and F1798, and wear testing per ASTM F2624. Biocompatibility was assessed per ISO 10993-1.

## Technological Characteristics

Materials: Titanium alloy, CoCr, UHMWPE. Components: 5.5 mm rods, springs, axial/polyaxial sliding bearings, C-clamps. Principle: Mechanical distraction via springs and sliding bearings to allow longitudinal growth. Connectivity: None (mechanical implant). Sterilization: Sterile, single-use. Standards: ASTM F1717 (vertebrectomy model), ASTM F1798 (interconnection properties), ASTM F2624 (wear), ISO 10993-1 (biocompatibility).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- SHILLA™ Growth Guidance System ([K140750](/device/K140750.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 15, 2026

BAAT Medical Products B.V.
Jasper Springer
Regulatory Affairs Officer
F. Hazemeijerstraat 800
Building A04
Hengelo, Overijssel 7555 RJ
Netherlands

Re: K251961
Trade/Device Name: SDS Growing Rod
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral Pedicle Screw System
Regulatory Class: Class II
Product Code: PGM
Dated: December 15, 2025
Received: December 16, 2025

Dear Jasper Springer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251961 - Jasper Springer
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251961 - Jasper Springer
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O’Neill -S

Colin O’Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251961

Device Name
SDS™ Growing Rod

Indications for Use (Describe)
The SDS Growing Rod is indicated for skeletally immature patients less than 10 years of age with the potential for additional spinal growth who require surgical treatment for correction and maintenance of the correction of severe, progressive, life-threatening early-onset deformities, including early-onset scoliosis, which are associated with or at risk of thoracic insufficiency syndrome.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 1 of 1
P2C Publishing Services (301) 443-6740

{4}

BAT MEDICAL
K251961

# 510(k) Summary

## 1. Applicant/Submitter

**Submitter Name:** BAAT Medical Products B.V.
**Submitter Address:** F. Hazemeijerstraat 800 - Building A04
7555 RJ Hengelo
The Netherlands

**Phone Number:** +31-(0)88-5656600
**Contact person:** Jasper Springer
**Date Prepared:** 26th of June, 2025

## 2. Device

**Device Trade name:** SDS Growing Rod
**Common Name:** Not Applicable
**Classification:** 888.3070 - Thoracolumbosacral pedicle screw system
**Product Code/s:** PGM - Growing Rod System
**Review Panel:** Orthopedic devices

## 3. Predicate Device

SHILLA™ Growth Guidance System (K140750)

## 4. Description of the Device

The SDS Growing Rod is an implant designed to correct and actively lengthen the spine in children with early onset spinal deformities while allowing growth, without the need for re-operations or other interventions. This is achieved by a spinal construct with one or two springs and sliding bearings. The spring provides a distraction force to correct the spinal curve while the sliding bearing is designed to enable growth by allowing displacement of one rod with respect to the other rod.

The SDS Growing Rod is available with 5.5 mm rods. Two configurations of the SDS Growing Rod are possible: hybrid and bilateral. The SDS Growing Rod is delivered sterile and is for single use only. The intended lifetime of the SDS Growing Rod while allowing growth is up to 5 years. The SDS Growing Rod is removed or locked after the treatment is completed.

The SDS Growing Rod is composed out of the following components:

- Titanium alloy springs with UHMWPE bushings
- Titanium alloy Axial Sliding Bearing with UHMWPE bearings

Page 1 of 3

{5}

BAT MEDICAL
K251961

- Titanium alloy Polyaxial Sliding Bearing with UHMWPE bearings
- Titanium alloy buttress
- CoCr rods
- Titanium alloy C-Clamp

The SDS Growing Rod is used in combination with:

- Dedicated instrument set
- Spring Compressor
- Spring Locker
- Torque Limiting Handle 9Nm
- C-Clamp Forceps
- SDS Tray

- Pedicle screws for 5.5 mm rods. The SDS Growing Rod is compatible with pedicle screws from the RESPONSE™ 5.5/6.0 mm System (Orthopediatrics) and their dedicated instrumentation for screw placement. Other components of the RESPONSE™ system, such as connectors, hooks, or ancillary fixation elements, may be used within the same spinal construct as clinically indicated. However, these components should not impede the sliding function of the growing rod or integrated spring mechanism.

# 5. Indication for Use Statement

The SDS Growing Rod is indicated for skeletally immature patients less than 10 years of age with the potential for additional spinal growth who require surgical treatment for correction and maintenance of the correction of severe, progressive, life-threatening early-onset deformities, including early-onset scoliosis, which are associated with or at risk of thoracic insufficiency syndrome.

# 6. Summary of Technological Characteristics of the Device Compared to the Predicate device

The technological characteristics of the subject device are similar to those of the predicate device, although they are not identical. Both devices consist of spinal rods and associated fixation components intended to maintain spinal correction while allowing controlled spinal growth over time, and both systems are designed to accommodate longitudinal adjustment during the growth period.

The primary differences between the subject and predicate devices relate to the specific configuration of components and the mechanism by which controlled motion and actuation are achieved. The subject device employs alternative interconnections and actuation mechanisms compared to the predicate device. These differences do not raise different questions of safety or effectiveness, as they are limited to the manner of implementation rather than the fundamental technological function.

Page 2 of 3

{6}

BAAT
MEDICAL
K251961

# 7. Summary of Performance Data, Clinical Data, and Design Controls

The SDS Growing Rod has undergone comprehensive non-clinical performance testing to verify mechanical integrity and functional equivalence to the predicate device.

Testing was performed in accordance with FDA-recognized consensus standards, including ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model), and ASTM F1798 (Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms). Wear was evaluated via a setup based on ASTM F1717 and a test procedure derived from applicable wear standards (ASTM F2624).

Biocompatibility was addressed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. A risk-based biological evaluation was performed considering the materials, manufacturing processes, and clinical use of the device. Based on this evaluation and supporting evidence, the device is considered biocompatible and safe for its intended use.

Retrospective clinical data on the subject device were provided in support of this submission. The clinical outcomes demonstrated the SDS Growing Rod to have a substantially equivalent safety and effectiveness profile compared to predicates when treating the same patient population. Clinical data analyses demonstrated safe use with adequate mitigation of risks associated with the metallic components of the device, including potential tissue ingrowth/encapsulation, local tissue reactions from wear or corrosion byproducts, and irritation from spring or sliding elements.

The results support the conclusion that the SDS Growing Rod performs at least as safely and effectively as predicate pedicle screw-based growing rod systems under similar conditions of use.

# 8. Conclusion of Substantial Equivalence

The SDS Growing Rod has the same intended use, similar indications for use, similar technology, comparable principles of operation, same range of size parameters, and similar materials as the identified predicate device.

The differences encountered do not negatively impact the overall safety and effectiveness of the SDS Growing Rod. Therefore, it can be concluded that the SDS Growing Rod is a safe and effective device and is substantially equivalent to the identified predicate system.

Page 3 of 3

---

**Source:** [https://fda-staging.innolitics.com/device/K251961](https://fda-staging.innolitics.com/device/K251961)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
