← Product Code [MYN](/productcode/MYN) · K251934

# qXR-Detect (K251934)

_Qure.Ai Technologies · MYN · Jan 16, 2026 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251934

## Device Facts

- **Applicant:** Qure.Ai Technologies
- **Product Code:** [MYN](/productcode/MYN.md)
- **Decision Date:** Jan 16, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2070
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device, PCCP

## Indications for Use

qXR-Detect is a computer-assisted detection (CADe) software device that analyzes chest radiographs and highlights suspicious regions of interest (ROIs). The device is intended to identify, highlight, and categorize suspicious regions of interest (ROI). Any suspicious ROI is highlighted by qXR-Detect and categorized into one of six categories (lung, pleura, bone, Mediastinum & Hila & Heart, hardware and other). The device is intended for use as a concurrent reading aid. qXR-Detect is indicated for adults only.

## Device Story

qXR-Detect is a CADe software device analyzing DICOM chest radiographs; identifies/categorizes suspicious ROIs into six categories (lung, pleura, bone, mediastinum/hila/heart, hardware, other). Operates as a concurrent reading aid for ER physicians, family medicine practitioners, and radiologists. Deployed on secure cloud servers or on-premises; receives de-identified images via HTTPS; outputs results to PACS or radiology software. Acts as an adjunct tool; not intended to replace clinical judgment or serve as primary interpretation. Benefits include improved detection/localization performance and reduced false positives, aiding clinical decision-making.

## Clinical Evidence

Multireader multicase study (n=301 samples, 18 readers: radiologists, ER physicians, family medicine practitioners). Primary endpoint: improvement in detection/localization (wAFROC). Overall wAFROC improved from 0.6894 (unaided) to 0.7505 (aided) (p<0.001). False positives per image reduced from 0.4182 to 0.3300. AUC improved from 0.8466 to 0.8720. Standalone performance testing also provided (AUCs 0.879-0.958 across categories).

## Technological Characteristics

Software-based CADe; deep learning (CNN) architecture. Inputs: DICOM chest X-rays. Outputs: ROI highlights/categorization. Deployment: cloud or on-premises. Connectivity: HTTPS, PACS integration. Standards: ISO 13485:2016, IEC 62304:2006+A1:2015. Locked algorithm with PCCP for future updates.

## Regulatory Identification

Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- Chest CAD ([K210666](/device/K210666.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 16, 2026

Qure.ai Technologies
% Sri Anusha Matta
Director- Regulatory and Clinical Affairs
6th Floor, Wing E, Times Square, Andheri- Kurla Road, Marol
Andheri (East)
MUMBAI, MAHARASHTRA 400059
INDIA

Re: K251934
Trade/Device Name: qXR-Detect
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN
Dated: June 24, 2025
Received: December 8, 2025

Dear Sri Anusha Matta:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251934 - Sri Anusha Matta
Page 2

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these

{2}

K251934 - Sri Anusha Matta
Page 3

requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Lu Jiang Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251934  |   |
|  Device Name qXR-Detect  |   |
|  Indications for Use (Describe) qXR-Detect is a computer-assisted detection (CADe) software device that analyzes chest radiographs and highlights suspicious regions of interest (ROIs). The device is intended to identify, highlight, and categorize suspicious regions of interest (ROI). Any suspicious ROI is highlighted by qXR-Detect and categorized into one of six categories (lung, pleura, bone, Mediastinum & Hila & Heart, hardware and other). The device is intended for use as a concurrent reading aid. qXR-Detect is indicated for adults only.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K251934

Qure.ai

510(k) Summary

510(k) SUMMARY

qXR-Detect

# 1 SUBMITTER

Qure.ai Technologies Pvt. Ltd.

6th Floor, Wing E, Times Square,

Andheri-Kurla Road, Marol, Andheri (East),

Mumbai - 400059, Maharashtra

Phone: +91-8867012858

Primary Contact Person: Sri Anusha Matta, Director of Regulatory and Clinical affairs.

Secondary Contact Person: Bunty Kundnani, Chief Regulatory Affairs Officer.

