Sleep Apnea Breathing Therapy Mask:JoyMask 20-B Full Face Mask, JoyMask 21-B Full Face Mask
K251920 · Dcstar, Inc. · BZD · May 21, 2026 · Anesthesiology
Device Facts
Record ID
K251920
Device Name
Sleep Apnea Breathing Therapy Mask:JoyMask 20-B Full Face Mask, JoyMask 21-B Full Face Mask
Applicant
Dcstar, Inc.
Product Code
BZD · Anesthesiology
Decision Date
May 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system. -The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment.
Device Story
Full face mask (JoyMask 20-B, 21-B) covers mouth and nose to deliver pressurized air from PAP device; comprises cushion, frame, elbow, and headgear. Cushion provides airtight seal; elbow acts as airflow conduit; includes exhalation port and anti-asphyxia valve (AAV) for CO2 flushing. Used in home environment by patients >30kg; prescription required. Clinician uses device to deliver prescribed therapy pressure; mask ensures effective airflow delivery to patient. Benefits include non-invasive airway support for sleep apnea management.
Clinical Evidence
No clinical data. Bench testing only, including dead space, exhaust flow, AAV pressure, resistance to flow, CO2 re-breathing, and sound levels per ISO 17510. Biocompatibility testing performed per ISO 10993 and ISO 18562 series.
Technological Characteristics
Full face mask with 22mm conical connector (ISO 5356-1). Pressure range 4-30 cmH2O. Non-sterile. Materials evaluated for biocompatibility per ISO 10993-5, -10, -23 and ISO 18562-2, -3, -4. Mechanical design includes AAV and exhalation port. No magnets.
Indications for Use
Indicated for adult patients (>30kg) requiring positive airway pressure therapy (CPAP or bi-level) for sleep apnea; intended for single-patient, multi-use in home environments.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2026
DCStar Inc.
Emma Chen
Regulatory Affairs Specialist
228 Park Ave S 30327
New York, New York 10003
Re: K251920
Trade/Device Name: Sleep Apnea Breathing Therapy Mask: JoyMask 20-B Full Face Mask, JoyMask 21-B Full Face Mask
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: April 16, 2026
Received: April 16, 2026
Dear Emma Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251920 - Emma Chen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251920 - Emma Chen
Page 3
Sincerely,
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251920 | |
| Device Name Sleep Apnea Breathing Therapy Mask:Joymask 20-B, Joymask 21-B | |
| Indications for Use (Describe) The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system. -The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## Type of submission
Traditional
## Date prepared
5/21/2026
## Submitter's Information
Company: DCSTAR INC
Address: 228 PARK AVE S 30327, NEW YORK, NY, UNITED STATES, 10003
Submitter: Emma Chen
Telephone: +1 (650) 672-9428
Email: compliance@complianceglobe.com
## Device Information
Device name: JoyMask 20-B Full Face Mask, JoyMask 21-B Full Face Mask
Classification name: Non Continuous Ventilator (IPPB)
Production code: BZD
Regulation number: 21 CFR 868.5905
Classification: Class II
## Predicate Device Information
510(k) Number: K241830
Manufacturer: Sleepnet Corporation
Device Name: Innova Full Face Vented Mask
Device Class: Class II
Product Code: BZD
{5}
510(k) Summary
# Reference Device Information
510(k) Number: K212371
Manufacturer: Fisher & Paykel Healthcare Ltd
Device Name: F&P Evora Full Face Mask
Device Class: Class II
Product Code: BZD
# Indications for use
The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system.
- The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment.
# Device Description
JoyMask 20-B and JoyMask 21-B are full face masks designed to cover the patient's entire mouth and nose, delivering therapeutic pressure through both airways. Each model comprises four main components: a cushion, frame, elbow, and headgear.
The headgear connects to the frame via a plastic clip, securing the cushion in place. The cushion ensures an airtight seal for effective airflow delivery, while the elbow serves as the conduit for the airflow itself.
