← Product Code [PSY](/productcode/PSY) · K251952

# Virasight (K251952)

_Virasoft Corporation · PSY · Jul 17, 2026 · Pathology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251952

## Device Facts

- **Applicant:** Virasoft Corporation
- **Product Code:** [PSY](/productcode/PSY.md)
- **Decision Date:** Jul 17, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.3700
- **Device Class:** Class 2
- **Review Panel:** Pathology
- **Attributes:** Software as a Medical Device

## Indications for Use

ViraSight is a web-based software-only device intended to aid pathologists in the viewing and management of digital images of scanned pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue. It is an aid to the pathologist to review these digital images for the purposes of pathology primary diagnosis. ViraSight is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using ViraSight.

## Device Story

ViraSight is a web-based, software-only image viewing and management system for digital pathology. It accepts whole-slide images (WSI) in SVS or NDPI formats, acquired from FDA-cleared Hamamatsu NanoZoomer S360MD or Leica Aperio GT 450 DX scanners. Operated by pathologists and lab technicians in clinical settings, the device provides a user interface for loading, navigating (panning/zooming), rotating, annotating, and measuring digital slides. It allows pathologists to enter and save diagnostic notes. The device does not perform automated image analysis or diagnostic functions; it serves as a digital workspace for primary diagnosis. By enabling remote or digital review of scanned slides, it facilitates efficient pathology workflows and diagnostic decision-making.

## Clinical Evidence

No clinical data. Evidence consists of bench testing: pixel-wise color comparison (CIEDE2000 < 3) against predicate viewers, turnaround time (loading, zooming, panning) validation, and measurement accuracy (length/area) testing (< 0.01% difference). Human factors validation was conducted with 15 representative users (pathologists/technicians) in simulated clinical environments, confirming safe and effective task performance.

## Technological Characteristics

Web-based software-only device; runs on Windows 11 via Chrome/Edge browsers. Supports SVS and NDPI file formats. Integrates with specific FDA-cleared scanners (Hamamatsu NanoZoomer S360MD, Leica Aperio GT 450 DX) and displays (JVC Kenwood, Barco, Dell). Features include image navigation, annotation, and distance/area measurement tools. No automated image analysis or diagnostic algorithms.

## Regulatory Identification

The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.

## Special Controls

A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.

## Predicate Devices

- NanoZoomer S360MD Slide scanner system ([K213883](/device/K213883.md))
- Aperio GT 450 DX ([K232202](/device/K232202.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG  
ADMINISTRATION

# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION  
DECISION SUMMARY**

# **I Background Information:**

# **A 510(k) Number**

K251952

# **B Applicant**

Virasoft Corporation

# **C Proprietary and Established Names**

ViraSight

# **D Regulatory Information**

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  QKQ | Class II | 21 CFR 864.3700 – Whole slide imaging system | PA - Pathology  |

# **II Submission/Device Overview:**

# **A Purpose for Submission:**

New device

# **B Type of Test:**

Software only device

# **III Intended Use/Indications for Use:**

# **A Intended Use(s):**

See Indications for Use below.

# **B Indication(s) for Use:**

ViraSight is a web-based software-only device intended to aid pathologists in the viewing and management of digital images of scanned pathology slides prepared from formalin-fixed, paraffin-

Food and Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20993-0002  
www.fda.gov

{1}

embedded (FFPE) tissue. It is an aid to the pathologist to review these digital images for the purposes of pathology primary diagnosis.

ViraSight is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using ViraSight.

### **C Special Conditions for Use Statement(s):**

Rx - For Prescription Use Only

## **IV Device/System Characteristics:**

### **A Device Description:**

ViraSight (version 1.0.6), is a web-based, software-only device intended to aid pathologists in the viewing and management of digital images of scanned pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue acquired from FDA-cleared digital pathology scanners on FDA-cleared displays, as specified in Tables 1 and 2 below. ViraSight is an image viewing software and does not include image analysis or perform diagnostic functions.