Date Prepared:

# 2 SUBJECT DEVICE

|  Name of Device: | qXR-Detect  |
| --- | --- |
|  Classification Name: | Medical image Analyzer  |
|  Regulatory Class: | Class II  |
|  Regulation Number: | 21 CFR 892.2070  |
|  Product Code: | MYN  |

# 3 PREDICATE DEVICES

|  Name of Device: | Chest CAD  |
| --- | --- |
|  Manufacturer: | Imagen Technologies, Inc  |
|  510(k) Number: | K210666  |
|  Regulatory Class: | Class II  |
|  Regulation Number: | 21 CFR 892.2070  |
|  Product Code | MYN  |

# 4 INTENDED USE / INDICATIONS FOR USE:

qXR-Detect is a computer-assisted detection (CADe) software device that analyzes chest radiographs and highlights suspicious regions of interest (ROIs). The device is intended to identify, highlight and categorize suspicious regions of interest (ROI). Any suspicious ROI is highlighted by qXR-Detect and categorized into one of six categories (lung, pleura, bone, Mediastinum &amp; Hila &amp; Heart, hardware and other). The device is intended for use as a concurrent reading aid. qXR-Detect is indicated for adults only.

# 5 DEVICE DESCRIPTION

qXR-Detect is a computer-assisted detection (CADe) software device that analyzes chest radiographs and highlights suspicious regions of interest (ROIs). The device is intended to identify, highlight and categorize suspicious regions of interest (ROI). Any suspicious ROI is highlighted by qXR-Detect and categorized into one of six categories (lung, pleura, bone, Mediastinum &amp;Hila &amp; Heart, hardware and other). qXR-Detect is indicated for adults only. qXR-Detect is an adjunct tool

{5}

Qure.ai

510(k) Summary

and is not intended to replace a clinician's review of the radiograph or his/her clinical judgment. The users must not use the qXR-Detect generated output as the primary interpretation.

The qXR-Detect device is intended to generate a secondary digital radiographic image that facilitates the confirmation of the presence of suspicious region of interest within the categories on a chest X-Ray. This device is intended to be used by trained professionals- ER physicians, family medicine practitioners, and radiologists.

The software works with DICOM chest X-ray images and can be deployed on a secure cloud server. De-identified chest X-rays are sent to qXR-Detect via HTTPS from the client's software. Results are fetched by the client's software and then forwarded to their PACS or any other systems including but not limited to specified radiology software database once the processing is complete or to the console of the digital radiographic processing system.

The underlying algorithm was trained on the large data set of  $\sim 2.5$  million Chest X-Ray scans which consisted of scans from the US, EU, India, Vietnam, Brazil. These scans consisted of various abnormal regions of interest.

# 6 COMPARISON WITH PREDICATE DEVICE

Table 1 Comparison between qXR-Detect and the Predicate Device

|  Components and Characteristics | Subject Device | Predicate Device  |
| --- | --- | --- |
|  Regulation Information  |   |   |
|  Device Name | qXR-Detect | Chest-CAD  |
|  Manufacturer | Qure.ai Technologies Pvt. Ltd. | Imagen Technologies, Inc.  |
|  510(k) Number | - | K210666  |
|  Regulation Number/Name | 21 CFR 892.2070 / Medical Image Analyzer | 21 CFR 892.2070 / Medical Image Analyzer  |
|  Product Code | MYN | MYN  |
|  Intended Use | qXR-Detect is a computer-assisted detection (CADe) software device that analyzes chest radiographs and highlights suspicious regions of interest (ROIs). The device is intended to identify, highlight and categorize suspicious regions of interest (ROI). Any suspicious ROI is highlighted by qXR-Detect and categorized into one of six categories (lung, pleura, bone, Mediastinum & Hila & Heart, hardware and other). The device is intended for use as a concurrent reading aid. qXR-Detect is indicated for adults only. | Chest-CAD is a computer-assisted detection (CADe) software device that analyses chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bone, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.  |
|  Patient Population | Adults with Chest Radiographs | Adults with Chest Radiographs  |
|  Indications for Use  |   |   |
|  Disease | Qure.ai Technologies Pvt. Ltd. | Imagen Technologies, Inc.  |
|  510(k) Number | - | K210666  |
|  Regulation Number/Name | 21 CFR 892.2070 / Medical Image Analyzer | 21 CFR 892.2070 / Medical Image Analyzer  |

{6}

Qure.ai
510(k) Summary

|  Image Modality | Chest Radiographs | Chest Radiographs  |
| --- | --- | --- |
|  Clinical Finding and Clinical Output | Suspicious regions of interest (ROI)
To inform the primary diagnostic and patient management decisions that are made by the clinical user. | Suspicious regions of interest (ROI)
To inform the primary diagnostic and patient management decisions that are made by the clinical user.  |
|  Intended Users | ER physicians, Family medicine practitioners and radiologists | Physicians  |
|  Software and Technical Information  |   |   |
|  Machine Learning Methodology | Deep Learning | Deep Learning  |
|  Image Source | DICOM Source (e.g., imaging device, intermediate DICOM node, PACS system, etc.) | DICOM Source (e.g., imaging device, intermediate DICOM node, PACS system, etc.)  |
|  Image Viewing | PACS system
Image annotations made on copy of original image. | PACS system
Image annotations made on copy of original image or image annotations toggled on/off  |
|  Deployment Platform | Deployment on-premises or on cloud and connection to several computing platforms and X-ray imaging platforms such as X-ray radiographic systems, or PACS | Deployment on-premises or on cloud and connection to several computing platforms and X-ray imaging platforms such as X-ray radiographic systems, or PACS  |
|  Privacy | HIPAA Compliant | HIPAA Compliant  |

