K251944 · Lipocosm, LLC · QPB · Jun 8, 2026 · General, Plastic Surgery
Device Facts
Record ID
K251944
Device Name
Lipocosm Harmonic System
Applicant
Lipocosm, LLC
Product Code
QPB · General, Plastic Surgery
Decision Date
Jun 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Lipocosm Harmonic System is intended for the removal of tissue and fluid from the body during general surgical procedures, when used with commercially available aspiration device, including suction lipoplasty for the purpose of aesthetic body contouring.
Device Story
Lipocosm Harmonic System is a powered suction lipoplasty device for aesthetic body contouring. System components include a control box, reusable handpiece, cannula, foot pedal, and power supply. The handpiece uses an internal eccentric mass to oscillate a 3mm stainless steel cannula in two dimensions (perpendicular to its axis) at frequencies up to 4,000 oscillations/minute. This motion facilitates the emulsification and fragmentation of adipose tissue, which is then removed via an external aspiration source connected to the cannula. The device is operated by trained healthcare professionals in a clinical setting. The foot pedal controls the start/stop function, while a potentiometer on the control box adjusts oscillation frequency. The output is the mechanical fragmentation of fat tissue to assist in aspiration. The system benefits patients by facilitating fat removal during lipoplasty procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 62366-1), cleaning validation, and steam sterilization validation (SAL 10⁻⁶).
Technological Characteristics
System components: Control box, handpiece, 3mm 304 stainless steel cannula, foot pedal. Oscillation: 0-4,000 oscillations/min via eccentric mass. Power: 100-240 VAC. Sterilization: Steam (pre-vacuum, 270°F for 4 min). Connectivity: Wired foot pedal and handpiece cable. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, ISO 10993-1.
Indications for Use
Indicated for aesthetic body contouring in patients requiring removal of tissue and fluid during general surgical procedures. Use limited to trained healthcare professionals.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}

06/08/2026
Lipocosm, LLC
Ercan Dilsen
VP Operations
580 Crandon Blvd. Suite 201
Key Biscayne, Florida 33149
Re: K251944
Trade/Device Name: Lipocosm Harmonic System
Regulation Number: 21 CFR 878.5040
Regulation Name: Suction Lipoplasty System
Regulatory Class: Class II
Product Code: QPB
Dated: January 30, 2026
Received: January 30, 2026
Dear Ercan Dilsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251944 - Ercan Dilsen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}
K251944 - Ercan Dilsen
Page 3
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alicia Hemphill -S
Digitally signed by Alicia
Hemphill -S
Date: 2026.06.08 19:31:04 -05'00'
Alicia L. Hemphill (Johnson), MS
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251944 | ? |
| Please provide the device trade name(s). | | ? |
| Lipocosm Harmonic System | | |
| Please provide your Indications for Use below. | | ? |
| Lipocosm Harmonic System is intended for aesthetic body contouring. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Lipocosm Harmonic System
Page 8 of 55
{4}
K251944, Page 1 of 9
# **510(K) Summary**
**Lipocosm Harmonic System**
# **1. Submission Sponsor**
Lipocosm LLC
580 Crandon Blvd, #201
Key Biscayne, FL, 33149 USA
Contact: Ercan Dilsen, Vice President Operations
# **2. Date Prepared**
April 7, 2026
# **3. Device Identification**
| Trade/Proprietary Name: | Lipocosm Harmonic System |
| --- | --- |
| Common/Usual Name: | Suction Lipoplasty System |
| Classification Name: | System, Suction, Lipoplasty for Removal without the Return of Adipose Tissue |
| Regulation Number: | 21 CFR 878.5040 |
| Product Code: | QBP |
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
# **4. Predicate Device(s)**
Device name: MicroAire® PAL® System (Console and vibration handpiece for the liposuction cannula)
510(k) number: K212024
Manufacturer: MicroAire® Surgical Instruments, LLC
The device of interest for comparison is the MicroAire® PAL® Console and Handpiece of the currently marketed and predicate MicroAire PAL® System. These two devices are controlled by the user during Lipoplasty procedures, the console for settings and the handpiece for the manipulation of the cannula for the fat removal. The predicate device and subject device are detailed further for comparison in Section 7 – Substantial Equivalence.