Glass slide scanning is performed upstream and independent of the use of ViraSight. Image acquisition is performed using the intended scanner(s), the Hamamatsu NanoZoomer S360MD Slide Scanner or the Leica Aperio GT 450 DX Scanner; with the scanning technician conducting quality control on the whole-slide images according to the scanner's instructions for use and laboratory specifications to determine if re-scans are needed.

The primary tasks within the ViraSight workflow completed by the user in the user interface are:

1. Logging into ViraSight
2. Loading and opening a digital slide
3. Navigating (including panning and zooming) the slide
4. Rotating the slide
5. Measuring distances and area in the digital image
6. Annotating regions of interest
7. Referring annotations to notes
8. Entering, finalizing and saving the pathologist's diagnosis as a note
9. Completing case review and logout

ViraSight is validated for use with the WSI scanner specified in Tables 1, display components specified in Table 2, and computer environment specified in Table 3.

**Table 1: WSI scanner**

|  Manufacturer | Device name  |
| --- | --- |
|  Hamamatsu | NanoZoomer S360MD Slide scanner  |
|  Leica | Aperio GT 450 DX  |

K251952 - Page 2 of 11

{2}

**Table 2: WSI display**

|  Manufacturer | Model  |
| --- | --- |
|  JVC Kenwood | JD-C240BN01A  |
|  Barco | MDPC-8127  |
|  Dell | UP3017  |
|  Dell | U3023E  |
|  Dell | U3223QE  |

**Table 3: Computer Environment Requirements**

|  Workstation Component | Specifications  |
| --- | --- |
|  Memory | Minimum 16 GB  |
|  CPU | Minimum 4 cores  |
|  Network Bandwidth | 50 Mbps bandwidth to the server  |
|  Operating System | Microsoft Windows 11  |
|  Supported browsers | Google Chrome (146.0.7680.178 and above) Microsoft Edge (146.0.3856.97 and above)  |

## **B Instrument Description Information:**

### **1. Instrument Name:**

ViraSight

### **2. Specimen Identification:**

The ViraSight device utilizes digital pathology images acquired from Hematoxylin and Eosin (H&E) stained glass slides using the Hamamatsu NanoZoomer S360MD Slide scanner or the Leica Aperio GT 450 DX scanner. A reading pathologist selects a case (patient) from an external worklist, and the subject device retrieves the corresponding images acquired by the WSI scanner. The scanned images are identified using the specimen identifier previously assigned to the case.

### **3. Specimen Sampling and Handling:**

Specimen sampling and handling are conducted independently and prior to the use of the subject device. This process involves obtaining biopsy or resection specimens, which are then processed using standard histology techniques. After H&E staining of the FFPE tissue sections, digital images are generated from the glass slides using the Hamamatsu NanoZoomer S360MD Slide scanner or the Leica Aperio GT 450 DX Scanner.

### **4. Calibration:**

Not applicable

### **5. Quality Control:**

Quality control procedures should be established by the laboratory in accordance with its standard operating procedures. Before using ViraSight for primary diagnosis, the user should verify that the

K251952 - Page 3 of 11

{3}

correct case has been selected and that all associated whole slide images have been successfully imported and are complete and of acceptable image quality for diagnostic review. Image quality should be assessed for factors such as focus, completeness of the scanned tissue, and the presence of artifacts that may interfere with interpretation. If image quality is considered inadequate for diagnostic purposes, the slide should be rescanned or reviewed using the laboratory's established procedures. ViraSight does not perform automated quality assessment of whole slide images and does not determine whether an image is acceptable for diagnosis. The final determination of image quality and suitability for diagnostic use remains the responsibility of the reviewing pathologist and the laboratory.