Page 3

{7}

Qure.ai

510(k) Summary

# 7 TESTING

# Software:

Software verification and validation testing was executed, and documentation was provided as recommended by FDA's Guidance for the Content of Premarket Submission for Device Software Functions (June 2023). During all verification and validation tests carried out for the qXR-Detect software, which included evaluating both the algorithmic functionality and the overall performance of the software and its components, qXR-Detect functioned as designed and successfully met the anticipated performance criteria.

Verification, validation, and testing activities were conducted to establish the performance, functionality, and reliability characteristics of the device. Unit Test, Integration Test, Regression Test and User Acceptance test were carried out to account towards the device's performance non-clinically. Functional testing is done to assess functional requirements of the product. The device passed all the tests based on determined acceptance criteria. Standards Regulatory references Used are ISO 13485: 2016 and IEC 62304:2006+A1:2015.

# Non-Clinical Performance Testing:

Nonclinical performance was conducted as a standalone study to assess the device's performance against the ground truth. Most of the scans for the study were obtained from across the US spanning 40 states and 5 regions in the US. The dataset was also well-balanced in terms of gender, with approximately a 50-50 male-to-female distribution. Age distribution spans a wide range, from 22 years to over 85 years.

The ground truth was established by 3 ground truthers annotated the chest X-ray scans for the presence of suspicious ROI categories. If there is at least one ground truth boundary for a particular category, the scan is considered to be positive for that category. A particular scan can be positive for multiple categories.

On comparison with the qXR-Detect output the device perform satisfactorily. The result of the performance is summaries in the table 2.

Table 2 Standalone Performance Testing Results for qXR-Detect

Standalone Performance metrics

|  Category | AUC (95%CI)  |
| --- | --- |
|  Lung | 0.893 (0.879-0.907)  |
|  Pleura | 0.95 (0.94-0.96)  |
|  Mediastinum /Hila | 0.891 (0.875-0.907)  |
|  Bone | 0.879 (0.854-0.905)  |

{8}

Qure.ai

510(k) Summary

|  Hardware | 0.958 (0.95-0.966)  |
| --- | --- |
|  Other | 0.915 (0.895-0.935)  |

Additionally, localization accuracy was also estimated as secondary analyses.

Table 3 Standalone Performance Testing Results for localization - qXR-Detect

|  Category | wAFROC (95% CI)  |
| --- | --- |
|  Lung | 0.831 (0.816-0.846)  |
|  Pleura | 0.89 (0.875-0.905)  |
|  Mediastinum & Hila & Heart | 0.867 (0.85-0.883)  |
|  Bone | 0.821 (0.789-0.852)  |
|  Hardware | 0.771 (0.759-0.782)  |
|  Others | 0.871 (0.845-0.897)  |
|  Aggregate | 0.839 (0.824-0.854)  |

It was observed that wAFROC was above 0.8 for most categories. This indicates that qXR-Detect is able to demonstrate a good performance in detecting and localizing ROI within the categories.

# Clinical Performance Testing:

A clinical performance study was conducted via a multireader multicase study conducted on 301 samples with 18 readers. The readers were radiologists, ER physicians and Family medicine practitioners. The study was designed as fully-crossed concurrent. This study protocol was designed in accordance with FDA's guidance document for industry and FDA staff titled "Clinical Performance Assessment: Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data in Premarket Notification (510(k)) Submissions".

The primary metric of the study was to determine the improvement in readers' detection and localization performance in identifying scans with suspicious chest ROIs in the 6 categories. The same was estimated using a wAFROC metric. The readers read scans in 2 sessions, with a washout period of 28 days between the 2 sessions. The scans were randomized in blocks such scans in one block were read aided and the next unaided in the first session. Scans that were initially read as aided; were read as unaided scans in the second session (after a minimum of 28 days of washout).

To compare reader performance across unaided and aided sessions, the area under the weighted alternative free-response receiver operating characteristic curve (wAFROC-FOM) was employed as the figure of merit for the primary analysis.