{5}
K251944, Page 2 of 9
## 5. Intended Use / Indication for Use Statement
Lipocosm Harmonic System is intended for the removal of tissue and fluid from the body during general surgical procedures, when used with commercially available aspiration device, including suction lipoplasty for the purpose of aesthetic body contouring.
## 6. Device Description
The main components of the Subject Lipocosm Harmonic System as follows:
- A reusable Handpiece that oscillates the Cannula in two dimensions in a plane perpendicular to its axis to facilitate lipoaspiration.
- A Control Box that controls the frequency of Cannula displacement.
- A reusable 3mm Cannula has a plastic hub proximally for attachment to the reusable Handpiece and employs a port on its proximal end for connecting it with the standard vacuum (suction) tubing. The distal end of the Cannula employs 12 holes that are used for the suction.
The oscillation of the Cannula is intended to help increase flowrate adipose tissue.
- A Foot Pedal that starts and stops the movement of the Cannula employs a cable that connects the Foot Pedal to the Control Box.
- A reusable Handpiece Cable that connects the control box to the Handpiece.
- A Power Supply for connecting the Control Box to an electric outlet.
## 7. Substantial Equivalence Discussion
The following table compares the Lipocosm Harmonic System to the currently marketed and predicate MicroAire® PAL System (K212024) with respect to indications for use, principles of operation, technological characteristics. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
Tables 1 and 2, respectively provides a summary comparison between the Subject Control Box and the Handpiece of the Lipocosm Harmonic System and the currently marketed and predicate MicroAire® PAL System.
{6}
K251944, Page 3 of 9
| Table 1 Device Comparison Table | | | |
| --- | --- | --- | --- |
| | Subject Device | Predicate Device | Device Comparison |
| 510(k) Number | K251944 | K212024 | NA |
| Applicant | Lipocosm, LLC | MicroAire Surgical Instruments, LLC | NA |
| Address | 580 Crandon Blvd. Suite 201, Key Biscayne FL 33149 | 3590 Grand Forks Road, Charlottesville, VA 22911 | NA |
| Device Name | Lipocosm Harmonic System (LHS) | PAL System | NA |
| Classification Regulation, Class | 21 CFR 878.5040, Class II | Identical | Same |
| Product Code | QPB | Identical | Same |
| | | | |
| Indications for Use | For aesthetic body contouring | For aesthetic body contouring | Same |
| Cannula | Material: Fabricated from medical Grade 304 Stainless Steel, and aluminum luer type connector. No. of apertures: 12 Length: 35cm Diameter: 3mm | PAL System Del Vecchi 12 Track Cannula: Material: Fabricated from medical-grade 304 stainless steel with a straight port connector. No. of apertures: 12 Length: 22 / 30cm Diameter: 3 / 4mm | Similar. Non-clinical testing was performed to address differences in cannula length. |
{7}
K251944, Page 4 of 9
| Device Overview Control Box | | | |
| --- | --- | --- | --- |
| **Device Description** | LHS is intended for the removal of tissue and fluid from the body during lipoplasty procedures, when used with commercially available aspiration devices, including suction lipoplasty for the purpose of aesthetic body contouring. The LHS consists of four main components: Electric instruments: Control Box, Hand Piece, Cable and Foot Pedal. The LHS Handpiece oscillates the LHS Cannula in liposuction procedures in two dimensions for aspiration of fat tissues. Note, LHS is connected to an independent aspiration source | The MicroAire® Power Assisted Liposuction (PAL) System is a medical device intended to be used for the aspiration of autologous adipose tissue in lipoplasty procedures. The PAL system consists of three components: electric instrument console, handpiece, and cable. The PAL-650 handpiece is powered by the MicroAire® 5020 Electric Instrument Control console via the 5006 PAL cable and connected to an independent aspiration source. The PAL-650 handpiece reciprocates a cannula used in liposuction procedures for aspiration of fat tissue. For more details on device see Section 11 – Device Description. | Similar. Subject and predicate devices include emulsification/fragmentation of tissue with differing mechanism of action. |
| **Image** | | | Similar with regards to functionality. |
| **User** | Only trained and experienced healthcare providers/professionals should use this medical equipment | Identical | Same |
