K251958 · Dynapulse Medical · NUH · Dec 23, 2025 · Neurology
Device Facts
Record ID
K251958
Device Name
VEINOPLUS Back
Applicant
Dynapulse Medical
Product Code
NUH · Neurology
Decision Date
Dec 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Veinoplus® Back is to be used for temporary relief of pain associated with sore and aching muscles in the back (shoulder, middle of the back, or lumbar area), due to strain from exercise or normal household and work activities.
Device Story
VEINOPLUS Back is a non-invasive, battery-operated, over-the-counter TENS device for back pain relief. Input consists of user-selected stimulation intensity via button controls; device transforms input into monophasic, rectangular asymmetric electrical pulses delivered through self-adhesive electrodes. Output includes electrical stimulation to muscles and visual feedback on an LCD screen (on/off status, battery, intensity). Used by patients in home or non-clinical settings. Device features automatic shut-off and patient override control. Healthcare providers do not operate the device; patients use it to manage muscle pain, potentially reducing reliance on other pain management methods.
Clinical Evidence
Bench testing only. Compliance with electrical safety and performance standards (IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10) demonstrated equivalence to the predicate device.
Technological Characteristics
Handheld TENS device; ABS and PC housing. Powered by 9V alkaline battery. Delivers monophasic, rectangular asymmetric pulses. Features LCD screen, button controls, and self-adhesive electrodes. Connectivity: standalone. Sterilization: not applicable. Software/firmware controlled. Compliant with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-10 (performance).
Indications for Use
Indicated for temporary relief of pain associated with sore and aching muscles in the back (shoulder, middle of the back, or lumbar area) due to strain from exercise or normal household and work activities in adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
DynaPulse Medical
% Danielle Besal
Principal Consultant
MRC Global
9085 E Mineral Circle, Suite 110
Centennial, Colorado 80112
Re: K251958
Trade/Device Name: VEINOPLUS Back
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NUH
Dated: November 19, 2025
Received: November 20, 2025
Dear Danielle Besal:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251958 - Danielle Besal
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251958 - Danielle Besal
Page 3
Sincerely,
JITENDRA V. VIRANI -S
CDR Jitendra Virani, MS MBA
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251958 | |
| Device Name VEINOPLUS Back | |
| Indications for Use (Describe) The Veinoplus® Back is to be used for temporary relief of pain associated with sore and aching muscles in the back (shoulder, middle of the back, or lumbar area), due to strain from exercise or normal household and work activities. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
1
510(k) Summary
K251958
VEINOPLUS Back
December 16, 2025
Company: DynaPulse Medical
13500 Wayzata Blvd
Minnetonka, MN 55305
Primary Contact: Danielle Besal
Principal Consultant | MRC Global
901.827.8670
Danielle.Besal@askmrcglobal.com
Company Contact: Steffen Magnell
CEO | DynaPulse Medical
612.464.3872
Steffen.Magnell@dynapulsemedical.com
Trade Name: VEINOPLUS Back
Common Name: Transcutaneous electrical nerve stimulator for pain relief
Classification Name: Stimulator, nerve, transcutaneous, over-the-counter
Classification: Class II
Regulation Number: 21 CFR 882.5890
Product Code: NUH
Predicate Device: K130802: OTC Electrical Stimulator Models MT900I, LT3060
Device Description:
VEINOPLUS Back is a non-invasive, battery-operated, over-the-counter device that provides transcutaneous electrical nerve stimulation (TENS). The VEINOPLUS Back consists of the handheld device with LCD screen and button controls, self-adhesive electrodes, cable to connect the device to the electrodes, 9V battery, and optional carrying accessories. When the device is in use, the on/off status, battery indicator, and stimulation intensity are displayed on the device’s screen. The user is able to control the stimulation intensity via the button controls on the device and the device includes safety features such as automatic shut-off and patient override control.
Indications for Use:
The Veinoplus® Back is to be used for temporary relief of pain associated with sore and aching muscles in the back (shoulder, middle of the back, or lumbar area), due to strain from exercise or normal household and work activities.
Indications for Use Comparison:
Both the subject and predicate devices are indicated for TENS (temporary relief of pain from sore and aching muscles). Although the subject device is designed for use on the
{5}
back, this target area is a subset of that indicated for the predicate and, therefore, the indications for use are considered substantially equivalent to those of the predicate.
## Technological Comparison:
The technological characteristics of the subject device are comparable to those of the predicate device. The subject device operates using the same mechanism of action as the predicate device. It also has a similar form factor of a handheld unit connected to self-adhesive electrodes, battery power source, user controls, and safety features. There are some differences between the output specifications of the subject device in comparison to the predicate; however, the specifications are within the range of those of the predicate and any differences were deemed negligible. Table 1 provides a complete comparison of the technological characteristics of the subject and predicate devices.
## Performance Testing:
Electrical safety and performance testing was completed on the device in accordance with the following standards: IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10. Results of the testing demonstrated equivalence with the predicate devices.