The overall wAFROC significantly improved from 0.6894 (0.6453 - 07335) in the unaided reads to 0.7505 (0.7029 - 0.8291) in reads aided by qXR-Detect with an improvement of 0.0611 (0.0320-0.0901) with a significant P value (&lt; 0.001). The wAFROC improved in all categories. The number of false positives per image was also presented. There was an improved in the aided reads, where the number of false positives per image reduced from 0.4182 (0.4012 - 0.4358) in the reads unaided - to

{9}

Qure.ai

510(k) Summary

0.3300 (0.3149 - 0.3457) in the reads aided by qXR-Detect. This indicates that the use of qXR-Detect does not cause the users to overcall conditions.

Table 4 wAFROC of reads unaided and aided by qXR-Detect

|  ROI | wAFROC with 95% CI - Unaided | wAFROC with 95% CI - Aided  |
| --- | --- | --- |
|  Overall | 0.6894 (0.6453 -0.7335) | 0.7505 (0.7029-0.7980)  |
|  Lung | 0.7270 (0.6857-0.7682) | 0.7908 (0.7525-0.8291)  |
|  Pleura | 0.7829 (0.7427-0.8230) | 0.8460 (0.8127-0.8794)  |
|  Mediastinum, Hila and Heart | 0.6857 (0.6364-0.7350) | 0.7773 (0.7298-0.8248)  |
|  Bone | 0.7117 (0.6712-0.7522) | 0.8133 (0.7745-0.8521)  |
|  Hardware | 0.7232 (0.6930-0.7533) | 0.7330 (0.7044-0.7616)  |
|  Other | 0.7975 (0.7512-0.8437) | 0.8329 (0.7929-0.8728)  |

17/18 readers showed improvements in wAFROC-FOM in reads aided by qXR-Detect across all categories. All 18 readers demonstrated improvement in detecting and localizing suspicious lung ROIs.

The accuracy of readers in detecting suspicious ROIs on chest X-rays was evaluated using the area under the curve (AUC). The AUC of the ROC curves were estimated and there was a significant improvement in the aided reads, as compared to the unaided reads. The overall AUC for aided reads was 0.8466 (0.8106 - 0.8826) whereas the aided reads showed AUC of 0.8720 (0.8339 - 0.9100). There was improvement observed across all categories.

Table 5 AUROC, Sensitivity and Specificity of reads unaided and aided by qXR-Detect

|  AUROC 95% CI - Unaided | AUROC with 95% CI - Aided | Sensitivity 95% CI - Unaided | Sensitivity with 95% CI - Aided | Specificity 95% CI - Unaided | Specificity with 95% CI - Aided  |
| --- | --- | --- | --- | --- | --- |
|  0.8466(0.8106-0.8826) | 0.8720(0.8339-0.9100) | 0.8896(0.8596-0.9195) | 0.9338(0.9096-0.9579) | 0.5556(0.4297-0.6814) | 0.6219(0.5065-0.7373)  |

We believe that the performance metrics indicate that the device is substantially equivalent to the predicate.

# 8 PREDETERMINED CHANGE CONTROL PLAN

The qXR-Detect submission contains a Predetermined Change Control Plan (PCCP). Modifications to the algorithm of qXR-Detect will be made in accordance with the guidelines outlines in the "Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions". The PCCP describes the modifications that can occur in a controlled manner ensuring the continued safety, effectiveness and performance of the device.

{10}

Qure.ai

510(k) Summary

The algorithm modifications will not include any adaptive algorithms that get continuously learn from the environment. There are two modifications outlined in the PCCP, and accordingly, any change will only occur after training, testing and validating the algorithm after which the change will be locked. When the modification is locked this will be communicated to all users through updated labelling via the Instructions for use. Verification and Validation activities will be conducted in accordance with the PCCP to compare the performance of the modified qXR-Detect algorithms to the original model to ensure the safety and effectiveness. The modification details are summarised as follows,

Table 6 Details of Modification M1 and M2

|  Modifications Summary  |   |
| --- | --- |
|  M1 – Data Driven Retraining  |   |
|  Description | Modification 1 (M1) will involve retraining the existing model using newly acquired, high-quality adult chest radiograph datasets to improve device performance without altering the underlying model architecture.  |
|  Rationale | Improve the algorithm performance  |
|  Retraining Triggers | The availability of a new, high-quality dataset obtained from different parts of US, that improves the diversity of the training data, offering a logical pathway to improve overall model performance.  |
|  Testing Methods | Retraining with newly acquired, high-quality adult chest radiograph datasets, followed by locked test-set evaluation and comparison to baseline model using predefined acceptance criteria. Performance evaluation for Modification 1 will ensure that the retrained CNN-based model continues to operate safely, effectively, and within the boundaries of the device's cleared indications for use. The updated model will be evaluated using a predefined set of clinically meaningful and statistically robust metrics to confirm that the performance is improved without introducing any regression across the six existing condition categories.  |
|  Update Procedures/Implementation | Before deployment, each update is reviewed to ensure it stays within the scope of the PCCP with no new outputs, indications, or workflow changes and is released only after all verification and validation confirm it meets predefined success criteria. If these criteria are  |

{11}

Qure.ai
510(k) Summary

|   | met, the update is implemented through a controlled, manual deployment process and made globally available to all end users.