| **Mechanism of Action** | The electronic Control Box of the Lipocosm Harmonic System that oscillates the Cannula in two dimensions in a plane perpendicular to its axis to facilitate lipoaspiration in lipoplasty applications. | MicroAire® Console is an electronic control system designed to reciprocate the cannula back and forth along the same path to facilitate lipoaspiration in lipoplasty applications. | Different, however, the mechanism of action of the subject and predicate devices facilitate lipoaspiration by emulsification/fragmentation of tissue. |
{8}
K251944, Page 5 of 9
| | | | |
| --- | --- | --- | --- |
| Motion | The cannula of LHS oscillates in two dimensions – side to side, perpendicular to the axis of the cannula. Oscillation refers to vibrating motion around a central point along the plane perpendicular to the axis of the cannula. | The cannula of the PAL System reciprocates both forward and backward, parallel to the axis of the cannula, and side to side, perpendicular to the axis of the cannula. Reciprocation is a straight-line back-and-forth motion along the same path. | Different. Non-clinical testing was performed to assess the equivalence of the fragmentation/emulsification of the subject device oscillating motion compared with the predicate device reciprocating motion. |
| Display | The LHS has a panel with an on/off switch, a potentiometer to control the frequency of the cannula, a receptacle to connect the Handpiece Cable, a receptacle that connects the Cable of the Foot Pedal, and a receptacle that connects to the power supply. | The color touch screen display of the Console provides an intuitive graphical user interface that allows users to view system status and set parameters with a touch of the screen. It also includes receptacle to connect the Cable and a receptacle that connects the control system to the power outlet | Different. The display of the two systems is different, however both systems provide the same functionality. |
| Speed Control | Potentiometer | Touch screen | Different. The speed controls of the two systems are different, however both systems provide the same functionality. |
| Sterility | Non-sterile | Identical | Same |
| Weight | 0.632 Kg | 6.35Kg | Different, however the intended purpose is the same. |
| AC Powered | 100 VAC – 240 VAC, 50/60 Hz | Identical | Same |
{9}
K251944, Page 6 of 9
| **No. of Attachable Devices** | One (1) Handpiece One (1) Foot Pedal | Two (2) Handpieces | Similar, both devices utilize Handpiece. |
| --- | --- | --- | --- |
| **Safety and Performance** | | | |
| **Electrical Safety Testing Passed** | IEC 60601-1-6:2010, IEC 60601-1-6:2010/AMD1:2013, IEC 60601-1-6:2010/AMD2:2020 for use in conjunction with IEC 62366-1:2015, IEC 62366-1:2015/AMD1:2020, and IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 IEC 60601-1-2 ed 4.1 (2020-09) | IEC 60601-1:2005 IEC 60601-1-6:2013 IEC 62366: 2015 IEC/EN 60601-1-2:2014 | Same |
| **Performance** | 0-4000 oscillations/min. | 4000 – 5000 strokes/min. | Similar, since intended purpose is the same |
| **Operating Environment** | | | |
| **Environment: Temperature Humidity** | Operation conditions 15°C to +24°C 20%-60% | Identical | Same |
| | **Subject Device** | **Predicate Device** | **Device Comparison** |
| --- | --- | --- | --- |
| **Image** | | | NA |
| **Mechanism of Action** | The Handpiece of the **LHS** utilizes an eccentric mass in the Handpiece that oscillates the Cannula in two-dimensional plane perpendicular to the Handle in the liposuction procedure | The MicroAire® PAL Handpiece is a powered surgical instrument that reciprocates a cannula back and forth in the liposuction procedure. | The Mechanism of Actions of the two systems are different, however the displacement and frequency of the Cannula of the two devices are similar. |
{10}
K251944, Page 7 of 9
| Table 2 Handpiece Comparison Table | | | |
| --- | --- | --- | --- |
| | Subject Device | Predicate Device | Device Comparison |
| **User** | Use of this device is limited to those physicians who, by means of formal professional training or sanctioned continuing medical education (including supervised operative experience), have attained proficiency in suction lipoplasty | Identical | Substantially Equivalent |