{6}
510(k) Summary
Substantial Equivalence Table
Table 1 Substantial Equivalence Comparison
| Item | Subject Device | Predicate Device | Reference Device | Remark |
| --- | --- | --- | --- | --- |
| Device | Sleep Apnea Breathing Therapy Mask - JoyMask 20-B and JoyMask 21-B | Innova Full Face Vented Mask | F&P Evora Full Face Mask | - |
| Product Code | BZD | BZD | BZD | Same |
| Regulation No. | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905 | Same |
| Indication for Use | The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system.-The mask is to be used by patients (weighing >30 kg), intended for single-patient reuse in the home environment. | Innova Full Face Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bilevel, operating at or above 3 cm H2O.The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.Single patient, multi-use in the home or hospital/institutional environment. | The F&P Evora Full Face mask is intended to be used by adults weighing ≥66lbs (30kgs) who have been diagnosed by a physician as requiring CPAP or Bi-Level therapy.The F&P Evora Full Face mask is intended for single patient use in the home and for multiple patient use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses. | Equivalent to predicate device |
| Population | Used on adult patients (>30kg) requiring positive airway pressure therapy. | Used on adult patients (>30kg) requiring positive airway pressure therapy. | Used on adult patients (≥30kg) requiring positive airway pressure therapy. | Equivalent to predicate device |
| Duration and Environment of Use | Single-patient, multi-use in the home environment. | Single patient, multi-use in the home or hospital/institutional environment. | Single patient use in the homeMultiple patient use in the hospital or other clinical setting | Equivalent to predicate device |
| Prescription for Use | Yes | Yes | Yes | Same |
| Principle of Operation | Provides a seal over the face (nose and mouth) to allow for delivery of pressurized air from a positive pressure | Provides a seal over the face (nose and mouth) to allow for delivery of pressurized air from a positive | Provides a seal over the face (nose and mouth) to allow for delivery of pressurized air from a positive pressure | Same |
{7}
510(k) Summary
| | device such as CPAP or bilevel device. Mask includes an exhalation port and AAV for flushing out exhaled CO2. | | | pressure device such as CPAP or bilevel device. Mask includes an exhalation port and AAV for flushing out exhaled CO2. | | | device such as CPAP or bilevel device. Mask includes an exhalation port and AAV for flushing out exhaled CO2. | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Non-sterile | Yes | | | Yes | | | Yes | | Same |
| Available Size | S, M, L | | | S, M, L | | | XS, S-M, L | | Equivalent to predicate device |
| Technical Specifications | | | | | | | | | |
| User Interface to administer therapy | 22mm conical connector (compliant with ISO 5356-1) | | | 22mm conical connector (compliant with ISO 5356-1) | | | 22mm conical connector (compliant with ISO 5356-1) | | Same |
| Dead Space | Size | Dead Space (ml) | | Size | Dead Space (ml) | | Size | Dead Space (ml) | Similar to predicate device The dead space are detailed in the IFU. |
| | Small | 175.6 | | Small | 155.3 | | Extra Small | 165.2 | |
| | Medium | 188.6 | | Medium | 181.7 | | Small-Medium | 162.6 | |
| | Large | 210.3 | | Large | 196.6 | | Large | 164.1 | |
| Pressure Range | 4 to 30 cmH2O | | | 3 to 20 cmH2O | | | 4 to 30 cmH2O | | Identical to reference device The pressure range, along with other technical specifications such as exhaust flow characteristics, is detailed in the IFU as a reference for clinicians. The difference from the predicate device do not introduce additional risks. |
| Therapy Pressure and its exhaust flow characteristics | Pressure (cmH2O) | | Flow (lpm) | Pressure (cmH2O) | Flow (lpm) | | Not Publicly Available | Similar to predicate device In accordance with ISO 17510, the therapy pressure and exhaust flow characteristics are detailed in | |
| | 4 | | 20.6 | 3 | 22.08 | | | | |
| | 11 | | 37.0 | 10 | 38.50 | | | | |
| | 17 | | 49.4 | 20 | 54.30 | | | | |
| | 24 | | 60.9 | | | | | | |
{8}
510(k) Summary
| | 30 | 67.8 | | | the IFU as a reference for the clinicians. | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | |
| AAV Opening/Closing Pressure | Opening – 1.89 cmH2O Closing – 2.22 cmH2O | | Opening – 2.10 cmH2O Closing – 2.28 cmH2O | | Not Publicly Available | Similar to predicate device Meet the acceptance criteria as per ISO 17510 (<4 cmH2O). The AAV Opening/Closing Pressure are detailed in the IFU as a reference for the clinicians. |
| Resistance to Flow | 50 lpm – 0.48 cmH2O 100 lpm – 0.78 cmH2O | | 50 lpm – 0.07 cmH2O 100 lpm – 0.39 cmH2O | | 50 lpm – 0.9 ± 0.3 cm H2O 100 lpm – 2.2 ± 0.3cm H2O | Similar to predicate device In accordance with ISO 17510, the resistance values are detailed in the IFU as a reference for the clinicians. |
| Magnet | No Magnet | | No Magnet | | No Magnet | Same |