# V Substantial Equivalence Information:

# A Predicate Device Name(s):

Aperio GT 450 DX; NanoZoomer S360MD Slide scanner system

# B Predicate 510(k) Number(s):

K232202, K213883

# C Comparison with Predicate(s):

|  Specification | Subject Device (K251952) | Predicate Device (K232202) | Predicate Device (K213883)  |
| --- | --- | --- | --- |
|  Device Trade Name | ViraSight | Aperio GT 450 DX | NanoZoomer S360MD Slide scanner system  |
|  Product Code | QKQ | PSY | PSY  |
|  Regulation | 21 CFR 864.3700 | 21 CFR 864.3700 | 21 CFR 864.3700  |
|  Regulation Name | Whole Slide Imaging System | Whole Slide Imaging System | Whole Slide Imaging System  |
|  Classification | II | II | II  |
|  General Device Characteristic Similarities  |   |   |   |
|  Indications for Use | ViraSight is a web-based software-only device intended to aid pathologists in the viewing and management of digital images of scanned pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue. It is an aid to the pathologist to review these digital images for the purposes of pathology primary diagnosis. ViraSight is not intended for use with frozen section, cytology, or non-FFPE hematopathology | The Aperio GT 450 DX is an automated digital slide creation and viewing system. The Aperio GT 450 DX is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio GT 450 DX is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. Aperio GT 450 DX is comprised of the Aperio GT 450 DX scanner, which generates images in the Digital Imaging and Communications in Medicine (DICOM) and in the ScanScope Virtual Slide (SVS) file formats, the Aperio WebViewer DX viewer, and the displays. The Aperio GT 450 DX is intended to be used with the interoperable components specified in Table 1. | The NanoZoomer S360MD Slide scanner system ("NanoZoomer System") is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded ("FFPE") tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.  |

K251952 - Page 4 of 11

{4}

|  Specification | Subject Device (K251952) | Predicate Device (K232202) |   |   |   | Predicate Device (K213883)  |
| --- | --- | --- | --- | --- | --- | --- |
|   | specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using ViraSight. | Table 1: Interoperable components of Aperio GT 450 DX |   |   |   | The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)- cleared display that has been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying additional compatible displays.  |
|   |  | Scanner Hardware | Scanner Output file format | Interoperable Viewing Software | Interoperable Displays |   |
|   |  | Aperio GT 450 DX scanner | SVS | Aperio WebViewer DX | Barco MDPC-8127 Dell UP3017 Dell U3023E Dell U3223QE |   |
|   |  | Aperio GT 450 DX scanner | SVS | Sectra Digital Pathology Module (3.3) | Dell U3223QE | The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System  |
|   |  | Aperio GT 450 DX scanner | DICOM | Sectra Digital Pathology Module (3.3) | Dell U3223QE |   |
|   |  | The Aperio GT 450 DX is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio GT 450 DX. |   |   |   |   |
|  Scanner | Hamamatsu NanoZoomer S360MD Slide scanner and Leica Aperio GT 450 DX | Leica Aperio GT 450 DX |   |   |   | Hamamatsu NanoZoomer S360MD Slide scanner  |
|  Interoperable Display | JVC Kenwood JD-C240BN01A Dell U3223QE Dell UP3017 Dell U3023E Barco MDPC-8127 | Barco MDPC-8127, Dell UP3017, Dell U3023E, Dell U3223QE |   |   |   | JVC Kenwood JD-C240BN01A, Barco MDPC-8127  |
|  Specimen Type | Surgical pathology slides prepared from FFPE tissue | Same |   |   |   | Same  |
|  Image File Format | SVS, NDPI | SVS, DICOM |   |   |   | NDPI  |
|  Image Manipulation Functions | Panning, zooming, image adjustments, annotations, and distance/area measurements | Panning, zooming, gamma function, annotations, and measurements |   |   |   | Panning, zooming, image adjustments, annotations, and distance/area measurements  |
|  Principle of Operation | ViraSight is a web-based, software-only device intended to aid pathologists in the viewing and management of digital images of scanned pathology slides prepared from | The Aperio GT 450 DX is a WSI system. The technician places the slides into the Aperio GT 450 DX scanner. The Aperio GT 450 DX scanner automatically loads the slides, takes the micro images, finds the tissues, and scans the slides. The scanner also automatically performs quality control (QC) and notifies the user of any image |   |   |   | After WSI are acquired by using NanoZoomer S360MD Slide scanner, the WSI are automatically saved to the hard disk during scanning and may be viewed later by using the included viewing software.  |