| **Technology** | The power assisted Cannula oscillation ranges from 0 to 4,000 oscillations/minute in two dimensions. The oscillation of the Cannula is approximately 3.5mm in the long axis. | The reciprocating output shaft of the handpiece drives the cannula through a stroke distance of 2.8 mm (±0.4 mm) at a rate of 4,000 to 5,000 strokes/minute. | Different, however, the function of the subject and predicate devices facilitate lipoaspiration by emulsification/fragmentation of tissue. |
| **Dimension** | 38mm x 40 mm x 248 mm (width x thickness x length). The section that is held in operators' hand is 30mm x 38 mm (width x thickness) | 25 mm x 38 mm x 250 mm (width x thickness x length) [45.5 mm x 250 mm ~ equivalent diameter x length envelope] | Similar. |
| **Weight** | 0.58 kg (Handpiece) 0.26 kg (Cable) | 0.5 kg (Handpiece) 0.28 kg (Cable) | Similar. |
| **Power Source** | Control Box, detachable cable | Identical | Same |
| **Performance** | | | |
| **Duty Cycle** | Duty cycle of 30 minutes continuous use, 2 hours off | Duty cycle of 2 hours continuous use, 2 hours off | Different, however, the use of the subject device, to facilitate lipoplasty, is ~25-30 minutes non-continuous. |
{11}
K251944, Page 8 of 9
| Table 2 Handpiece Comparison Table | | | |
| --- | --- | --- | --- |
| | Subject Device | Predicate Device | Device Comparison |
| Surface Temperature during Operation | < 30°C | < 41°C | Similar |
| Reuse | | | |
| Cleaning | Manual | • Manual • Automated | Same |
| Sterilization | Handpiece and Handpiece Cable are steam sterilized by the end user | Identical | Same |
### 8. Non-Clinical Performance Data
To demonstrate substantial equivalence of the LHS to the currently marketed and predicate, MicroAire® PAL® System, Lipocosm LLC completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The LHS passed the tests in accordance with internal requirements, national standards, and international standards provided below, supporting its substantial equivalence to the currently marketed and predicate MicroAire® PAL® System.
- Biological evaluation per ISO 10993-1 – The results of the testing demonstrate that the Cannula was biocompatible.
- Medical electrical equipment per IEC 60601-1:2005/AMD 2:2020, AAMI ES60601-1:2005/AMD2:2021, EN 60601-1:2006/A2:2021 entitled, “Part 1: General requirements for basic safety and essential performance.” The results of this test demonstrate the basic safety of the LHS.
- Medical electrical equipment per IEC 60601-1-6:2010, IEC 60601-1-6:2010/AMD1:2013, IEC 60601-1-6:2010/AMD2:2020 for use in conjunction with IEC 62366-1:2015, IEC 62366-1:2015/AMD1:2020, and IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2: 2020 entitled, “General requirements for basic safety and essential performance - Collateral standard: Usability.” The results of this test demonstrate the electromagnetic compatibility of the LHS.
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 ed 4.1 (2020-09) entitled, “Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests.” The results of this test demonstrate the basic safety of the LHS.
{12}
K251944, Page 9 of 9
- Bench / Design Validation testing was performed on the LHS. Test results demonstrate that the LHS met the acceptance criteria and perform as intended.
- Cleaning validation study was performed by MycoScience on the reusable components of the Subject LHS using the manual cleaning procedure specified in the Instruction for Use of the Subject Lipocosm Harmonic System. The cleaning validation demonstrates that the cleaning method is adequate.
- Sterilization validation – The results of the sterilization validation of the reusable components of the LHS; Cannula, Handpiece, and Handpiece Cable demonstrate that a SAL of 10⁻⁶ is achieved when these reusable components are Steam sterilization pre-vacuum cycle of 4 minutes at 270⁰F.
## 9. Clinical Performance Data
There was no human clinical testing required to support the substantial equivalence of the LHS as the indications for use is equivalent to the currently marketed and predicate device. These types of devices, including the predicate devices, have been on the market for many years.
## 10. Statement of Substantial Equivalence
The subject device, the LHS, has been compared with the currently marketed and predicate device, MicroAire® PAL® System. Through intended use, technology, and performance testing, the LHS is substantially equivalent to the currently marketed and predicate MicroAire® PAL® System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.