| CO2 re-breathing performance | Setting | Relative CO2 increase | Setting | Relative CO2 increase | Not Publicly Available | Similar to predicate device Meet the acceptance criteria as per ISO 17510 (<20% under normal condition and <60% under single fault condition). |
| | 4 cmH2O | 11.6% | 3 cmH2O | 16% | | |
| | 5 cmH2O | 9.4% | 5 cmH2O | 16% | | |
| | 10 cmH2O | 5.0% | 10 cmH2O | 12% | | |
| | Single Fault1 | 23.9% | Single Fault1 | 25% | | |
| | Single Fault2 | 30.7% | Single Fault2 | 45% | | |
| AAV Inspiratory/Expiratory Resistance in Single Fault Condition | Inspiratory – 0.89 cm H2O Expiratory – 0.63 cm H2O | | Inspiratory – 0.6 cm H2O Expiratory – 0.5 cm H2O | | Not Publicly Available | Similar to predicate device Meet the acceptance criteria as per ISO 17510 (<10 |
{9}
510(k) Summary
| | | | | cmH2O). |
| --- | --- | --- | --- | --- |
| Sound Pressure and Sound Power Level | Sound Pressure – 27.0 dBA Sound Power – 34.1 dBA | Sound Pressure – 29.98 dBA Sound Power – 32.99 dBA | Sound Pressure: 20.2 dBA, with uncertainty 2.5 dBA Sound Power: 28.2 dBA, with uncertainty 2.5 dBA | Similar to predicate device In accordance with ISO 17510, the values are detailed in the IFU. |
## Substantial Equivalence Discussion
## Indications for Use
Both the subject devices and the predicate device are intended for use with Positive Airway Pressure (PAP) devices to provide airway pressure therapy for adult patients weighing over $30\mathrm{kg}$. Both devices are available by prescription only and are intended for single-patient, multi-use in the home environment. The minor differences between them are strictly limited to language expression.
Therefore, the indications for use of the subject devices (JoyMask 20-B and JoyMask 21-B) are considered identical to those of the predicate device (Innova Full Face Vented Mask). No new risks regarding the indications for use are introduced.
## Technical Characteristic and Performance
Both the subject devices and the predicate device are full-face masks designed to provide a seal over the user's face (covering both the nose and mouth) to deliver pressurized air from a positive airway pressure source.
In terms of technical characteristics, the subject devices are demonstrated to be similar to the predicate device and reference device. The minor technical differences have been fully evaluated in accordance with applicable standard and introduce no new risks. Therefore, the technical characteristics and performance of the subject devices (JoyMask 20-B and JoyMask 21-B) are substantially equivalent to those of the predicate device (Innova Full Face Vented Mask), raising no new questions of safety or effectiveness.
{10}
510(k) Summary
## Performance Testing - Clinical
Not Applicable
## Performance Testing - Animal
Not Applicable
## Non-clinical testing
The following tests were conducted to verify the safety and effectiveness of subject devices.
## Performance Testing
The following technical characteristic have been verified:
Dead Space, Exhaust Flow Characteristics, AAV Pressure, Resistance to Flow, CO2 Re-breathing Performance, AAV Inspiratory/Expiratory Resistance, Sound Pressure and Sound Power Level
## Biocompatibility Testing
Biocompatibility testing was performed in accordance with the following standards:
- ISO 10993-5 – Cytotoxicity
- ISO 10993-10 – Sensitization
- ISO 10993-23 – Intracutaneous Reactivity
- ISO 18562-2 – Particulate Matter Release
- ISO 18562-3 – VOCs Emissions with Toxicological Risk Assessment
- ISO 18562-4 – Leachates in Condensates with Toxicological Risk Assessment
## Other Testing
- Transportation Testing
- Cleaning Validation
- Shelf Life and User Life Verification
## Recognized Standards/Guidance
The subject devices comply with the following standards/guidance
- Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff
- Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process Guidance for Industry and Food and Drug Administration Staff
{11}
ISO 17510:2015 - Medical devices - Sleep apnoea breathing therapy - Masks and application accessories
ISO 5356: Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets
ISO 17664: Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices
ISO 10993-1:2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021 - Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-17:2023 - Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents
ISO 10993-23:2021 - Biological evaluation of medical devices - Part 23: Tests for irritation
ISO 18562-1:2024 - Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
ISO 18562-2:2024 - Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
ISO 18562-3:2024 - Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic substances
ISO 18562-4:2024 - Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 4: Tests for leachables in condensate
## Conclusion
Based on the information above, the subject devices (JoyMask 20-B and JoyMask 21-B) and the predicate device (Innova Full Face Vented Mask) have the same indication for use and similar technical characteristic.
The SE discussion and relevant data have demonstrated that the subjects devices are substantially equivalent to the predicate device and the differences do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.