## Conclusion:
The subject device has the same intended use as the predicate device and operates through the same mechanism of action. Although there are some differences in the technical output specifications, the subject device specifications are within the range of those of the predicate and there were no different questions of safety or effectiveness raised by the differences. Testing on the device demonstrated compliance with applicable safety and performance standards and, thus, with the predicate device. In conclusion, VEINOPLUS Back is considered substantially equivalent to the identified predicate device.
2
{6}
Table 1. Comparison of Subject and Predicate Devices
| | SUBJECT | PREDICATE |
| --- | --- | --- |
| | VEINOPLUS Back | OTC Electrical Stimulator Models MT900I, LT3060 K130802 |
| Product code | NUH | NUH, NGX |
| Indications for Use | The Veinoplus® Back is to be used for temporary relief of pain associated with sore and aching muscles in the back (shoulder, middle of the back, or lumbar area), due to strain from exercise or normal household and work activities. | MT9001-TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance. LT3060 -TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. |
| Rx or OTC | OTC | OTC |
| Functions | Electrical stimulation | Electrical stimulation |
| Power source | 9V alkaline battery | 9V battery |
| Method of line current isolation | Complete; no connection (uses battery power only) | N/A |
| Patient leakage current | N/A (normal and single fault conditions) | 0.61μA (normal), 0.68uA (single fault condition) |
| Average DC current through electrodes when device is on, but no pulses being applied | 0 | 0 |
| Number of output modes | 1 | 1 TENS, 1 EMS |
| Number of output channels | 1 | Alternating |
| Synchronous / Alternating | N/A | |
| Method of channel isolation | N/A | By electrical circuit and software |
| Regulated current or voltage | Voltage | Current |
| Software / Firmware / Microprocessor Control | Yes | Yes |
| Automatic Overload Trip | No | Yes |
| Automatic No-Load Trip | No | Yes |
| Automatic Shut Off | Yes | Yes |
| Patient Override Control | Yes | Yes |
| User Display | On/off status, battery indicator, stimulation intensity (equal to | On/off status, low battery, voltage/current level |
{7}
| | | SUBJECT | PREDICATE |
| --- | --- | --- | --- |
| | | VEINOPLUS Back | OTC Electrical Stimulator
Models MT900I, LT3060
K130802 |
| | | peak value of output voltage on
500Ω load with 5% accuracy) | |
| Battery cut-off voltage | | 5.6 V ±10% | Unknown |
| Timer | | 45 min | 1-60 min |
| Weight | | 160g | 128g |
| Dimensions (WxHxD) | | 150x60x21mm | 117x60x34mm |
| Housing materials | | ABS & PC | ABS |
| Electrical safety | | IEC 60601-1 compliant | IEC 60601-1 compliant |
| EMC | | IEC 60601-1-2 compliant | IEC 60601-1-2 compliant |
| Performance | | IEC 60601-2-10 compliant | IEC 60601-2-10 compliant |
| Waveform | | Monophasic, rectangular
asymmetric | Biphasic, square |
| Maximum Output Voltage (V) | | 50±5% @ 500 Ω
50±5% @ 2 kΩ
50±5% @ 10 kΩ | 96±20% @ 500 Ω
200±20%; 228±20% @ 2 kΩ
210±20%; 230±20% @ 10 kΩ |
| Maximum Output Current (mA) | | 100±0% @ 500 Ω
25±10% @ 2 kΩ
5±10% @ 10 kΩ | 96±20% @ 500 Ω
50±20%; 57±20% @ 2 kΩ
10.5±20%; 11.5±20% @ 10 kΩ |
| Pulse width (μs) | | 25-240 | 50-300 |
| Frequency (Hz) | | 1-120 | 1-150 |
| Net charge DC /
microcoulombs
(μC) per pulse | | < 0.6 | 0 |
| Maximum phase charge (μC) | | 24 @500 Ω | 28.8 @500 Ω |
| Maximum current density
(mA/cm²) | | 1 @500 Ω | 1.15 @500 Ω |
| Maximum power density
(using smallest electrode
conductive surface area)
(mW/cm²) | | 50 @500 Ω | 0.373 @500 Ω |
| Burst
mode
(pulse
trains) | Pulses/burst | 60 | 7 |
| | Bursts/sec | 1.25 | 0.5/1/2/3/4/5 Hz |
| | Burst duration | 0.1 seconds | 70 ms |
| | Duty cycle
(bursts/sec x
burst duration) | 0.125 | 0.1 |
| On time (s) | | 0.1 | N/A; 1-30 |
| Off time (s) | | 0.9 | N/A; 1-60 |
{8}
| | SUBJECT | PREDICATE |
| --- | --- | --- |
| | VEINOPLUS Back | OTC Electrical Stimulator
Models MT900I, LT3060
K130802 |
| Additional feature | Each train of pulses is preceded by a ramp (pulse width modulation from 65μs-135μs) | Unknown |
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.