Thulium Laser System (Deluxlase)

K251919 · Nanjing Bestview Laser S&T Co., Ltd. · GEX · Sep 26, 2025 · General, Plastic Surgery

Device Facts

Record IDK251919
Device NameThulium Laser System (Deluxlase)
ApplicantNanjing Bestview Laser S&T Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Deluxase is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).

Device Story

Thulium laser system (Deluxase) emits pulsed 1927 nm coherent near-infrared light for dermatological applications. System includes laser source, footswitch activation, optical fiber, and handpiece. Integrated LED touch screen allows user control of laser parameters; 650 nm aiming beam assists targeting. Operated by clinicians in clinical settings. Device delivers energy to treatment zone to achieve soft tissue coagulation or lesion treatment. Benefits include precise tissue interaction for specified dermatological conditions.

Clinical Evidence

Bench testing only. No clinical data presented. Safety and effectiveness supported by non-clinical testing including EMC, electrical, mechanical, thermal, and biocompatibility (cytotoxicity, irritation, sensitization) evaluations, and adherence to recognized consensus standards.

Technological Characteristics

Thulium laser; 1927 nm wavelength; 650 nm aiming beam. Optical fiber delivery with handpiece. LED touch screen interface. Standards: IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2, IEC 60825-1, ISO 10993-5, ISO 10993-10, ISO 10993-23.

Indications for Use

Indicated for dermatological procedures requiring soft tissue coagulation, treatment of actinic keratosis, and benign pigmented lesions (e.g., lentigos, solar lentigos, ephelides).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 26, 2025 Nanjing Bestview Laser S&T Co., Ltd. Jing Wang Management Representative 1st & 2nd Floor, Building 5, Area 1, Phase 2 Liandong U Vly Science and Tech Innovation Pk, No.1 Hengyi Rd Nanjing, Jiangsu 210000 China Re: K251919 Trade/Device Name: Thulium Laser System (Deluxlase) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 23, 2025 Received: June 23, 2025 Dear Jing Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K251919 - Jing Wang Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K251919 - Jing Wang Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2025.09.26 10:27:03 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Thulium Laser System Page 9 of 38 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251919 | ? | | Please provide the device trade name(s). | | ? | | Thulium Laser System (Deluxase) | | | | Please provide your Indications for Use below. | | ? | | Deluxase is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} 510(k) #: K251919 510(k) Summary Prepared on: 2025-06-23 | Contact Details | 21 CFR 807.92(a)(1) | | --- | --- | | Applicant Name | Nanjing Bestview Laser S&T Co., Ltd. | | Applicant Address | 1st & 2nd Floor, Building 5, Area 1, Phase 2, Liandong U Valley Science and Technology Innovation Park, No.1 Hengyi Road, Nanjing Economic and Technological Development Zone, Nanjing 210000, Jiangsu, P.R. China Nanjing Jiangsu 210000 China | | Applicant Contact Telephone | +86-15824831075 | | Applicant Contact | Ms. Jing Wang | | Applicant Contact Email | jingwang@bestviewlaser.com | | Device Name | 21 CFR 807.92(a)(2) | | Device Trade Name | Thulium Laser System (Deluxase) | | Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology | | Classification Name | Powered Laser Surgical Instrument | | Regulation Number | 878.4810 | | Product Code(s) | GEX | | Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | K223727 | Lavieen | | Product Code | GEX | | Device Description Summary | 21 CFR 807.92(a)(4) | | Thulium Laser System is a thulium laser, producing a pulsed beam of coherent near-infrared light (1927 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LED touch screen gives the user control over the necessary laser system parameters. The Thulium Laser System is equipped with a 650 nm aiming beam. | | | Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | Deluxase is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). | | | Indications for Use Comparison | 21 CFR 807.92(a)(5) | | The indications for use of the proposed device are the same as the predicate device. | | | Technological Comparison | 21 CFR 807.92(a)(6) | {5} The proposed device has no difference in significant technological characteristics such as wavelength, max pulse power and pulse width with the predicate device. There are only slight differences in scan area size, machine size and weight which are unimportant and can not result in a negative effect on safety and effectiveness. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Nonclinical Tests: (1) EMC (2) Electrical, Mechanical, and Thermal (3) Biocompatibility: Cytotoxicity, Irritation, Sensitization (4) Bench testing Recognized Consensus Standards: (1) IEC 60601-1: 2005(R)2012; (2) IEC 60601-2-22:2019; (3) IEC 60601-1-2: 2020; (4) IEC 60825-1:2014; (5) ISO 10993-5:2009; (6) ISO 10993-10: 2021; (7) ISO 10993-23: 2021 Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...