Users will be notified of updates implemented under the PCCP through formal communication channels, including revised labelling, updated user manuals, and detailed release notes. These materials will clearly outline any changes to the device’s performance, functionality, or indications for use, and will be disseminated in accordance with established quality system procedures to ensure timely, accurate, and comprehensive stakeholder awareness.  |
| --- | --- |
|  Impact Assessment | **Benefits:** Improved detection and localization performance across six clinically relevant categories;

**Risks:** Potential overfitting and unintended bias if new data are imbalanced; risk of minor performance variability.

**Risk Mitigation:** The modified model will be tested for non-inferiority on the performance study test dataset which will contain new unseen data.  |
|  **M2 – Architectural modification**  |   |
|  Description | Modification 2 (M2) will involve a controlled architectural change through modular replacement of the existing Convolutional Neural Network (CNN) encoder with a Vision Transformer (ViT)-based encoder, while maintaining the remaining components and the workflow of the model pipeline.  |
|  Rationale | Enhance feature extraction and global context understanding to further improve diagnostic performance  |
|  Retraining Triggers | Given the established potential of ViT models, Modification 2 will be explored if a superior Vision encoder becomes available. If the updated architecture demonstrates improvement as described in the Acceptance Criteria, the modification will be accepted.  |
|  Testing Methods | For testing the M2 modification, candidate models (trained on ViT architecture) will undergo rigorous evaluation against the baseline CNN model. The updated model will be evaluated using a predefined set of clinically meaningful and statistically robust metrics to confirm that the performance is improved without introducing any regression across the six existing condition categories. Testing  |

Page 8

{12}

Qure.ai
510(k) Summary

|   | includes assessment of performance metrics- (AUC, Box level sensitivity, specificity, subgroup analysis). This ensures that candidate model continues to operate safely, effectively, and within the boundaries of the device’s cleared indications for use.  |
| --- | --- |
|  Update Procedures/Implementation | Before deployment, each update is reviewed to ensure it stays within the scope of the PCCP with no new outputs, indications, or workflow changes and is released only after all predefined verification and validation confirm it meets predefined success criteria. If these criteria are met, the update is implemented through a controlled, manual deployment process and made globally available to all end users.

Users will be notified of updates implemented under the PCCP through formal communication channels, including revised labelling, updated user manuals, and detailed release notes. These materials will clearly outline any changes to the device’s performance, functionality, or indications for use, and will be disseminated in accordance with established quality system procedures to ensure timely, accurate, and comprehensive stakeholder awareness.  |
|  Impact Assessment | **Benefits:** Improved ROI classification accuracy and localization precision
**Risks:** Possibility of new failure modes due to architectural change and model instability; risk of decoder-compatibility errors.
**Risk Mitigation:** Full integration verification; performance comparison to baseline using identical metrics and datasets  |

# 9 CONCLUSION

The comparison in Table 1 as well as the software &amp; performance testing presented above demonstrate that the qXR-Detect device is substantially equivalent to the predicate device. Both devices are computer-assisted detection (CADe) software devices that analyze chest radiographs using machine learning techniques to identify and localise suspicious regions of interest (ROI). The new device does not introduce fundamentally new scientific technology, and the clinical tests demonstrate that the device is safe and effective.

The qXR-Detect is a software device with similar indications, technological characteristics, and principles of operation as the predicate devices. The comparison of intended purpose, technological

Page 9

{13}

Qure.ai
510(k) Summary

characteristics and performance demonstrates that the qXR-Detect device performs as intended and can be considered as substantially equivalent to the predicate device, Chest CAD (K210666).

Based on the information provided in this premarket notification, including the indications for use, technological characteristics, and performance testing, qXR-Detect is determined to be substantially equivalent to the predicate device. The device does not raise new questions of safety or effectiveness and demonstrates comparable safety, efficacy, and performance.

Page 10

---

**Source:** [https://fda-staging.innolitics.com/device/K251934](https://fda-staging.innolitics.com/device/K251934)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