K251952 - Page 5 of 11

{5}

|  Specification | Subject Device (K251952) | Predicate Device (K232202) | Predicate Device (K213883)  |
| --- | --- | --- | --- |
|   | formalin-fixed, paraffin-embedded (FFPE) tissue from Hamamatsu NanoZoomer S360MD Slide scanner or Leica Aperio GT 450 DX scanner. Glass slide scanning is performed upstream and independent of the use of ViraSight. The scanning technician should perform quality control of all glass slides and WSIs following the scanner Instructions for Use and other laboratory procedures. Prior to using a WSI for diagnosis, the pathologist should ensure that all scanned slide images have been imported for every case and the images are of acceptable quality for diagnostic purposes. The pathologist reviews scanned images from all the slides associated with a case before rendering a diagnosis. ViraSight supports navigation and visualization of whole-slide images, annotation and measurement of regions of interest, creation of notes referring to annotations, color adjustment of image to edit brightness and contrast. | quality issue during the image acquisition. The image data is sent to end-user-provided image storage attached to the local network. During the review, the pathologist opens WSI images acquired with the WSI scanner from the image storage, performs further QC, and reads WSI images of the slides to make a diagnosis. | During review, the pathologist opens WSI from the image storage attached to local network, performs further QC, and reads WSI of the slides to make a diagnosis.  |
|  Device Components | ViraSight Image Management System and Monitor display | Scanner, Image Management Software, and Display | Scanner, Image Management Software, and Display  |
|  **General Device Characteristic Differences**  |   |   |   |
|  User Interface | ViraSight | Aperio WebViewer DX | NZViewMD  |

## VI Standards/Guidance Documents Referenced:

1. FDA Guidance “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices”. April 20, 2016.
2. FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices”. February 3,

K251952 - Page 6 of 11

{6}

2016.

3. FDA Guidance “Content of Premarket Submissions for Device Software Functions”. June 14, 2023.
4. FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”. September 27, 2023.
5. IEC 62304:2006 + A1:2015 – Medical Device Software – Software Life Cycle Processes
6. ISO 13485:2016 Medical Devices – Quality Management Systems – Requirements for Regulatory Purposes
7. ISO 14971:2019 – Medical Devices – Application of Risk Management to Medical Devices
8. IEC 62366-1:2015 + A1:2020 – Application of Usability Engineering to Medical Devices

**VII Performance Characteristics (if/when applicable):**

**A Analytical Performance:**

1. Precision/Reproducibility:
Not applicable
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Accuracy (Instrument)
Not applicable
5. Carry-Over:
Not applicable

**B Other Supportive Instrument Performance Characteristics Data:**

Technical performance testing was conducted with the subject device, ViraSight as specified below.

**1. Bench Testing - Pixelwise comparison test**

Pixel-wise comparison testing was conducted to compare images reproduced by ViraSight and the comparators; the predicate device's Image Review Manipulation Software (IRMS, as defined in the FDA guidance document Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices, dated April 20, 2016) — as described below for the same whole slide image generated from an interoperable scanner, to validate equivalent image reproduction. The equivalence between ViraSight and the comparators was evaluated by bench testing based on pixel-wise comparison for both supported scanner file formats: SVS files generated by the Leica Aperio GT 450 DX scanner, and NDPI files generated by the Hamamatsu NanoZoomer S360MD scanner.

Pixelwise comparison testing was performed with the assistance of the “Image Viewer Integrity Evaluation System (IVIES)” application, which is an FDA-qualified Medical Device Development Tool [MDDT; Image Viewer Integrity Evaluation System (IVIES)]. ViraSight was tested as operating with its intended components, including the scanner and monitor display, to evaluate the pixel-wise differences. A total of 48 formalin-fixed, paraffin-embedded (FFPE) pathology tissue glass slides,

K251952 - Page 7 of 11

{7}

covering a range of tissue types and diagnostic categories, were scanned using both slide scanner systems. For each WSI dataset, three rectangular regions of interest (ROIs) were identified per slide. Screenshots of these ROIs were captured at two magnification levels, 20x and 40x, on the Dell U3023E display. For SVS files, ViraSight screenshots were compared against Leica Aperio WebViewer DX screenshots under matched browser conditions in Google Chrome and Microsoft Edge. For NDPI files, screenshots of ViraSight running in Google Chrome and Microsoft Edge were compared against Hamamatsu NZViewMD screenshots. The original 640×480 ROI screenshots were cropped to 630×470 pixels to remove annotation boundary visuals, and every pixel of each cropped screenshot pair was sampled to calculate the pixel-wise color difference using the CIEDE2000 (ΔE₀₀) metric through IVIES. The acceptance criterion required that the 95th percentile of the pixel-wise color differences in every ROI image pair between ViraSight and the comparator must be less than 3 CIEDE2000 (i.e., < 3 ΔE₀₀).

The test results demonstrated that the maximum 95th percentile of pixel-wise color differences between ViraSight and any comparator was less than 3 CIEDE2000 across all comparison sets (e.g., all scanners, file formats, browsers, and magnification levels). No tested ROI image pair exceeded the ΔE₀₀ = 3 threshold at the 95th percentile, indicating that ViraSight output images were visually consistent with the comparators for both supported scanner file formats across all tested browser conditions and viewer zoom levels. The detailed pixel-wise comparison test results, including the maximum, minimum, and mean 95th percentile ΔE₀₀ values across all ROI pairs for each scanner, file format, subject device/browser, comparator, and magnification level, are provided in the Table 4 below.

Therefore, testing results demonstrated that WSIs reproduced by ViraSight are identical to images reproduced by the predicate devices.

Table 4. ViraSight Pixelwise Comparison Testing Results

|  Scanner | Image File Format | Subject Device/ Browser | Comparator (Predicate device IRMS /Browser) | Magnification Level | Results  |
| --- | --- | --- | --- | --- | --- |
|  Hamamatsu NanoZoomer S360MD Scanner | NDPI | ViraSight / Chrome | Hamamatsu NZViewMD | 40x | max (95th percentile ΔE00) = 0.69 min (95th percentile ΔE00) = 0.36 mean (95th percentile ΔE00) = 0.47  |
|   |   |   |   |  20x | max (95th percentile ΔE00) = 1.61 min (95th percentile ΔE00) = 0.37 mean (95th percentile ΔE00) = 0.68  |
|   |   |  ViraSight / Edge |   | 40x | max (95th percentile ΔE00) = 0.69 min (95th percentile ΔE00) = 0.36 mean (95th percentile ΔE00) = 0.47  |
|   |   |   |   |  20x | max (95th percentile ΔE00) = 1.60 min (95th percentile ΔE00) = 0.37 mean (95th percentile ΔE00) = 0.67  |
|  Leica Aperio GT 450 DX Scanner | SVS | ViraSight / Chrome | Leica Aperio WebViewer DX / Chrome | 40x | max (95th percentile ΔE00) = 2.96 min (95th percentile ΔE00) = 0.94 mean (95th percentile ΔE00) = 2.08  |
|   |   |   |   |  20x | max (95th percentile ΔE00) = 2.93 min (95th percentile ΔE00) = 1.23 mean (95th percentile ΔE00) = 2.27  |

K251952 - Page 8 of 11

{8}

|   |  | ViraSight /Edge | Leica Aperio WebViewer DX /Edge | 40x | max (95th percentile ΔE00) = 2.98 min (95th percentile ΔE00) = 0.95 mean (95th percentile ΔE00) = 2.11  |
| --- | --- | --- | --- | --- | --- |
|   |   |   |   |  20x | max (95th percentile ΔE00) = 2.93 min (95th percentile ΔE00) = 1.23 mean (95th percentile ΔE00) = 2.27  |

## 2. Turnaround Time Testing

Turnaround time (TAT) testing was performed to demonstrate that ViraSight provides acceptable image rendering performance for the intended use of the subject device. A total of 30 H&E-stained FFPE tissue glass slides were scanned using the intended scanners to generate 30 WSIs in SVS format and 30 WSIs in NDPI format. TAT testing was performed across all supported browsers — Google Chrome and Microsoft Edge — and both supported scanner file formats. Each browser/file-format configuration was tested in two independent runs using WSIs over a range of file sizes.

Turnaround time values were recorded for WSI loading, 20X zooming, 40X zooming, and panning. All tested WSIs were loaded successfully, and zooming and panning operations were completed without rendering failure, application crash, loss of image data or interruption preventing intended use.

The test results demonstrated acceptable turnaround time performance across all tested browser environments and supported file formats with respect to the intended use of the subject device, as shown in Table 5 below.

**Table 5. ViraSight Turnaround Time Testing Results**

|  Scanner | Image File Format | Subject Device / Browser | Results (Sec)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Loading | 20X Zooming | 40X Zooming | Panning  |
|  Leica Aperio GT 450 DX scanner | SVS | ViraSight / Chrome | Mean = 1.032 Range = 0.594–1.717 | Mean = 1.655 Range = 0.553–3.343 | Mean = 0.403 Range = 0.014–3.595 | Mean = 0.172 Range = 0.062–0.504  |
|   |   |  ViraSight / Edge | Mean = 0.952 Range = 0.653–1.839 | Mean = 1.084 Range = 0.081–2.304 | Mean = 0.192 Range = 0.057–2.018 | Mean = 0.104 Range = 0.020–0.363  |
|  Hamamatsu NanoZoomer S360MD Slide scanner | NDPI | ViraSight / Chrome | Mean = 0.833 Range = 0.664–1.178 | Mean = 1.970 Range = 0.845–3.137 | Mean = 0.224 Range = 0.076–0.942 | Mean = 0.116 Range = 0.062–0.286  |
|   |   |  ViraSight / Edge | Mean = 0.869 Range = 0.675–1.407 | Mean = 1.727 Range = 0.409–2.960 | Mean = 0.248 Range = 0.058–0.807 | Mean = 0.145 Range = 0.075–1.345  |

K251952 - Page 9 of 11

{9}

Therefore, ViraSight has been found to have acceptable turnaround time performance with respect to its intended use.

### 3. Measurement – distance and area

Measurement accuracy testing was conducted to demonstrate that ViraSight accurately represents length and area measurements across the supported scanner file formats, intended browser environments, viewer magnification levels, and slide orientations. A total of three H&E-stained FFPE tissue slides, and one calibration/reference slide were used as the test input. The slides were represented in SVS and NDPI formats using the intended scanners. All supported browsers — Google Chrome and Microsoft Edge — two viewer magnification levels (20x and 40x), horizontal and vertical slide orientations, and two types of measurements — length and area — were tested. Area measurements included rectangle/square, circle/ellipse, and freehand region measurements where supported.

Measurement accuracy was evaluated by comparing the ViraSight-generated measurement values against the applicable reference values, which were based on known reference dimensions, reference screenshots, or measurements from the applicable reference viewer where technically supported. The acceptance criterion required less than 0.01% difference between the ViraSight-generated measurement and the applicable reference value.

The test results demonstrated that the pre-specified acceptance criterion of < 0.01% difference was met across all tested configurations. All completed measurement records were exact matches with no error, as listed in Table 6 below.

**Table 6. ViraSight Measurement Results**

|  Scanner | Image Format | Subject Device / Browser | Percent Difference (%)  |   |
| --- | --- | --- | --- | --- |
|   |   |   |  Length Measurement | Area Measurement  |
|  Leica Aperio GT 450 DX scanner | SVS | ViraSight / Chrome | 0 | 0  |
|   |   |  ViraSight / Edge | 0 | 0  |
|  Hamamatsu NanoZoomer S360MD Slide scanner | NDPI | ViraSight / Chrome | 0 | 0  |
|   |   |  ViraSight / Edge | 0 | 0  |

Therefore, ViraSight has been found to perform accurate length and area measurements with respect to its intended use.

### 4. Human Factor (Usability) Testing

Human factors validation testing was conducted per FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016), using task-based use scenarios performed by representative users across two sites under simulated use conditions representative of the intended

K251952 - Page 10 of 11

{10}

clinical environment. A systematic evaluation of critical tasks required for the intended operation of ViraSight was performed by 15 representative users, including pathologists as primary users and laboratory technicians as secondary users. Overall, the results of the human factors testing were acceptable.

### VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

### IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K251952 - Page 11 of 11

---

**Source:** [https://fda-staging.innolitics.com/device/K251952](https://fda-staging.innolitics.com/device/K251952)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
